← Product Code [IYN](/productcode/IYN) · K153408

# HM70A Diagnostic Ultrasound System (K153408)

_Samsung Medison Co., Ltd. · IYN · Dec 21, 2015 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K153408

## Device Facts

- **Applicant:** Samsung Medison Co., Ltd.
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Dec 21, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric, 3rd-Party Reviewed

## Indications for Use

The HM70A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-esophageal (non-Cardiac), Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal (Cardiac), Peripheral vessel.

## Device Story

HM70A is a hand-held, software-controlled diagnostic ultrasound system. It acquires ultrasound data via various transducers and displays it in modes including 2D, M-mode, Color/Power Doppler, Spectral Doppler, Harmonic imaging, 3D/4D, and Elastoscan. Used in clinical settings by healthcare professionals for anatomical measurement and diagnostic analysis. The system provides real-time acoustic output display (mechanical/thermal indices). It supports advanced features like Stress Echo, Strain, and Needle Mate (biopsy guidance). Output is displayed on-screen for clinician interpretation to aid in diagnosis and clinical decision-making. Benefits include portable diagnostic imaging and fluid flow analysis.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including acoustic output measurements (NEMA UD-2, UD-3), biocompatibility (ISO 10993-1), and electrical/electromagnetic safety (IEC 60601 series).

## Technological Characteristics

Hand-held ultrasound system; supports linear, curved, endocavity, phased array, and static probes (1.0–20.0 MHz). Energy source: ultrasonic. Connectivity: DICOM, Wireless LAN. Software-controlled with features like MultiVision (spatial compounding) and ClearVision (noise reduction). Biocompatibility per ISO 10993-1. Safety compliance: IEC 60601-1, IEC 60601-2-37.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- UGEO H70c Diagnostic Ultrasound System ([K130803](/device/K130803.md))
- RS80A Diagnostic Ultrasound System ([K151663](/device/K151663.md))
- EKO7 Diagnostic Ultrasound System ([K101455](/device/K101455.md))
- ACCUVIX XG Diagnostic Ultrasound System ([K103397](/device/K103397.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles forming the staff and a ribbon-like design representing the snakes.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 21, 2015

Samsung Medison Co., Ltd. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street. NW BUFFALO MN 55313

Re: K153408

Trade/Device Name: HM70A Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: November 23, 2015 Received: November 25, 2015

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

**Robert Oclo**

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

# K153408

Device Name

HM70A Diagnostic Ultrasound System

Indications for Use (Describe)

The HM70A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-esophageal (non-Cardiac), Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal (Cardiac), Peripheral vessel.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) No.: Device Name: HM70A Diagnostic Ultrasound System Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                               |  |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|-------------------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)              |  |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                               |  |
|                           | Fetal/Obstetrics (See Note 3) | N                                                 | N | N   | N   | N                 | Note 1               | Notes 2, 4, 7, 8, 9, 11       |  |
|                           | Abdominal (See Note 12)       | N                                                 | N | N   | N   | N                 | Note 1               | Notes 2, 4, 7, 8, 9, 10, 11   |  |
|                           | Intra-operative (See Note 6)  | N                                                 | N | N   | N   | N                 | Note 1               | Note 7, 8, 9, 11              |  |
|                           | Intra-operative (Neuro.)      | N                                                 | N | N   | N   | N                 | Note 1               | Note 7, 8, 9, 11              |  |
| Fetal Imaging<br>& Other  | Laparoscopic                  |                                                   |   |     |     |                   |                      |                               |  |
|                           | Pediatric                     | N                                                 | N | N   | N   | N                 | Note 1               | Note 2, 5, 6, 7, 8, 9, 11     |  |
|                           | Small Organ (See Note 5)      | N                                                 | N | N   | N   | N                 | Note 1               | Note 2, 5, 6, 7, 8, 9, 10, 11 |  |
|                           | Neonatal Cephalic             | N                                                 | N | N   | N   | N                 | Note 1               | Note 2, 7, 8, 9               |  |
|                           | Adult Cephalic                | N                                                 | N | N   | N   | N                 | Note 1               | Note 4, 7                     |  |
|                           | Trans-rectal                  | N                                                 | N | N   | N   | N                 | Note 1               | Note 2, 7, 8, 9, 10, 11       |  |
|                           | Trans-vaginal                 | N                                                 | N | N   | N   | N                 | Note 1               | Note 2, 7, 8, 9, 10, 11       |  |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                               |  |
|                           | Trans-esoph. (non-Cardiac)    | N                                                 | N | N   | N   | N                 | Note 1               | Note 4, 7                     |  |
|                           | Musculo-skel. (Convent.)      | N                                                 | N | N   | N   | N                 | Note 1               | Note 2, 5, 6, 7, 9, 11        |  |
|                           | Musculo-skel. (Superfic.)     | N                                                 | N | N   | N   | N                 | Note 1               | Note 2, 5, 6, 7, 9, 11        |  |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                               |  |
|                           | Other (spec.) (See Note 13)   | N                                                 | N | N   | N   | N                 | Note 1               | Note 2, 7, 8, 10, 11          |  |
| Cardiac                   | Cardiac Adult                 | N                                                 | N | N   | N   | N                 | Note 1               | Note 4, 7                     |  |
|                           | Cardiac Pediatric             | N                                                 | N | N   | N   | N                 | Note 1               | Note 4, 7                     |  |
|                           | Trans-esophageal (Cardiac)    | N                                                 | N | N   | N   | N                 | Note 1               | Note 4, 7                     |  |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                               |  |
| Peripheral<br>Vessel      | Peripheral vessel             | N                                                 | N | N   | N   | N                 | Note 1               | Note 2, 5, 6, 7, 8, 9, 11     |  |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                               |  |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+C+PW, B+C+PW, B+PD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Note10: ElastoScan

