Safe Secure Sterilization Pouch with Steam and Ethylene Oxide Process Indicators
K153540 · Safe Secure Packaging Co., Ltd. · FRG · Aug 4, 2016 · General Hospital
Device Facts
Record ID
K153540
Device Name
Safe Secure Sterilization Pouch with Steam and Ethylene Oxide Process Indicators
Applicant
Safe Secure Packaging Co., Ltd.
Product Code
FRG · General Hospital
Decision Date
Aug 4, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
Safe Secure Paper Sterilization Pouches and Rolls are intended to be used to enclose another medical device, in a single or double pouch configuration, that is to be sterilized by a health care provider using: - Gravity steam at 121°C (250°F) for 30 minutes; 25 minutes dry time . - Pre-vacuum steam at 132°C (270°F) for 4 minutes; 20 minutes dry time . - . Pre-vacuum steam at 134°C (273°F) for 3 minutes; 20 minutes dry time - . Pre-vacuum steam at 135°C (275°F) for 3 minutes; 16 minutes dry time - . Ethylene Oxide (EO) with a concentration of 735 mg/L at 55°C (131°F) and 50% relative humidity for 60 minutes. Aeration time of 8 hours at 60℃ (140°F). The device is not intended and has not been validated for sterilization of devices that contain lumens. The external chemical indicators on the pouches/rolls are intended to demonstrate that the device has been exposed to the steam or EO sterilization process and to distinguish between processed devices. The chemical indicators change from green to purple after exposure to steam and from yellow to brown after exposure to ethylene oxide. The Tyvek version of the pouch is for EO sterilization only. If stored according to the recommended conditions, the products before sterilization have a maximum shelf life of 5 years from the date of manufacture. The pouches are intended to allow sterilization of the enclosed medical device(s) and also to maintain sterility (SAL=10 4). The subject device is intended and has been validated to maintain sterlity of the enclosed devices for 6 months after steam sterilization and 24 months after EO sterilization. The maximum validated pouch load is 2.64 pounds (1.2kg).
Device Story
Sterilization pouches/rolls constructed of medical-grade paper or Tyvek thermally sealed to plastic film; used by healthcare providers to enclose non-lumened medical devices for sterilization. Pouches feature external chemical indicators (steam: green-to-purple; EO: yellow-to-brown) to verify exposure to sterilization cycles. Subject device supports gravity steam, pre-vacuum steam, and EO sterilization. Device maintains sterility (SAL 10^-6) for 6 months post-steam or 24 months post-EO. Maximum load 1.2kg. Output is a sterile-packaged device; indicators provide visual confirmation of processing to assist clinical staff in distinguishing processed from unprocessed items.
Clinical Evidence
Bench testing only. Performance validated via visual inspection, seal integrity, dimensional analysis, seal leak, seal strength, sterilization indicator testing, biological indicator testing, accelerated aging/shelf-life, and full/half-cycle sterilization validation using worst-case loads. Biocompatibility testing included cytotoxicity, sensitization, and irritation. No clinical data.
Technological Characteristics
Materials: Porous paper or Tyvek, PET/CPP high-temperature blue compound film, high-temperature adhesive paper, medical double-sided tape. Indicators: Steam (green-to-purple), EO (yellow-to-brown). Dimensions: 2-20 inches length, 8-30 inches width. Sterilization: Gravity steam, pre-vacuum steam, EO. Testing per ASTM/ISO standards.
Indications for Use
Indicated for healthcare providers to enclose non-lumened medical devices for sterilization via gravity steam, pre-vacuum steam, or ethylene oxide (EO) gas. Not for devices with lumens.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Safe Secure Sterilization Pouch with Steam and Ethylene Oxide Process Indicators (K112591)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a flowing design that suggests movement or progress.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 4, 2016
Safe Secure Packaging Co., Ltd % Mr. Brian Edwards Senior Medical Research Manager, Regulatory NAMSA 4050 Olson Memorial Highway Golden Valley, MN 55422
Re: K153540
Trade/Device Name: Safe Secure Sterilization Pouch with Steam and Ethylene Oxide Process Indicators Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG Dated: July 11, 2016 Received: July 12, 2016
Dear Mr. Edwards:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runne, DDS, MA
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153540
### Device Name
Safe Secure Sterilization Pouch with Steam and Ethylene Oxide Process Indicators
### Indications for Use (Describe)
Safe Secure Paper Sterilization Pouches and Rolls are intended to be used to enclose another medical device, in a single or double pouch configuration, that is to be sterilized by a health care provider using:
- Gravity steam at 121°C (250°F) for 30 minutes; 25 minutes dry time .
