K153509 · Implant Direct Sybron Manufacturing, LLC · NHA · Aug 26, 2016 · Dental
Device Facts
Record ID
K153509
Device Name
GPS Angled Abutment
Applicant
Implant Direct Sybron Manufacturing, LLC
Product Code
NHA · Dental
Decision Date
Aug 26, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
GPS® Angled Abutments are designed to be used in support of a dental implant(s) to provide support for prosthetic restorations in a partially or completely edentulous patient. These abutments are designed to only receive a fabricated multi-unit bridge or overdenture. Abutments are intended for use in the mandible or maxilla. Implants can be indicated for immediate loading when good primary stability has been achieved and with appropriate occlusal loading.
Device Story
GPS® Angled Abutments are two-piece dental abutments designed to provide prosthetic support for multi-unit bridges or overdentures. The device consists of an angled base (15° or 30°) and a coronal GPS top that interfaces with snap-on cap attachments. The abutment is secured to a dental implant via a fixation screw. Used by clinicians in dental clinics, the device allows for prosthetic restoration in patients with dental implants. The output is a stable platform for overdentures or bridges, benefiting patients by restoring function and aesthetics in edentulous or partially edentulous cases. The device is provided non-sterile and requires steam sterilization by the user.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including static compression and fatigue testing per ISO 14801, biocompatibility testing per ISO 10993-5, and steam sterilization validation per ISO 17665-1.
Technological Characteristics
Two-piece dental abutment; materials: titanium alloy with TiN coating; angles: 15° and 30°; sterilization: steam (non-sterile provided); connectivity: mechanical interface with specific dental implant systems (Legacy, SwishTapered, SwishPlus, SwishActive, InterActive).
Indications for Use
Indicated for partially or completely edentulous patients requiring support for multi-unit bridge or overdenture prosthetic restorations in the mandible or maxilla. Compatible with specific dental implant systems. Immediate loading indicated if primary stability and occlusal loading requirements are met.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the symbol in a circular fashion. The logo is presented in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 26, 2016
Implant Direct Sybron Manufacturing, LLC Ms. Renee Bennett Regulatory Affairs Specialist 3050 East Hillcrest Drive Thousand Oaks, California 91362
Re: K153509
Trade/Device Name: GPS® Angled Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: July 27, 2016 Received: July 28, 2016
Dear Ms. Bennett:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang
-s
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K153509
Device Name GPS® Angled Abutments
### Indications for Use (Describe)
GPS® Angled Abutments are designed to be used in support of a dental implant(s) to provide support for prosthetic restorations in a partially or completely edentulous patient. These abutments are designed to only receive a fabricated multi-unit bridge or overdenture. Abutments are intended for use in the mandible or maxilla. Implants can be indicated for immediate loading when good primary stability has been achieved and with appropriate occlusal loading.
### Compatibility:
· Legacy System: Prosthetically compatible with Zimmer Dental Tapered Screw-Vent system implants (3.7mmD, 4.1mmD, 8mm-16mm Length), 4.5mm platform implants(4.7mmD, 8mm-16mm Length), and 5.7mm platform implants (6.0mmD, 8mm-16mm Length).
•SwishTapered System: Prosthetically compatible with Straumann Standard Plus system RN platform implants (3.3mmD-4.8mmD, 6mm-16mm Length) and WN platform implants(4.8mmD, 6mm-12mm Length). •SwishPlus System: Prosthetically compatible with Straumann Standard Plus system RN platform implants (3.3mmD-4.8mmD, 6mm-16mm Length) and WN platform implants (4.8mmD, 6mm-12mm Length). •SwishActive Implants: SwishActive implants are prosthetically compatible with InterActive 3.0 and 3.4mm abutments and Nobel Biocare conical connection NobelActive™ NP (Narrow Platform - 3.0mm diameter) and NobelActive™ RP (Regular Platform - 3.4mm diameter) titanium abutments. InterActive 3.0 and 3.4mm abutments are prosthetically compatible with Nobel Biocare connection NobelActive™ NP (Narrow Platform-3.0mm diameter) and NobelActive™ RP (Regular Platform - 3.4mm diameter) (3.5-5.0mmD. 8.5-18mmLength). • InterActive System: InterActive implants are prosthetically compatible with InterActive 3.0 and 3.4mm abutments and Nobel Biocare conical connection NobelActive™ NP (Narrow Platform - 3.0mm diameter) and NobelActive™ RP (Regular Platform - 3.4mm diameter) titanium abutments. InterActive 3.0 and 3.4mm abutments are prosthetically compatible with Nobel Biocare conical connection Nobel Active™ NP (Narrow Platform- 3.0mm diameter) and NobelActive™ RP (Regular Platform - 3.4mm diameter) (3.5-5.0mmD, 8.5-18mmLength) implants.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY for GPS® Angled Abutments
#### 1. Submitter Information:
2.
