Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System
K153250 · Renovis Surgical Technologies · OVE · Mar 16, 2016 · Orthopedic
Device Facts
Record ID
K153250
Device Name
Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System
Applicant
Renovis Surgical Technologies
Product Code
OVE · Orthopedic
Decision Date
Mar 16, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System is indicated for intervertebral body fusion procedures in skeletally mature patients with cervical degenerative disc disease at one level from C2-T1. Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System implants are to be used with autogenous bone graft. Patients should be skeletally mature and have at least six weeks of non-operative treatment prior to implantation. The Renovis Tesera SC ACF System is a stand-alone system when used with the cover plate and screws provided, and requires no additional supplemental fixation. When used as a stand-alone system, the use of three (3) screws and the cover plate assembly. When used without the cover plate and three screws the Renovis Tesera SC ACF System is a non-stand-alone system and requires additional supplemental fixation cleared by the FDA for use in the cervical spine to augment stability.
Device Story
System comprises Tesera Trabecular Titanium (T3) cages, bone screws, and anterior cover plate assembly; used for cervical intervertebral body fusion. Cages are additively manufactured from Ti-6Al-4V alloy; screws and cover plate also titanium-based (cover plate includes Nitinol). Device implanted by surgeons in clinical settings. When used with cover plate and three screws, system acts as stand-alone fixation; without these, requires supplemental fixation. Cages feature side windows for bone graft visualization. Implants provided sterile via gamma irradiation. System includes reusable instrumentation for implantation. Benefits include vertebral stabilization and reduced expulsion risk.
Clinical Evidence
Bench testing only. Performed static/dynamic compression, shear, and torsion (ASTM F2077-14); subsidence (ASTM F2267-04); expulsion testing; and cover plate corrosion testing (ASTM F2129-08). Finite Element Analysis also conducted. No clinical data presented.
Technological Characteristics
Materials: Ti-6Al-4V (ASTM F136-13) and Nitinol (ASTM F2063-12). Additively manufactured porous titanium cages. Stand-alone fixation via cover plate and screws. Sterilization: Gamma irradiation. No software or electronic components.
Indications for Use
Indicated for intervertebral body fusion in skeletally mature patients with cervical degenerative disc disease at one level from C2-T1. Requires autogenous bone graft and at least six weeks of prior non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, overlapping each other to create a sense of depth and connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 16, 2016
Renovis Surgical Technologies, Incorporated % Sharyn Orton, Ph.D. Senior Consultant MEDIcept, Incorporated 200 Homer Avenue Ashland, Massachusetts 01721
Re: K153250
Trade/Device Name: Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVE, ODP Dated: February 13, 2016 Received: February 16, 2016
Dear Dr. Orton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{2}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K153250
#### Device Name
Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System
#### Indications for Use (Describe)
The Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System is indicated for intervertebral body fusion procedures in skeletally mature patients with cervical degenerative disc disease at one level from C2-T1. Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System implants are to be used with autogenous bone graft. Patients should be skeletally mature and have at least six weeks of non-operative treatment prior to implantation.
The Renovis Tesera SC ACF System is a stand-alone system when used with the cover plate and screws provided, and requires no additional supplemental fixation. When used as a stand-alone system, the use of three (3) screws and the cover plate assembly.
When used without the cover plate and three screws the Renovis Tesera SC ACF System is a non-stand-alone system and requires additional supplemental fixation cleared by the FDA for use in the cervical spine to augment stability.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### Traditional 510(k) Summary as required by 21 CFR 807.92(a) K153250
- A ) Submitted by: Renovis Surgical Technologies Inc. 1901 W. Lugonia Ave. Ste 340 Redlands, CA 92374 Phone: 909-557-2360 Fax: 909-307-8571
- Official Contact: Anthony DeBenedictis Vice President of Quality Assurance
- Consultant: Sharyn Orton, Ph.D. MEDIcept, Inc. 200 Homer Ave Ashland, MA 01721
Date prepared: 1-11-2016
B) Device Name: Intervertebral body fusion device with integrated fixation, cervical
- Proprietary Name: Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System Device Class: Class II
Regulation number: 21 CFR 888.3080
Intervertebral body fusion device Regulation name:
- Product code: OVE - intervertebral fusion device with integrated fixation, cervical ODP - intervertebral fusion device with bone graft, cervical
Classification panel: Orthopedic
- C) Primary Predicate: K102547 Nuvasive CoRoent Small Interlock
D) Device Description:
The Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System includes the Tesera Trabecular Titanium (T3) cages, bone screws (self-drilling and self-tapping options) and anterior cover plate assembly. When used with the cover plate and screws, the Tesera SC ACF System requires no supplementary fixation systems. The screws protrude through the interbody portion of the device into the adjacent vertebral bodies, stabilize the vertebral body and reduce the likelihood of expulsion of the implant. The Tesera Trabecular Titanium (T3) cages have
Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System Traditional 510k Premarket Notification
{5}------------------------------------------------
## NOWIE
windows in the side for visualization of the bone graft. The accompanying cover plate is designed to resist screw back-out and must be used when the screws are implanted. When used without the cover plate and screws, the Tesera SC ACF cages require supplemental fixation.
