K153353 · Embla Systems · OLZ · Jun 23, 2016 · Neurology
Device Facts
Record ID
K153353
Device Name
SANDMAN ELITE
Applicant
Embla Systems
Product Code
OLZ · Neurology
Decision Date
Jun 23, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Attributes
Software as a Medical Device, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K153353 · Jun 23, 2016
SANDMAN ELITE
Embla Systems
Retrospective de-identified PSG sleep study records
Retrospective clinical PSG data were used to validate the performance of the device's computer-assisted scoring modules (respiratory, limb movement, and snore events) by comparing software output against a reference standard derived from expert manual scoring.
Retrospective validation study; Follow-up/Duration: > 411.99 hours of data
Adult patients (>18 years old) with a clinical indication for a sleep study; Sample Size: 56 subjects (54 evaluated for events)
Reference standard defined by majority rule of three expert PSG specialists
Positive Percent Agreement (PPA) and False Detection Rate per hour for event detection modules
Indications for Use
The Sandman Elite software is intended for Polysomnography studies on pediatric and adult patients, and allows recording, displaying, analysis, printing and storage of physiological signals to assist in the diagnosis of various sleep disorders. The Sandman software also allows: • Sleep report templates which summarize recorded and scored sleep data using simple measures including count, average, maximum and minimum values as well as data ranges for trended values; • An optional automated analysis of physiological signals that is intended for use only in adults; • An optional audio/visual alert for user defined threshold on calibrated DC input. These alerts are not intended for use as life support such as vital signs monitoring or continuous medical surveillance in intensive care units. The Sandman software is intended for use only by qualified and trained medical practitioners in research and clinical environments, who evaluate the software output with their clinical experience and judgment to provide diagnostic conclusions about the patient's condition.
Device Story
Sandman Elite is a desktop/laptop-based software application for sleep laboratories. It collects/digitizes physiological signals (EEG, EOG, EMG, ECG, airflow, respiratory effort, SpO2, body position, snoring) from external proprietary amplifiers. The software provides real-time data collection, display, storage, and analysis tools. It includes computer-assisted scoring modules for respiratory events (apnea/hypopnea), limb movements, and snoring, intended for adult data. Clinicians use the software to review, score, and generate reports. The system does not control life-sustaining functions or drug delivery. By automating event detection, the device assists trained practitioners in analyzing large volumes of polysomnography data, potentially improving diagnostic efficiency and consistency in sleep disorder evaluation.
Clinical Evidence
Clinical validation study of 56 adult PSG studies (54 subjects evaluated for events). Performance of assisted-scoring modules compared to a reference standard (majority rule of 3 expert scorers). Results (Mean PPA): Central apnea 95.1%, Mixed apnea 97.6%, Obstructive apnea 92.0%, Hypopnea 85.9%, Limb movement 86.0%, Snoring 82.0%. False detection rates per hour ranged from 0.8 (Mixed apnea) to 21.6 (Snoring). Study concludes performance is clinically equivalent to expert manual scoring and predicate devices.
Technological Characteristics
Software-only device running on standard desktop/laptop hardware. Collects digitized physiological signals via external proprietary amplifiers. Signal processing includes band-pass and notch filters (e.g., 10-100 Hz band-pass, 60 Hz notch). Connectivity: local or network server storage. Software modules utilize threshold-based and signal-processing algorithms for event detection. No patient-contacting materials or energy delivery components.
Indications for Use
Indicated for pediatric and adult patients undergoing polysomnography studies to assist in the diagnosis of sleep disorders. Automated analysis features are indicated for adult patients only.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 23, 2016
Embla Systems Shane Sawall Regulatory Affairs Manager 1 Hines Road Suite 202 Kanata, Ontario Canada K2K 3C7
Re: K153353
Trade/Device Name: Sandman Elite Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLZ Dated: Mav 20, 2016 Received: May 24, 2016
Dear Shane Sawall:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{2}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Hoffmann -A
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K153353
Device Name Sandman Elite
#### Indications for Use (Describe)
The Sandman Elite software is intended for Polysomnography studies on pediatric and allows recording, displaying, analysis, printing and storage of physiological signals to assist in the diagnosis of various sleep disorders. The Sandman software also allows:
- •Sleep report templates which summarize recorded and scored sleep data using simple measures including count, average, maximum and minimum values as well as data ranges for trended values;
- •An optional automated analysis of physiological signals that is intended for use only in adults;
- · An optional audio/visual alert for user defined threshold on calibrated DC input. These alerts are not intended for use as life support such as vital signs monitoring or continuous medical surveillance in intensive care units.
The Sandman software is intended for use only by qualified and trained medical practitioners in research and clinical environments, who evaluate the software output with their clinical experience and judgment to provide diagnostic conclusions about the patient's condition.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Natus Neurology. The word "natus" is in a teal sans-serif font, with a small registered trademark symbol to the right of the "s". Below "natus" is the word "neurology" in a smaller, black, italicized sans-serif font. The logo is simple and clean, with a focus on the company name.
