The Volcano s5i®/CORE™ and CORE™ Mobile Precision Guided Therapy System is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature. It is also indicated as an adjunct to conventional angiographic procedures to provide an image of vessel lumen and wall structures. ChromaFlo® is indicated for qualitative blood flow information from peripheral and coronary vasculature; flow information can be an adjunct to other methods of estimating blood flow and blood perfusion. The VH® IVUS System is intended to be used in conjunction with imaging catheters during diagnostic ultrasound imaging of the peripheral and coronary vasculature. The Volcano VH IVUS System is intended to semi-automatically visualize boundary features and perform spectral analysis of vascular features that the user may wish to examine more closely during routine diagnostic ultrasound imaging examinations. The pressure feature is intended for use in all blood vessels, including coronary and peripheral arteries, to measure intravascular blood pressure during diagnostic angiography and/or interventional procedures. The iFR® Modality of the s5/s5//CORE and CORE Mobile Precision Guided Therapy System is indicated in all blood vessels, including coronary and peripheral arteries, to measure intravascular blood pressure angiography and/or interventional procedures. The iFR® Modality is intended to be used in conjunction with currently marketed Volcano pressure wires. Rotational 45MHz feature is intended for the qualitative evaluation of vascular morphology in the coronary arteries and vasculature as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and the wall structures. The Pullback feature of the PIMr withdraws the imaging core within the protective sheath for a maximum of 15 cm.
Device Story
System uses IVUS catheters with distal transducers to emit/receive high-frequency sound waves; transforms acoustic impedance data into 360° cross-sectional grayscale images. Features include VH IVUS for semi-automatic border detection and plaque classification; ChromaFlo for color-mapped blood flow velocity; FFR/iFR modalities for pressure measurement using guide wires and aortic pressure data. Used in cath labs by physicians; CORE Control Pad provides touch-screen GUI for case navigation and measurements. iFR modality calculates pressure gradients during resting cardiac cycle without hyperemic agents; iFR Pullback generates pressure maps along vessel length. System aids clinical decision-making by visualizing vessel morphology, plaque composition, and hemodynamic significance of lesions.
Clinical Evidence
Bench testing only. Performance data included software verification and simulated use/usability validation to demonstrate equivalence to predicate devices.
Technological Characteristics
Ultrasonic pulsed echo imaging system; IVUS catheters with distal transducers; pressure guide wires; CORE Control Pad touch-screen accessory. Connectivity: CPU-based system with GUI. Software: Windows 7 OS. Features: VH IVUS (spectral analysis), ChromaFlo (color flow mapping), FFR/iFR pressure modalities.
Indications for Use
Indicated for patients requiring qualitative/quantitative evaluation of vascular morphology in coronary and peripheral arteries. Used as an adjunct to angiography for imaging vessel lumen/wall structures, blood flow/perfusion assessment, and intravascular blood pressure measurement during diagnostic or interventional procedures.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
CORE/CORE Mobile Precision Guided Therapy Systems (K133641)
s5/s5i Intravascular Ultrasound Imaging & Pressure Systems (K140291)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in a sans-serif font and is black. To the left of the text is a symbol that appears to be three lines that are curved.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 22, 2015
Volcano Corporation Brian Park Sr. Regulatory Affairs Specialist 3721 Valley Centre Drive San Diego, CA 92130
Re: K153369
Trade/Device Name: Volcano s5/s5i/CORE and CORE Mobile Precision Guided Therapy Systems Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic Pulsed Echo Imaging System Regulatory Class: Class II Product Code: IYO, DSK, DSA Dated: November 20, 2015 Received: November 23, 2015
Dear Brian Park:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mitchell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153669
#### Device Name
Volcano s5/s5i/CORE and CORE Mobile Precision Guided Therapy Systems
#### Indications for Use (Describe)
The Volcano s5i®/CORE™ and CORE™ Mobile Precision Guided Therapy System is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature. It is also indicated as an adjunct to conventional angiographic procedures to provide an image of vessel lumen and wall structures.