3.
| Company: | Implant Direct Sybron Manufacturing LLC |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 3050 East Hillcrest Drive, Thousand Oaks<br>CA 91362 USA |
| Contact Person:<br>Telephone Number: | Renee Bennett<br>818-444-3348 |
| Date Prepared: | August 16, 2016 |
| Device Name: | |
| ● Proprietary Name: | GPS® Angled Abutments |
| ● Classification Name: | Endosseous Dental Implant Abutment |
| ● CFR Number: | 872.3630 |
| ● Device Class: | II |
| ● Product Code: | NHA |
| Predicate Device: | Primary: GoDirect implants cleared under<br>K090234 - Spectra System Dental Implants<br>2008<br>Reference: Straight GPS abutments cleared<br>under K130572 - InterActive/SwishPlus2<br>Implant System<br>Reference: Screw Receiving abutment with<br>Ball top cleared under K143011- 2014<br>InterActive / SwishActive System |
{5}------------------------------------------------
#### Description of Device: 4.
GPS® Angled Abutments are designed to be used in conjunction with dental implant(s) to provide support for prosthetic restorations in a partially or completely edentulous patient. Abutments are intended for use in the mandible or maxilla. Implants can be indicated for immediate loading when good primary stability has been achieved and with appropriate occlusal loading.
The proposed GPS® Angled Abutments are intended to provide extra prosthetic options currently unavailable to the clinician. The proposed abutments are intended to complement the cleared systems listed in Table 2 below. The proposed abutments have the same compatible implant interface specific to each implant system and platform size.
| # | System Name | 510(k) | Proposed<br>Compatible GPS®<br>Abutments |
|---|------------------------------------------|---------|-----------------------------------------------------|
| 1 | InterActive SwishPlus2<br>Implant System | K130572 | InterActive®<br>SwishActive GPS<br>Angled Abutments |
| 2 | Legacy Abutment System | K060063 | Legacy GPS® Angled<br>Abutments |
| 3 | Spectra System<br>(RePlant) | K061319 | Replant GPS® Angled<br>Abutments |
| 4 | Swissplant Dental Implant<br>System | K081396 | Swish GPS® Angled<br>Abutments |
TABLE 1: List of cleared Implant Direct systems and the proposed compatible GPS® angled abutments
From Table 1, the worst case GPS® Angled abutment is the InterActive® /SwishActive GPS® Angled Abutment, 3.0 mm Platform. The 3.0 mm platform assembly represents the smallest size abutment from the InterActive system. Since the 3.4 mm implant and screw size is larger, the abutments are stronger than the 3.0 mm platform abutments. Hence, the 3.0 mm platform represents the worst case.
New fatigue testing has been completed on the worst case proposed GPS® Angled Abutment assembly..
The GPS® Angled abutments have the same coronal GPS Angled Top for all proposed GPS® Angled abutments. The GPS top has a superior outer radius at its coronal region that is identical to the primary predicate GoDirect implants (K090234 - Spectra System Dental Implants 2008). Since the proposed GPS® Angled abutments have the same coronal top design as the primary predicate GoDirect implants (K090234 - Spectra System Dental Implants 2008), the same compatibility will be achieved with the Zest
{6}------------------------------------------------
Locator® liner type attachments manufactured by Zest Anchor, Inc.
- 5. Indications for Use:
GPS® Angled Abutments are designed to be used in support of a dental implant(s) to provide support for prosthetic restorations in a partially or completely edentulous patient. These abutments are designed to only receive a fabricated multi-unit bridge or overdenture. Abutments are intended for use in the mandible or maxilla. Implants can be indicated for immediate loading when good primary stability has been achieved and with appropriate occlusal loading.
