K153543 · Whill, Inc. · ITI · Feb 12, 2016 · Physical Medicine
Device Facts
Record ID
K153543
Device Name
WHILL Model M
Applicant
Whill, Inc.
Product Code
ITI · Physical Medicine
Decision Date
Feb 12, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The intended use of the Model M powered wheelchair is to provide outdoor and indoor mobility to persons limited to a seated position that are capable of operating a powered wheelchair.
Device Story
WHILL Model M is a battery-operated, 4-wheel drive powered wheelchair for indoor/outdoor use. User operates device via directional controller mounted on arm supports to control speed and movement; R-net 120 amp motor controller manages drive system. Device features welded steel frame, two 12V 50Ah batteries, and all-directional front wheels. User releases controller to trigger controlled stop. Designed for operation on various surfaces including carpet, wood, concrete, and dirt. Provides mobility for individuals with physical limitations; intended for prescription use. Benefits include enhanced maneuverability and independent mobility for seated users.
Clinical Evidence
Bench testing only. No clinical data provided. Device performance was validated against ANSI/RESNA WC-1, WC-2, and WC-4 standards, including static/dynamic stability, braking effectiveness, obstacle climbing, electromagnetic compatibility, and flammability (ISO 8191-1/2, UL94). Biocompatibility testing conducted per ISO 10993-1, -5, and -10.
Technological Characteristics
4-wheel drive powered wheelchair; welded steel frame. Powered by two 12V 50Ah batteries; R-net 120 amp motor controller. Features all-directional front wheels. Dimensions: 23.6" width, 37"-42" length. Max weight capacity 220 lb. Connectivity: None. Standards: ANSI/RESNA WC-1/2/4, ISO 7176-16, ISO 8191-1/2, UL94, ISO 10993.
Indications for Use
Indicated for persons limited to a seated position capable of operating a powered wheelchair, requiring indoor and outdoor mobility.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing right, suggesting a focus on people and health.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 12, 2016
Whill, Inc. % Mark Job Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, Minnesota 55313
Re: K153543
Trade/Device Name: Whill Model M Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: January 28, 2016 Received: January 29, 2016
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Hoffmann -A
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153543
Device Name WHILL Model M
Indications for Use (Describe)
The intended use of the Model M powered wheelchair is to provide outdoor mobility to persons limited to a seated position that are capable of operating a powered wheelchair.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 5 510(k) Summary
| Submitter's Name: | WHILL, Inc. |
|--------------------------|-------------------------------------------|
| Address: | 285 Old County Rd, Suite 6, San Carlos CA |
| Contact Person: | Terese Bogucki |
| Title: | Regulatory Consultant |
| Telephone Number: | 650-488-7799 |
| Fax Number: | 650-227-2264 |
| Email: | terri@decusbiomedical.com |
| Date Summary Prepared: | August 24, 2015 |
| Device Proprietary Name: | WHILL Model M Powered Wheelchair |
| Model Number: | Model M |
| Common Name: | Powered Wheelchair |
| Regulation Number: | 21 CFR 890.3860 |
| Product Code: | ITI |
| Device Class: | II |
# Predicate Device
| Trade Device Name: | F3 Corpus |
|------------------------|------------------------------------------------------------|
| Manufacturer: | Permobil AB |
| Address: | Per Uddens Vag 20 Timra, SE 86123 Vasternorrland<br>Sweden |
| Regulation Number: | 21 CFR 890.3860 |
| Regulation Name: | Wheelchair, Powered |
| Regulatory Class: | Class II |
| Product Code: | ITI |
| 510(k) Number: | 143180 |
| 510(k) Clearance Date: | March 27, 2015 |
{4}------------------------------------------------
#### 5.1 Description of the Device
The WHILL Model M is an indoor/outdoor battery-operated 4 wheel drive powered wheelchair. It is powered by two 12 VDC 50Ah batteries and controlled by the R-net 120 amp motor controller. As with all commerciallyavailable powered wheelchairs, the user sits in the wheelchair seat and uses controls positioned on the arms to turn the chair on, control the speed, and direct the movement. The directional controller can be mounted on the left or right arm. When the user activates the directional controller the brakes are released and the motors rotate to move the device in the desired direction. When the user releases the directional controller the device is brought to a controlled stop.