ChromaFlo® is indicated for qualitative blood flow information from peripheral and coronary vasculature; flow information can be an adjunct to other methods of estimating blood flow and blood perfusion.
The VH® IVUS System is intended to be used in conjunction with imaging catheters during diagnostic ultrasound imaging of the peripheral and coronary vasculature. The Volcano VH IVUS System is intended to semi-automatically visualize boundary features and perform spectral analysis of vascular features that the user may wish to examine more closely during routine diagnostic ultrasound imaging examinations.
The pressure feature is intended for use in all blood vessels, including coronary and peripheral arteries, to measure intravascular blood pressure during diagnostic angiography and/or interventional procedures.
The iFR® Modality of the s5/s5//CORE and CORE Mobile Precision Guided Therapy System is indicated in all blood vessels, including coronary and peripheral arteries, to measure intravascular blood pressure angiography and/or interventional procedures. The iFR® Modality is intended to be used in conjunction with currently marketed Volcano pressure wires.
Rotational 45MHz feature is intended for the qualitative evaluation of vascular morphology in the coronary arteries and vasculature as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and the wall structures. The Pullback feature of the PIMr withdraws the imaging core within the protective sheath for a maximum of 15 cm.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
| SPONSOR: | Volcano Corporation<br>3721 Valley Center Drive<br>San Diego, CA 92130 |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT/SUBMITTER: | Brian Park<br>Sr. Regulatory Affairs Specialist<br>Volcano Corporation<br>3721 Valley Center Drive<br>San Diego, CA 92130<br>Tel: (858) 720-4176<br>Fax: (858) 481-1027 |
| DATE PREPARED: | November 20, 2015 |
| DEVICE: | Volcano s5/s5i/CORE® and CORE® Mobile Precision<br>Guided Therapy Systems |
| TRADE NAME: | Volcano s5/s5i/CORE® and CORE® Mobile Precision<br>Guided Therapy Systems |
| COMMON NAME: | Ultrasonic Pulsed Echo Imaging System |
| CLASSIFICATION: | 892.1560, Ultrasonic Pulsed Echo Imaging System, II, IYO<br>870.1110, Blood Pressure Computer, II, DSK<br>870.2900, Patient Transducer and Electrical Cable, II, DSA |
| PREDICATE DEVICE: | CORE/CORE Mobile Precision Guided Therapy Systems<br>(K133641)<br>s5/s5i Intravascular Ultrasound Imaging & Pressure<br>Systems (K140291)<br>Volcano iFR Modality (K133323)<br>Volcano iFR Scout (K150441) |
| DEVICE DESCRIPTION: | The Volcano s5i®/CORE™ and CORE™ Mobile Precision<br>Guided Therapy System provides qualitative and<br>quantitative evaluation of vascular morphology in the<br>coronary arteries and vessels of the peripheral vasculature.<br>It is also indicated as an adjunct to conventional<br>angiographic procedures to provide an image of vessel<br>lumen and wall structures. Intravascular ultrasound (IVUS)<br>utilizes the acoustic impedance of vascular structures to |
{4}------------------------------------------------
provide cross sectional images from inside the vessel. The IVUS catheter uses a transducer near the distal tip to emit and receive high frequency sound waves.
The system is then able to analyze the signal that is received by the transducer to differentiate between vessel structures to produce a 360° cross sectional image.
These grayscale images can then be enhanced using VH IVUS. VH analysis provides automatic border detection for the vessel and lumen borders, as well as plaque composition. Plaque is automatically classified into four categories in order to simplify interpretation of the IVUS image
Alternatively, the ChromaFlo feature can be used to identify the blood flow. The ChromaFlo feature uses patented technology to provide a visual depiction of blood flow through the vessel. This is accomplished by overlaying a two-dimensional color mapping of relative blood flow velocity onto the grayscale ultrasound image.
In the FFR (or pressure) mode, the system acquires intraluminal data from a pressure guide wire while simultaneously taking aortic pressure data from the established ECG/EKG catheterization laboratory equipment. The system will measure pressure from the transducer on the guide wire both proximal and distal to a lesion and will calculate the fractional flow reserve (FFR).