Compatibility:
·Legacy System: Prosthetically compatible with Zimmer Dental Tapered Screw-Vent system 3.5mm platform implants (3.7mmD, 4.1mmD, 8mm-16mm Length), 4.5mm platform implants(4.7mmD, 8mm-16mm Length), and 5.7mm platform implants (6.0mmD, 8mm-16mm Length).
·SwishTapered System: Prosthetically compatible with Straumann Standard and Standard Plus system RN platform implants (3.3mmD-4.8mmD, 6mm-16mm Length) and WN platform implants(4.8mmD, 6mm-12mm Length).
·SwishPlus System: Prosthetically compatible with Straumann Standard and Standard Plus system RN platform implants (3.3mmD-4.8mmD, 6mm-16mm Length) and WN platform implants (4.8mmD, 6mm-12mm Length).
·SwishActive Implants: SwishActive implants are prosthetically compatible with InterActive 3.0 and 3.4mm abutments and Nobel Biocare conical connection NobelActive™ NP (Narrow Platform - 3.0mm diameter) and NobelActive™ RP (Regular Platform - 3.4mm diameter) titanium abutments. InterActive 3.0 and 3.4mm abutments are prosthetically compatible with Nobel Biocare conical connection NobelActive™ NP (Narrow Platform- 3.0mm diameter) and NobelActive™ RP (Regular Platform - 3.4mm diameter) (3.5-5.0mmD. 8.5-18mmLength).
·InterActive System: InterActive implants are prosthetically compatible with InterActive 3.0 and 3.4mm abutments and Nobel Biocare conical connection NobelActive™ NP (Narrow Platform - 3.0mm diameter) and NobelActive™ RP (Regular Platform - 3.4mm diameter) titanium abutments. InterActive 3.0 and 3.4mm abutments are prosthetically compatible with Nobel Biocare conical connection NobelActive™ NP (Narrow Platform- 3.0mm diameter) and NobelActive™ RP (Regular Platform - 3.4mm diameter) (3.5-5.0mmD, 8.5-18mmLength) implants.
- 6. Description of Safety and Substantial Equivalence:
## Technological Characteristics
The proposed Implant Direct Sybron Manufacturing LLC GPS® Angled abutments have the same intended use and technological characteristics and hence are shown to be substantially equivalent to the primary predicate device GoDirect implants (cleared under K090234 - Spectra System Dental Implants 2008). The proposed GPS Angled abutments and primary predicate GoDirect implants
{7}------------------------------------------------
(K090234) system contains both implants and abutments, and the intended use reflects this. The difference in the indications for use of the proposed devices and the primary predicate was due to the GoDirect implant being cleared with other Implant Direct implants (K090234), as part of the Spectra-System Dental Implants 2008. Also, the primary predicate was originally cleared with implants that are intended for both single unit and multi-unit restorations, while the proposed devices are only indicated for multi-unit bridge or overdentures and not for single unit restorations. Thus, the proposed devices have more limited indications for use than the primary predicate that does not alter the intended use for demonstration of substantial equivalence.
In addition, the reference predicate GPS Straight Abutments (cleared under K143011- 2014 InterActive / SwishActive System) and the reference predicate InterActive® Screw Receiving Overdenture abutment with a ball top (cleared under K130572 - InterActive/SwishPlus2 Implant System) are also considered substantially equivalent. The worst case proposed angled GPS abutment uses the same compatible implant interfaces, have the same thread size, same materials, same coatings, and same manufacturing processes as reference predicate abutments (K143011 - 2014 InterActive / SwishActive System).
The proposed GPS Angled abutments have the same coronal top features as the primary predicate GoDirect Implant (K090234 - Spectra System Dental Implants 2008) and reference predicate GPS Straight Abutments (cleared under K143011 -2014 InterActive / SwishActive System). The proposed GPS® Angled abutments are a two-piece design having an identical lower piece with same interface features and angles as the reference predicate InterActive screw receiving angled abutments with a ball top (K130572 - InterActive/SwishPlus2 Implant System).
The proposed GPS® Angled Abutments differ from the primary predicate by having a two piece design with an angled body that utilizes a GPS top piece in order to have the same GPS compatibility with removable attachments.
These changes do not affect substantial equivalence of the proposed devices as demonstrated by internal performance and biocompatibility testing. GPS® Angled Abutment are substantially equivalent to the primary predicate GoDirect device cleared under (K090234 - Spectra System Dental Implants 2008) based on composition, performance and testing comparisons.
