Elysiom Polychromatic LED Light Therapy (PLLT) System (Elysiom)
K153389 · In Light Wellness Systems, Inc. · ILY · Mar 30, 2016 · Physical Medicine
Device Facts
Record ID
K153389
Device Name
Elysiom Polychromatic LED Light Therapy (PLLT) System (Elysiom)
Applicant
In Light Wellness Systems, Inc.
Product Code
ILY · Physical Medicine
Decision Date
Mar 30, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
To provide LED (light-emitting diode) light associated gentle warmth (limited 1. temperature elevation) therapy to temporarily relieve minor pain, stiffness and muscle spasms; and To temporarily increase the local blood circulation. 2.
Device Story
Elysiom Polychromatic LED Light Therapy (PLLT) System is a pulsed LED device for topical use. It consists of a controller unit and flexible LED pads applied directly to the skin. The device emits light in red (600-660nm), blue (430-520nm), and infrared (750-910nm) spectra. A microprocessor-controlled system delivers a timed sequence of light pulses to provide gentle warmth. The device is used to temporarily relieve minor pain, stiffness, and muscle spasms, and to increase local blood circulation. It features an auto-shutoff timer and safety fuses. The system is designed for clinical or home use, operated by the user or a healthcare provider. The output is intended to provide symptomatic relief through localized thermal and light-based stimulation.
Clinical Evidence
Bench testing only. No clinical data provided. Performance was verified through electrical safety and electromagnetic compatibility testing per IEC 60601-1, IEC 60601-1-2, and IEC 62471:2008 standards. Software verification and validation were performed by a third party, confirming all requirements met with pass/fail criteria.
Technological Characteristics
Pulsed LED light therapy device. Wavelengths: 430-880nm (Red, Blue, IR). Power: 100-240 VAC input, 12VDC output. Features flexible pads with 50-264 LEDs. Safety: Resistor current limiting per row, timer, fuses, auto-shutoff. Connectivity: 5-pin DIN jack. Standards: IEC 60601-1 (Safety), IEC 60601-1-2 (EMC), IEC 62471:2008 (Photobiological).
Indications for Use
Indicated for temporary relief of minor pain, stiffness, and muscle spasms, and for temporary increase of local blood circulation in patients requiring topical light therapy.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 30, 2016
In Light Wellness Systems, Inc. % Mr. Fred Ma President And CEO Medical Quality International, LLC 7195 Longview Drive Cleveland, Ohio 44139
Re: K153389
Trade/Device Name: Elysiom Polychromatic LED Light Therapy System Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: ILY Dated: November 18, 2015 Received: November 23, 2015
Dear Mr. Ma:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K153389
Device Name
Elysiom TM Polychromatic LED Light Therapy System, In Light Wellness Systems, Inc.
Indications for Use (Describe)
To: 1) temporarily relieve minor pain, stiffness and muscle spasms; and 2) temporarily increase the local blood circulation.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
K153389 Page 1 of 4
#### 6.0 510(k) Summary
## Submission Application and Correspondence
| Name of Device: | Elysiom™ Polychromatic LED Light Therapy (PLLT)<br>System, In Light Wellness Systems, Inc. | |
|-------------------------------|--------------------------------------------------------------------------------------------|--|
| Trade/Proprietary/Model Name: | Elysiom™ Polychromatic LED Light Therapy (PLLT)<br>System (Elysiom™) | |
| Common or Usual Name: | LED Light Therapy Device | |
| Classification Names: | Physical Medicine Therapeutic Devices | |
#### Table 1. Devices to Which New Device is Substantially Equivalent
| Company | Device | K Number |
|-------------------------------------------|---------------------|----------|
| Bioremedi<br>Therapeutic Systems,<br>Inc. | HealthLight™ System | K101894 |
## Device Description
Elysiom™ Polychromatic LED Light Therapy (PLLT) System is a pulsed light emitting diode (LED) device providing a temporary increase in local circulation and temporary relief of minor pain. The LED lights provide gentle warmth. Elysiom™ uses a unique timed sequence of emissions, known as pulses, to create an environment in which change occurs regularly and rapidly.
## Intended Use of the Device
- To provide LED (light-emitting diode) light associated gentle warmth (limited 1. temperature elevation) therapy to temporarily relieve minor pain, stiffness and muscle spasms; and
- To temporarily increase the local blood circulation. 2.
