PowerLoc MAX Power Injectable Infusion Set, SafeStep Huber Needle Set
K153440 · C.R. Bard, Inc. · FPA · Aug 18, 2016 · General Hospital
Device Facts
Record ID
K153440
Device Name
PowerLoc MAX Power Injectable Infusion Set, SafeStep Huber Needle Set
Applicant
C.R. Bard, Inc.
Product Code
FPA · General Hospital
Decision Date
Aug 18, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PowerLoc MAX® Power-Injectable Infusion Set is an intravascular administration set with a non-coring right angle needle and manually activated needle stick prevention safety mechanism which reduces the risk of accidental needlestick injuries by shielding the needle. The needle is used to access surgically implanted vascular ports. The PowerLoc MAX® Power-Injectable Infusion Set is indicated for use in the administration of fluids and drugs, as well as blood sampling through surgically implanted vascular ports. When used with ports that are indicated for power injection of contrast media into the central venous system, the PowerLoc MAX® Power-Injectable Infusion Set is also indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate at 11.8 cPs is 5 ml/s for 19 gauge and 20 gauge needles, and 2 ml/s for 22 gauge needles. The SafeStep® Huber Needle Set is a device intended for insertion into the septum of a subcutaneously implanted port and for the infusion of fluids into the port. The SafeStep® Huber Needle safety feature is manually activated during needle removal, and is designed to aid in the prevention of accidental needlesticks.
Device Story
PowerLoc MAX and SafeStep Huber needle sets are intravascular administration sets used to access surgically implanted vascular ports. Devices feature non-coring right-angle needles and manually activated safety mechanisms to shield needles post-removal, reducing needlestick risk. PowerLoc MAX supports power injection of contrast media at specified flow rates (up to 5 ml/s for 19/20G, 2 ml/s for 22G). An optional pre-loaded stabilization device accessory aids in port access by stabilizing the port during needle insertion. Devices are used by healthcare practitioners in clinical settings. Output is the delivery of fluids/drugs or contrast media, or blood withdrawal. Benefits include reduced needlestick injury risk and reliable port access.
Clinical Evidence
Bench testing only. Performance validated via internal protocols (slide force, removal force, clinical acceptability) and adherence to ISO/BS EN standards including ISO 23908 (sharps protection), ISO 8536-4 (infusion equipment), ISO 10993 (biocompatibility), and ISO 594 (luer fittings). No clinical trial data required.
Technological Characteristics
Non-coring right-angle Huber needles; manually activated safety shielding mechanism. Materials: biocompatible medical grade. Connectivity: none. Sterilization: Ethylene Oxide (ISO 11135) or Radiation (ISO 11137-1). Standards: ISO 23908, ISO 8536-4, ISO 594-1/2, ISO 9626, ISO 10555-1, ISO 7864.
Indications for Use
Indicated for patients requiring access to surgically implanted vascular ports for fluid/drug administration, blood sampling, or power injection of contrast media (PowerLoc MAX only). Contraindicated for ports not rated for power injection when performing power injection.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
SafeStep® MAX Power-Injectable Infusion Set (K073050)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 18, 2016
Darlene Hull Regulatory Affairs Specialist C.R. Bard, Inc. 605 North 5600 West Salt Lake City, Utah 84116
Re: K153440
Trade/Device Name: PowerLoc Max® Power-Injectable Infusion Set, SafeStep® Huber Needle Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: July 19, 2016 Received: July 20, 2016
Dear Darlene Hull:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K153440
Device Name PowerLoc MAX® Power-Injectable Infusion Set
#### Indications for Use
The PowerLoc MAX® Power-Injectable Infusion Set is an intravascular administration set with a non-coring right angle needle and manually activated needle stick prevention safety mechanism which reduces the risk of accidental needlestick injuries by shielding the needle is used to access surgically implanted vascular ports.
