GC Reline II

K153253 · GC America, Inc. · EBI · Mar 3, 2016 · Dental

Device Facts

Record IDK153253
Device NameGC Reline II
ApplicantGC America, Inc.
Product CodeEBI · Dental
Decision DateMar 3, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3760
Device ClassClass 2
AttributesTherapeutic

Intended Use

GC RELINE II is a resilient material for making soft relining of dentures.

Device Story

GC Reline II is a vinyl polysiloxane (VPS) silicone material used for chairside soft relining of dentures. The system consists of base and catalyst silicone pastes, which are extruded from a cartridge and automixed via a mixing tip. The device is available in three viscosities: Soft, Extra Soft, and Extra Extra Soft. A resin primer is used for adhesion, and a modifier is available to activate existing reline surfaces. The device is intended for use by dental professionals in a clinical setting. By providing a resilient lining, the device improves denture comfort and fit for the patient. The material cures via chemical polymerization of vinyl polysiloxane initiated by a platinum catalyst system.

Clinical Evidence

Bench testing only. No clinical data provided. Compliance with ISO 10139-2:2009 (soft lining materials) and ISO 10993 (biocompatibility) demonstrated through testing of physical properties (hardness, bond strength, sorption, solubility) and biological safety (cytotoxicity, irritation, sensitization).

Technological Characteristics

VPS silicone-based resilient denture liner. Materials: high density polyethylene (cartridge/remover), polypropylene (cap/mixing tip/modifier bottle), glass (primer bottle). Curing mechanism: chemical polymerization of vinyl polysiloxane via platinum catalyst. Conforms to ISO 10139-2:2009 and ISO 10993-1, -2, -5, -10. Non-sterile, 2-year shelf life.

Indications for Use

Indicated for use as a resilient material for soft relining of dentures.

