K153247 · Medtronic Navigation, Inc. · PGW · Feb 18, 2016 · Neurology
Device Facts
Record ID
K153247
Device Name
Fusion Compact Navigation System
Applicant
Medtronic Navigation, Inc.
Product Code
PGW · Neurology
Decision Date
Feb 18, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The Medtronic Fusion Compact computer-assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous ENT procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to images of the anatomy.
Device Story
FUSION Compact is an electromagnetic (EM) image-guided surgical navigation platform for ENT procedures (e.g., skull base, FESS). It inputs pre-operative CT or MR images and tracks specialized surgical probes in real-time using an EM localization system (AxiEM II). The system consists of an all-in-one PC/monitor, EM field transmitter, and navigation software. During surgery, the system continuously updates the probe position on the displayed anatomical images. Surgeons use this real-time positional feedback to validate their judgment and guide instruments relative to patient anatomy. Used in medical offices or operating rooms by surgeons to improve precision in locating anatomical structures.
Clinical Evidence
Bench testing only. No clinical data. Testing included AAMI/ANSI ES 60601-1 and IEC 60601-1-2 compliance, software/hardware verification, and 3D positional accuracy testing. Accuracy was validated in simulated surgical office and OR environments, demonstrating a maximum error ≤ 3.0 mm.
Technological Characteristics
Electromagnetic navigation system; Intel-based PC/monitor all-in-one; AxiEM II localization; Class D amplifier; USB 1.1/2.0 connectivity. Software runs on Ubuntu Linux (C++). Emitter stand/clamp mounting. Supports 4-coil ports (up to 4 coils per port). Standards: AAMI/ANSI ES 60601-1, IEC 60601-1-2, IEC 62366.
Indications for Use
Indicated for patients undergoing open or percutaneous ENT procedures where stereotactic surgery is appropriate and rigid anatomical structures (e.g., skull) can be referenced to imaging.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Predicate Devices
StealthStation System GoldenEye Micro-Magnetic Tracking System Option (K001284)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of three human profiles facing to the right, arranged in a way that suggests forward movement or progress. The profiles are depicted in a simple, line-art style. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 18, 2016
Medtronic Navigation, Inc. Rishi Sinha Principal Regulatory Affairs Specialist 826 Coal Creek Circle Louisville, Colorado 80027
Re: K153247
Trade/Device Name: Fusion Compact Navigation System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: PGW Dated: January 15, 2016 Received: January 19, 2016
Dear Mr. Sinha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet
address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go
to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet
address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Eric A. Mann -S
for Malvina Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) N/A
Device Name Fusion Compact Navigation System
#### Indications for Use (Describe)
The Medtronic Fusion Compact computer-assisted surgery system and its associated applications are intended as an aid for precisely locating anatomical structures in either open or percutaneous ENT procedures. Their use is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the skull, can be identified relative to images of the anatomy.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
| Submitter: | Medtronic Navigation, Inc.<br>826 Coal Creek Circle<br>Louisville, CO 80027 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Rishi Sinha (Primary)<br>Principal Regulatory Affairs Specialist<br>Phone: (269) 903-4373<br>Fax: (720) 890-3500<br>E-mail: rishi.k.sinha@medtronic.com |
| | Michael Blasco (Alternate)<br>Senior Regulatory Affairs Manager<br>Phone: (720)890-3391<br>Fax: (720) 890-3500<br>E-mail: michael.blasco@medtronic.com |
| Date Summary Prepared: | January 15, 2016 |
| Device Trade Name: | Medtronic FUSION Compact™ Navigation System |
| Common Name: | Ear Nose and Throat Stereotaxic Instrument |
| Device Classification: | Class II |
| Product Code: | PGW |
| Classification Name: | 882.4560 - Stereotaxic Instrument |
| Predicate Device: | K001284 - StealthStation System GoldenEye Micro-Magnetic<br>Tracking System Option |
| Device Description: | The FUSION Compact™ system is a surgical guidance platform<br>that supports use of special application software and associated<br>instruments. The system reformats patient-specific CT or MR<br>images acquired before surgery and displays them on-screen from<br>a variety of perspectives (axial, sagittal, coronal). During surgery,<br>the system tracks the position of specialized surgical probes in or<br>on the patient anatomy and continuously updates the probe<br>position on these images. |
| | While the surgeon's judgment remains the ultimate authority,<br>real-time positional information obtained through the FUSION<br>Compact™ system can serve to validate this judgment as well as<br>guide. |
