K153417 · Tosoh Bioscience, Inc. · JIT · Dec 18, 2015 · Clinical Chemistry
Device Facts
Record ID
K153417
Device Name
ST AIA-PACK PROG III Calibrator Set
Applicant
Tosoh Bioscience, Inc.
Product Code
JIT · Clinical Chemistry
Decision Date
Dec 18, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Intended Use
The ST AIA-PACK PROG III Calibrator Set is intended for In Vitro Diagnostic Use Only for the calibration of the ST AIA-PACK PROG III assay.
Device Story
ST AIA-PACK PROG III Calibrator Set consists of six liquid levels of bovine protein matrix containing progesterone (0 to 45 ng/mL approx.) and sodium azide preservative. Used in clinical laboratories on Tosoh AIA System Analyzers to calibrate the ST AIA-PACK PROG III assay. Calibrator values are assigned gravimetrically using USP reference standards and verified on AIA-2000 instruments. Healthcare providers use the resulting assay calibration to quantify progesterone levels in patient serum or heparinized plasma samples. Accurate calibration ensures the assay provides reliable progesterone measurements, which aids clinicians in assessing reproductive health and endocrine function.
Clinical Evidence
Bench testing only. Stability studies (real-time and open-vial) confirmed recovery within 100 +/- 10% and reproducibility (CV) <= 10%. Value assignment was validated by analyzing 5 replicates on two Tosoh AIA-2000 instruments, demonstrating precision (CV) between 2.0% and 4.8% across calibrator levels.
Technological Characteristics
Liquid calibrator set; bovine protein matrix; sodium azide preservative. Six levels (0, 0.5, 1.5, 5.0, 15, 45 ng/mL). Traceable to USP reference standard (Lot #11J239). Designed for use on Tosoh AIA System Analyzers. Storage: 2-8°C.
Indications for Use
Indicated for in vitro diagnostic calibration of the ST AIA-PACK PROG III assay. No specific patient population, age, or gender restrictions are defined for the calibrator itself, as it is a laboratory reagent used by clinical laboratory professionals.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Related Devices
K143144 — ST AIA-PACK PROG II Calibrator Set · Tosoh Bioscience, Inc. · Dec 3, 2014
K163546 — Lumipulse G Progesterone-N Calibrators · Fujirebio Diagnostics,Inc. · Jan 13, 2017
K113373 — DIMENSION VISTA LOCI 9 CALIBRATOR- LOCI 9 CAL · Siemens Healthcare Diagnostics, Inc. · Dec 30, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a series of human profiles, rendered in a dark color.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
TOSOH BIOSCIENCE, INC. ROBERT WICK REGULATORY SPECIALIST 6000 SHORELINE COURT, SUITE 101 SOUTH SAN FRANCISCO, CA 94080
December 18, 2015
Re: K153417
Trade/Device Name: ST AIA-PACK PROG III Calibrator Set Regulation Number: 21 CFR 862.1150 Regulation Name: Calibrator Regulatory Class: II Product Code: JIT Dated: November 24, 2015 Received: November 25, 2015
Dear Robert Wick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Katherine Serrano -S
For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K153417
Device Name ST AIA-PACK PROG III Calibrator Set
Indications for Use (Describe) The ST AIA-PACK PROG III Calibrator Set is intended for In Vitro Diagnostic Use Only for the calibration of the ST AIA-PACK PROG III assay.
Type of Use (Select one or both, as applicable)
| <label><input type="checkbox"/> For activities that do NOT disturb State-owned submerged lands</label> |
|--------------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> For Construction (that disturb State-owned submerged lands)</label> |
|× | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary ST AIA-PACK PROG III Calibrator Set
## SAFETY AND EFFECTIVENESS AS REQUIRED BY 1. 21 CFR 807.92
This Summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirement 21 CFR 807.92.
#### 2. SUBMITTER NAME AND ADDRESS
Date of Summary Preparation: Submitter:
November 23, 2015 Tosoh Bioscience, Inc 3600 Gantz Road Grove City, OH 43123
Contact Person:
Robert L. Wick Requlatory Specialist 6000 Shoreline Ct., Ste. 101 South San Francisco, CA 94080 Phone: 650-636-8117 Fax: 650-636-8121 Email: Robert.Wick@Tosoh.com
## 510K NUMBER, DEVICE PROPRIETARY NAME, 3. COMMON NAME, PURPOSE FOR THE SUBMISSION, REGULATORY CLASSIFICATION, PANEL, PRODUCT CODE, AND 21 CFR NUMBER.