- Pre-vacuum steam at 132°C (270°F) for 4 minutes; 20 minutes dry time .
- . Pre-vacuum steam at 134°C (273°F) for 3 minutes; 20 minutes dry time
- . Pre-vacuum steam at 135°C (275°F) for 3 minutes; 16 minutes dry time
- . Ethylene Oxide (EO) with a concentration of 735 mg/L at 55°C (131°F) and 50% relative humidity for 60 minutes. Aeration time of 8 hours at 60℃ (140°F).
The device is not intended and has not been validated for sterilization of devices that contain lumens.
The external chemical indicators on the pouches/rolls are intended to demonstrate that the device has been exposed to the steam or EO sterilization process and to distinguish between processed devices. The chemical indicators change from green to purple after exposure to steam and from yellow to brown after exposure to ethylene oxide.
The Tyvek version of the pouch is for EO sterilization only.
If stored according to the recommended conditions, the products before sterilization have a maximum shelf life of 5 years from the date of manufacture. The pouches are intended to allow sterilization of the enclosed medical device(s) and also to maintain sterility (SAL=10 4). The subject device is intended and has been validated to maintain sterlity of the enclosed devices for 6 months after steam sterilization and 24 months after EO sterilization.
The maximum validated pouch load is 2.64 pounds (1.2kg).
The following tables list the available model numbers of the Safe Secure Sterilization Pouches and Rolls.
| Model | Type | Sterilization Cycle(s) | Dimensions(mm) | |
|-------------|-----------------------------------|-------------------------|----------------|-------------|
| | | | Width (mm) | Length (mm) |
| ABHSP100001 | POUCH PAPER HEAT SEAL 3.5X10 | Steam or Ethylene Oxide | 89 | 254 |
| ABHSP100002 | POUCH PAPER HEAT SEAL 7.5X13 | Steam or Ethylene Oxide | 191 | 330 |
| ABHSP100003 | POUCH PAPER HEAT SEAL 6X10 | Steam or Ethylene Oxide | 152 | 254 |
| ABHSP100004 | POUCH PAPER HEAT SEAL 4.5X11 | Steam or Ethylene Oxide | 114 | 279 |
| ABHSP100005 | POUCH PAPER HEAT SEAL 5X15 | Steam or Ethylene Oxide | 127 | 381 |
| ABHSP100006 | POUCH PAPER HEAT SEAL 10X15 | Steam or Ethylene Oxide | 254 | 381 |
| ABHSP100007 | POUCH PAPER HEAT SEAL 18X22 | Steam or Ethylene Oxide | 457 | 559 |
| ABHSP100008 | POUCH PAPER HEAT SEAL 12X18 | Steam or Ethylene Oxide | 305 | 457 |
| ABHSP100009 | POUCH PAPER HEAT SEAL 4X8 | Steam or Ethylene Oxide | 102 | 203 |
| ABHSP100010 | POUCH PAPER HEAT SEAL 3X8 | Steam or Ethylene Oxide | 76 | 203 |
| ABHSP100011 | POUCH PAPER HEAT SEAL 16X16 | Steam or Ethylene Oxide | 406 | 406 |
| ABHSP100012 | POUCH PAPER HEAT SEAL 4X22 | Steam or Ethylene Oxide | 102 | 559 |
| ABHSP100013 | POUCH PAPER HEAT SEAL 13X15 | Steam or Ethylene Oxide | 330 | 381 |
| ABHSP100014 | POUCH PAPER HEAT SEAL 8X10 | Steam or Ethylene Oxide | 203 | 254 |
| ABHSP100015 | POUCH PAPER HEAT SEAL 8X16 | Steam or Ethylene Oxide | 203 | 406 |
| ABHSP110001 | TUBING PAPER 6IN WIDTH 100FT ROLL | Steam or Ethylene Oxide | 151 | 30500 |