Note11: MultiVision (old name: Spatial Compound Imaging)

Note12: Includes Renal, Gynecology/Pelvis

Note13: Includes Urology/Prostate

{4}------------------------------------------------

# 510(k) No.:

### Device Name: SC1-6 for use with HM70A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               |   | Mode of Operation (*includes simultaneous B-mode) |     |     |                   |                      |                  |
|---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B | M                                                 | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                    |   |                                                   |     |     |                   |                      |                  |
|                           | Fetal/Obstetrics (See Note 3) | P | P                                                 | P   |     | P                 | Note 1               | Notes 4, 7, 8, 9 |
|                           | Abdominal (See Note 12)       | P | P                                                 | P   |     | P                 | Note 1               | Notes 7, 8, 9    |
|                           | Intra-operative (See Note 6)  |   |                                                   |     |     |                   |                      |                  |
|                           | Intra-operative (Neuro.)      |   |                                                   |     |     |                   |                      |                  |
| Fetal Imaging<br>& Other  | Laparoscopic                  |   |                                                   |     |     |                   |                      |                  |
|                           | Pediatric                     | P | P                                                 | P   |     | P                 | Note 1               | Notes 7, 8, 9    |
|                           | Small Organ (See Note 5)      |   |                                                   |     |     |                   |                      |                  |
|                           | Neonatal Cephalic             |   |                                                   |     |     |                   |                      |                  |
|                           | Adult Cephalic                |   |                                                   |     |     |                   |                      |                  |
|                           | Trans-rectal                  |   |                                                   |     |     |                   |                      |                  |
|                           | Trans-vaginal                 |   |                                                   |     |     |                   |                      |                  |
|                           | Trans-urethral                |   |                                                   |     |     |                   |                      |                  |
|                           | Trans-esoph. (non-Cardiac)    |   |                                                   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Convent.)      |   |                                                   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Superfic.)     |   |                                                   |     |     |                   |                      |                  |
|                           | Intra-luminal                 |   |                                                   |     |     |                   |                      |                  |
|                           | Other (spec.) (See Note 13)   |   |                                                   |     |     |                   |                      |                  |
|                           | Cardiac Adult                 |   |                                                   |     |     |                   |                      |                  |
| Cardiac                   | Cardiac Pediatric             |   |                                                   |     |     |                   |                      |                  |
|                           | Trans-esophageal (Cardiac)    |   |                                                   |     |     |                   |                      |                  |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                  |
| Peripheral<br>Vessel      | Peripheral vessel             |   |                                                   |     |     |                   |                      |                  |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                  |

N= new indication; P= previously cleared by FDA K130803; E= added under Appendix E

### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+C+PW, B+C+PW, B+PD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Note10: ElastoScan

Note11: MultiVision (old name: Spatial Compound Imaging)

Note12: Includes Renal, Gynecology/Pelvis

Note13: Includes Urology/Prostate

{5}------------------------------------------------

510(k) No.:

### Device Name: C2-6 for use with HM70A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               |   | Mode of Operation (*includes simultaneous B-mode) |     |     |                   |                      |                     |  |
|---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|---------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B | M                                                 | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)    |  |
| Ophthalmic                | Ophthalmic                    |   |                                                   |     |     |                   |                      |                     |  |
|                           | Fetal/Obstetrics (See Note 3) | P | P                                                 | P   | P   | P                 | Note 1               | Notes 2, 4, 7, 8, 9 |  |
|                           | Abdominal (See Note 12)       | P | P                                                 | P   | P   | P                 | Note 1               | Notes 2, 7, 8, 9    |  |
|                           | Intra-operative (See Note 6)  |   |                                                   |     |     |                   |                      |                     |  |
|                           | Intra-operative (Neuro.)      |   |                                                   |     |     |                   |                      |                     |  |
| Fetal Imaging<br>& Other  | Laparoscopic                  |   |                                                   |     |     |                   |                      |                     |  |
|                           | Pediatric                     | P | P                                                 | P   | P   | P                 | Note 1               | Notes 2, 7, 8, 9    |  |
|                           | Small Organ (See Note 5)      |   |                                                   |     |     |                   |                      |                     |  |
|                           | Neonatal Cephalic             |   |                                                   |     |     |                   |                      |                     |  |
|                           | Adult Cephalic                |   |                                                   |     |     |                   |                      |                     |  |
|                           | Trans-rectal                  |   |                                                   |     |     |                   |                      |                     |  |
|                           | Trans-vaginal                 |   |                                                   |     |     |                   |                      |                     |  |
|                           | Trans-urethral                |   |                                                   |     |     |                   |                      |                     |  |
|                           | Trans-esoph. (non-Cardiac)    |   |                                                   |     |     |                   |                      |                     |  |
|                           | Musculo-skel. (Convent.)      |   |                                                   |     |     |                   |                      |                     |  |
|                           | Musculo-skel. (Superfic.)     |   |                                                   |     |     |                   |                      |                     |  |
|                           | Intra-luminal                 |   |                                                   |     |     |                   |                      |                     |  |
|                           | Other (spec.) (See Note 13)   |   |                                                   |     |     |                   |                      |                     |  |
|                           | Cardiac Adult                 |   |                                                   |     |     |                   |                      |                     |  |
| Cardiac                   | Cardiac Pediatric             |   |                                                   |     |     |                   |                      |                     |  |
|                           | Trans-esophageal (Cardiac)    |   |                                                   |     |     |                   |                      |                     |  |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                     |  |
| Peripheral<br>Vessel      | Peripheral vessel             |   |                                                   |     |     |                   |                      |                     |  |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                     |  |

N= new indication; P= previously cleared by FDA K130803; E= added under Appendix E

### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+CW, B+C+PW, B+PD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Note10: ElastoScan

Note11: MultiVision (old name: Spatial Compound Imaging)

Note12: Includes Renal, Gynecology/Pelvis

Note13: Includes Urology/Prostate

Note 2: Includes imaging for guidance of biopsy

{6}------------------------------------------------

### 510(k) No.: Device Name: CF4-9 for use with HM70A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                  |  |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |  |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                  |  |
|                           | Fetal/Obstetrics (See Note 3) | P                                                 | P | P   | P   | P                 | Note 1               | Notes 2, 7, 8, 9 |  |
|                           | Abdominal (See Note 12)       | P                                                 | P | P   | P   | P                 | Note 1               | Notes 2, 7, 8, 9 |  |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                  |  |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                  |  |
| Fetal Imaging             | Laparoscopic                  |                                                   |   |     |     |                   |                      |                  |  |
| &<br>Other                | Pediatric                     | P                                                 | P | P   | P   | P                 | Note 1               | Notes 2, 7, 8, 9 |  |
|                           | Small Organ (See Note 5)      | P                                                 | P | P   | P   | P                 | Note 1               | Notes 2, 7, 8, 9 |  |
|                           | Neonatal Cephalic             | P                                                 | P | P   | P   | P                 | Note 1               | Notes 2, 7, 8, 9 |  |
|                           | Adult Cephalic                |                                                   |   |     |     |                   |                      |                  |  |
|                           | Trans-rectal                  |                                                   |   |     |     |                   |                      |                  |  |
|                           | Trans-vaginal                 |                                                   |   |     |     |                   |                      |                  |  |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                  |  |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                  |  |
|                           | Musculo-skel. (Convent.)      |                                                   |   |     |     |                   |                      |                  |  |
|                           | Musculo-skel. (Superfic.)     |                                                   |   |     |     |                   |                      |                  |  |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                  |  |
|                           | Other (spec.) (See Note 13)   |                                                   |   |     |     |                   |                      |                  |  |
|                           | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                  |  |
| Cardiac                   | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                  |  |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                  |  |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                  |  |
| Peripheral                | Peripheral vessel             | P                                                 | P | P   | P   | P                 | Note 1               | Notes 2, 7, 8, 9 |  |
| Vessel                    | Other (spec.)                 |                                                   |   |     |     |                   |                      |                  |  |