| | The FUSION Compact™ is an Electromagnetic-based navigation<br>system, also known as an Image Guided System (IGS), which<br>consists of clinical navigation software, a referencing system and |
| | platform/computer hardware. Navigation tracks the position of<br>navigated instruments in relation to the surgical anatomy and<br>identifies this position on diagnostic or intraoperative images of<br>the patient. The FUSION Compact™ system is modular and<br>designed for use in medical office or operating room environments<br>to conduct ENT procedures including Endoscopic Skull Base,<br>Lateral Skull Base, and Functional Endoscopic Sinus Surgery<br>(FESS) procedures. The FUSION Compact™ will utilize a monitor<br>and PC all-in-one computer and a free-standing emitter holder<br>option. The system supports the FUSION Compact™ 1.0<br>software application. The FUSION Compact™ will interface with<br>the existing Electromagnetic (EM) localization system and existing<br>ENT instruments. |
| Indications for Use: | The Medtronic FUSION Compact™ computer-assisted surgery<br>system and its associated applications are intended as an aid for<br>precisely locating anatomical structures in either open or<br>percutaneous ENT procedures. Their use is indicated for any<br>medical condition in which the use of stereotactic surgery may be<br>appropriate, and where reference to a rigid anatomical structure,<br>such as the skull, can be identified relative to images of the<br>anatomy. |
{4}------------------------------------------------
| <b>Substantial Equivalence:</b> | The FUSION Compact™ Navigation System is substantially equivalent to the following device: |
|---------------------------------|--------------------------------------------------------------------------------------------|
|---------------------------------|--------------------------------------------------------------------------------------------|
- K001284 StealthStation System GoldenEye Micro-. Magnetic Tracking System Option
| | Subject Device | Predicate |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | FUSION Compact™<br>Navigation System | StealthStation GoldenEye<br>(K001284) |
| Classification | Class II | Class II |
| Product Code | PGW; 882.4560 - Stereotaxic<br>Instrument | HAW; 882.4560 - Stereotaxic<br>Instrument |
| Indications for<br>Use | The Medtronic FUSION<br>Compact™ computer-assisted<br>surgery system and its<br>associated applications are<br>intended as an aid for precisely<br>locating anatomical structures in<br>either open or percutaneous<br>ENT procedures. Their use is<br>indicated for any medical<br>condition in which the use of<br>stereotactic surgery may be<br>appropriate, and where | The StealthStation®<br>GoldenEye System is intended<br>as an aid for precisely locating<br>anatomical structures in either<br>open or percutaneous<br>procedures. The<br>StealthStation® System is<br>indicated for any medical<br>condition in which the use of<br>stereotactic surgery may be<br>appropriate, and where<br>reference to a rigid anatomical |
| | Subject Device | Predicate |
| | FUSION Compact™<br>Navigation System | StealthStation GoldenEye<br>(K001284) |
| | reference to a rigid anatomical<br>structure, such as the skull, can<br>be identified relative to a CT- or<br>MR-based model or digitized<br>landmarks of the anatomy. | structure, such as the skull, a<br>long bone, or vertebra, can be<br>identified relative to a CT-<br>based or MR-based model,<br>fluoroscopy images, or<br>digitized landmarks of the<br>anatomy. |
| | | Example procedures include<br>but are not limited to: |
| | | Cranial procedures:<br>■ Cranial biopsies<br>■ Tumor resections<br>■ Craniotomies/craniectomies<br>■ Skull base procedures<br>Thalamotomies/pallidotomies |
| | | Spinal implant procedures,<br>such as pedicle screw<br>placement |
| | | ENT procedures:<br>■ Transphenoidal procedures<br>■ Intranasal procedures<br>■ Orbital nerve decompression<br>procedures<br>Sinus procedures, such as<br>maxillary antrostomies,<br>ethmoidectomies,<br>sphenoidotomies/sphenoid<br>explorations, turbinate<br>resections, and frontal<br>sinusotomies |
| Operating<br>Principle | The electromagnetic system<br>collects measurements,<br>determines position and<br>orientation from each coil<br>independently. The system<br>transmits multiple magnetic<br>fields within which the receive<br>coils can be measured. A<br>measurement for each coil within<br>each transmitted field is used to | The electromagnetic system<br>collects measurements,<br>determines position and<br>orientation from each coil<br>independently. The system<br>transmits multiple magnetic<br>fields within which the receive<br>coils can be measured. A<br>measurement for each coil<br>within each transmitted field is |
| | Subject Device | Predicate |
| | FUSION Compact™<br>Navigation System | StealthStation GoldenEye<br>(K001284) |