Table 1: Safe Secure Medical Grade Paper Sterilization Pouch/Roll Products
{3}------------------------------------------------
| Model | Type | Sterilization Cycle(s) | Dimensions(mm) | |
|-------------|-----------------------------------------|-------------------------|----------------|-------------|
| | | | Width (mm) | Length (mm) |
| ABHSP110002 | TUBING PAPER 9IN WIDTH 100FT ROLL | Steam or Ethylene Oxide | 228 | 30500 |
| ABHSP110003 | TUBING PAPER 4IN WIDTH 100FT ROLL | Steam or Ethylene Oxide | 101 | 30500 |
| ABHSP110004 | TUBING PAPER 3IN WIDTH 100FT ROLL | Steam or Ethylene Oxide | 75 | 30500 |
| ABHSP110005 | TUBING PAPER 12IN WIDTH 100FT<br>ROLL | Steam or Ethylene Oxide | 305 | 30500 |
| ABHSP110006 | TUBING PAPER 2IN WIDTH 100FT ROLL | Steam or Ethylene Oxide | 51 | 30500 |
| ABHSP120001 | POUCH MULTIPLE INNER 8 POUCHES<br>SHEET | Steam or Ethylene Oxide | 230 | 415 |
| ABHSP120002 | POUCH MULTIPLE INNER 4 POUCHES<br>SHEET | Steam or Ethylene Oxide | 230 | 415 |
| ABSSP100001 | POUCH PAPER SELF SEAL 7.5X13 | Steam or Ethylene Oxide | 191 | 353 |
| ABSSP100002 | POUCH PAPER SELF SEAL 5.25X10 | Steam or Ethylene Oxide | 140 | 277 |
| ABSSP100003 | POUCH PAPER SELF SEAL 13X18 | Steam or Ethylene Oxide | 330 | 480 |
| ABSSP100004 | POUCH PAPER SELFSEAL 12X15 | Steam or Ethylene Oxide | 305 | 404 |
| ABSSP100005 | POUCH PAPER SELF SEAL 3.5X8.75 | Steam or Ethylene Oxide | 89 | 245 |
| ABSSP100006 | POUCH PAPER SELF SEAL 5X15 | Steam or Ethylene Oxide | 127 | 404 |
| ABSSP100007 | POUCH PAPER SELF SEAL 3.5X9.875 | Steam or Ethylene Oxide | 89 | 274 |
| ABSSP100008 | POUCH PAPER SELF SEAL 8X16 | Steam or Ethylene Oxide | 203 | 429 |
| ABSSP100009 | POUCH PAPER SELF SEAL 3.5X22 | Steam or Ethylene Oxide | 89 | 582 |
| ABSSP100010 | POUCH PAPER SELF SEAL 9.125X25.5 | Steam or Ethylene Oxide | 232 | 671 |
| ABSSP100011 | POUCH PAPER SELF SEAL 4X11 | Steam or Ethylene Oxide | 102 | 302 |
### Table 2: Safe Secure Tyvek Sterilization Pouch/Roll Products
| Model | Type | Sterilization Cycle(s) | Dimensions(mm) | |
|-------------|----------------------------------|------------------------|----------------|-------------|
| | | | Width (mm) | Length (mm) |
| ABSST100001 | POUCH TYVEK SELF SEAL 8X12 | Ethylene Oxide Only | 203 | 345 |
| ABSST100002 | POUCH TYVEK SELF SEAL 6X10 | Ethylene Oxide Only | 152 | 294 |
| ABSST100003 | POUCH TYVEK SELF SEAL 12X18 | Ethylene Oxide Only | 305 | 497 |
| ABSST100004 | POUCH TYVEK SELF SEAL 10X15 | Ethylene Oxide Only | 254 | 421 |
| ABSST100005 | POUCH TYVEK SELF SEAL 4X9 | Ethylene Oxide Only | 102 | 269 |
| ABSST100006 | POUCH TYVEK SELF SEAL 3X7 | Ethylene Oxide Only | 76 | 218 |
| ABSST100007 | POUCH TYVEK SELF SEAL 4X12 | Ethylene Oxide Only | 102 | 345 |
| ABSST100008 | POUCH TYVEK SELF SEAL 4X22 | Ethylene Oxide Only | 102 | 599 |
| ABHST100001 | POUCH TYVEK HEAT SEAL 8X12 | Ethylene Oxide Only | 203 | 305 |
| ABHST100002 | POUCH TYVEK HEAT SEAL 6X10 | Ethylene Oxide Only | 152 | 254 |
| ABHST100003 | POUCH TYVEK HEAT SEAL 10X15 | Ethylene Oxide Only | 254 | 381 |
