K153237 · Orthalign, Inc. · OLO · Mar 1, 2016 · Neurology
Device Facts
Record ID
K153237
Device Name
OrthAlign Plus System
Applicant
Orthalign, Inc.
Product Code
OLO · Neurology
Decision Date
Mar 1, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The OrthAlign Plus® System is a computer-controlled system intended to assist the surgeon in determining reference alignment axes in relation to anatomical and instrumentation structures during stereotactic orthopedic surgical procedures. The OrthAlign Plus® System facilitates the accurate positioning of implants, relative to these alignment axes. The system aids the surgeon in controlling leg length and offset discrepancies in Total Hip Arthroplasty: Posterior. Example orthopedic surgical procedures include but are not limited to: - Total Knee Arthroplasty - Total Hip Arthroplasty: Anterior/Posterior - Unicompartmental Knee Arthroplasty: Tibial transverse resection
Device Story
Computer-assisted surgical navigation system for orthopedic procedures (knee/hip arthroplasty). Uses palm-sized computer module and reference sensor to detect/measure angular and positional changes in triaxial format. Surgeon registers anatomical landmarks; system calculates mechanical axes and guides instrument positioning relative to these axes. Used in OR by surgeons to improve implant placement accuracy, control leg length, and manage offset discrepancies. Provides real-time graphical/numerical feedback on display screen. Benefits include improved surgical precision and alignment during arthroplasty.
Clinical Evidence
Bench testing only. Included system hardware verification/validation, cleaning/sterilization/shipping validations, and biocompatibility per ISO 10993-1. System accuracy verified via bench testing with mechanical fixtures and foam models to confirm angular and depth accuracy of navigated resection planes. Usability validated in cadaver study with advising surgeon to confirm design inputs in simulated environment.
Technological Characteristics
System comprises single-use computer module and reusable instrumentation (femoral/tibial/hip jigs, reference sensor). Uses sensors to generate spatial coordinates; provides graphical/numerical display. Biocompatibility per ISO 10993-1. Navigation functionality for unicompartmental knee arthroplasty added via software/instrumentation update. Principles of operation identical to predicate.
Indications for Use
Indicated for patients undergoing stereotactic orthopedic surgical procedures, including Total Knee Arthroplasty, Total Hip Arthroplasty (Anterior/Posterior), and Unicompartmental Knee Arthroplasty (tibial transverse resection), to assist in determining reference alignment axes and positioning implants.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized representation of three human profiles facing to the right, stacked on top of each other. The profiles are depicted in a simple, flowing line style.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 1, 2016
OrthAlign, Incorporated Mr. David Vancelette Director of Quality Assurance & Regulatory Affairs 120 Columbia, Suite 500 Aliso Viejo, California 92656
Re: K153237
Trade/Device Name: OrthAlign Plus® System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: January 29, 2016 Received: February 2, 2016
Dear Mr. Vancelette:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{2}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### K153237
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K153237
Device Name OrthAlign Plus® System
#### Indications for Use (Describe)
The OrthAlign Plus® System is a computer-controlled system intended to assist the surgeon in determining reference alignment axes in relation to anatomical and instrumentation structures during stereotactic orthopedic surgical procedures. The OrthAlign Plus® System facilitates the accurate positioning of implants, relative to these alignment axes. The system aids the surgeon in controlling leg length and offset discrepancies in Total Hip Arthroplasty: Posterior.
Example orthopedic surgical procedures include but are not limited to:
- · Total Knee Arthroplasty
- · Total Hip Arthroplasty: Anterior/Posterior
- · Unicompartmental Knee Arthroplasty: Tibial transverse resection
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
{4}------------------------------------------------
## 3. 510(K) SUMMARY
This 510(k) summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92(c).