510k No.: Device Proprietary Name: Purpose of Submission: Regulatory Classification Common Name: Classification: Product Code: Panel: 21 CFR Number:
ST AIA-PACK PROG III Calibrator Set New Product Calibrator, Secondary Progesterone Test System Class II JIT Clinical Chemistry 21 CFR 862.1150
{4}------------------------------------------------
## PREDICATE DEVICE PROPRIETARY NAMES AND 4. 510 (K) NUMBERS
Predicate Device Number: Predicate: Manufacturer:
K933269 ST AIA-PACK PROG (Calibrator Set) Tosoh Bioscience, Inc. (previously known as Tosoh Medics, Inc.)
#### INTENDED USE 5.
The ST AIA-PACK PROG III Calibrator Set is intended for In Vitro Diagnostic Use Only for the calibration of the ST AIA-PACK PROG III assay.
#### 6. DEVICE DESCRIPTION
According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.
## Device Description:
- 2 x 1 mL ST AIA-PACK PROG III Calibrator (1) 0 nq/mL
Bovine protein matrix containing no detectable concentration of progesterone with sodium azide as a preservative (Liquid).
| 2 x 1 mL ST AIA-PACK PROG III Calibrator (2) | | 0.5 | ng/mL (approx.) |
|----------------------------------------------|--|-----|-----------------|
| ST AIA-PACK PROG III Calibrator (3) | | 1.5 | ng/mL (approx.) |
| ST AIA-PACK PROG III Calibrator (4) | | 5.0 | ng/mL (approx.) |
| ST AIA-PACK PROG III Calibrator (5) | | 15 | ng/mL (approx.) |
| ST AIA-PACK PROG III Calibrator (6) | | 45 | ng/mL (approx.) |
Bovine protein matrix containing the assigned concentration of progesterone (described on each vial) with sodium azide as a preservative.
ST AIA-PACK PROG III Calibrator Set
P/N # 025340
The ST AIA-PACK PROG III Calibrator Set is designed specifically for use on the Tosoh AIA System Analyzers which have been previously cleared as a family of instruments under K971103. Only materials obtained from Tosoh should be used. Materials obtained elsewhere should not be substituted since assay performance is characterized based strictly on Tosoh materials.
The ST AIA-PACK PROG III Calibrator Set is designed for use with ST AIA-PACK PROG III and ST AIA-PACK PROG III Sample Diluting Solution.
{5}------------------------------------------------
#### PREDICATE COMPARISON TABLE 7.
Substantial Equivalence:
## Comparison between the Tosoh ST AIA-PACK PROG III Calibrator Set and the Tosoh ST AIA-PACK PROG Calibrator Set
#### Similarities
| Characteristic | | Predicate |
|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| | Tosoh AIA-PACK PROG III<br>Calibrator Set | Tosoh AIA-PACK PROG<br>Calibrator Set (K933269) |
| Intended Use | The ST AIA-PACK PROG III<br>Calibrator Set is intended for In Vitro<br>Diagnostic Use Only for the<br>calibration of the ST AIA-PACK<br>PROG III assay. | The AIA-PACK PROG Calibrator Set is intended for In Vitro<br>Diagnostic Use Only for the<br>calibration of the ST AIA-PACK<br>PROG Assays. |
| Indications for Use (same<br>as Intended Use) | The ST AIA-PACK PROG III<br>Calibrator Set is intended for In Vitro<br>Diagnostic Use Only for the<br>calibration of the ST AIA-PACK<br>PROG III assay. | The AIA-PACK PROG Calibrator Set is intended for In Vitro<br>Diagnostic Use Only for the<br>calibration of the ST AIA-PACK<br>PROG Assays. |
| Analyte | Progesterone | Same |
| Format | Liquid<br>Six bottles, one for<br>each of the six calibrator<br>levels. | Same |
| Matrix | Serum and heparinized plasma | Same |
| Storage | Store upright and refrigerate at 2 to<br>8°C. | Same |
| Stability<br>(unopened vial) | Stable until the expiration date<br>stated on the label when stored at 2<br>- 8°C. | Same |
| Calibration Stability | Stable up to 90 days | Same |
| Stability<br>(opened vial) | 24 hours (1 day) | Same |