| ABHST100004 | POUCH TYVEK HEAT SEAL 12X18 | Ethylene Oxide Only | 305 | 457 |
| ABHST100005 | POUCH TYVEK HEAT SEAL 4X12 | Ethylene Oxide Only | 102 | 305 |
| ABHST100006 | POUCH TYVEK HEAT SEAL 4X9 | Ethylene Oxide Only | 102 | 229 |
| ABHST100007 | POUCH TYVEK HEAT SEAL 4X22 | Ethylene Oxide Only | 102 | 559 |
| ABHST100008 | POUCH TYVEK HEAT SEAL 3X7 | Ethylene Oxide Only | 76 | 178 |
| ABHST110001 | TUBING TYVEK 9IN WIDTH 75FT ROLL | Ethylene Oxide Only | 228 | 22900 |
| ABHST110002 | TUBING TYVEK 6IN WIDTH 75FT ROLL | Ethylene Oxide Only | 151 | 22900 |
| ABHST110003 | TUBING TYVEK 4IN WIDTH 75FT ROLL | Ethylene Oxide Only | 101 | 22900 |
| ABHST110004 | TUBING TYVEK 3IN WIDTH 75FT ROLL | Ethylene Oxide Only | 75 | 22900 |
{4}------------------------------------------------
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 3 of 3
PSC Publishing Services (301) 443-6740 EF
{5}------------------------------------------------
#### 510(k) Summary 5.0
#### Submission Sponsor 5.1
Safe Secure Packaging (Shenzhen) Co., Ltd No.4 Chengguang Industrial Park Guanlan Street Baoan District, Shenzhen Guangdon Province, 518110 China Tel: 86-755-27040498 Fax: 86-755-27040488 Howard Jia, President howard.jia@anbaopack.com
#### 5.2 Submission Correspondent
NAMSA 4050 Olson Memorial Hwy Suite 450 Minneapolis, MN 55422
Brian J. Edwards Principal Medical Research Manager, Regulatory Phone: +1-763-390-8409 Fax: +1-763-287-3836 bedwards@namsa.com
#### 5.3 Date Prepared
August 1, 2016
#### 5.4 Device Identification
Trade/Proprietary Name: Safe Secure Sterilization Pouches with Steam and Ethylene Oxide Process Indicators Common/Usual Name: Sterilization Pouch Classification Name: Sterilization Wrap Classification Regulation: 21 CFR Part 880.6850 Product Code: FRG Device Class: II Classification Panel: General Hospital Model Numbers: ABHSP, ABSSP, ABHST, ABSST
#### ર્સ્ડ Predicate Devices
K112591 - Safe Secure Sterilization Pouch with Steam and Ethylene Oxide Process Indicators
{6}------------------------------------------------
#### 5.6 Device Description
The Safe Secure Sterilization Pouches are made from either a medical grade porous paper or Tyvek thermally sealed to a plastic film on the left, right, and bottom of the pouch. The top side is open in order to receive the medical device to be sterilized and to seal the pouch. They are available either pre-manufactured self-seal or heat-seal sizes or in a roll of heat-seal tubing for the customer to cut the pouches to their required size.
The base model numbers for the subject device are described as follows, where the last three letters corresponds to the type of material:
- ABSST - Self seal Tyvek
- ABHST Heat seal Tyvek .
- ABHSP Heat seal paper ●
- ABSSP Self seal paper .
The pouches also contain external chemical indicators used to indicate that the pouches have been processed by either a steam or ethylene oxide sterilization process.
The medical grade paper versions can be used in either gravity steam, pre-vacuum steam, or ethylene oxide (EO) sterilization cycles. The Tyvek version is only for EO sterilization.