The PowerLoc MAX® Power-Iniectable Infusion Set is indicated for use in the administration of fluids and drugs, as well as blood sampling through surgically implanted vascular ports.
When used with ports that are indicated for power injection of contrast media into the central venous system, the PowerLoc MAX® Power-Injectable Infusion Set is also indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate at 11.8 cPs is 5 ml/s for 19 qauge and 20 gauge needles, and 2 ml/s for 22 gauge needles.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C) 区
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration
Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
| FORM FDA 3881 (8/14) | Page 1 of 1 | PSC |
|------------------------------------|-------------|-----|
| Publishing Services (301) 443-6740 | EF | |
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31. 2017 See PRA Statement below.
510(k) Number (if known)
K153440
Device Name SafeStep® Huber Needle Set
Indications for Use
The SafeStep® Huber Needle Set is a device intended for insertion into the septum of a subcutaneously implanted port and for the infusion of fluids into the port. The SafeStep® Huber Needle safety feature is manually activated during needle removal, and is designed to aid in the prevention of accidental needlesticks.
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response. including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration
Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
| FORM FDA 3881 (8/14) |
|------------------------------------|
| Publishing Services (301) 443-6740 |
Page 1 of 1
PSC
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for BARD ACCESS SYSTEMS. The word "BARD" is in large, bold, sans-serif font. Below that, in a smaller font, are the words "ACCESS SYSTEMS".
# 510(k) Summary
21 CFR § 807.92(a)
| General<br>Provisions | Submitter Name:<br>Address: | Bard Access Systems, Inc.<br>605 North 5600 West<br>Salt Lake City, UT 84116 |
|-----------------------|-------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact Person:<br>Telephone Number:<br>Fax Number:<br>Date of Preparation: | Darlene Hull<br>Regulatory Affairs Specialist<br>(801) 522-5613<br>(801) 522-5425<br>23 November 2015 |
| Subject<br>Devices | Trade Name:<br>Common Name:<br>Classification Name:<br>Product Code/<br>Regulation: | PowerLoc MAX® Power-Injectable Infusion Set<br>Huber Needle Intravascular Administration Set<br>Intravascular Administration Set<br>FPA<br>21 CFR § 880.5440 |
| | Trade Name:<br>Common Name:<br>Classification Name:<br>Product Code/<br>Regulation: | SafeStep® Huber Needle Set<br>Huber Needle Intravascular Administration Set<br>Intravascular Administration Set<br>FPA<br>21 CFR § 880.5440 |
| Predicate<br>Devices | Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer: | SafeStep® MAX Power-Injectable Infusion Set<br>Intravascular Administration Set<br>K073050<br>Bard Access Systems, Inc. |
| | Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer: | Luther Safety Huber Needle Set<br>Intravascular Administration Set<br>K040527<br>Bard Access Systems, Inc. |
{5}------------------------------------------------
| | PowerLoc MAX® Power-Injectable Infusion Set | |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | The PowerLoc MAX® Power-Injectable Infusion Set is a standard non-<br>coring intravascular administration set with a non-coring Huber type<br>right angle needle and a manually activated needle-stick prevention<br>safety mechanism which reduces the risk of accidental needlestick<br>injuries by shielding the needle. The device is used to access<br>surgically implanted vascular ports and is indicated for use in the<br>administration of fluids and drugs, as well as blood sampling. | |
| | The PowerLoc MAX® Power-Injectable Infusion Set is also indicated<br>for power injection of contrast media into the central venous system<br>only through an implanted port that is also indicated for power<br>injection. The maximum recommended infusion rate at 11.8 cPs is 5<br>ml/s for 19 gauge and 20 gauge needles and 2 ml/s for 22 gauge<br>needles. The PowerLoc MAX® Power-Injectable Infusion Set is<br>offered with and without a Y-site. | |
| Device<br>Descriptions | SafeStep® Huber Needle Set | |
