← Product Code [LNH](/productcode/LNH) · K153343

# MAGNETOM Aera, MAGNETOM Skyra, MAGNETOM Prisma, MAGNETOM Prisma fit (K153343)

_Siemens Medi Cal Solutions, Inc. · LNH · Apr 15, 2016 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K153343

## Device Facts

- **Applicant:** Siemens Medi Cal Solutions, Inc.
- **Product Code:** [LNH](/productcode/LNH.md)
- **Decision Date:** Apr 15, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The MAGNETOM systems [MAGNETOM Aera, MAGNETOM Skyra, MAGNETOM Prisma and MAGNETOM Prisma™] are indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra, when interpreted by a trained physician, yield information that may assist in diagnosis. The MAGNETOM systems may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room display and MR-Safe biopsy needles.

## Device Story

syngo MR E11C software update for MAGNETOM Aera, Skyra, and Prisma/Prismafit MRI systems. Inputs: RF signals/gradient data from patient anatomy. Processing: Reconstruction of transverse, sagittal, coronal, oblique images/spectra; includes new sequences (qDWI, FAST_TSE) and features (SMS EPI, GOBrain, Auto Bolus Detection). Output: Diagnostic images/spectra displayed for physician interpretation. Used in clinical settings by trained professionals. Enhances workflow/acquisition speed (e.g., GOBrain) and image quality (e.g., fat saturation, segmentation). Supports interventional procedures with MR-safe devices. Benefits: Improved diagnostic information, reduced acoustic noise, and optimized workflow for brain/breast/abdominal imaging.

## Clinical Evidence

No clinical trials conducted. Evidence consists of bench testing, including acoustic noise measurements for quiet sequences, software verification/validation per FDA guidance, and sample clinical images demonstrating performance of new/modified sequences and coils.

## Technological Characteristics

MRI system software (syngo MR E11C). Conforms to IEC 62304, IEC 60601-1 series, and NEMA standards. Features include gradient-based imaging, multi-channel head coils (20, 32, 64 channel), and sequence-based signal processing. Hardware includes cosmetic front panel updates and specific coil configurations for Skyra-24.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- syngo MR E11B for MAGNETOM Aera (24) (1.5T), MAGNETOM Avantofit (1.5T), MAGNETOM Skyrafit (3T) and MAGNETOM Prisma/ Prismafit (3T) ([K151597](/device/K151597.md))
- Software syngo MR E11A for the MAGNETOM systems Aera/Skyra ([K141977](/device/K141977.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around a symbol. The symbol consists of three stylized human profiles facing to the right, with flowing lines connecting them. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 15, 2016

Siemens Medical Solutions, Inc. % Mr. John Urtz Regulatory Affairs Specialist 40 Liberty Boulevard Mail Code 65-1A MALVERN PA 19355

Re: K153343

Trade/Device Name: MAGNETOM Aera, MAGNETOM Skyra, MAGNETOM Prisma, MAGNETOM Prismafit Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: March 14, 2016 Received: March 14, 2016

Dear Mr. Urtz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D. O'Hara

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

For

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known)

k153343

Device Name

MAGNETOM Aera, MAGNETOM Skyra, MAGNETOM Prisma/Prismafit

#### Indications for Use (Describe)

The MAGNETOM systems are indicated for use as magnetic resonance diagnostic devices (MRDD) that produce transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that display the internal structure and/or function of the head, body or extremities.

Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician, yield information that may assist in diagnosis.

The MAGNETOM systems described above may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room display and MR-Safe biopsy needles.

Type of Use (Select one or both, as applicable)

| <input checked="true" type="checkbox"/> Prescription Use (Part 84 CFR 201 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 201 Subpart C) |
|--------------------------------------------------------------------------------------|----------------------------------------------------------------------|
|--------------------------------------------------------------------------------------|----------------------------------------------------------------------|

|× Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The color of the text is a bright teal. The letters are evenly spaced and the word is horizontally oriented.