{8}------------------------------------------------
| Technological<br>Characteristics | Proposed<br>Device: GPS®<br>Angled<br>Abutment | Primary Predicate<br>Device:<br>GoDirect Implants<br>(K090234 – Spectra<br>System Dental<br>Implants 2008) | Reference Predicate<br>Device: Straight<br>GPS Abutment<br>(K143011 - 2014<br>InterActive/<br>SwishActive<br>System) | Reference<br>Predicate Device:<br>InterActive®<br>Swishplus2<br>Implant System<br>(Abutment with<br>Ball Top)<br>(K130572 -<br>InterActive/<br>SwishPlus2<br>Implant System) |
|----------------------------------|-----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Implant Direct<br>LLC | Implant Direct LLC | Implant Direct LLC | Implant Direct LLC |
| Regulation No. | 21 CFR 872. 3630 | 21 CFR 872. 3640 | 21 CFR 872. 3640 | 21 CFR 872. 3640 |
| Regulation<br>Class | II | II | II | II |
| Product Code | NHA | DZE, NHA | DZE, NHA | DZE, NHA |
| Technological<br>Characteristics | Proposed<br>Device: GPS®<br>Angled<br>Abutment | Primary Predicate<br>Device:<br>GoDirect Implants<br>(K090234 - Spectra<br>System Dental<br>Implants 2008) | Reference Predicate<br>Device: Straight<br>GPS Abutment<br>(K143011 - 2014<br>InterActive/<br>SwishActive<br>System) | Reference<br>Predicate Device:<br>InterActive®<br>Swishplus2<br>Implant System<br>(Abutment with<br>Ball Top)<br>(K130572 –<br>InterActive/<br>SwishPlus2<br>Implant System) |
| Indications<br>for Use | GPS® Angled<br>Abutments are<br>designed to be<br>used in support<br> | Spectra-System<br>Dental implants 2008<br>are comprised of dental<br>implant fixtures and<br>prosthetic devices that<br>compose a two- piece<br>implant system. The<br>implants are intended<br>for the use in the<br>mandible and maxilla,<br>in support of single unit<br>or multiple unit cement<br>or screw receiving<br>restorations and for<br>retention and support<br>of<br>overdentures.<br>The implants are<br>intended for immediate<br>placement and function<br>for the support of<br>single tooth or<br>multiple-tooth<br>restorations,<br>recognizing bone<br>stability and<br>appropriate occlusal<br>load requirements. | InterActive®<br>/SwishActive<br>Implant System<br>consists of two-<br>piece implants for<br>one-stage or two-<br>stage surgical<br>procedures. These<br>implants are<br>intended for use in<br>partially and fully<br>edentulous upper<br>and lower jaws in<br>support of single or<br>multiple-unit<br>restorations and<br>terminal or<br>intermediate<br>abutment support<br>for fixed<br>bridgework.<br>Implants can be<br>indicated for<br>immediate loading<br>when good primary<br>stability has been<br>achieved and with<br>appropriate<br>occlusal loading. | InterActive®<br>/SwishPlus2<br>Implant System<br>consists of two-<br>piece implants for<br>one-stage or two-<br>stage surgical<br>procedures. These<br>implants are<br>intended for use in<br>partially and fully<br>edentulous upper<br>and lower jaws in<br>support of single or<br>multiple-unit<br>restorations and<br>terminal or<br>intermediate<br>abutment support<br>for fixed |
## TABLE 2: Substantial Equivalence: Comparison of Proposed Device with Predicate on Technological Characteristics
{9}------------------------------------------------
{10}------------------------------------------------
| Compatibility: |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| •Legacy System:<br>Prosthetically<br>compatible with<br>Zimmer Dental<br>Tapered Screw-Vent<br>system 3.5mm<br>platform implants<br>(3.7mmD, 4.1mmD,<br>8mm-16mm<br>Length), 4.5mm<br>platform<br>implants(4.7mmD,<br>8mm-16mm<br>Length), and 5.7mm<br>platform implants<br>(6.0mmD, 8mm-<br>16mm Length).<br>•SwishTapered<br>System:<br>Prosthetically<br>compatible with<br>Straumann Standard<br>and Standard Plus<br>system RN platform<br>implants (3.3mmD-<br>4.8mmD, 6mm-<br>16mm Length) and<br>WN platform<br>implants(4.8mmD,<br>6mm-12mm<br>Length).<br>•SwishPlus System:<br>Prosthetically<br>compatible with<br>Straumann Standard<br>and Standard Plus<br>system RN platform<br>implants (3.3mmD-<br>4.8mmD, 6mm-<br>16mm Length) and<br>WN platform<br>implants (4.8mmD,<br>6mm-12mm<br>Length). |