# 510(K) Summary
Submitted by: Embla Systems 1 Hines Road Suite 202 Kanata, Ontario Canada K2K 3C7
Contact Person: Shane Sawall Regulatory Affairs Manager Tel.: (800) 356-0007 x8673 E-mail: shane.sawall@natus.com
Date Prepared: June 22, 2016
Proprietary Name: Sandman Elite
Common Name: Polysomnograph software
Classification Name: Automatic Event Detection Software for Polysomnograph with Electroencephalograph
Product code: OLZ
Device Class: II
Requlation Number: 21 CFR 882.1400
Predicate Device: Sandman (K934599); Natus SleepWorks (K090277)
#### Description:
#### 1. Overview Sandman Elite Software
The Sandman Elite Application is a software product that runs on a desktop or laptop computer and requires no specialized hardware. It is a Windows based application used at sleep labs by trained clinicians to investigate sleep disorders.
#### 2. Main Applications
The Sandman Elite application collects and digitizes the electrical voltages of patient physiological signals. After collecting and saving the signals, it provides tools and modules to analyze the signals, which aid in the interpretation of a sleep study. The software consists of four main applications:
{5}------------------------------------------------
- Collection Real time collection of EEG and other physiological ● parameters.
- . Analysis - Real time analysis of data to identify events which may require special attention from qualified medical professional in attendance.
- Data Management - Allows the user to copy, move, back up, and delete collected patient files
- . Configuration - Allows the user to change configuration settings such as your site information, the media used for file storage and to create and manage user preferences.
The Sandman Elite application also includes a separate independent Report Builder module. The Report Builder is a tool developed to assist the user in creating a customized report. All data are stored either locally or on a remote hard disk (network server). Provisions exist for archiving to several appropriate types of digital media, most frequently CD ROM.
## 3. Typical work flow using the Sandman Elite Application software
During the Collection phase, the software collects data based on user selected sampling rates, amplifier set-up, and amplifier calibrations. These three factors define how the software collects and displays the data in real-time. The recording is saved on the hard disk of the user's computer in a raw data format. The user can customize displays, observations, event markers, and tags.
## 4. Computer-assisted scoring modules
The Sandman Elite application software contains a number of computerassisted scoring modules. All computer assisted-scoring modules are provided to assist trained medical practitioners in the review and analysis of vast amounts of polysomnography data. The computer assisted-scoring modules are for analysis of adult patient data only. Each computer-assisted scoring module runs a specific type of event scoring on the patient file. The scoring rule parameters used in the computer-assisted scoring modules depend on the montage type associated with the study.
The ten computer-assisted scoring modules are:
- · Respiratory module
- · Desaturation module
- · pH module
- · ECG module
- · Associations module
- EtC02 module
- · PLM module
{6}------------------------------------------------
-
- Snoring module
- Bad Data module • XactTrace module
-
# 5. Diagnosis
The Sandman Elite application software is intended for use by qualified and trained medical practitioners in research and clinical environments. who evaluate the software output with their clinical experience and judgement to provide diagnostic conclusions about the patient's condition. The Sandman Elite software does not control the delivery of energy, the administration of drugs, or any form of life sustaining function to the patient.
# Indications for Use:
The Sandman Elite software is intended for Polysomnography studies on pediatric and adult patients, and allows recording, displaying, analysis, printing and storage of physiological signals to assist in the diagnosis of various sleep disorders. The Sandman software also allows:
- . Sleep report templates which summarize recorded and scored sleep data using simple measures including count, average, maximum and minimum values as well as data ranges for trended values;
- An optional automated analysis of physiological signals that is intended for use only in adults:
- An optional audio/visual alert for user defined threshold on calibrated DC ● input. These alerts are not intended for use as life support such as vital signs monitoring or continuous medical surveillance in intensive care units.
The Sandman software is intended for use only by qualified and trained medical practitioners in research and clinical environments, who evaluate the software output with their clinical experience and judgment to provide diagnostic conclusions about the patient's condition.