# 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92.

- I. General Information

| Establishment  | Siemens Medical Solutions USA. Inc.<br>40 Liberty Boulevard<br>Mail Code 65-1A<br>Malvern, PA 19355, USA<br>Registration Number: 2240869                                                                                                                                                                                     |
|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared  | April 13, 2016                                                                                                                                                                                                                                                                                                               |
| Manufacturer   | Siemens AG / Siemens Healthcare GmbH<br>Henkestrasse 127<br>Erlangen Bayern, Germany 91052<br>Registration Number: 3002808157<br>SIEMENS SHENZHEN MAGNETIC RESONANCE LTD.<br>Siemens MRI Center<br>Hi-Tech Industrial park (middle)<br>Gaoxin C. Ave., 2nd<br>Shenzhen 518057, P.R. CHINA<br>Registration Number: 3004754211 |
| Contact Person | Mr. John Urtz<br>Regulatory Affairs Specialist<br>Siemens Healthcare<br>Siemens Medical Solutions USA, Inc.<br>Customer Solutions Group<br>40 Liberty Boulevard<br>Mail Code 65-1A<br>Malvern, PA 19355, USA<br>Phone: (610) 448-6002<br>Fax: (610) 640-4481                                                                 |
| Device Name    | syngo MR E11C software for MAGNETOM Aera,<br>MAGNETOM Skyra and MAGNETOM Prisma/Prismafit MR<br>Systems                                                                                                                                                                                                                      |

#### Siemens 510(k) Traditional Premarket Notification

April 13, 2016

syngo MR E11C Software for Siemens MR Systems: MAGNETOM Aera (1.5T), MAGNETOM Skyra (3T) and MAGNETOM Prisma/ Prismafit (3T)

{4}------------------------------------------------

Trade Names: MAGNETOM Aera MAGNETOM Skyra MAGNETOM Prisma MAGNETOM Prisma™

| Classification Name:  | Magnetic Resonance Diagnostic Device (MRDD) |
|-----------------------|---------------------------------------------|
| Classification Panel: | Radiology                                   |
| CFR Code:             | 21 CFR § 892.1000                           |
| Classification:       | Class II                                    |
| Product Code:         | Primary: LNH, Secondary: LNI, MOS           |

#### II. Safety and Effectiveness Information Supporting Substantial Equivalence

#### Intended Use

The indications for use for the subject device are the same as the predicate device and are as follows:

The MAGNETOM systems [MAGNETOM Aera, MAGNETOM Skyra, MAGNETOM Prisma and MAGNETOM Prisma™] are indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities.

Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra, when interpreted by a trained physician, yield information that may assist in diagnosis.

The MAGNETOM systems may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room display and MR-Safe biopsy needles.

#### Device Description

The subject device, syngo MR E11C system software, is being made available for the following MAGNETOM MR Systems:

- MAGNETOM Aera,
- MAGNETOM Skyra, ●
- MAGNETOM Prisma and
- MAGNETOM Prisma™ ●

#### Siemens 510(k) Traditional Premarket Notification

April 13, 2016

syngo MR E11C Software for Siemens MR Systems: MAGNETOM Aera (1.5T), MAGNETOM Skyra (3T) and MAGNETOM Prisma/ Prismafit (3T)

{5}------------------------------------------------

## SIEMENS

The syngo MR E11C SW includes new sequences. new features and minor modifications of already existing features. A high level summary of the new sequences and features is included below. Further, modified and migrated sequences and features are also listed.

#### New Sequences

- qDWI a new Quiet Diffusion Sequence for Brain Imaging .
A noise reduced sequence for diffusion weighted imaging (called "qDWI") is provided with syngo MR E11C. The sequence is technically based on a readout segmented EPI sequence (RESOLVE, currently available on the predicate device software). Gradients have been stretched to minimize the slew rate in order to reduce acoustic noise.

