The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Device Story
SUNCO Mechanical Wheelchair (model SKW-9003) is a manual, foldable, four-wheeled indoor/outdoor mobility device. It features a sling-style seat and backrest, removable desk-length armrests, and swing-away detachable footrests. Operated by the user or a caregiver, the device provides mobility for individuals unable to walk. It is designed for manual propulsion. The device is not intended for use on sand, wet/icy surfaces, steep inclines (>10 degrees), or escalators. It includes a curb stepper for minor obstacles. The device is constructed with resistance-ignitability fabric upholstery and a metal frame. It is intended for use in home or clinical environments to assist with patient transport and mobility.
Clinical Evidence
Bench testing only. No clinical data provided. Device safety and effectiveness were demonstrated through biocompatibility testing (ISO 10993-5, -10) and mechanical performance testing per ISO 7176 series standards (stability, braking, dimensions, strength, and ignition resistance).
Technological Characteristics
Manual wheelchair; foldable metal frame; sling-style seat/backrest; PVC solid casters (7.9"x1"); PU solid rear wheels (24"x1"); resistance-ignitability fabric (EN 1021-1/2); pull-to-lock brakes. Complies with ISO 7176-1, -3, -5, -7, -8, -11, -13, -15, -16, -22. No electronic components or software.
Indications for Use
Indicated for persons restricted to a sitting position requiring mobility assistance.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 20, 2016
Danyang Sunco Machinery Co., Ltd Jen Ke-Min Official Correspondent Taojing Road, Optics Industrial Park, Situ Town Danyang City, 212300 CN
Re: K153328
Trade/Device Name: SUNCO Mechanical Wheelchair, model SKW-9003 Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I Product Code: IOR Dated: March 4. 2016 Received: July 11, 2016
Dear Jen Ke-Min:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{2}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Hoffmann -A
- Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K153328
#### Device Name
SUNCO Mechanical Wheelchair, model SKW-9003
Indications for Use (Describe)
The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Taojing Road, Optics Industrial Park, Situ Town, Danyang City, Jiangsu, China 212300 Tel: +86-511-86733686 Fax:+86-511-86733116
## SECTION D
## 510(k) Summary of Safety and Effectiveness
(per 21 CFR 807.92)
{5}------------------------------------------------
Taojing Road, Optics Industrial Park, Situ Town, Danvang Citv. Jiangsu. China 212300 Tel: +86-511-86733686 Fax:+86-511-86733116
#### I. SUBMITTER
Danyang Sunco Machinery Co., Ltd. Taojing Road, Optics Industrial Park, Situ Town, Danyang City, Jiangsu, China 212300
Phone: +86-511-86733686 Fax: +86-511-86733116 Contact Person: Jen, Ke-Min, ceirs.jen(@msa.hinet.net Date Prepared: July 20, 2016
#### DEVICE II.
| Name of Device: | SUNCO Mechanical Wheelchair, model SKW-9003 |
|-----------------------|---------------------------------------------|
| Common or Usual Name: | Mechanical Wheelchair |
| Classification Name: | Wheelchair, Mechanical (21 CFR 890.3850) |
| Regulatory Class: | Class I |
| Product Code: | IOR |
### III. PREDICATE DEVICE
Valentine International Ltd. Valentine Steel Wheelchair, model 1000, K130017
### IV. DEVICE DESCRIPTION
The SUNCO Mechanical Wheelchair, model SKW-9003 is an indoor / outdoor wheelchair that has a base with four-wheeled with a seat. The device can be disassembled for transport and it is foldable easily. The device uses a standard sling type back and seat, and the upholstery fabric meets the EN 1021-1:2006 Furniture --Assessment of ignitability of upholstered furniture - Part 1: Ignition source smouldering cigarette & EN 1021-2:2006 Furniture -- Assessment of ignitability of upholstered furniture - Part 2: Ignition source match flame equivalent
{6}------------------------------------------------
Taojing Road, Optics Industrial Park, Situ Town,
Danyang Citv. Jiangsu. China 212300
Tel: +86-511-86733686 Fax:+86-511-86733116
The features of the subject device include
- The back upholstery material is resistance-ignitability fabric.