Regulatory Classification

Identification

A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 3, 2016 GC America, Inc. Mr. Mark Heiss Director, Regulatory & Academic Affairs 3737 W. 127th St. Alsip. Illinois 60803 Re: K153253 Trade/Device Name: GC Reline II Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, or Rebasing Resin Regulatory Class: II Product Code: EBI Dated: February 1, 2016 Received: February 3, 2016 Dear Mr. Heiss: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Page 2 - Mr. Mark Heiss Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Tina Kiang -S for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K153253 Device Name GC Reline II Indications for Use (Describe) GC RELINE II is a resilient material for making soft relining of dentures. Type of Use (Select one or both, as applicable) | <span style="font-size:10pt;"><b> <label><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> </b></span> | |---------------------------------------------------------------------------------------------------------------------------------------------------| | <span style="font-size:10pt;"><b> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> </b></span> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the letters 'GC' in a stylized, teal-colored font. The letters are slightly italicized and have a bold appearance. There are quotation marks before and after the letters. GC AMERICA INC. 3737 W 127 STREET ALSIP, ILLINOIS 60803 TEL (708) 597-0900 FAX (708) 926-9100 www.gcamerica.com ## K153253 510(k) Summary February 29, 2016 Date Prepared: - Submitter Information: 1. GC AMERICA INC. 3737 W. 127th Street Alsip, IL 60803 | Contact Person: | Mark Heiss, D.D.S. | |--------------------|--------------------| | Phone: | (708) 926-3090 | | Alternate Contact: | Lori Rietman | | Phone: | (708) 926-3092 | | Fax: | (708) 926-9100 | - 2. Device Name: Proprietary Name: Classification Name: CFR Regulation: Device Classification: Product Code: GC RELINE II Denture relining, repairing or rebasing resin 21CFR 872.3760 Class II EBI - 3. Predicate Device: | Company | Device | 510(k) | Date Cleared | |------------------|---------------------------|---------|---------------| | GC America, Inc. | GC Reline Soft/Extra Soft | K990376 | March 5, 1999 | {4}------------------------------------------------ - 4. Description of Device: GC RELINE II is a VPS silicone for relining of dentures. The components consist of the base silicone paste and the catalyst silicone paste that are extruded from a cartridge and are automixed with a mixing tip. GC Reline II is available in the following viscosities: Soft, Extra Soft, and Extra Extra Soft. The cartridge is made of high density polyethylene and the cap is made of polypropylene. The mixing tip is made of polypropylene. GC Reline Primer for Resin is a resin adhesive that is packaged in a glass bottle. GC Reline Modifier is available in Liquid A and Liquid B and is similar to the Base and Catalyst of the subject device. Liquid A and Liquid B are dispensed in equal amounts onto the mixing pad and mixed together. It is used to modify (activate) the existing reline surface to accept additional Reline II material. న. Indications for Use: GC RELINE II is a resilient material for making soft relining of dentures. - 6. Technological characteristics: The components of GC RELINE II and the predicate device are similar. The subject device GC Reline II contains a new ingredient, Polydimethylsiloxane, which is used in each of the three viscosities. Bench testing demonstrates that the addition of the new ingredient does not impact technological characteristics when compared to the predicate K990376. The curing mechanism of the applicant and predicate devices is by polymerization of uncured vinyl polysiloxane. This reaction is caused by chemical polymerization initiator systems. Bench top testing indicates that even with different formula, applicant device and predicate devices, meet specifications listed in ISO 10139-2. Therefore, the applicant and predicate device are substantially equivalent. There are no statistically significant differences between GC RELINE II and GC RELINE when comparing performance data. {5}------------------------------------------------ | | Applicant device | Primary Predicate | |---------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product category | CHAIRSIDE<br>VINYL<br>POLYSILOXANE<br>RESILIENT DENTURE LINER | CHAIRSIDE<br>VINYL<br>POLYSILOXANE<br>RESILIENT DENTURE LINER | | Trade name | GC RELINE II | GC RELINE (Soft/ Extra Soft) | | 510(k) Number | K153253 | K990736 | | Indications for<br>Use | GC RELINE II is a resilient material for<br>making soft relining of dentures. | GC RELINE Soft Extra Soft is a resilient<br>material for making soft relining of dentures. | | Device Description | GC RELINE II is a VPS silicone for relining<br>of dentures. The device consists of the base<br>paste and the catalyst paste Including VPS.<br>The material sets by mixing the catalyst paste<br>with the base paste. | GC RELINE is a VPS silicone for relining of<br>dentures. The device consists of the base paste<br>and the catalyst paste Including VPS. The material<br>sets by mixing the catalyst paste with the base<br>paste. | | Device<br>Components | Base:<br>Silicon dioxide<br>Vinyldimethylpolysiloxane<br>Methylhydrogen dimethylpolysiloxane<br>Polydimethylsiloxane*<br>Colorant<br>Catalyst: Silicon dioxide<br>Vinyldimethylpolysiloxane<br>Polydimethylsiloxane*<br>Platinum Catalyst | Base:<br>Silicon dioxide<br>Vinyldimethylpolysiloxane<br>Methylhydrogen dimethylpolysiloxane<br>Methylphenylpolysiloxane<br>Colorant<br>Catalyst: Silicon dioxide<br>Vinyldimethylpolysiloxane<br>Platinum Catalyst | | Similar Physical<br>Properties<br>(tested in<br>accordance with<br>ISO 10139-2) | Shore A hardness, 24h<br>Shore A hardness, 28d<br>Bond strength (MPa)<br>Sorption( $\[\mu$ /mm3)<br>Solubility ( $\[\mu$ /mm3) | Shore A hardness, 24h<br>Shore A hardness, 28d<br>Bond strength (MPa)<br>Sorption( $\[\mu$ /mm3)<br>Solubility ( $\[\mu$ /mm3) | | Composition of<br>Materials | Cartridge - high density polyethylene<br>Cap - polypropylene<br>Mixing tip - polypropylene.<br>GC Reline II Primer for Resin-8g glass bottle.<br>GC Reline II Modifier-polypropylene bottle.<br>GC Reline II Remover-high density<br>polyethylene. | Cartridge - high density polyethylene<br>Cap-polypropylene<br>Mixing tip - polypropylene.<br>GC Reline II Primer for Resin - 8 g glass bottle.<br>GC Reline II Modifier-polypropylene bottle. | | Standards of<br>Conformity | ISO 10139:2009, Sections 5.1, 5.2, 5.3<br>ISO10993-1:2009, Sections 6.2.2.2, 6.2.2.3,<br>6.2.2.4<br>ISO 10993-2:2006<br>ISO 10993-5:2009 Sections 8.5<br>ISO 10993-10:2010, Sections 6, 7 | Not available | | Biocompatibility | 10993-5 Cytotoxicity - pass<br>10993-10 irritation - pass<br>10993-10 sensitization - pass | Not available | {6}------------------------------------------------ ### Biocompatibility Testing | Test | Test Standard | Test Results<br>(pass or fail) | |---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------| | Cytotoxicity Test | Under the conditions of ISO 10993-1:2009, Biological evaluation of medical<br>devices, sections 6.2.2.2, 6.2.2.3, 6.2.2.4, the test articles should meet the test<br>requirements | Pass | | Cytotoxicity Test<br>In Vitro | Under the conditions of ISO 10993-5:2009, Biological evaluation of medical<br>devices, section 8.5 Determination of cytotoxicity, the test articles should meet<br>the test requirements | Pass | | Irritation and<br>Sensitization | Under the conditions of ISO 10993-10:2010, biological evaluation of medical<br>devices, Sections 6, 7, the test articles should meet the<br>test requirements | Pass | #### 7. Nonclinical Testing: Benchtop testing demonstrates that the subject device GC Reline II, like its predicate device GC Reline Soft/Extra Soft, is in compliance with ISO 10139-2:2009 Dentistry-Soft lining materials for removable dentures-Part 2: Materials for long-term use, and as such, the subject device meets specific requirements for softness, adhesion, water sorption and water solubility as well as for packaging, marking and manufacturer's instructions for soft denture lining materials suitable for long-term use. The subject device GC Reline II was also tested and met the requirement for cytotoxicity, irritation and sensitivity as shown in the biocompatibility table above. #### 8. Substantial equivalence: The subject device GC RELINE II and its predicate device GC RELINE (Soft/ Extra Soft) share the same function and intended use. The curing mechanism of the new and predicate device is equivalent. Bench testing demonstrates that the addition of a new ingredient in the subject device does not impact technological characteristics when compared to the predicate K990376. The subject device and predicate devices encompass the same range of physical dimensions, including hardness, bond strength, sorption, solubility, and also share equivalent biocompatibility. Both the subject and predicate devices are provided nonsterile and have a shelf life of 2 years. Any differences in the technological characteristics do not raise new questions. GC AMERICA INC. has demonstrated that, for the purposes of FDA's regulation of medical devices, the subject device GC Reline II is substantially equivalent to the predicate device in intended use, material composition, fundamental scientific technology, principals of operation, and basic design.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%