- Improvements in BLADE Imaging (fast TSE) .
With this software version, an optimized variant of the TSE for abdominal imaging is released and is called FAST_TSE. The main characteristics of FAST TSE are the same as for TSE which is one of the most basic sequences available on all MAGNETOM Systems.

#### New Features

- SMS EPI a new feature for Brain Diffusion and BOLD Imaging . In Simultaneous Multi Slice imaging several slices are excited simultaneously and separated during image reconstruction. The feature is intended for EPI diffusion and EPI BOLD brain imaging.
- . GOBrain – a new feature which supports brain examination in short acquisition time for MAGNETOM Aera/Skyra

With syngo MR E11C a complete brain examination in short acquisition time is provided for MAGNETOM Aera and MAGNETOM Skyra systems with a 20, 32, or 64 channel head coil. The examination consists of a localizer, T1w, T2w, dark fluid (FLAIR), DWI, and T2*w protocol.

- New features for Breast Dot Engine Dynamic Breast Evaluation and . Auto Bolus Detection
With synqo MR VD13A (clearance K133435) the feature 'Breast Dot Engine', a workflow solution for breast MRI, was cleared. Following the FDA clearance of the contrast agent Gadavist® for breast MRI, this software version adds the Automatic Bolus Detection feature for support of contrast exams.

#### Modified Sequences or Features

- । SPAIR fat sat improvements (applicable for 3T MR Systems addressed by syngo MR E11C)
- । Improvements in Liver Seqmentation
- Improvement in Multi Echo Dixon for Fat Iron Quantification

April 13, 2016

syngo MR E11C Software for Systems: MAGNETOM Aera (1.5T), MAGNETOM Skyra (3T) and MAGNETOM Prisma/ Prismafit (3T)

{6}------------------------------------------------

## SIEMENS

- A flow compensated SE sequence for 3T systems (Skyra and । Prisma/Prisma™) only
- । Improvements in Inline Ventricular function
- Improvements in VIBE ।

### Migrated Features

- QISS, MyoMaps and LiverLab for MAGNETOM Skyra 24 only
- The Body 18 and Body 18 long coil will be made available for MAGNETOM Skyra (24-channel configuration only ("Skyra-24")). These coils are cleared on other Skyra systems with the primary predicate device with exception of Skyra-24.

#### Other Modifications

- Front cover panel refresh for MAGNETOM Aera and MAGNETOM Skyra
- Provide secure MR scanner setup for DoD (Department of Defense) Information Assurance compliance.

### Technological Characteristics

syngo MR E11C for MAGNETOM Aera MAGNETOM Skyra, Software MAGNETOM Prisma and MAGNETOM Prismall systems ("syngo MR E11C") has the same technological characteristics as the predicate devices MR systems (K151579; cleared September 29, 2015 and K141977, cleared November 19, 2014).

The subject device is substantially equivalent to the predicate devices with regard to the operational environment, programming language, operating system and performance.

syngo MR E11C SW conforms to the standard for software medical devices (IEC 62304:2006) and IEC as well as NEMA standards.

While syngo MR E11C SW offers new and modified SW features, modified SW features with respect to the predicate device systems, the syngo MR E11C software on the subject device MR Systems has the same technological characteristics as the predicate device systems. Further, this submission includes hardware modifications including new coils for the Skyra-24 system (already cleared with other Skyra systems) and a cosmetic modification to the front cover panel for Aera and Skyra systems. These do not represent a change in technological characteristics.

#### Nonclinical Tests

The following performance testing was conducted on the subject device

- Sample clinical images were taken for particular new and modified sequences. ●
- . Acoustic noise measurements were performed for quiet sequences

syngo MR E11C Software for Systems: MAGNETOM Aera (1.5T), MAGNETOM Skyra (3T) and MAGNETOM Prisma/ Prismafit (3T)

{7}------------------------------------------------

- . Image quality assessments of all new/modified sequences and algorithms, were completed.
- Software verification and validation testing was completed in accordance with the ● FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"

The results from each set of tests demonstrate that the device performs as intended and is thus substantially equivalent to the predicate devices to which it has been compared.