- ������������������������������������������������������������������������������������������������������������������������������������������������������������������������������ The removable desk-length armrest and swing-away detachable footrest..
Due to the device design of the body structure the following surfaces are recommended NOT to operate on:
- . Sand surface
- . 1Wet or icy surface
- . Road maintenance hole metal cover
- . Avoid going up multiple steps.
- . Avoid using escalators. Use the elevator.
- . Too steep incline over 10 degrees.
- . Turning Diameter 31"
- Ground clearance: 2.3" .
- . Curb climbing ability: 0.8"
#### v. INDICATIONS FOR USE
The device is intended for medical purposes to provide mobility to persons restricted to a sitting position.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE
- Comparison table
| ITEMS | Predicate device | Subject device |
|---------------|---------------------------------|------------------------------------------|
| Manufacturer | Valentine International Ltd. | DANYANG SUNCO<br>Machinery Co., Ltd. |
| Brand name | Valentine | SUNCO |
| Device name | Steel Wheelchair,<br>model 1000 | Mechanical Wheelchair,<br>model SKW-9003 |
| 510(k) Number | K130017 | K153328 |
{7}------------------------------------------------
Taojing Road, Optics Industrial Park, Situ Town,
| Similarities | | |
|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | The device is intended for<br>medical purposes to provide<br>mobility to persons<br>restricted to a sitting<br>position. | Same indications for use |
| Operating environments | For Indoor / Outdoor use | Same operating environments |
| Technological<br>characteristics | Testing per ISO<br>7176-1/-3/-5/-7/-8/-11/-13/<br>-15/-16 series standards | Same technological standards |
| Overall dimensions<br>Length<br>Width<br>Height | 42"<br>25.2"<br>36.2" | 42"<br>25.2"<br>36" |
| FRAME<br>Cross brace<br>Backrest height<br>Reclining backrest<br>Seat sling<br>Frame color | YES<br>Fixed<br>Fixed<br>Padded Nylon<br>Blue Powder Coating | YES<br>Fixed<br>Fixed<br>Padded Nylon<br>Silver hammer tone |
| HANGERS<br>Swing-away<br>Elevating leg rest<br>Articulating leg rest<br>Footplate style<br>Heel loop<br>Footrest angle | YES<br>YES<br>YES<br>Padded<br>No<br>10~15° | Same hangers |
| REAR AXLE<br>Offset axle<br>Quick-release axle | YES<br>YES | Same rear axle |
| REAR WHEEL<br>Size<br>Tire type<br>Handrim Diameter /<br>material | 24"*1"<br>PU Solid Material<br>21" / Steel Composite | Same rear wheel |
| Wheel Lock | Pull-to-Lock | Same wheel lock |
| Ground Clearance | 2.3" | Same ground clearance |
| Climbing Angle | 10 degrees | Same climbing angel |
| Curb climbing ability | 0.8" | Same curb climbing ability |
| Minimum turning<br>diameter | 31" | Same minimum turning radius |
| Curb Stepper | YES | YES |
| ARMREST | | |
| Arm pad | Padded | Padded |
| Flip back | YES | YES |
| Height-adjustable | NO | NO |
| Optional accessory | | |
| Anti-tipper | YES | YES |
| Seat belt | YES | YES |
| Differences | | |
| Seat dimensions | | |
| Depth | 16" | 16" |
| Height | 20" | 20" |
| Width | 18" | 20" |
| Casters | | |
| Size | 8"*1" | 7.9"*1" |
| Tire type | PU solid material | PVC solid material |
| Weight of wheelchair | 40 lb / 18.1 kg | 38.6 lb / 17.5 kg |
| Weight Capacity | 250 lb / 113.4 kg | 220 lb / 100 kg |
| Warranty | 12 months for the main parts<br>(footrest, wheel locks, armrest,<br>cross braces, backrest canes,<br>front fork, fork stem house)<br>The chair side frames are<br>guaranteed for 5 years from the<br>date of purchase. | 12 months for the main parts<br>(chair side frames, footrest, wheel<br>locks, armrest, cross braces,<br>backrest canes/Push handle tube,<br>front fork, fork stem house) |
{8}------------------------------------------------