| | determine the position and<br>orientation of the coil.<br>Measurement position andcombined to determine the<br>instrument position and<br>orientation. Two measured<br>positions, one from each coil,<br>are utilized to determine position<br>and orientation of a two coiled<br>instrument. Three measured<br>positions, one from each coil,<br>are utilized to determine position<br>and orientation of a three coiled<br>instrument. | used to determine the position<br>and orientation of the coil.<br>Measurement position and<br>orientation for the coils are<br>combined to determine the<br>instrument position and<br>orientation. Two measured<br>positions, one from each coil,<br>are utilized to determine position<br>and orientation of a two coiled<br>instrument. |
| System Accuracy | Maximum error less than 3.0 mm<br>at 95% confidence and 99.5%<br>reliability. | Mean error is less than 2.0 mm<br>with 99% confidence level of<br>4.0 mm |
| Computer and Software Technology | | |
| Computer | Intel-Based PC<br>Computer/Monitor all-in-one | Intel-Based PC |
| Operating<br>System | Linux-Based: Ubuntu | Linux-based IRIX |
| Programming<br>Language | C++ | C++ |
| Clinical Software<br>Features | Imaging modalities, registration<br>features, views (axial, coronal,<br>sagittal) | Imaging modalities, registration<br>features, views (axial, coronal,<br>sagittal) |
| Electromagnetic Technology | | |
| Electromagnetic<br>Localization<br>System | Medtronic Navigation AxiEM II<br>Integrated | Medtronic Navigation AxiEM I<br>Integrated |
| EM Configuration | EM field transmitter creates an<br>EM field that is used to track<br>and triangulate on the receive<br>coils | EM field transmitter creates an<br>EM field that is used to track<br>and triangulate on the receive<br>coils |
| Communication | USB 1.1 and 2.0 | Serial Line (RS-232) Connection |
| Amplifier<br>Technology | Class D | Class AB |
| Digitizer Box | CAC and NPI integrated into a<br>single unit, portable and<br>mountable | CAC and NPI provided in<br>separate components as cart on<br>wheels |
| Instrument ports | 4-coils per port, 8 ports in total | 2-coils per port, 6 ports in total |
| | Subject Device | Predicate |
| | FUSION Compact™<br>Navigation System | StealthStation GoldenEye<br>(K001284) |
| Mobile Emitter | Emitter generates a low-<br>intensity magnetic field to locate<br>coils mounted within the tracker | Emitter generates a low-<br>intensity magnetic field to locate<br>coils mounted within the tracker |
| EM Mounting<br>apparatus | Emitter Stand or Emitter Clamp | No mounting capability as this<br>unit is incorporated into the cart<br>and mounting is not needed. |
| Interfacing | Navigation Probe Interface (NPI)<br>is configured for up to 4 coils per<br>connection port. The<br>instruments used are designed<br>with 2 or 3 receive coils per<br>instrument. Each of the receive<br>coils is connected via the<br>instrument cable to the<br>instrument. | Navigation Probe Interface (NPI)<br>is configured for up to 2 coils per<br>connection port. The<br>instruments used are designed<br>with 2 receive coils per<br>instrument. Each of the receive<br>coils is connected via the<br>instrument cable to the<br>instrument connector, which is<br>then connected to the NPI<br>through a connection port. |
## Table 1: FUSION Compact™ Predicate Device Comparison Table
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
Performance Testing: Testing conducted demonstrates the product will perform as intended according to the outlined design requirements. The following testing was conducted on the FUSION Compact™ Navigation System to establish substantial equivalence of the system and verify that the device will perform as intended meeting all of the design inputs:
- AAMI/ANSI ES 60601-1:2012 Medical Electrical ● Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD)
- . IEC 60601-1-2:2014 - Medical Electrical Equipment -Part 1-2: General requirements for safety; Electromagnetic Compatibility - Requirements and Tests (2/2014)
- Software Verification and Validation testing verifying the . software requirements are met and software performs as intended
- . Hardware Verification testing ensuring the hardware requirements identified for the system are met and hardware performs as intended
- Accuracy Testing was conducted to ensure that the ● FUSION Compact met the prescribed requirement for accuracy. The FUSION Compact System demonstrates performance in 3D positional accuracy with a maximum error ≤ 3.0 mm. This performance was determined using a peqboard in a surqical office and OR simulated environment with all of the typical instruments and equipment seen in those environments. The upper bound
{8}------------------------------------------------
of the tip error data for both the surgical office and OR simulated environments was determined by statistical data analysis.
- Usability Testing was conducted in accordance to IEC . 62366 demonstrating that the usability and human factors requirements were adequately met.
Conclusion: The FUSION Compact™ Navigation System is similar in technological characteristics, imaging performance and indications for use as the predicate devices listed. These aspects, along with the functional testing conducted to the FDA recognized standards, demonstrate that FUSION Compact™ Navigation System does not raise new risks of safety and effectiveness when compared to the predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.