#### Clinical Tests

No clinical tests were conducted to support the subject device and the substantial equivalence argument; however, clinical images were provided to support the new coils as well as the new software features of the subject device.

#### Safety and Effectiveness

The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.

Risk management is ensured via a risk analysis in compliance with ISO 14971:2007 to identify and provide mitigation to potential hazards in a risk analysis beginning early in the design phase and continuing throughout the development of the product. These risks are controlled via measures realized in software development, SW testing and product labeling. To minimize risks, Siemens adheres to recognized and established industry practices and standards, such as the IEC 60601-1 series, to minimize electrical and mechanical risk. Furthermore, the operators are healthcare professionals familiar with and responsible for the acquisition and post processing of magnetic resonance images.

The syngo MR E11C software for the MAGNETOM Aera, MAGNETOM Skyra, MAGNETOM Prisma and MAGNETOM Prismall systems, conforms to the applicable FDA recognized and international IEC, ISO and NEMA standards with regards to performance and safety as recommended by the respective MR FDA Guidance Document.

#### Substantial Equivalence

E11C SW for the MAGNETOM Aera, MAGNETOM Skyra, svnao MR MAGNETOM Prisma and MAGNETOM Prisma™ Systems, includes all of the features of the primary predicate device and additional sequences and features as noted above.

syngo MR E11C Software for Siemens MR Systems: MAGNETOM Aera (1.5T), MAGNETOM Skyra (3T) and MAGNETOM Prisma/ Prismafit (3T)

{8}------------------------------------------------

### Predicate Device Information

| Primary Predicate Device                                                                                                                       | FDA Clearance<br>Number and Date          | Product<br>code | Manufacturer                                  |
|------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|-----------------|-----------------------------------------------|
| syngo MR E11B for MAGNETOM<br>Aera (24) (1.5T), MAGNETOM<br>Avantofit (1.5T), MAGNETOM<br>Skyrafit (3T) and MAGNETOM<br>Prisma/ Prismafit (3T) | K151597, cleared<br>September 29,<br>2015 | LNH<br>LNI,MOS  | Siemens AG /<br>Siemens<br>Healthcare<br>GmbH |
| Secondary Predicate Device                                                                                                                     | FDA Clearance<br>Number and Date          | Product<br>code | Manufacturer                                  |
| Software syngo MR E11A for the<br>MAGNETOM systems<br>Aera/Skyra                                                                               | K141977 cleared<br>November 19,<br>2014   | LNH<br>LNI,MOS  | Siemens AG /<br>Siemens<br>Healthcare<br>GmbH |

#### Conclusion as to Substantial Equivalence

syngo MR E11C software for the MAGNETOM Aera, MAGNETOM Skyra, MAGNETOM Prisma and MAGNETOM Prismal" Systems has the same intended use and the same basic technological characteristics as the predicate device systems, MAGNETOM Aera, MAGNETOM Skyra, MAGNETOM Prisma and MAGNETOM Prismat" Systems with syngo MR E11B, with respect to the magnetic resonance features and functionalities. While there are some minor technical features that vary with respect to the predicate device MR Systems, the conclusions from the non-clinical data suggest that the features with different technological characteristics from the predicate devices bear an equivalent safety and performance profile as that of the predicate and secondary predicate devices. Therefore the subject device is substantially equivalent to the predicate device.

syngo MR E11C Software for Siemens MR Systems: MAGNETOM Aera (1.5T), MAGNETOM Skyra (3T) and MAGNETOM Prisma/ Prismafit (3T)

---

**Source:** [https://fda.innolitics.com/device/K153343](https://fda.innolitics.com/device/K153343)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