The chair frame is a welded nut and steel construction and includes two rear wheels with drive units (motor, gear and brake) connected by belts to the front all directional wheels. Adjustments can be made to the two arm supports, foot support, and seat height and depth to position the user correctly in the device. The device supports a maximum weight of 220 lb, and has an approximate driving range of 12 miles. The device can be operated on carpet, tile, wood, vinyl, concrete, blacktop, dirt, gravel, grass, and wet (<1" standing water) surfaces.
#### 5.2 Indications for Use
The intended use of the Model M powered wheelchair is to provide outdoor and indoor mobility to persons limited to a seated position that are capable of operating a powered wheelchair.
#### 5.3 Summary of Technological Characteristics Comparison
Table 5-1 shows the similarities and differences between the two products. The key difference is that the subject device has the front all-directional wheel and the device controls on two arms instead of one. While these characteristics are different, the testing demonstrates that they do not raise new questions of safety or effectiveness.
| Item | Predicate | Subject Device |
|---------------|-------------|----------------|
| General | | |
| Manufacturer | Permobil AB | WHILL, Inc. |
| Model | F3 Corpus | Model M |
| 510(k) Number | K143180 | unknown |
Table 5-1 Summary of Technological Characteristics Comparison
{5}------------------------------------------------
| Item | Predicate | Subject Device | | |
|----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Intended Use | To provide indoor and outdoor<br>mobility to persons restricted<br>to a sitting position that are<br>capable of operating a powered<br>wheelchair. | Same | | |
| Indications for<br>Use | The intended use of the F3<br>powered wheelchair is to<br>provide indoor and outdoor<br>mobility to persons limited to a<br>seating position that are<br>capable of operating a powered<br>wheelchair. | The intended use of the Model<br>M powered wheelchair is to<br>provide outdoor and indoor<br>mobility to persons limited to<br>a seated position that are<br>capable of operating a<br>powered wheelchair. | | |
| Rx/OTC<br>Designation | Rx or OTC | Rx | | |
| Physical Characteristics | | | | |
| Device Width | 24" | 23.6" | | |
| Device Length | Drive base only:<br>40" with anti-tippers<br>36" without anti-tippers | 37"- 42" (depending on foot<br>plate setting) | | |
| Weight (including<br>batteries) | 386 lb with Group 34 batteries<br>405 1b with Group 24 batteries | 240 lb | | |
| Device<br>Construction | Steel | Welded nut and steel | | |
| Batteries (2) | 12VDC 60Ah (Group M34) or<br>12VDC 73Ah (Group M24) | 12V 50Ah | | |
| Operating Characteristics | | | | |
| Maximum Weight<br>Capacity | 300 lb (136 kg) | 220lb (100kg) | | |
| Maximum Speed | 6 mph<br>5 mph (without anti-tippers) | 5.5 mph | | |
| Turning Radius | 26.5" | 28" | | |
| Obstacle Climbing<br>Height | 3" | 3" | | |
| Drive System | Front wheel drive | 4 Wheel Drive | | |
| Dynamic Stability<br>(incline) | 6 degrees | 10 degrees | | |
| Item | Predicate | Subject Device | | |
| Driving Range<br>(full battery<br>charge) | 13 miles | 12 miles | | |
| Battery Charging<br>Time | 8 hours | ~8 hours to 80% | | |
| Design Features | | | | |
| Power Controller | R-net 120 amp controller | R-net 120 amp controller | | |
| Speed Settings | 5 (maximum) | 3 | | |
| Front Wheel Type | Pneumatic or foam-filled | All directional wheel | | |
| Rear Wheel Type | Solid polyurethane | Pneumatic or solid | | |
| Anti-tip Wheels | Front anti-tip wheels (optional) | Rear anti-tip wheels | | |
| Pressure relief<br>handles | None | Yes | | |
| Seat slide | No | Yes | | |
| Tail lamps (2) | Reflectors<br>Lights (optional) | Red LED lights (always on) | | |
| Chair Tilt/<br>Recline/ Leg Lift/<br>Seat Elevator | Yes | No | | |
| Tie Downs | Yes | Yes (optional) | | |
| Non-Clinical Performance Testing | | | | |
| Wheel<br>Performance<br>Testing | Detailed performance testing<br>conducted to ISO 7176<br>standard | Detailed performance testing<br>conducted to RESNA WC-1,<br>WC-2 and WC-4 standards | | |