#### 5.7 Indication for Use
Safe Secure Paper Sterilization Pouches and Rolls are intended to be used to enclose another medical device, in a single or double pouch configuration, that is to be sterilized by a health care provider using:
- Gravity Steam at 121°C (250°F) for 30 minutes: 25 minutes dry time .
- . Pre-vacuum Steam at 132°C (270°F) for 4 minutes; 20 minutes dry time
- . Pre-vacuum Steam at 134°C (273°F) for 3 minutes; 20 minutes dry time
- . Pre-vacuum steam at 135°C (275°F) for 3 minutes; 16 minutes dry time
- . Ethylene Oxide (EO) with a concentration of 735 mg/L at 55°C (131°F) and 50% to 80% relative humidity for 60 minutes. Aeration time of 8 hours at 60°C (140°F).
The device is not intended and has not been validated for sterilization of devices that contain lumens.
The external chemical indicators on the pouches/rolls are intended to demonstrate that the device has been exposed to the Steam or EO sterilization process and to distinguish between processed and unprocessed devices. The chemical indicators change from green to purple after exposure to steam and from yellow to brown after exposure to ethylene oxide.
{7}------------------------------------------------
The Tyvek version of the pouch is for EO sterilization only.
If stored according to the recommended conditions, the products before sterilization have a maximum shelf life of 5 years from the date of manufacture. The pouches are intended to allow sterilization of the enclosed medical device(s) and also to maintain sterility (SAL=10 %). The subject device is intended and has been validated to maintain sterility of the enclosed devices for 6 months after steam sterilization and 24 months after EO sterilization.
The maximum validated pouch load is 2.64 pounds (1.2kg).
These products are to be used for medical products without lumens.
The follow table list the available model numbers of the Safe Secure Sterilization Pouches and Rolls.
| | | | Dimensions (mm) | |
|-------------|---------------------------------------|-------------------------|-----------------|----------------|
| Model | Type | Sterilization Cycle(s) | Width<br>(mm) | Length<br>(mm) |
| ABHSP100001 | POUCH PAPER HEAT SEAL 3.5X10 | Steam or Ethylene Oxide | 89 | 254 |
| ABHSP100002 | POUCH PAPER HEAT SEAL 7.5X13 | Steam or Ethylene Oxide | 191 | 330 |
| ABHSP100003 | POUCH PAPER HEAT SEAL 6X10 | Steam or Ethylene Oxide | 152 | 254 |
| ABHSP100004 | POUCH PAPER HEAT SEAL 4.5X11 | Steam or Ethylene Oxide | 114 | 279 |
| ABHSP100005 | POUCH PAPER HEAT SEAL 5X15 | Steam or Ethylene Oxide | 127 | 381 |
| ABHSP100006 | POUCH PAPER HEAT SEAL 10X15 | Steam or Ethylene Oxide | 254 | 381 |
| ABHSP100007 | POUCH PAPER HEAT SEAL 18X22 | Steam or Ethylene Oxide | 457 | 559 |
| ABHSP100008 | POUCH PAPER HEAT SEAL 12X18 | Steam or Ethylene Oxide | 305 | 457 |
| ABHSP100009 | POUCH PAPER HEAT SEAL 4X8 | Steam or Ethylene Oxide | 102 | 203 |
| ABHSP100010 | POUCH PAPER HEAT SEAL 3X8 | Steam or Ethylene Oxide | 76 | 203 |
| ABHSP100011 | POUCH PAPER HEAT SEAL 16X16 | Steam or Ethylene Oxide | 406 | 406 |
| ABHSP100012 | POUCH PAPER HEAT SEAL 4X22 | Steam or Ethylene Oxide | 102 | 559 |
| ABHSP100013 | POUCH PAPER HEAT SEAL 13X15 | Steam or Ethylene Oxide | 330 | 381 |
| ABHSP100014 | POUCH PAPER HEAT SEAL 8X10 | Steam or Ethylene Oxide | 203 | 254 |
| ABHSP100015 | POUCH PAPER HEAT SEAL 8X16 | Steam or Ethylene Oxide | 203 | 406 |
| ABHSP110001 | TUBING PAPER 6IN WIDTH 100FT ROLL | Steam or Ethylene Oxide | 151 | 30500 |
| ABHSP110002 | TUBING PAPER 9IN WIDTH 100FT ROLL | Steam or Ethylene Oxide | 228 | 30500 |
| ABHSP110003 | TUBING PAPER 4IN WIDTH 100FT ROLL | Steam or Ethylene Oxide | 101 | 30500 |
| ABHSP110004 | TUBING PAPER 3IN WIDTH 100FT ROLL | Steam or Ethylene Oxide | 75 | 30500 |
| ABHSP110005 | TUBING PAPER 12IN WIDTH 100FT<br>ROLL | Steam or Ethylene Oxide | 305 | 30500 |