| | The SafeStep® Huber Needle Set is a standard right angle Huber<br>needle and administration set with a needlestick prevention feature,<br>designed for use with a vascular access infusion system. It is<br>manufactured with conventional medical grade, biocompatible<br>materials. The SafeStep® Huber Needle Set operates as a standard<br>Huber needle with the addition of a safety feature to aid in the<br>prevention of needlestick injuries to the health practitioner. The<br>SafeStep® Huber Needle Set is offered with and without a Y-site. | |
| | Stabilization Device | |
| | The stabilization device is intended for use as an accessory to the<br>subject PowerLoc MAX® and SafeStep® infusion sets, and is supplied<br>pre-loaded with the subject devices. It is placed centrally over the<br>implanted port so that the base surrounds the implanted port under the<br>skin. The infusion set needle handle is pressed down until the needle<br>has entered the port septum, then the stabilization device is lifted off of<br>the needle. The stabilization device is then discarded per hospital<br>protocol. | |
| Intended Use | Both the subject PowerLoc MAX® Power-Injectable Infusion Set and<br>SafeStep® Huber Needle Sets, with and without the stabilization<br>device, are intended for use in the administration of fluids and drugs,<br>as well as blood sampling through surgically implanted vascular ports. | |
| Indications For<br>Use | PowerLoc MAX® Power-Injectable Infusion Set | |
| | The PowerLoc MAX® Power-Injectable Infusion Set is an intravascular<br>administration set with a non-coring right angle needle and manually<br>activated needle stick prevention safety mechanism which reduces the<br>risk of accidental needlestick injuries by shielding the needle. The | |
| | needle is used to access surgically implanted vascular ports. | |
| | The PowerLoc MAX® Power-Injectable Infusion Set is indicated for<br>use in the administration of fluids and drugs, as well as blood sampling<br>through surgically implanted vascular ports. | |
| | When used with ports that are indicated for power injection of contrast<br>media into the central venous system, the PowerLoc MAX® Power-<br>Injectable Infusion Set is also indicated for power injection of contrast<br>media. For power injection of contrast media, the maximum<br>recommended infusion rate at 11.8 cPs is 5 ml/s for 19 gauge and 20<br>gauge needles, and 2 ml/s for 22 gauge needles. | |
| | SafeStep® Huber Needle Set | |
| | The SafeStep® Huber Needle Set is a device intended for insertion<br>into the septum of a subcutaneously implanted port and for the infusion<br>of fluids into the port. The SafeStep® Huber Needle safety feature is<br>manually activated during needle removal, and is designed to aid in the<br>prevention of accidental needlesticks. | |
| Technological<br>Characteristics | Technological characteristics including design and function of the<br>subject devices, the PowerLoc MAX® Power-Injectable Infusion Set<br>and SafeStep® Huber Needle Set without stabilization device, are<br>similar with respect to those of the predicate devices, SafeStep® MAX<br>Power-Injectable Infusion Set and Luther Safety Huber Needle Set. | |
| | The subject devices may be used with a pre-loaded stabilization device<br>accessory that is intended to stabilize the port during access and aid<br>insertion of the needle into the port. The stabilization device utilizes<br>materials that are a different formulation than materials used in the<br>predicate devices, and have not been used in another legally marketed<br>device within the same classification regulation with the same intended<br>use. The stabilization device constitutes a change to the ergonomics of<br>the patient/user interface of the subject devices, but does not change<br>the design or function of the subject devices when used as an<br>accessory to these devices. | |
| Performance<br>Tests | Verification and validation tests were performed in accordance with<br>Design Controls per 21 CFR §820.30. The following performance data<br>were referenced in support of the substantial equivalence<br>determination. Testing was performed based on Bard Access Systems<br>(BAS) internal test protocols as well as on existing standards that are<br>relevant to the subject devices. | |