| Flammability | Padded parts: EN 1021-1, EN<br>1021-2 and ISO 7176-16<br>Plastic parts: UL94 | Upholstered parts: ISO 8191-<br>1 and ISO 8191-2 and covers:<br>ISO 7176-16<br>Plastic parts: UL94 | | |
{6}------------------------------------------------
#### Performance Data 5.4
In addition to design verification and validation testing WHILL conducted performance testing to demonstrate the subject device is in compliance with the following test standards:
{7}------------------------------------------------
- ANSI/RESNA WC-1:2009 Section 1: Determination of Static ● Stability
- ANSI/RESNA WC-2:2009 Section 2: Determination of Dynamic ● Stability
- ANSI/RESNA WC-2:2009 Section 3: Determination of Effectiveness . of Brakes
- ANSI/RESNA WC-2:2009 Section 4: Determination of Static . Stability Determination of Dynamic Stability Determination of Effectiveness of Brakes Energy Consumption for Determination of Theoretical Distance
- ANSI/RESNA WC-1:2009 Section 5: Maneuvering Space ● Determination of Dimensions, Mass and Maneuvering Space
- ANSI/RESNA WC-2:2009 Section 6: Determination of Maximum . Speed, Acceleration and Deceleration
- ANSI/RESNA WC-1:2009 Section 7: Method of Measurement of . Seating and Wheel Dimensions
- ANSI/RESNA WC-1:2009 Section 8: Requirements and Test . Methods for Static, Impact and Fatigue Strengths
- ANSI/RESNA WC-2:2009 Section 9: Climatic Tests ●
- ANSI/RESNA WC-2:2009 Section 10: Determination of Obstacle ● Climbing
- . ANSI/RESNA WC-1:2009 Section 11: Test Dummies
- ANSI/RESNA WC-1:2009 Section 13: Determination Of Coefficient . Of Friction Of Test Surfaces
- ANSI/RESNA WC-2:2009 Section 14: Power And Control Systems . For Electrically Powered Wheelchairs - Requirements And Test Methods
- ANSI/RESNA WC-1:2009 Section 15: Requirements for Information . Disclosure, Documentation and Labeling
- ANSI/RESNA WC-1:2009 Section 16: Resistance to ignition of . upholstered parts - Requirements and test methods (tests to ISO 8191-1 and ISO 8191-2)
- ANSI/RESNA WC-4:2012 Section 19: Wheelchairs used as seats in . motor vehicles
- ANSI/RESNA WC-2:2009 Section 21: Requirements and Test . Methods for Electromagnetic Compatibility of Electrically Powered Wheelchairs and Motorized Scooters
{8}------------------------------------------------
- ASTM D4169, Standard Practice for Performance Testing of ● Shipping Containers and Systems
- ISO 7176-16: 2012 Wheelchairs -- Part 16: Resistance to ignition of ● postural support devices
- . ISO 8191-1: 1987 Furniture -- Assessment of the ignitability of upholstered furniture -- Part 1: Ignition source: smouldering cigarette
- ISO 8191-2: 1988 Furniture -- Assessment of ignitability of . upholstered furniture -- Part 2: Ignition source: match-flame equivalent
- . UL 94 Standard for Tests for Flammability of Plastic Materials for Parts in Devices and Appliances
- ISO 10993-1 Biological evaluation of medical devices -- Part 1: . Evaluation and testing within a risk management process
- . ISO 10993-5 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity
- . ISO 10993-10 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization
- ISO 7176-14: 2008 Wheelchairs -- Part 14: Power and control . systems for electrically powered wheelchairs and scooters --Requirements and test methods (for R-Net Power Module)
Based upon the results of this testing, it was determined the Model M performance was substantially equivalent to the predicate device.
#### ર. ર Substantial Equivalence Conclusion
The July 28, 2014 FDA Guidance entitled "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]'' was used to determine substantial equivalence. The analysis shows that the Intended Use and Indications for Use, principles of operation, and conditions of use are identical, and that differences in technical characteristics do not raise different questions of safety and effectiveness. There are technological differences between the subject and predicate device however, the results of performance testing demonstrate that these differences do not raise any new questions of safety or effectiveness compared to other similar powered wheelchairs currently on the market. Therefore one can conclude that the Model M is substantially equivalent to the predicate Permobil F3 Corpus device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.