N= new indication; P= previously cleared by FDA K130803; E= added under Appendix E

### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+CW, B+C+PW, B+C+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Note10: ElastoScan

Note11: MultiVision (old name: Spatial Compound Imaging)

Note12: Includes Renal, Gynecology/Pelvis

Note13: Includes Urology/Prostate

Note 2: Includes imaging for guidance of biopsy

{7}------------------------------------------------

## 510(k) No.: Device Name: EVN4-9 for use with HM70A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                         |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|-------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)        |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                         |
|                           | Fetal/Obstetrics (See Note 3) | N                                                 | N | N   | N   | N                 | Note 1               | Note 2, 7, 11           |
|                           | Abdominal (See Note 12)       | N                                                 | N | N   | N   | N                 | Note 1               | Note 2, 7, 10, 11       |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                         |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                         |
| Fetal Imaging<br>& Other  | Laparoscopic                  |                                                   |   |     |     |                   |                      |                         |
|                           | Pediatric                     |                                                   |   |     |     |                   |                      |                         |
|                           | Small Organ (See Note 5)      |                                                   |   |     |     |                   |                      |                         |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                         |
|                           | Adult Cephalic                |                                                   |   |     |     |                   |                      |                         |
|                           | Trans-rectal                  | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 8, 9, 10, 11 |
|                           | Trans-vaginal                 | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 8, 9, 10, 11 |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                         |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                         |
|                           | Musculo-skel. (Convent.)      |                                                   |   |     |     |                   |                      |                         |
|                           | Musculo-skel. (Superfic.)     |                                                   |   |     |     |                   |                      |                         |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                         |
|                           | Other (spec.) (See Note 13)   | N                                                 | N | N   | N   | N                 | Note 1               | Note 2, 7, 10, 11       |
|                           | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                         |
| Cardiac                   | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                         |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                         |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                         |
| Peripheral<br>Vessel      | Peripheral vessel             |                                                   |   |     |     |                   |                      |                         |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                         |

N= new indication; P= previously cleared by FDA K130803; E= added under Appendix E

### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+CW, B+C+PW, B+PD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Note10: ElastoScan

Note11: MultiVision (old name: Spatial Compound Imaging)

Note12: Includes Renal, Gynecology/Pelvis

Note13: Includes Urology/Prostate

Note 2: Includes imaging for guidance of biopsy

{8}------------------------------------------------

510(k) No.:

### Device Name: L4-7 for use with HM70A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                        |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)       |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                        |
|                           | Fetal/Obstetrics (See Note 3) |                                                   |   |     |     |                   |                      |                        |
|                           | Abdominal (See Note 12)       | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11 |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                        |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                        |
| Fetal Imaging<br>& Other  | Laparoscopic                  |                                                   |   |     |     |                   |                      |                        |
|                           | Pediatric                     | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11 |
|                           | Small Organ (See Note 5)      | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11 |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                        |
|                           | Adult Cephalic                |                                                   |   |     |     |                   |                      |                        |
|                           | Trans-rectal                  |                                                   |   |     |     |                   |                      |                        |
|                           | Trans-vaginal                 |                                                   |   |     |     |                   |                      |                        |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                        |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                        |
|                           | Musculo-skel. (Convent.)      | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11 |
|                           | Musculo-skel. (Superfic.)     | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11 |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                        |
|                           | Other (spec.) (See Note 13)   |                                                   |   |     |     |                   |                      |                        |
|                           | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                        |
| Cardiac                   | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                        |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                        |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                        |
| Peripheral<br>Vessel      | Peripheral vessel             | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11 |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                        |

N= new indication; P= previously cleared by FDA K130803; E= added under Appendix E

### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+CW, B+C+PW, B+PD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Note10: ElastoScan

Note11: MultiVision (old name: Spatial Compound Imaging)