Table 1: Safe Secure Medical Grade Paper Sterilization Pouch/Roll Products
{8}------------------------------------------------
# Safe Secure Packaging Co. Ltd.
Sterilization Pouch
| Model | Type | Sterilization Cycle(s) | Dimensions (mm) | |
|-------------|-----------------------------------------|-------------------------|-----------------|-------------|
| | | | Width (mm) | Length (mm) |
| ABHSP110006 | TUBING PAPER 2IN WIDTH 100FT ROLL | Steam or Ethylene Oxide | 51 | 30500 |
| ABHSP120001 | POUCH MULTIPLE INNER 8 POUCHES<br>SHEET | Steam or Ethylene Oxide | 230 | 415 |
| ABHSP120001 | POUCH MULTIPLE INNER 4 POUCHES<br>SHEET | Steam or Ethylene Oxide | 230 | 415 |
| ABSSP100001 | POUCH PAPER SELF SEAL 7.5X13 | Steam or Ethylene Oxide | 191 | 353 |
| ABSSP100002 | POUCH PAPER SELF SEAL 5.25X10 | Steam or Ethylene Oxide | 140 | 277 |
| ABSSP100003 | POUCH PAPER SELF SEAL 13X18 | Steam or Ethylene Oxide | 330 | 480 |
| ABSSP100004 | POUCH PAPER SELFSEAL 12X15 | Steam or Ethylene Oxide | 305 | 404 |
| ABSSP100005 | POUCH PAPER SELF SEAL 3.5X8.75 | Steam or Ethylene Oxide | 89 | 245 |
| ABSSP100006 | POUCH PAPER SELF SEAL 5X15 | Steam or Ethylene Oxide | 127 | 404 |
| ABSSP100007 | POUCH PAPER SELF SEAL 3.5X9.875 | Steam or Ethylene Oxide | 89 | 274 |
| ABSSP100008 | POUCH PAPER SELF SEAL 8X16 | Steam or Ethylene Oxide | 203 | 429 |
| ABSSP100009 | POUCH PAPER SELF SEAL 3.5X22 | Steam or Ethylene Oxide | 89 | 582 |
| ABSSP100010 | POUCH PAPER SELF SEAL 9.125X25.5 | Steam or Ethylene Oxide | 232 | 671 |
| ABSSP100011 | POUCH PAPER SELF SEAL 4X11 | Steam or Ethylene Oxide | 102 | 302 |
### Table 2: Safe Secure Tyvek Sterilization Pouch/Roll Products
| Model | Type | Sterilization Cycle(s) | Dimensions (mm) | |
|-------------|----------------------------------|------------------------|-----------------|-------------|
| | | | Width (mm) | Length (mm) |
| ABSST100001 | POUCH TYVEK SELF SEAL 8X12 | Ethylene Oxide Only | 203 | 345 |
| ABSST100002 | POUCH TYVEK SELF SEAL 6X10 | Ethylene Oxide Only | 152 | 294 |
| ABSST100003 | POUCH TYVEK SELF SEAL 12X18 | Ethylene Oxide Only | 305 | 497 |
| ABSST100004 | POUCH TYVEK SELF SEAL 10X15 | Ethylene Oxide Only | 254 | 421 |
| ABSST100005 | POUCH TYVEK SELF SEAL 4X9 | Ethylene Oxide Only | 102 | 269 |
| ABSST100006 | POUCH TYVEK SELF SEAL 3X7 | Ethylene Oxide Only | 76 | 218 |
| ABSST100007 | POUCH TYVEK SELF SEAL 4X12 | Ethylene Oxide Only | 102 | 345 |
| ABSST100008 | POUCH TYVEK SELF SEAL 4X22 | Ethylene Oxide Only | 102 | 599 |
| ABHST100001 | POUCH TYVEK HEAT SEAL 8X12 | Ethylene Oxide Only | 203 | 305 |
| ABHST100002 | POUCH TYVEK HEAT SEAL 6X10 | Ethylene Oxide Only | 152 | 254 |
| ABHST100003 | POUCH TYVEK HEAT SEAL 10X15 | Ethylene Oxide Only | 254 | 381 |
| ABHST100004 | POUCH TYVEK HEAT SEAL 12X18 | Ethylene Oxide Only | 305 | 457 |
| ABHST100005 | POUCH TYVEK HEAT SEAL 4X12 | Ethylene Oxide Only | 102 | 305 |
| ABHST100006 | POUCH TYVEK HEAT SEAL 4X9 | Ethylene Oxide Only | 102 | 229 |
| ABHST100007 | POUCH TYVEK HEAT SEAL 4X22 | Ethylene Oxide Only | 102 | 559 |