| | Testing performed based on BAS internal test protocols are as follows: | |
| | Slide Force testing was performed to ensure that the force<br>required to slide the needle down the length of the stabilization<br>device did not exceed the force required to insert the needle in | |
{6}------------------------------------------------
{7}------------------------------------------------
| the port septum.<br>Stabilization Device Removal Force testing was performed to<br>ensure that it was not difficult to remove the stabilization device<br>once the subject needle was inserted into the port.<br>Acceptability of the Stabilization Device to verify that the | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| stabilization device was clinically acceptable in the placement<br>of a port access needle. | |
| Acceptability of GuardIVa® with Stabilization Device to<br>validate the use as directed in the IFU of the antimicrobial<br>dressing with the stabilization device. | |
| In addition to BAS internal protocols, the following guidance<br>documents and standards were used to determine appropriate<br>methods for evaluating the performance of the device: | |
| • ISO 23908 First edition 2011-06-11 Sharps injury protection -<br>Requirements and test methods - Sharps protection features for<br>single-use hypodermic needles, introducers for catheters and<br>needles used for blood sampling | |
| • ISO 8536-4 Fifth edition 2010-10-01 Infusion equipment for<br>medical use - Part 4: Infusion sets for single use, gravity feed<br>[Including: Amendment 1 (2013)]10. | |
| • BS EN ISO 8536-10:2015 Infusion equipment for medical use -<br>Part 10: Accessories for fluid lines for single use with pressure<br>infusion equipment. | |
| • ISO 10993-1 Fourth edition, 2009-10-15 - Biological Evaluation<br>of Medical Devices Part 1: Evaluation and Testing within a risk<br>management process. | |
| • ISO 10993-7: 2008 - Biological evaluation of medical devices -<br>Part 7: Ethylene oxide sterilization residuals. | |
| • ISO 11135: 1994 - Medical Devices - Validation and Routine<br>Control of Ethylene Oxide Sterilization. | |
| • BS EN ISO 11607-1 Packaging for terminally sterilized medical<br>devices Part 1: Requirements for materials, sterile barrier<br>systems and packaging systems. | |
| • ISO 594-1: 1986, Conical Fittings with a 6% (Luer) Taper for<br>Syringes, Needles, and Certain Other Equipment. | |
| • ISO 594-2: 1998, Conical Fittings with a 6% (Luer) Taper for<br>Syringes, Needles and Certain Other Medical Equipment - Part<br>2: Lock Fittings. | |
| • ISO 9626 AMD 1: 2001 - Stainless Steel Needle Tubing for<br>Manufacture of Medical Devices. | |
| • ISO 10555-1: 2013 – Intravascular catheters – Sterile and | |
| | single-use catheters – Part 1: General requirements.<br>• ISO 11137-1: 2006 – Sterilization of health care products –<br>Radiation - Part 1: Requirements for development, validation<br>and routine control of a sterilization process for medical<br>devices. |
| | • ISO 7864: 1993 – Sterile hypodermic needles for single use. |
| | • FDA Guidance Medical Devices with Sharps Injury Prevention<br>Features issued August 9, 2005. |
| | • FDA Guidance Establishing Safety and Compatibility of Passive<br>Implants in the Magnetic Resonance (MR) Environment issued<br>December 11, 2014. |
| | The subject devices met all predetermined acceptance criteria derived<br>from the above listed references and demonstrated substantially<br>equivalent performance as compared to the cited predicate devices. |
| | Risk management, including a failure modes and effects analysis<br>(FMEA), of the subject device was conducted in accordance with BS<br>EN ISO 14971:2012, Medical Devices – Risk Management for Medical<br>Devices. |
| Summary of<br>Substantial<br>Equivalence | Based on the intended use, technological characteristics, and safety<br>and performance testing, the subject PowerLoc MAX® Power-<br>Injectable Infusion Set and SafeStep® Huber Needle Set used with<br>and without the stabilization device meet the requirements that are<br>considered sufficient for its intended use as compared to the predicate<br>devices cited. Therefore, the subject devices are substantially<br>equivalent to the predicates. |
{8}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.