The iFR® Modality is similar to the FFR Modality in that both modalities are capable of taking pressure measurements and both modalities have the identical indications for use and use the identical pressure wires. The primary difference is a new software algorithm that evaluates a portion of the cardiac wave cycle. The FFR Modality is calculated based on the entire cardiac wave cycle, whereas the iFR® Modality is calculated by isolation of the cardiac wave cycle where intracoronary resistance is naturally constant and minimized and where intracoronary flow is maximized. This results in the ability to measure pressure without administration of a hyperemic agent with the iFR® Modality whereas the FFR Modality is calculated after administration of a hyperemic agent.
{5}------------------------------------------------
iFR® Pullback and Live iFR are new features added to the iFR® Modality that will allow the assessment of a lesion (spot measurement) or vessel (pullback measurement) in a resting condition without the requirement of a hyperemic agent. The iFR® Pullback feature will allow the user to assess a length of vessel by placing the pressure sensor distally, beginning recording/measuring, and pulling the sensor back through the vessel to a stopping point. This generates a "map" of the wave-free pressure gradient iFR® values along the vessel and a distal iFR® value that represents the condition of the vessel at the most distal point of the recording.
The rotational mode is available on the system with the Revolution catheter connected to the system via the PIMr (SpinVision). Cross-sectional images from inside the vessel are displayed on the monitor allowing the user to make measurements.
The CORE™ Control Pad is a touch screen accessory that is a secondary controller in the Volcano s5i®/CORE™ and CORE™ Mobile Precision Guided Therapy System (integrated and mobile systems). Images, data, and case navigation controls are relayed to and from the CORE™ Control Pad display via the system central processing unit (CPU). These images, data and controls are presented in a graphical user interface (GUI) displayed on the touch screen of the CORE™ Control Pad. The user may use the GUI to navigate IVUS and FFR cases and to make measurements on intravascular ultrasound images that are presented on the primary system display.
The Volcano s5i®/CORE™ and CORE™ Mobile Precision INDICATIONS FOR USE: Guided Therapy System is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature. It is also indicated as an adjunct to conventional angiographic procedures to provide an image of vessel lumen and wall structures.
> ChromaFlo® is indicated for qualitative blood flow information from peripheral and coronary vasculature: flow information can be an adjunct to other methods of estimating blood flow and blood perfusion.
{6}------------------------------------------------
The VH® IVUS System is intended to be used in conjunction with imaging catheters during diagnostic ultrasound imaging of the peripheral and coronary vasculature. The Volcano VH IVUS System is intended to semi-automatically visualize boundary features and perform spectral analysis of RF ultrasound signals of vascular features that the user may wish to examine more closely during routine diagnostic ultrasound imaging examinations.
The pressure feature is intended for use in all blood vessels, including coronary and peripheral arteries, to measure intravascular blood pressure during diagnostic angiography and/or interventional procedures.
The iFR® Modality of the s5/s5i/CORE and CORE Mobile Precision Guided Therapy System is indicated in all blood vessels, including coronary and peripheral arteries, to measure intravascular blood pressure during diagnostic angiography and/or interventional procedures. The iFR® Modality is intended to be used in conjunction with currently marketed Volcano pressure wires.
Rotational 45MHz feature is intended for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vasculature as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and the wall structures. The Pullback feature of the PIMr withdraws the imaging core within the protective sheath for a maximum of 15 cm.
## COMPARISON OF CHARACTERISTICS:
Via this submission, Volcano proposes the following software changes:
- · Update the operating system of the Systems and CCP from Windows XP to Windows 7
- Update the software of the currently cleared CCP accessory to allow the user to use the CCP to switch the System easily between the FFR and iFR Modalities and activate iFR Scout on Systems that contain iFR and iFR Scout software
## PERFORMANCE DATA:
Performance testing completed for a determination of substantial equivalence included the following:
{7}------------------------------------------------
- . Software Verification
- . Simulated Use / Usability Validation
The results of the performance data demonstrate equivalence to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.