| ABHST100008 | POUCH TYVEK HEAT SEAL 3X7 | Ethylene Oxide Only | 76 | 178 |
| ABHST110001 | TUBING TYVEK 9IN WIDTH 75FT ROLL | Ethylene Oxide Only | 228 | 22900 |
{9}------------------------------------------------
## Safe Secure Packaging Co. Ltd. Sterilization Pouch
| Model | Type | Sterilization Cycle(s) | Dimensions (mm) | |
|-------------|----------------------------------|------------------------|-----------------|-------------|
| | | | Width (mm) | Length (mm) |
| ABHST110002 | TUBING TYVEK 6IN WIDTH 75FT ROLL | Ethylene Oxide Only | 151 | 22900 |
| ABHST110003 | TUBING TYVEK 4IN WIDTH 75FT ROLL | Ethylene Oxide Only | 101 | 22900 |
| ABHST110004 | TUBING TYVEK 3IN WIDTH 75FT ROLL | Ethylene Oxide Only | 75 | 22900 |
#### Comparison to Predicates 5.8
The Safe Secure Sterilization Pouches are substantially equivalent to the previous pouch version cleared under K112591 – Safe Secure Sterilization Pouch with Steam and Ethylene Oxide Process Indicators.
## Safe Secure Sterilization Pouch Substantial Equivalence Table
| Item<br># | | Safe Secure Sterilization Pouch<br>(Predicate Device) | Safe Secure Sterilization Pouch<br>(Subject Device) | |
|-----------|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| 1 | 510(k) Number | K112591 | TBD | |
| 2 | Manufacturer | Safe Secure Packaging<br>(Shenzhen) Co., Ltd<br>No.4 Chengguang Industrial<br>Park,Guanlan Street, Baoan<br>District,Shenzhen,518110 China | Safe Secure Packaging<br>(Shenzhen) Co., Ltd<br>No.4 Chengguang Industrial<br>Park, Guanlan Street, Baoan<br>District,Shenzhen,518110 China | Same |
| 3 | Model<br>Numbers | ABHSP, ABSSP | ABHSP, ABSSP, ABHST,<br>ABSST | Substantial<br>Equivalent |
| 4 | Classification | II | II | Same |
| 5 | Product Code | FRG | FRG | Same |
| 6 | Regulation | 21 CFR 880.6850 | 21 CFR 880.6850 | Same |
| Item | | Safe Secure Sterilization Pouch<br>(Predicate Device) | Safe Secure Sterilization Pouch<br>(Subject Device) | |
| # | | | | |
| 7 | Indications for<br>Use | Safe Secure Sterilization Pouch<br>with Steam and Ethylene Oxide<br>Process Indicators (Heat-seal<br>and Self-seal) is intended to be<br>used to enclose another medical<br>device that is to be sterilized by a<br>health provider by gravity steam<br>and ethylene oxide (EO). The<br>recommended steam<br>sterilization cycle parameters is<br>30 minutes at 121°C (250°F).<br>The recommended EO gas<br>sterilization cycle is greater<br>than or equal to 735mg/L of<br>ethylene oxide (EO) for 1 hour at<br>55°C (130 °F)) and 50% to<br>80%RH. The pouch's external<br>chemical ink indicators on the<br>pouches are intended to<br>demonstrate that the device has<br>been expose to the steam<br>or EO sterilization process and<br>to distinguish between processed<br>and unprocessed devices. The<br>pouch is intended to allow<br>sterilization of the enclosed<br>medical device and also to<br>maintain sterility (SAL= 10-6) of<br>the enclosed device until used. | Safe Secure Paper Sterilization<br>Pouches and Rolls are intended<br>to be used to enclose another<br>medical device, in a single or<br>double