K153464 · Fujifilm Medical Systems U.S.A, Inc. · MQB · Apr 8, 2016 · Radiology
Device Facts
Record ID
K153464
Device Name
FDR D-EVO flat Panel Detector System
Applicant
Fujifilm Medical Systems U.S.A, Inc.
Product Code
MQB · Radiology
Decision Date
Apr 8, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Wireless/Wired FDR D-EVO flat panel detector system is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications including pediatric and neonatal exams wherever conventional film/screen or CR systems may be used. The FDR D-EVO is not intended for mammography, fluoroscopy, tomography, and angiography applications. The FDR D-EVO can be used with Virtual Grid Software, which is optional image processing software installed on Fujifilm's FDX Console. Virtual Grid Software can be used in lieu of an anti-scatter grid to improve image contrast in general radiographic images by reducing the effects of scatter radiation.
Device Story
FDR D-EVO is a portable digital flat panel detector system capturing X-ray exposures for display. Used in general radiography (adults, pediatrics, neonates) in clinical environments; replaces conventional film/screen or CR systems. Detectors (GOS or CsI scintillators) interface with X-ray sources via wired/wireless connection or automatic X-ray detection (SmartSwitch). Optional 'Virtual Grid Software' on FDX Console processes images to improve contrast by reducing scatter radiation, allowing use without physical anti-scatter grids. Clinicians use the software to potentially reduce patient radiation dose by up to 50% while maintaining image quality comparable to hardware grids. Software is applied post-acquisition; users toggle processing ON/OFF based on visual assessment. Device supports upright (wall bucky) and recumbent (table bucky) exams.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing and image quality evaluations. Testing confirmed Virtual Grid Software achieves image contrast equivalent to hardware grids with up to 50% dose reduction.
Indicated for general projection radiographic imaging of human anatomy, including pediatric and neonatal patients. Not for mammography, fluoroscopy, tomography, or angiography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
FDR D-EVO Flat Panel Detector System (DR-ID600) (K141765)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized caduceus, which is a staff with two snakes entwined around it, representing medicine and healing. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 8, 2016
FUJIFILM Medical Systems U.S.A., Inc. % Mr. Peter Altman Regulatory and Quality Consultant 419 West Avenue STAMFORD CT 06902
Re: K153464
Trade/Device Name: FDR D-EVO Flat Panel Detector System (DR-ID600) with Improved Virtual Grid Software Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: March 11, 2016 Received: March 14, 2016
Dear Mr. Altman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|-------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| <b>Indications for Use</b> | | |
| 510(k) Number (if known) | K153464 | |
| Device Name | FDR D-EVO Flat Panel Detector System (DR-ID600) with Improved Virtual Grid Software | |
| Indications for Use (Describe) | The Wireless/Wired FDR D-EVO flat panel detector system is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications including pediatric and neonatal exams wherever conventional film/screen or CR systems may be used. The FDR D-EVO is not intended for mammography, fluoroscopy, tomography, and angiography applications. The FDR D-EVO can be used with Virtual Grid Software, which is optional image processing software installed on Fujifilm's FDX Console. Virtual Grid Software can be used in lieu of an anti-scatter grid to improve image contrast in general radiographic images by reducing the effects of scatter radiation. | |
| Type of Use (Select one or both, as applicable) | ☐ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
элемдог рисг гэтэй хэрэгтэй үүл (009Дг-ЯД) ОЛЭ-Д ЯДЭ
E-レ
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Fujifilm logo. The logo is black, except for a small red square with a white triangle cut out of it. Below the logo, in a smaller font, is the text "Value from Innovation."
