← Product Code [NUC](/productcode/NUC) · K153372

# Repagyn vaginal suppositories (K153372)

_Farma-Derma S.R.L. · NUC · Jun 30, 2016 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K153372

## Device Facts

- **Applicant:** Farma-Derma S.R.L.
- **Product Code:** [NUC](/productcode/NUC.md)
- **Decision Date:** Jun 30, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5300
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

Repagyn is a personal lubricant, for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms.

## Device Story

Repagyn Vaginal Suppositories are semi-solid, non-sterile preparations containing hyaluronic acid and semi-synthetic glycerides; packaged in PVC/PE blisters. Designed for vaginal application; inserted manually into vaginal cavity. Upon contact with vaginal mucosa, body temperature causes suppository to melt into viscous mass; provides lubrication and moisture. Used by patients for personal care; intended to enhance comfort during intimate sexual activity. Benefits include supplementation of natural lubrication. Not compatible with natural rubber latex, polyurethane, or polyisoprene condoms.

## Clinical Evidence

Bench testing only. Biocompatibility testing performed per ISO 10993 standards: acute systemic toxicity (ISO 10993-11:2006), cytotoxicity (ISO 10993-5:2009), vaginal irritation (ISO 10993-10:2010), and guinea pig maximization test (ISO 10993-10:2010). Endotoxin levels verified via USP <85>. Shelf life of 24 months established via real-time aging. Condom compatibility tested per ASTM D7661-10.

## Technological Characteristics

Hyaluronic acid and semi-synthetic glycerides formulation. 2g weight per suppository. PVC/PE blister packaging. Melting point corresponds to body temperature. Biocompatibility per ISO 10993. Condom compatibility tested per ASTM D7661-10.

## Regulatory Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

## Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

## Predicate Devices

- Hyalo Gyn ([K094039](/device/K094039.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of three stylized human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the largest and the bottom profile being the smallest.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 30, 2016

Farma-Derma s.r.1. Mara Calzolari Business Development Manager Via Dei Bersaglieri, 10 40010 Sala Bolognese Bologna, Italy

Re: K153372

> Trade/Device Name: Repagyn Vaginal Suppositories Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: Class II Product Code: NUC Dated: Mav 31, 2016 Received: June 2, 2016

Dear Mara Calzolari,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known)

K153372

Device Name

Repagyn vaginal suppositories

Indications for Use (Describe)

Reparyn is a personal lubricant, for vaginal application, intended to moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms.

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary for Repagyn Vaginal Suppositories

This 510(k) Summary is being submitted in accordance with the requirements of 21 C.F.R. 807.92.

#### General Information 1.

| Submitter:                 | FARMA-DERMA S.r.l.<br>Via dei Bersaglieri, 10<br>40010 Sala Bolognese (BO)<br>ITALY                                                                                                                                     |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person:            | Ms. Mara Calzolari<br>FARMA-DERMA srl<br>Via dei Bersaglieri 10<br>40010 Sala Bolognese (BO)<br>Italy<br>Mobile: +39 3486914895<br>Office: +39 051221606<br>Fax: +39 0516814833<br>Email: mara.calzolari@farmaderma.it  |
| Additional contact person: | Ms. Giulia Canzano<br>FARMA-DERMA srl<br>Via dei Bersaglieri 10<br>40010 Sala Bolognese (BO)<br>Italy<br>Mobile: +39 3486914892<br>Office: +39 0516814181<br>Fax: +39 0516814833<br>Email: giulia.canzano@farmaderma.it |
| Summary Preparation Date:  | November 19, 2015                                                                                                                                                                                                       |
| 2. Names                   |                                                                                                                                                                                                                         |
| Device Name:               | Repagyn Vaginal Suppositories                                                                                                                                                                                           |
| Regulation number:         | 21 C.F.R. 884.5300                                                                                                                                                                                                      |
| Classification:            | II                                                                                                                                                                                                                      |
| Product Code:              | NUC                                                                                                                                                                                                                     |
| Common Name:               | Lubricant Personal                                                                                                                                                                                                      |

{4}------------------------------------------------

#### 3. Predicate Devices

Repagyn Vaginal Suppositories are substantially equivalent to Hyalo Gyn, K094039, registered by Fidia Pharmaceuticals.

## 4. Device Description

The Farma-Derma Repagyn Vaginal Suppositories are a non-sterile hyaluronic acid and glycerides based semi-solid preparation presented in the form of vaginal suppositories of 2 g in weight individually packed in a polyvinyl chloride (PVC)/polyethylene (PE) blister. The device is available for sale in a pack containing 10 vaginal suppositories (two blisters of 5 vaginal suppositories each) while the sample pack contains 3 vaginal suppositories (one blister).