pouch configuration, that<br>is to be sterilized by a health<br>care provider using:<br>• Gravity Steam at 121°C<br>(250°F) for 30 minutes<br>• Pre-vacuum Steam at 132°C<br>(270°F) for 4 minutes<br>• Pre-vacuum Steam at 134°C<br>(273°F) for 3 minutes<br>• Pre-vacuum steam at 135°C<br>(275°F) for 3 minutes<br>• Ethylene Oxide (EO) with a<br>concentration of 735 mg/L at<br>55°C (131°F) and 50% to 80%<br>relative humidity for 60<br>minutes. Aeration time<br>of 8 hours at 60°C<br>(140°F).<br>The external chemical indicators<br>on the pouches/rolls are intended<br>to demonstrate that the device<br>has been exposed to the Steam<br>or EO sterilization process and<br>to distinguish between processed<br>and unprocessed devices.<br>The Tyvek version of the pouch<br>is for EO sterilization only.<br>If stored according to the<br>recommended conditions, the<br>products before sterilization<br>have a maximum shelf life of 5<br>years from the date of<br>manufacture. The pouches are<br>intended to allow sterilization of<br>the enclosed medical device(s)<br>and also to maintain sterility<br>(SAL=10-6).<br>The maximum validated pouch<br>load is 2.64 pounds (1.2kg).<br>These products are to be used for | Substantially<br>Equivalent |
| 8 | Material<br>Composition | Porous paper or Tyvek,<br>PET/CPP high-temperature<br>blue compound film, High<br>temperature adhesive paper,<br>Medical Double Side Tape, EO<br>gas indicator ink, Steam<br>indicator ink. | Porous paper or Tyvek,<br>PET/CPP high-temperature<br>blue compound film, High<br>temperature adhesive paper,<br>Medical Double Side Tape, EO<br>gas indicator ink, Steam<br>indicator ink. | Same |
| Item | | Safe Secure Sterilization Pouch<br>(Predicate Device) | Safe Secure Sterilization Pouch<br>(Subject Device) | |
| # | | | | |
| 9 | Pouch<br>Dimensions* | Length: 2 to 20 inches<br>Width: 8 to 30 inches | Length: 2 to 20 inches<br>Width: 8 to 30 inches | Same |
| 10 | Roll<br>Dimensions | Width: 8 to 30 inches | Width: 8 to 30 inches | Same |
| 11 | Air Permeance | The maximum equivalent pore<br>size diameter shall not exceed<br>50um. | The maximum equivalent pore<br>size diameter shall not exceed<br>50um. | Same |
| 12 | Shelf Life | 5 years before sterilization; 6<br>months after gravity steam<br>sterilization or 24 months after<br>EOsterilization | 5 years before sterilization; 6<br>months after steam sterilization<br>(gravity or pre-vacuum) or 24<br>months after EOsterilization | Same |
| 13 | Paper and Film<br>Thickness &<br>Density | Paper: ±2 gm/m²<br>Film: +0.002 mm<br>Tyvek: ±2 gm/m² | Paper: ±2 gm/m²<br>Film: +0.002 mm<br>Tyvek: ±2 gm/m² | Same |
| 14 | Seal Strength<br>before<br>sterilization | 2.5 to 5.5 N/15mm | 2.5 to 5.5 N/15mm | Same |
| 15 | Seal strength<br>after steam<br>sterilization | 2.5 to 5.5 N/15mm | > 1.5 N / 15mm | Substantially<br>Equivalent |
| 16 | Seal strength<br>after EO<br>sterilization | 2.5 to 5.5 N/15mm | > 1.5 N / 15mm | Substantially<br>Equivalent |
| 17 | Burst Strength | 2.5 to 5.5 N/15mm…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.