# 510(k) Summary
# FDR D-EVO Flat Panel Detector System (DR-ID600) With Improved Virtual Grid Software
# Date: April 7, 2016
# Submitter's Information:
FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue Stamford, CT, 06902, USA
# Contact Person:
| Name: | Peter Altman |
|------------|-----------------------------------|
| Title: | Regulatory and Quality Consultant |
| Telephone: | (203) 602-3576 |
# Identification of the Device:
| Proprietary/Trade Name: | FDR D-EVO Flat Panel Detector System (DR-ID600)<br>with Improved Virtual Grid Software |
|-------------------------|----------------------------------------------------------------------------------------|
| Classification Name: | Stationary X-ray system |
| Regulations Number: | 21 CFR 892.1680 |
| Product Codes: | MQB |
| Device Class: | Class II |
| Review Panel: | Radiology |
| Common Name: | Flat Panel Digital Detector System |
# Identification of the Legally Marketed Device:
FDR D-EVO Flat Panel Detector System (DR-ID600), K141765 cleared 10/3/2014
| Proprietary/Trade Name: | FDR D-EVO Flat Panel Detector System (DR-ID600) |
|-------------------------|-------------------------------------------------|
| Classification Name: | Stationary X-ray system |
| Regulations Number: | 21 CFR 892.1680 |
| Product Codes: | MQB |
| Device Class: | Class II |
| Review Panel: | Radiology |
| Common Name: | Flat Panel Digital Detector System |
#### l. DEVICE DESCRIPTION
Fujifilm's FDR D-EVO Flat Panel Detector System (DR-ID600) is a portable digital detector system that interfaces with, and acquires and digitizes x-ray exposures, from
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Fujifilm logo. The logo is black, except for a small red square with a white triangle cut out of it, which is located above the "I" in Fujifilm. Below the logo, the words "Value from Innovation" are written in a smaller, black font.
standard radiographic systems. The FDR D-EVO is designed to be used in any environment that would typically use a radiographic cassette for examinations of adults, pediatrics and neonates. The detector models support both wireless and wired/tethered data communication between the detector and the system. Detectors can be placed in a wall bucky for upright exams, a table bucky for recumbent exams, or removed from the bucky for non-grid exams.
The FDR D-EVO can be used with or without Virtual Grid Software, which is optional image processing software installed on Fujifilm's FDX Console. The Virtual Grid Software was originally designed to improve image contrast in general radiographic images by reducing the effects of scatter radiation, primarily for exams acquired without a grid. Initially, the anatomical regions of the software were for chests and abdomens only, where the effects of scatter radiation can be prominent if a grid is not used. The user can decide whether or not to apply the Virtual Grid image processing by turning it ON or OFF as they see fit based on the displayed image. The same software is also offered with other legally marketed Fujifilm DR Systems that use the same FDX Console as the acquisition workstation since this post-processing algorithm does not depend on how the image is acquired, or with what acquisition device.
Since the K141765 clearance, however, several improvements have been made in order to expand the effectiveness of the Virtual Grid Software, which is the reason of the new submission. The improvements are 1) expansion of target body parts to cover the whole anatomy, 2) target image format to include images obtained by Fujifilm's CR systems, and 3) adding a new claim that Virtual Grid Software can be used in lieu of an antiscatter grid.
An additional benefit of using the Virtual Grid Software involves X-ray radiation dose. It is a well-known practice that exposure conditions should be increased when a hardware grid is used in order to compensate for the primary X-ray energy absorbed by the grid strips. On the other hand, since the Virtual Grid processing can be applied to images taken without a grid. Virtual Grid can achieve image contrast equivalent to that of a hardware grid, but at lower dose levels than are needed when using a hardware grid. Based on Fujifilm's performance testing and image quality evaluation results, a maximum dose reduction of 50% when using VG with Fuiifilm DR detectors can achieve comparable image quality when compared to using a real grid.
Virtual Grid Software comes with some limitations such as VG should not be applied to images that were acquired with a real grid. VG cannot be applied to images that were acquired prior to the VG software installation. VG can only be used for selected exposure menus listed in the manual. VG should not be used for images acquired with additional beam filtration.