The Renovis Tesera SC ACF System cages are available in a variety of sizes (widths, height, depths, and bone screw sizes) to suit the individual pathology and anatomical conditions of the patient. The cages are additively manufactured then machined from Titanium alloy (Ti-6A1-4V). The bone screws are manufactured from Titanium alloy (Ti-6Al-4V) and the cover plate assembly is manufactured from Titanium alloy (Ti-6Al-4V) and Nitinol alloy. The implants are provided to the end user sterilized through gamma irradiation.
The system also consists of reusable instrumentation used to implant the cages, screws, and cover plate assembly.
The Renovis Tesera SC ACF System implants comply with the following material standards:
- . ASTM F136-13 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications
- ASTM F2063-12 Standard Specification for Wrought Nickel-Titanium Shape Memory ● Alloys for Medical Devices and Surgical Implants
E) Indications For Use:
The Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System is indicated for intervertebral body fusion procedures in skeletally mature patients with cervical degenerative disc disease at one level from C2-T1. Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System implants are to be used with autogenous bone graft. Patients should be skeletally mature and have at least six weeks of non-operative treatment prior to implantation.
The Renovis Tesera SC ACF System is a stand-alone system when used with the cover plate and screws provided, and requires no additional supplemental fixation. When used as a stand-alone system, the cages require the use of three (3) screws and the cover plate assembly.
When used without the cover plate and three screws the Renovis Tesera SC ACF System is a non-stand-alone system and requires additional supplemental fixation cleared by the FDA for use in the cervical spine to augment stability.
- F) Substantial Equivalence Discussion
Additional predicate devices include the Depuy/Synthes Zero P (K072981) and Centinel Spine Stalif-C (K072415). The Renovis Tesera SC ACF System cage is manufactured from different material than the predicate devices, but has a similar intended use, sizes, basic design, and function. The Renovis Tesera SC ACF System cage is manufactured using the same method (additive manufacturing) as the Renovis FDA cleared S128 ALIF (K131122; K140106) and S134 ALIF (K142095) which are reference devices.
{6}------------------------------------------------
### G) Performance Testing - Bench
Mechanical testing of the Renovis Tesera SC ACF System included:
- . Subsidence testing without screws per ASTM F2267-04 (reapproved 2011) Standard test method for measuring load induced subsidence of intervertebral body fusion device under static axial compression
- Static Compression, Dynamic Compression, Static Shear Compression, Dynamic Shear . Compression. Static Torsion and Dynamic Torsion Testing per ASTM F2077-14 Test methods for intervertebral body fusion devices
- Expulsion testing with and without screws per ASTM Draft Standard F04.25.02.02 Static ● Expulsion
- . Cover plate assembly corrosion per ASTM F2129-08 Standard test method for conducting cyclic potentiodynamic polarization measurements to determine the corrosion susceptibility of small implant devices
- Finite Element Analysis (cage) ●
Sterilization, shelf life, and packaging studies have been previously conducted.
#### Conclusion
The Renovis Tesera SC ACF System has a similar Indication for Use, sizes, design, and function as the predicate devices. The Renovis Tesera SC ACF System has the same manufacturing process as the reference device.
Differences between the Renovis Tesera SC ACF System and the predicate devices do not raise new issues of safety or effectiveness. The Renovis Tesera SC ACF System is substantially equivalent to the predicate devices.
- H) Other Standards and Guidance
- ASTM F983-86 (Reapproved 2013) Standard Practice for Permanent Marking of Orthopaedic Implant Components
- ASTM F565-04 (Reapproved 2013) Standard Practice for Care and Handling of ● Orthopedic Implants and Instruments
- . ASTM A564 - Standard Specification for Hot-Rolled and Cold-Finished Age-Hardening Stainless Steel Bars and Shapes
- . Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Intervertebral Body Fusion Device, October 2007
- Guidance for Industry and FDA Staff: Spinal System 510(k)s", May 2004 ●
- . Accepted into previous testing
- ISO 11607-2: 2009 Packaging for terminally sterilized medical devices part 1: o requirements for materials, sterile barrier systems and packaging systems
- Per ASTM F1980-07 Standard guide for accelerated aging of sterile . barrier systems for medical devices
Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System Traditional 510k Premarket Notification
{7}------------------------------------------------
# NOWIS
- ISO 11137-1:2006 Sterilization of health care products Radiation Part 1: o Requirements for development, validation and routine control of a sterilization process for medical devices
- ISO 11137-2:2013 Sterilization of health care products Radiation Part 2: O Establishing the sterilization dose
- ISTA 2A, 2011, Partial-Simulation Performance Test Procedure: Packaged o Products 150lb (68 kg) or Less
- ASTM D4169:2009, Standard Practice for Performance Testing of Shipping O Containers and Systems
- ISO 17665-1:2006 Sterilization of health care products Moist heat -- Part 1: O Requirements for the development, validation, and routine control of a sterilization process for medical devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.