{11}------------------------------------------------
| Technological<br>Characteristics | Proposed<br>Device: GPS®<br>Angled<br>Abutment | Primary Predicate<br>Device:<br>GoDirect Implants<br>(K090234 – Spectra<br>System Dental<br>Implants 2008) | Reference Predicate<br>Device: Straight<br>GPS Abutment<br>(K143011 – 2014<br>InterActive/<br>SwishActive<br>System) | Reference<br>Predicate Device:<br>InterActive®<br>Swishplus2<br>Implant System<br>(Abutment with<br>Ball Top)<br>(K130572 –<br>InterActive/<br>SwishPlus2<br>Implant System) |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | SwishActive Implants:<br>SwishActive implants are prosthetically compatible with InterActive 3.0 and 3.4mm abutments and Nobel Biocare conical connection NobelActiveTM NP (Narrow Platform – 3.0mm diameter) and NobelActiveTM RP (Regular Platform – 3.4mm diameter) titanium abutments. InterActive 3.0 and 3.4mm abutments are prosthetically compatible with Nobel Biocare conical connection NobelActiveTM NP (Narrow Platform– 3.0mm diameter) and NobelActiveTM RP (Regular Platform – 3.4mm diameter) (3.5–5.0mmD, 8.5-18mmLength). | | | |
{12}------------------------------------------------
| InterActive System:<br>InterActive<br>implants are<br>prosthetically<br>compatible with<br>InterActive 3.0<br>and 3.4mm<br>abutments and<br>Nobel Biocare<br>conical connection<br>NobelActive™ NP<br>(Narrow Platform<br>– 3.0mm<br>diameter) and<br>NobelActive™ RP<br>(Regular Platform<br>– 3.4mm<br>diameter)<br>titanium<br>abutments.<br>InterActive 3.0<br>and 3.4mm<br>abutments are<br>prosthetically<br>compatible with<br>Nobel Biocare<br>conical connection<br>NobelActive™ NP<br>(Narrow Platform–<br>3.0mm diameter)<br>and NobelActive™<br>RP (Regular<br>Platform – 3.4mm<br>diameter) (3.5-<br>5.0mmD, 8.5-<br>18mmLength)<br>implants. | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
{13}------------------------------------------------
| | General<br>Design | 2-piece abutment<br>with a GPS top for<br>compatibility with a<br>snap-on cap<br>attachment system<br>and a implant<br>engaging feature<br>with 15 degree and<br>30 degree angled<br>base. | GPS top integrated into<br>coronal end of implant<br>utilizing snap-on cap<br>attachment systems. | 1-piece abutment with<br>a GPS top for<br>compatibility with<br>snap-on cap<br>attachment system<br>and with a threaded<br>outer body to attach to<br>implant. | 2-piece abutment with<br>a ball top for<br>compatibility with a<br>snap-on cap<br>attachment system<br>and a implant<br>engaging feature with<br>15 degree and 30<br>degree angled base. |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| a) | Fixation<br>screw<br>Design | Same | N/A | Different | Same |
| | b) Lower<br>piece<br>design | Same | N/A | Different | Same |
| c) | Coronal<br>top design<br>(Outer<br>Features) | Same | Same | Same | Different |
| Technological<br>Characteristics | Proposed<br>Device: GPS®<br>Angled<br>Abutment | Primary Predicate<br>Device:<br>GoDirect Implants<br>(K090234 - Spectra<br>System Dental<br>Implants 2008) | Reference Predicate<br>Device: Straight<br>GPS Abutment<br>(K143011 - 2014<br>InterActive/<br>SwishActive<br>System) | Reference<br>Predicate Device:<br>InterActive®<br>Swishplus2<br>Implant System<br>(Abutment with<br>Ball Top)<br>(K130572 -<br>InterActive/<br>SwishPlus2<br>Implant System) | |
| d) Coronal<br>top design<br>(Internal<br>Features) | Same | Same | Different | Different | |
| e) Abutment<br>Angle | 15° and 30° | 0° | 0° | 15° and 30° | |
| f) Abutment /<br>Implant<br>Interface<br>Type | Engaging | N/A | Non-engaging | Engaging | |
| g) Attachment<br>-Abutment<br>Interface<br>Platform | 3.9mm | 3.9mm | 3.9mm | 2.54mm | |
| Material | Titanium alloy<br>base with TiN<br>coating | Titanium alloy base<br>with TiN coronal<br>surface | Titanium alloy<br>base with TiN<br>coating | Titanium alloy<br>base | |