Device Integration: The FDR D-EVO can be used with any legally-marketed and appropriately certified X-ray source. If FDR D-EVO is used with the optional Hand Switch Interface Box, a connection to the X-ray source/qenerator's timing signals (prep and exposure signals) is necessary. Even if such timing signals are not available, the
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Fujifilm logo. The logo consists of the word "FUJIFILM" in a bold, sans-serif font, with the "I" in "FUJI" having a red square at the top. Below the logo is the tagline "Value from Innovation" in a smaller, sans-serif font.
FDR D-EVO can acquire an image using Automatic x-ray detection function (known as 'SmartSwitch' in the US), so the limitation is neqligible.
#### .. INDICATIONS FOR USE
The Wireless/Wired FDR D-EVO flat panel detector system is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications including pediatric and neonatal exams wherever conventional film/screen or CR systems may be used. The FDR D-EVO is not intended for mammography, fluoroscopy, tomography, and angiography applications.
The FDR D-EVO can be used with Virtual Grid Software, which is optional image processing software installed on Fujifilm's FDX Console. Virtual Grid Software can be used in lieu of an anti-scatter grid to improve image contrast in general radiographic images by reducing the effects of scatter radiation.
#### = SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
FDR D-EVO FPD System (DR-ID600) continues to offer 600 series detectors with two different scintillator materials - GOS (Gadolinium oxysulfide) or Csl (Cesium lodide) and various detector sizes to choose from. All detector characteristics including indirect conversion method using Fujifilm's unique Irradiated Side Sampling (ISS) technology, pixel pitch (150µm), resolution (16 bit), DQE (Detective Quantum Efficiency) measurements, MTF (Modulation Transfer Function) performance remains the same. The wireless communication feature, and automatic x-ray detection function (known as 'SmartSwitch' in the US) also stay unaffected by the changes. The same acquisition workstation, FDX Console will continue to support FDR D-EVO FPD System (DR-ID600), but FDX Console v9.0 or later will release the Improved Virtual Grid Software.
#### IV. SUBSTANTIAL EQUIVALENCE
The FDR D-EVO FPD System (DR-ID600) with Improved Virtual Grid Software is substantially equivalent to the following legally marketed device:
| Legally Marketed Device | 510(k) # | Clearance Date |
|-------------------------------------------------|----------|----------------|
| FDR D-EVO Flat Panel Detector System (DR-ID600) | K141765 | 10/3/2014 |
The Virtual Grid Software was cleared with FDR D-EVO in K141765 as optional software installed in the acquisition workstation, FDX Console. The FDR D-EVO's detector characteristics as well as other major features such as wireless communication and automatic x-ray detection function ('SmartSwitch') remain unchanged. The FDR D-EVO can be used with or without Virtual Grid Software, and the improvements made to the Virtual Grid Software do not affect the intended use or alter the fundamental scientific technology of the legally marketed device.
#### V. SUMMARY OF STUDIES
Non-clinical Performance Data: The conformity to the voluntary standards such as AAMI/ANSI ES60601-1, IEC 60601-1, IEC 60601-1-2, IEC 62304, IEC 62366, IEC 62494-1 and DICOM remains unaffected. As required by the risk analysis, all verification and validation activities related to the improvements made to the Virtual Grid Software were performed and the results were satisfactory. Additionally the bench
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Fujifilm logo. The logo is black, except for a red square with a white triangle cut out of the top right corner. Below the logo, the words "Value from Innovation" are written in black.
testing and image quality evaluation demonstrated that Virtual Grid Software can achieve image contrast equivalent to that of a hardware grid, but at lower dose levels than are needed when using a hardware grid. The results of the bench testing and image evaluation are provided in the submission.
Clinical Performance Data: The changes do not require clinical studies. The substantial equivalence has been demonstrated by non-clinical studies.
#### VI. CONCLUSION
Based upon the supporting data summarized above, we concluded the Improved Virtual Grid Software offered with FDR D-EVO Flat Panel Detector System (DR-ID600) is as safe and effective as the legally marketed device K141765 and do not raise different questions of safety and effectiveness than K141765.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.