{6}------------------------------------------------
| Characteristic | | Predicate |
|----------------|-------------------------------------------------------------|-------------------------------------------------------------|
| | Tosoh AIA-PACK PROG III<br>Calibrator Set | Tosoh AIA-PACK PROG<br>Calibrator Set (K933269) |
| Shelf-life | 8 months when stored unopened<br>and refrigerated at 2-8°C | 12 months when stored unopened<br>and refrigerated at 2-8°C |
| Traceability | USP (United States Pharmacopeia)<br>Standard (Lot #11J239). | Internal reference standards. |
| Base | Bovine serum | Human serum |
| Levels | Six<br>(0, 0.5, 1.5, 5.0, 15, 45<br>ng/mL approximately) | Six<br>(0, 0.5, 1.5, 5.0, 20, 60<br>ng/mL approximately) |
| Analyzer | Tosoh AIA-2000 | Tosoh AIA-1200 |
#### Differences
#### 8. SUMMARY OF STABILITY STUDIES
## Summary of Stability Studies
## Real Time Testing
ST AIA-PACK Progesterone III Calibrator Set were stored at refrigerated temperatures and assayed at 3, 6 and 9 months after the day of the first assay.
The acceptance criteria for recovery was within 100 +/- 10%.
The criterion for reproducibility (CV %) was </= 10%.
Current Real Time Studies support a 8 month Shelf life at 2-8°C.
## Open Vial Stability
Open vial stability of the ST AIA-PACK Progesterone III Calibrator Set was assessed by reconstituting the material according to the package insert. Samples were reconstituted and stored at refrigerated temperatures for 2 days and tested for progesterone.
The criterion for recovery was within 100 +/-10%.
The criterion for reproducibility (CV %) was </= 10%.
Current open vial studies support a reconstituted claim of 1 day when stored at 2-8°C.
#### SUMMARY OF VALUE ASSIGNMENT 9.
The ST AIA-PACK PROG III Calibrator Set contains assigned concentrations of progesterone. The assigned value is determined on a lot-by-lot basis and is designed to provide an assay calibration range of 0.1 to 40 ng/mL of progesterone. The calibrators for use with the ST AIA-PACK PROG III were referred to through the USP (United States Pharmacopeia) Standard (Lot #I1J239).
{7}------------------------------------------------
The primary reference material was prepared by diluting the USP reference material with calibrator base and the progesterone value as reference material was assigned gravimetrically.
The values of the product calibrator were assigned using the AIA-2000 instruments with the secondary reference material as calibrator. The values were verified by comparing measured results with those obtained with the previous lot for control materials.
The value assignment was determined by analyzing 5 replicates of the test calibrator on two Tosoh AIA-2000 instruments and the mean, SD and CV calculated for each lot before release to the user. The acceptance criteria states the precision measured by the CV should be within 10%.
| Progesterone<br>(Calibrator Levels) | N | Reference<br>Value<br>(ng/mL) | Mean<br>(ng/mL) | CV % |
|-------------------------------------|---|-------------------------------|-----------------|------|
| Cal (2) | 5 | 0.5 | .504 | 4.8 |
| Cal (3) | 5 | 1.5 | 1.53 | 2.2 |
| Cal (4) | 5 | 5.0 | 5.08 | 2.2 |
| Cal (5) | 5 | 15 | 15.0 | 2.0 |
| Cal (6) | 5 | 45 | 45.5 | 2.9 |
#### TRACEABILITY 10.
| ANALYTE | SUPPLIER | PRODUCT<br>NUMBER | ORIGIN | SOURCE |
|--------------|----------|-------------------|----------------------------------------------------------|----------------------|
| Progesterone | Sigma | 11J239 | Steroid stripped<br>and lipid<br>stripped human<br>serum | Chemical<br>Compound |
## 11. CONCLUSION
Testing results indicate that the proposed device is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.