Note12: Includes Renal, Gynecology/Pelvis

Note13: Includes Urology/Prostate

Note 2: Includes imaging for guidance of biopsy

{9}------------------------------------------------

### 510(k) No.: Device Name: L5-13 for use with HM70A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                            |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|----------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)           |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                            |
|                           | Fetal/Obstetrics (See Note 3) |                                                   |   |     |     |                   |                      |                            |
|                           | Abdominal (See Note 12)       |                                                   |   |     |     |                   |                      |                            |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                            |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                            |
| Fetal Imaging<br>& Other  | Laparoscopic                  |                                                   |   |     |     |                   |                      |                            |
|                           | Pediatric                     | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11     |
|                           | Small Organ (See Note 5)      | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 5, 6, 7, 9, 10, 11 |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                            |
|                           | Adult Cephalic                |                                                   |   |     |     |                   |                      |                            |
|                           | Trans-rectal                  |                                                   |   |     |     |                   |                      |                            |
|                           | Trans-vaginal                 |                                                   |   |     |     |                   |                      |                            |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                            |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                            |
|                           | Musculo-skel. (Convent.)      | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11     |
|                           | Musculo-skel. (Superfic.)     | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11     |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                            |
|                           | Other (spec.) (See Note 13)   |                                                   |   |     |     |                   |                      |                            |
|                           | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                            |
| Cardiac                   | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                            |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                            |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                            |
| Peripheral<br>Vessel      | Peripheral vessel             | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11     |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                            |

N= new indication; P= previously cleared by FDA K130803; E= added under Appendix E

### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+CW, B+C+PW, B+C+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Note10: ElastoScan

Note11: MultiVision (old name: Spatial Compound Imaging)

Note12: Includes Renal, Gynecology/Pelvis

Note13: Includes Urology/Prostate

Note 2: Includes imaging for guidance of biopsy

{10}------------------------------------------------

### 510(k) No.: Device Name: L7-16 for use with HM70A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                        |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)       |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                        |
|                           | Fetal/Obstetrics (See Note 3) |                                                   |   |     |     |                   |                      |                        |
|                           | Abdominal (See Note 12)       |                                                   |   |     |     |                   |                      |                        |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                        |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                        |
| Fetal Imaging<br>& Other  | Laparoscopic                  |                                                   |   |     |     |                   |                      |                        |
|                           | Pediatric                     | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11 |
|                           | Small Organ (See Note 5)      | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11 |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                        |
|                           | Adult Cephalic                |                                                   |   |     |     |                   |                      |                        |
|                           | Trans-rectal                  |                                                   |   |     |     |                   |                      |                        |
|                           | Trans-vaginal                 |                                                   |   |     |     |                   |                      |                        |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                        |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                        |
|                           | Musculo-skel. (Convent.)      | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11 |
|                           | Musculo-skel. (Superfic.)     | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11 |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                        |
|                           | Other (spec.) (See Note 13)   |                                                   |   |     |     |                   |                      |                        |
|                           | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                        |
| Cardiac                   | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                        |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                        |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                        |
| Peripheral<br>Vessel      | Peripheral vessel             | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11 |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                        |

N= new indication; P= previously cleared by FDA K130803; E= added under Appendix E

### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+CW, B+C+PW, B+PD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Note10: ElastoScan

Note11: MultiVision (old name: Spatial Compound Imaging)

Note12: Includes Renal, Gynecology/Pelvis

Note13: Includes Urology/Prostate

Note 2: Includes imaging for guidance of biopsy

{11}------------------------------------------------

### 510(k) No.: Device Name: PE2-4 for use with HM70A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                  |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                  |
|                           | Fetal/Obstetrics (See Note 3) |                                                   |   |     |     |                   |                      |                  |
|                           | Abdominal (See Note 12)       | P                                                 | P | P   | P   | P                 | Note 1               | Note 4, 7        |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                  |
| Fetal Imaging<br>& Other  | Laparoscopic                  |                                                   |   |     |     |                   |                      |                  |
|                           | Pediatric                     |                                                   |   |     |     |                   |                      |                  |
|                           | Small Organ (See Note 5)      |                                                   |   |     |     |                   |                      |                  |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                  |
|                           | Adult Cephalic                | P                                                 | P | P   | P   | P                 | Note 1               | Note 4, 7        |
|                           | Trans-rectal                  |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-vaginal                 |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Convent.)      |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Superfic.)     |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.) (See Note 13)   |                                                   |   |     |     |                   |                      |                  |
|                           | Cardiac Adult                 | P                                                 | P | P   | P   | P                 | Note 1               | Note 4, 7        |
| Cardiac                   | Cardiac Pediatric             | P                                                 | P | P   | P   | P                 | Note 1               | Note 4, 7        |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                  |
| Peripheral<br>Vessel      | Peripheral vessel             |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                  |