Taojing Road, Optics Industrial Park, Situ Town, Danyang Citv. Jiangsu. China 212300 Tel: +86-511-86733686 Fax:+86-511-86733116
#### ● Discussion
From the above comparison table, we knew that the indications for use of both devices are the same. Both mainframes of two devices are foldable. The castor tires are PU 8" solid tires for predicate device and PVC 7.9" solid tires for subject device. The PU tires can absorb more vibrational impact from the unsmooth ground than PVC tires. The difference is not much as to raise any safety and effectiveness concerns. The weights of wheelchairs (40 lb vs. 38.6 lb) differ not so much to raise any safety and effectiveness concerns. This difference means the user needs less power to move the wheelchair for the subject device. The front/rear tire sizes PVC solid 7.9"*1"/ PU solid 24"*1" and weight capacity (220 lbs.) are indicated on the product durable label and user manual for the
{9}------------------------------------------------
Taojing Road, Optics Industrial Park, Situ Town. Danvang Citv. Jiangsu. China 212300 Tel: +86-511-86733686 Fax:+86-511-86733116
subject device. The users can be informed of those limitations. The chair side frames are guaranteed for 5 years for predicate device and 1 year for the subject device. The difference in the chair side frames warranty periods just involves the more cost paid by the users of the subject device, and it does not raise any safety and effectiveness concerns. Overall dimensions are the same, and seat width has the difference of 2". No safety and effectiveness concerns are raised about this. The other safety and effectiveness concerns of the subject device have been considered and mitigated by complying with the ISO 7176 series standards.
### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility testing
The biocompatibility evaluation for the SUNCO Mechanical Wheelchair SKW-9003 was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The testing included the following tests:
- Cytotoxicity: ISO 10993-5:2009
- Sensitization: ISO 10993-10: 2009
- Irritation .: ISO 10993-10:2009
### Safety testing
To ensure the safety and effectiveness of the device, the following ISO 7176 series standards were complied with:
- 1) ISO 7176-1 Wheelchairs Part 1: Determination of static stability, 2014. (FDA
Recognition Number: 16-158 )
- 2) ISO 7176-3 Wheelchairs Part 3: Determination of effectiveness of brakes, 2012.
(FDA Recognition Number: 16-192)
{10}------------------------------------------------
Taojing Road, Optics Industrial Park, Situ Town, Danvang Citv. Jiangsu. China 212300 Tel: +86-511-86733686 Fax:+86-511-86733116
- 3) ISO 7176-5 Wheelchairs Part 5: Determination of overall dimensions, mass and maneuvering space, 2008. (FDA Recognition Number: 16-163)
- 4) ISO 7176-7 Wheelchairs Part 7: Determination of seating dimensions Definitions and measuring method, 1998. (FDA Recognition Number: NA)
- 5) ISO 7176-8 Wheelchairs Part 8: Static, impact and fatigue strength for manual wheelchairs, 2014. (FDA Recognition Number: NA)
- 6) ISO 7176-11 Wheelchairs Wheelchairs Part 11: Test dummies, 2012. (FDA Recognition Number: 16-190)
- 7) ISO 7176-13 Wheelchairs Part 13: Determination of coefficient of friction of test surfaces, 1989. (FDA Recognition Number: 16-25)
- 8) ISO 7176-15 Wheelchairs Part 15: Requirements for information disclosure, documentation and labelling, 1996. (FDA Recognition Number: 16-27)
- 9) ISO 7176-16 Wheelchairs Part 16 Requirements and test methods for resistance to ignition of upholstered parts, 2012. (FDA Recognition Number: 16-191)
- 10) ISO 7176-22 Wheelchairs Part 22 Set-up procedures, 2014 (FDA Recognition Number: NA)
#### VIII. CONCLUSIONS
The conclusions drawn from the nonclinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device identified in paragraph III of this section. They are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.