Due to its specific shape the device can be introduced into the vaginal cavity using a finger.

Repagyn Vaginal Suppositories are a Hyaluronic acid (HA) based product; the HA is mixed in with semi-synthetic glycerides that represent the major quantity of material used in the formulation.

When the vaginal suppository is in contact with the vaginal mucosa it melts due to body temperature and becomes a viscous mass which remains in contact with the vaginal mucosa.

It acts similarly to other vaginal creams or gels for topical application with Hyaluronic acid based formulations.

The specifications for Repagyn Vaginal Suppositories are described in Table 1.

| Specifications                                                        |                                           |
|-----------------------------------------------------------------------|-------------------------------------------|
| Appearance                                                            | Off-white opaque vaginal<br>suppositories |
| Disintegration time                                                   | According to Eu. Ph.                      |
| Average weight                                                        | 2 g                                       |
| Uniformity of weight                                                  | According to Eu. Ph.                      |
| Content of HA                                                         | 5 mg                                      |
| TAMC                                                                  | ≤100 UFC/g                                |
| TYMC                                                                  | ≤10 UFC/g                                 |
| Pseudomonas Aeruginosa,<br>Staphilococcus Aureus, Candida<br>Albicans | Absent/g                                  |

#### Table 1: Repagyn Vaginal Suppositories specifications

Furthermore the endotoxins level will be checked according to the kinetic test (LAL test) on each lot of finished product.

{5}------------------------------------------------

#### 5. Indications for Use

The intended use for Repagyn vaginal suppositories is:

"Repagyn is a personal lubricant, for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms."

#### 6. Technological Characteristics

comparison between the technological characteristics of Repagyn vaginal র suppositories and the predicate device Hyalo Gyn is provided in Table 2.

|                      | Repagyn Vaginal<br>Suppositories                                                       | Hyalo Gyn<br>(Predicate Device<br>K094039)                                                                            |
|----------------------|----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Base type            | Glycerides                                                                             | Water                                                                                                                 |
| Primary ingredient   | Glycerides                                                                             | Propylene glycol and water                                                                                            |
| Condom compatibility | Not compatible with natural<br>rubber latex, polyisoprene,<br>and polyurethane condoms | Compatible with<br>lubricated/non-lubricated<br>latex, lubricated<br>polyurethane, lubricated<br>natural skin condoms |
| Packaging            | PVC/PE blister of 3 or 5<br>vaginal suppositories                                      | Tube                                                                                                                  |

Table 2: Technological characteristics - Comparison between Repagyn vaginal suppositories and Hyalo Gyn

Repagyn Vaginal Suppositories and Hyalo Gyn differ in their shape, Repagyn is in the form of vaginal suppositories and Hyalo Gyn is an aqueous gel. However, when the suppositories are introduced in the vaginal cavity, they melt because of body temperature and become a viscous mass that act in the same way as the predicate device.

The components used for the predicate device are similar to the ones used for Repagyn vaginal suppositories.

The area of application for both medical devices is the vaginal mucosa. The only difference is that the predicate device Hyalo Gyn (K094039) can be applied also on the penis during sexual intercourses while Repagyn is designed for vaginal application only.

{6}------------------------------------------------

#### 7. Performance Data

#### Biocompatibility:

Biocompatibility studies including acute system toxicity, vaginal irritation, cytotoxicity, and guinea maximization sensitization were performed according to the ISO standards as described below:

- Acute Systemic Toxicity: ISO 10993-11:2006 ●
- Cytotoxicity: ISO 10993-5:2009
- Vaginal Irritation Testing: ISO 10993-10:2010 ●
- Guinea Pig Maximization Test (GPMT): ISO 10993-10:2010. ●
- Endotoxin levels determination: USP <85> ●

#### Shelf Life:

The shelf life of Repagyn vaginal suppositories is 24 months. This is based on the results of real time aging studies that demonstrated that the device maintains its specifications over the duration of its shelf life.

#### Condom Compatibility:

Condom compatibility was performed using the FDA recognized consensus standard ASTM D7661-10 Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms. The Repagyn Vaginal Suppositories were found to be not compatible with natural rubber latex, polyisoprene and polyurethane condoms.

#### 8. Conclusion

Repagyn Vaginal Suppositories are substantially equivalent to the predicate device, Hyalo Gyn.

---

**Source:** [https://fda.innolitics.com/device/K153372](https://fda.innolitics.com/device/K153372)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