N= new indication; P= previously cleared by FDA K130803; E= added under Appendix E

### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+CW, B+C+PW, B+C+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

- Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Note10: ElastoScan

Note11: MultiVision (old name: Spatial Compound Imaging)

Note12: Includes Renal, Gynecology/Pelvis

Note13: Includes Urology/Prostate

{12}------------------------------------------------

510(k) No.:

### Device Name: P3-8 for use with HM70A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                  |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                  |
| Fetal Imaging<br>& Other  | Fetal/Obstetrics (See Note 3) |                                                   |   |     |     |                   |                      |                  |
|                           | Abdominal (See Note 12)       | P                                                 | P | P   | P   | P                 | Note 1               | Note 4, 7        |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                  |
|                           | Laparoscopic                  |                                                   |   |     |     |                   |                      |                  |
|                           | Pediatric                     |                                                   |   |     |     |                   |                      |                  |
|                           | Small Organ (See Note 5)      |                                                   |   |     |     |                   |                      |                  |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                  |
|                           | Adult Cephalic                | P                                                 | P | P   | P   | P                 | Note 1               | Note 4, 7        |
|                           | Trans-rectal                  |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-vaginal                 |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Convent.)      |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Superfic.)     |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.) (See Note 13)   |                                                   |   |     |     |                   |                      |                  |
| Cardiac                   | Cardiac Adult                 | P                                                 | P | P   | P   | P                 | Note 1               | Note 4, 7        |
|                           | Cardiac Pediatric             | P                                                 | P | P   | P   | P                 | Note 1               | Note 4, 7        |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                  |
| Peripheral<br>Vessel      | Peripheral vessel             |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                  |

N= new indication; P= previously cleared by FDA K130803; E= added under Appendix E

### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+CW, B+C+PW, B+C+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Note10: ElastoScan

Note11: MultiVision (old name: Spatial Compound Imaging)

Note12: Includes Renal, Gynecology/Pelvis

Note13: Includes Urology/Prostate

Note 2: Includes imaging for guidance of biopsy

{13}------------------------------------------------

### 510(k) No.: Device Name: VN4-8 for use with HM70A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                  |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                  |
|                           | Fetal/Obstetrics (See Note 3) | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 8     |
|                           | Abdominal (See Note 12)       | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 8     |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                  |
| Fetal Imaging<br>& Other  | Laparoscopic                  |                                                   |   |     |     |                   |                      |                  |
|                           | Pediatric                     | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 8     |
|                           | Small Organ (See Note 5)      |                                                   |   |     |     |                   |                      |                  |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                  |
|                           | Adult Cephalic                |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-rectal                  |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-vaginal                 |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Convent.)      |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Superfic.)     |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.) (See Note 13)   |                                                   |   |     |     |                   |                      |                  |
|                           | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                  |
| Cardiac                   | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                  |
| Peripheral<br>Vessel      | Peripheral vessel             |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                  |

N= new indication; P= previously cleared by FDA K130803; E= added under Appendix E

### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+CW, B+C+PW, B+PD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Note10: ElastoScan

Note11: MultiVision (old name: Spatial Compound Imaging)

Note12: Includes Renal, Gynecology/Pelvis

Note13: Includes Urology/Prostate

Note 2: Includes imaging for guidance of biopsy

{14}------------------------------------------------

### 510(k) No.: Device Name: CW2.0 for use with HM70A Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                  |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                  |
| Fetal Imaging<br>& Other  | Fetal/Obstetrics (See Note 3) |                                                   |   |     |     |                   |                      |                  |
|                           | Abdominal (See Note 12)       |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                  |
|                           | Laparoscopic                  |                                                   |   |     |     |                   |                      |                  |
|                           | Pediatric                     |                                                   |   |     |     |                   |                      |                  |
|                           | Small Organ (See Note 5)      |                                                   |   |     |     |                   |                      |                  |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                  |
|                           | Adult Cephalic                |                                                   |   |     | P   |                   |                      |                  |
|                           | Trans-rectal                  |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-vaginal                 |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Convent.)      |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Superfic.)     |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.) (See Note 13)   |                                                   |   |     |     |                   |                      |…

---

**Source:** [https://fda.innolitics.com/device/K153408](https://fda.innolitics.com/device/K153408)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).