# Comparison to Predicate Device:
| | Predicate | Predicate | Subject Device |
|--------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <b>Sandman</b> K934599 | <b>SleepWorks</b> K090277 | <b>Sandman Elite</b> |
| Device Class | Class II | Class II | Class II |
| Class Name | Electroencephalograph | Electroencephalograph | Electroencephalograph |
| Product Code | OLV = Standard<br>Polysomnograph with<br>Electroencephalograph | OLZ = Automatic Event<br>Detection Software for<br>Polysomnograph with | OLZ = Automatic Event<br>Detection Software for<br>Polysomnograph with |
| | Predicate | Predicate | Subject Device |
| | Sandman K934599 | SleepWorks K090277 | Sandman Elite |
| | | Electroencephalograph | Electroencephalograph |
| Intended User<br>Indications for<br>Use | Medical Professional<br>The SANDMAN<br>provides paperless<br>digital recording and<br>handling of<br>physiological signals<br>intended for a sleep<br>EEG laboratory. | Medical Professional<br>The Sleepworks software<br>works in conjunction with<br>Connex, Trex or Netlink<br>amplifiers intended for<br>polysomnography studies.<br>The software allows<br>recording, displaying,<br>analysis, printing and<br>storage of physiological<br>signals to assist in the<br>diagnosis of various<br>sleep disorders and sleep<br>related respiratory<br>disorders.<br>The Sleepworks allows:<br>Automated analysis of<br>physiological signals that<br>is intended for use only in<br>adults.<br>An optional Audio /<br>visual alert for user<br>defined threshold on<br>calibrated DC input.<br>These alerts are not<br>intended for use as life<br>support such as vital<br>signs monitoring<br>or continuous medical<br>surveillance in intensive<br>care units.<br>Sleep report templates<br>are provided which<br>summarize recorded and<br>scored sleep<br>data using simple<br>measures including count,<br>average, maximum and<br>minimum<br>values as well as data<br>ranges for trended values;<br>Sleep Works software<br>does not provide any<br>diagnostic conclusion | Medical Professional<br>The Sandman Elite<br>software is intended for<br>Polysomnography<br>studies on pediatric and<br>adult patients, and<br>allows recording,<br>displaying, analysis,<br>printing and storage of<br>physiological signals to<br>assist in the diagnosis<br>of various sleep<br>disorders. The<br>Sandman software also<br>allows:<br><br>Sleep report<br>templates which<br>summarize recorded<br>and scored sleep data<br>using simple measures<br>including count,<br>average, maximum and<br>minimum values as well<br>as data ranges for<br>trended values;<br><br>An optional<br>automated analysis of<br>physiological signals<br>that is intended for use<br>only in adults;<br><br>An optional<br>audio/visual alert for<br>user defined threshold<br>on calibrated DC input.<br>These alerts are not<br>intended for use as life<br>support such as vital<br>signs monitoring or<br>continuous medical<br>surveillance in intensive<br>care units. The |
| | Predicate | Predicate | Subject Device |
| | Sandman K934599 | SleepWorks K090277 | Sandman Elite |
| | | condition and is intended<br>to be used only by<br>qualified and trained<br>medical practitioners;<br>in research and clinical<br>environments. | intended for use only by<br>qualified and trained<br>medical practitioners in<br>research and clinical<br>environments, who<br>evaluate the software<br>output with their clinical<br>experience and<br>judgment to provide<br>diagnostic conclusions<br>about the patient's<br>condition. |
| User input | Mouse/keyboard | Mouse/keyboard | Mouse/keyboard |
| Acquire,<br>display, store,<br>and archive<br>PSG data | Yes | Yes | Yes |
| Signal digitized | Amplifier included as<br>part of the system | By separate proprietary<br>amplifier | By separate proprietary<br>amplifier |
| Software<br>Modules | | | |
| Respiratory<br>event marking | Yes (Manual) | Yes (Manual & Computer<br>Assisted) | Yes (Manual &<br>Computer Assisted) |
| Sleep staging/<br>scoring | Yes. Manual | Yes. Manual/computer<br>assisted | Yes. Manual |
| Arousal Event<br>Marking | Yes (Manual) | Yes. Manual/computer<br>assisted | Yes. Manual |
| Limb<br>movements<br>event marking | Yes (Manual) | Yes (Manual & Computer<br>Assisted) | Yes (Manual &<br>Computer Assisted) |
| Snore event<br>marking | Yes (Manual) | Yes (Manual & Computer<br>Assisted) | Yes (Manual &<br>Computer Assisted) |
| Oxygen<br>Desaturation<br>event marking | Yes (Manual) | Yes (Manual & Computer<br>Assisted) | Yes (Manual &<br>Computer Assisted) |
| Heart Rate data<br>trend &<br>summary | No | Yes (Manual & Computer<br>Assisted) | Yes (Manual &<br>Computer Assisted) |
| PH data trend<br>and summary | No | Yes (Manual & Computer<br>Assisted) | Yes (Manual &<br>Computer Assisted) |
| EtCO2 data<br>trend &<br>summary | No | Yes (Manual & Computer<br>Assisted) | Yes (Manual &<br>Computer Assisted) |
| Associate<br>related events | No | Yes (Manual & Computer<br>Assisted) | Yes (Manual &<br>Computer Assisted) |
| | Predicate | Predicate | Subject Device |
| | Sandman K934599 | SleepWorks K090277 | Sandman Elite |
| data "Bad Data" | | Assisted) | Computer Assisted)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.