| Sterilization: | Provided non-<br>sterile with<br>validated steam<br>sterilization<br>instructions for<br>use. Product<br>label indicates<br>sterility<br>according to ISO<br>standard. | Provided sterile. No<br>sterilization<br>instructions needed for<br>sterile product.<br>Sterilization process is<br>validated. Product label<br>indicates sterility<br>according to ISO<br>standard. | Provided non-<br>sterile with<br>validated steam<br>sterilization<br>instructions for use.<br>Product label<br>indicates sterility<br>according to ISO<br>standard. | Provided non-<br>sterile with<br>validated steam<br>sterilization<br>instructions for use.<br>Product label<br>indicates sterility<br>according to ISO<br>standard. | |
| Packaging | Mounted to inner<br>vial, inside an<br>outer vial and<br>sealed with a cap | Mounted to inner vial,<br>inside an outer vial and<br>sealed with a cap | Mounted to inner<br>vial, inside an outer<br>vial and sealed<br>with a cap | Mounted to inner<br>vial, inside an outer<br>vial and sealed<br>with a cap | |
{14}------------------------------------------------
{15}------------------------------------------------
## Non-Clinical Performance Data
Non-clinical testing was performed on the proposed GPS® Angled abutments.
Mechanical testing was performed on the proposed devices following FDA "Class II Special Control Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Abutments" and ISO 14801 in static compression and fatigue. The worst case abutment testing was performed on the proposed abutments with Implant Direct implants and with compatible third party implants. The implant/abutment device combinations were able to withstand an equivalent or higher load than the primary predicate device.
In addition, validations were conducted on non-sterile abutments for steam sterilization and biocompatibility testing.
A steam sterilization validation was conducted using the same sterilization parameters specified in the instructions for use which is used for the proposed devices. A sterilization validation was conducted on a non-sterile abutment conforming to AAMI/ANSI/ISO 17665-1 and ANSVAAMI ST79 to a sterility assurance level (SAL) of 10 -6. In addition, the compatible GPS comfort cap was validated to be steam sterilizable.
Biocompatibility testing was conducted on the devices made with the same material and processes as the GPS components. Testing was conducted on the proposed Angled GPS Abutments, primary predicate GoDirect implant and compatible comfort caps according to ISO 10993-5. The results did not show a toxic reaction from the proposed finished devices.
The non-clinical performance data demonstrates that the proposed GPS® Angled abutments are mechanically equivalent, can be steam sterilized, and are biocompatible for the intended use.
Clinical Performance Data [N/A]
Clinical data is not needed to adequately characterize substantial equivalence.
## Conclusion as to Substantial Equivalence
The differences between the proposed devices and the predicate device were reviewed to evaluate the substantial equivalency. The following features are modifications to the primary predicate device.
- · The proposed Angled GPS abutments are a multiple piece design using an angled abutment, fixation screw, and a GPS top to make the complete assembly. In comparison, the primary predicate GoDirect implant (K090234 - Spectra System Dental Implants 2008) is a solid
{16}------------------------------------------------
device where the coronal end of the implant is machined with GPS features. The coronal end of the GoDirect implant has the same features (including coating) as the GPS top (P/N: 1000-90S).
- The proposed Angled GPS abutments are manufactured at angles of 15° and 30°. The primary predicate device GPS top is straight and aligned with the implant axis.
The technical characteristics and non - c l i n i c a 1 performance data demonstrate that the proposed GPS® Angled Abutments are substantially equivalent to the primary predicate device (GoDirect implant cleared under K090234 - Spectra System Dental Implants 2008).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.