K153384 · United Therapeutics Corporation · KDN · Feb 19, 2016 · Gastroenterology, Urology
Device Facts
Record ID
K153384
Device Name
Low Potassium Dextran Solution with Tris Diluent
Applicant
United Therapeutics Corporation
Product Code
KDN · Gastroenterology, Urology
Decision Date
Feb 19, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5880
Device Class
Class 2
Indications for Use
Low Potassium Dextran Solution with Tris Diluent is indicated for flushing, storage and transportation of isolated lungs after removal from the donor in preparation for eventual transplantation into a recipient.
Device Story
Low Potassium Dextran (LPD) Solution is a sterile, non-pyrogenic, physiological salt solution used for hypothermic flushing, storage, and transport of isolated human lungs. The solution is acidic and requires the addition of Tris Diluent to adjust the pH to 7.4 immediately prior to use. The device is intended for use by clinical professionals during the organ procurement and transplantation process. It serves to preserve the organ outside the body, maintaining viability for eventual transplantation. The system is supplied as single-use components in LDPE bottles.
Clinical Evidence
Bench testing only. Evidence includes chemical composition analysis, sterility (SAL 10^-6), pyrogenicity testing, stability testing (shelf life), and biocompatibility testing per ISO 10993-1. Additional testing included USP <161> conformance, pH adjustment verification, fill volume verification, mechanical requirements, and summative usability testing.
Technological Characteristics
Sterile, non-pyrogenic physiological salt solution. Supplied in LDPE bottles. Requires manual pH adjustment with Tris Diluent to 7.4. Biocompatibility per ISO 10993-1. Sterility Assurance Level (SAL) of 10^-6.
Indications for Use
Indicated for flushing, storage, and transportation of isolated human lungs post-donor removal, in preparation for transplantation into a recipient.
Regulatory Classification
Identification
An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name encircling a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to suggest community and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 7, 2016
United Therapeutics Corporation Rita Lee, Ph.D., RAC Senior Scientist, Regulatory Affairs 55 TW Alexander Drive, PO Box 14186 Research Triangle Park, NC 27709
Re: K153384
Trade/Device Name: Low Potassium Dextran Solution with Tris Diluent Regulation Number: 21 CFR§ 876.5880 Regulation Name: Isolated Kidney Perfusion and Transport System and Accessories Regulatory Class: II Product Code: KDN Dated: (Date on orig SE ltr): November 20, 2015 Received: (Date on orig SE ltr): November 23, 2015
Dear Rita Lee,
This letter corrects our substantially equivalent letter of February 19, 2016.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Rita Lee, Ph.D., RAC
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Image /page/1/Picture/8 description: The image shows the name "Benjamin R. Fisher -S" in a large, bold font. The text is black and appears to be centered on a white background. The letters are clearly legible, and the overall impression is clean and professional.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
## K153384
### Device Name
Low Potassium Dextran Solution with Tris Diluent
Indications for Use (Describe)
Low Potassium Dextran Solution with Tris Diluent is indicated for flushing, storage and transportation of isolated lungs after removal from the donor in preparation for eventual transplantation into a recipient.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------|-----------------------------------------------------------------------|
| <div align="center">Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div align="center">Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
| <div align="center">☒</div> | <div align="center">☐</div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
#### I. SUBMITTER
United Therapeutics Corporation 55 TW Alexander Drive Research Triangle Park, NC 27709, United States
Phone: 919-485-8350 Fax: 919-313-1298
Contact: Rita Lee, PhD RAC, Senior Scientist, Regulatory Affairs Date Prepared: November 19, 2015
#### II. DEVICE
Name of Device: Low Potassium Dextran Solution with Tris Diluent Common or Usual Name: Solution for lung preservation Classification Name: Isolated kidney perfusion and transport system and accessories (21 CFR §876.5880)
Regulatory Class: II Product Code: KDN
#### PREDICATE DEVICE III.
Perfadex® with THAM, K091989
#### IV. DEVICE DESCRIPTION
Low Potassium Dextran (LPD) Solution is a sterile, clear, non-pyrogenic, physiological salt solution for hypothermic flushing, storage, and transportation of human lungs outside the human body. LPD Solution is acidic and is adjusted shortly before use to pH 7.4 by the addition of Tris Diluent.
LPD Solution with Tris Diluent is supplied as 1 liter of LPD Solution in a low density polyethylene (LDPE) bottle (8 per pack), co-packed with one single-use bottle of Tris Diluent. Tris Diluent is also supplied separately in an eight-unit pack. Each container of LPD Solution and Tris Diluent is single-use.
#### INDICATIONS FOR USE / INTENDED USE V.
LPD Solution with Tris Diluent is indicated for flushing, storage and transportation of isolated lungs after removal from the donor in preparation for eventual transplantation into a recipient. The indications for use and intended use for LPD Solution with Tris Diluent are identical. Both the subject and predicate devices have indications for use statement and intended use.
{4}------------------------------------------------
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
LPD Solution with Tris Diluent is substantially equivalent to Perfadex with THAM. LPD Solution with Tris Diluent and Perfadex with THAM have identical intended use. Both LPD Solution and Perfadex are clear, sterile, non-pyrogenic solutions. Perfadex is supplied with THAM. which is added for pH adjustment before use. Similarly. LPD Solution is supplied with Tris Diluent, which is added for pH adjustment before use.
Differences in technological characteristics include packaging, expanded storage conditions, and a change to solution buffering instructions. All technological differences have been evaluated and do not raise any new questions of the safety or effectiveness when compared to the predicate. nor do they affect the safety and effectiveness when compared to the predicate.
#### PERFORMANCE DATA VII.
Chemical composition testing results have confirmed that LPD Solution and Tris Diluent have the same chemical composition as Perfadex and THAM, respectively.
Results of sterility and pyrogenicity testing on LPD Solution and Tris Diluent have shown that both solutions are sterile with a Sterility Assurance Level (SAL) of 106 and nonpyrogenic, similar to Perfadex and THAM. Stability testing, including analysis of chemical characteristics, particulate matter, and sterility assessment, has been conducted on LPD Solution and Tris Diluent to verify the shelf life. The results have verified the shelf life for LPD Solution and Tris Diluent, as manufactured and sterilized.
Biocompatibility test results have shown that LPD Solution and LPD Solution after pH adjustment by Tris Diluent meet the requirements of ISO 10993-1.
Results of testing for USP <161> conformance, pH adjustment of LPD Solution with Tris Diluent, verification of fill volume, mechanical requirements, and summative usability testing did not raise any new questions of the safety or effectiveness.
## VII. CONCLUSION
LPD Solution with Tris Diluent has the identical intended use compared to Perfadex with THAM. Evaluation of differences in technological characteristics did not raise any new questions of the safety or effectiveness of LPD Solution with Tris Diluent compared to the predicate device. Therefore, LPD Solution with Tris Diluent is substantially equivalent to Perfadex with THAM.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.