{4}------------------------------------------------
| Product Name | HealthLight™ System | Elysium™ |
|----------------------------|--------------------------------------------------|---------------------------------------------|
| Method Of Use | Ready to Use | Ready to Use |
| US FDA Classification | CFR 21 | CFR 21 |
| Regulation Applied | Part 890.5500: Physical | Part 890.5500: Physical Medicine |
| | Medicine Therapeutic<br>Devices, Lamp, Infrared, | Therapeutic Devices, Lamp, |
| | Therapeutic Heating | Infrared, Therapeutic Heating |
| Device Code | ILY | ILY |
| Claim | Symptomatic Relief | Symptomatic Relief |
| Area of Use | All body parts | All body parts |
| Intended use | Minor pain, stiffness, | Minor pain, stiffness, muscle spasm |
| | muscle spasm and | and impaired circulation |
| | impaired circulation | |
| Type of Product | Topical Use Device | Topical Use Device |
| Presentation | LED Pads and controllers | LED Pads and controllers |
| Output Power<br>Increments | 12v DC | 12v DC |
| Output Port | 5 Pin Din Jack | 5 Pin Din Jack |
| Power Requirements | 100-240 VAC | 100-240 VAC |
| Increments of Power | 10.4-12 VDC | 10.4-12 VDC |
| Energy of Power | .5-5 Amps | .5-5 Amps |
| Beam Diameter | 5mm | 5mm |
| Number of LEDs | Varies per application<br>(50-132) | Varies per application (50-264) |
| Output of Blue LED | 430nm-520nm | 430nm-520nm |
| Output of Red LED | 600nm-660nm | 600nm-660nm |
| Output of IR LED | 750nm-910nm | 750nm-910nm |
| Pattern of LED | Red/Blue alternating IR<br>LEDs | Red/Blue alternating IR LEDs |
| Configuration of LEDs | Varies per application | Varies per application |
| Wave Length | 475nm, 640nm, 880nm | 430nm-880nm |
| Duty Cycle | 50.00% | 30.00-50.00% |
| Controls | Push button | Push button |
| Cooling Method | Resistor current limiter in | Resistor current limiter in each row |
| | each row of LEDs | of LEDs |
| Accessories | Power supply, goggles, | Power supply, goggles, sterile |
| | sterile sleeves | sleeves |
| Safety Features | Timer, fuses to prevent<br>shorting of pads | Timer, fuses to prevent shorting of<br>pads |
#### Table 2. Summary of Characteristics of the Device Compared to the Predicate
{5}------------------------------------------------
| Audio Warning Signal | Time out beep | Time out beep |
|--------------------------------------|----------------------------------|---------------|
| Recommended distance<br>from patient | Contact skin | Contact skin |
| Timer Modes | 5, 10, 15, 20, 25, 30<br>minutes | 20 minutes |
# 10. Substantial Equivalence Comparison :
HealthLight and Elysiom™ are both pulsed infrared devices using a separate controller to power flexible pads attached by a cable. HealthLight and Elysiom™ contain an array of light emitting diodes from the same LED supplier. HealthLight and Elysiom™ controllers are designed to power more than one pad at a time. HealthLight and Elysiom™ are powered by power supplies converting power to 12VDC. HealthLight and Elysiom™ both operate with a microprocessor, a timer and an auto shut-off feature. HealthLight and Elysiom™ both have designs to allow the controller to fail in the SAFE MODE, i.e. power down completely and shut off. HealthLight and Elysiom™ both use 5 pin DIN plugs that are removable from the device, as commonly is the case for electronic devices.
HealthLight offers a two, three or four port controller configurations while Elysiom™ offers a two and six port configurations. HealthLight offers seven flexible light therapy pads to use in combination with the controller, while Elysiom™ offers nine.
The differences identified above do not adversely impact the safety and effectiveness of the In Light Wellness Systems Elysiom™ device.
As a conclusion, the Elysiom™ PLLT System Device is substantially equivalent to the predicate devices, i.e. the BioRemedi HealthLight System, (K101894).
## Non-Clinical Performance Tests:
## 14. Software verification and validation:
Software was verified and validated by a third-party professional organization (Business & Decision, Hosting and Consulting Services). The actual results simply using Pass/Fail standards were utilized for the evaluation and they demonstrated as:
2/Port Reconciliation -All Actual Results matched the Expected Results. There were no deviations/failures. All tests passed indicating all 2/Port software requirements were met.
6/Port Reconciliation –All Actual Results matched the Expected Results. There were no deviations/failures. All tests passed indicating all 2/Port software requirements were met.
{6}------------------------------------------------
# 15. Electromagnetic Compatibility and Electrical Safety (IEC 60601-1 and IEC 60601-1-2)
ElysiomTM Polychromatic LED Light Therapy (PLLT) System has tested for IEC standards tests following FDA requirements and International Electronic Commission standards for Electronic Compatibility (IEC 6060-1-2) and Electronic Safety (IEC 6060-1-1, 3rd Edition/IEC 62471:2008 for Photobiological Safety of Lamps and Lamp System) by UL, LLC the one of most creditable world class Underwriters Laboratories located in Research Triangle Park, North Carolina (UL LLC, 12 Laboratory Drive Research Triangle Park, NC 27709 USA). The test results demonstrated that both 2/Ports and 6/Ports ElysiomTM Polychromatic LED Light Therapy (PLLT) System met IEC Electronic Safety and Electromagnetic Compatibility criteria as Polychromatic LED Light Therapy (PLLT) device coded under ILY by the US FDA. i.e. IEC 6060-1-1/IEC62471:2008 and IEC6060-1-2, 3rd Edition; while the risks for the device are extremely low and can always be mitigated.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.