| DATE | January 29, 2016 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT | OrthAlign, Inc.<br>120 Columbia<br>Suite 500<br>Aliso Viejo, CA 92656<br>Tel: (949) 715-2424<br>Fax: (949) 831-9500 |
| OFFICIAL<br>CORRESPONDENT | David Vancelette<br>OrthAlign, Inc.<br>120 Columbia,<br>Suite 500<br>Aliso Viejo, CA 92656<br>dvancelette@orthalign.com<br>Tel: (858) 692-0335<br>Fax: (949) 831-9500 |
| TRADE NAME | OrthAlign Plus® System |
| COMMON NAME | Stereotaxic Instrument |
| DEVICE<br>CLASSIFICATION | Class II, 21 CFR §882.4560 |
| PRODUCT CODES | OLO: Orthopedic Stereotaxic Instrument |
| PREDICATE<br>DEVICES | OrthAlign Plus® System (K140331)<br>Aesculap OrthoPilot Next Generation (K141694) |
| SUBMISSION TYPE | Traditional 510(k). The subject device is a modification<br>to the previously cleared OrthAlign Plus® System<br>(K140331). |
### SUBSTANTIALLY EQUIVALENT TO:
The OrthAlign Plus® System is substantially equivalent to the previously cleared OrthAlign Plus® System (K140331) and Aesculap OrthoPilot Next Generation (K141694).
# DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The OrthAlign Plus® System is a non-invasive computer assisted surgical navigation system for use in knee and hip arthroplasty procedures. The OrthAlign Plus® System is configured to detect, measure, and display angular and positional measurement changes in a triaxial format.
{5}------------------------------------------------
The OrthAlign Plus® System utilizes a palm-sized computer module and reference sensor to generate positional information in orthopedic procedures providing a sequence of steps for registration of anatomical landmarks, calculation of mechanical axes, and positioning of instruments relative to the mechanical axes.
In total knee arthroplasty procedures, the device assists the surgeon in:
- Establishing the mechanical axis of the femur, determining the varus/valgus angle and the flexion/extension angle of the cutting block relative to the femur.
- Establishing the mechanical axis of the tibia, determining the varus/valgus angle ● and the posterior slope angle of the cutting block relative to the tibia.
In hip arthroplasty procedures, the device assists the surgeon in:
- Establishing the orientation of the anterior pelvic plane and determining the ● inclination angle and the anteversion angle of the shell impactor relative to the anterior pelvic plane.
- . Measuring the intraoperative change in leg length and offset (posterior approach).
In unicompartmental knee arthroplasty procedures, the device assists the surgeon in:
- Establishing the mechanical axis of the tibia, determining the varus/valgus angle ● and the posterior slope angle of the cutting block relative to the tibia, for the transverse resection.
The OrthAlign Plus® System comprises a single use computer module and reusable instrumentation.
### INDICATIONS FOR USE:
The OrthAlign Plus® System has the same indications for use as the previously cleared OrthAlign Plus® System (K140331). Additional functionality has been added to the predicate device to enable the navigation of tibial transverse resections in unicompartmental knee arthroplasty. Also, Indications for Use are common to the Aesculap OrthoPilot Next Generation (K141694). Thus, the Indications for Use are as follows:
### OrthAlign Plus® System:
The OrthAlign Plus® System is a computer-controlled system intended to assist the surgeon in determining reference alignment axes in relation to anatomical and instrumentation structures during stereotactic orthopedic surgical procedures. The OrthAlign Plus® System facilitates the accurate positioning of implants, relative to these alignment axes. The system aids the surgeon in controlling leg length and offset discrepancies in Total Hip Arthroplasty: Posterior.
Example orthopedic surgical procedures include but are not limited to:
- Total Knee Arthroplasty
- . Total Hip Arthroplasty: Anterior/Posterior
- Unicompartmental Knee Arthroplasty: Tibial transverse resection
{6}------------------------------------------------
# TECHNICAL CHARACTERISTICS (COMPARED TO PREDICATE):
The OrthAlign Plus® System was cleared under K140331. The OrthAlign Plus® System comprises a single use computer module, a reusable reference sensor, a reusable femoral jig, a reusable tibial jig, a reusable posterior hip jig and a reusable anterior hip jig. The device utilizes algorithms to convert sensor outputs into spatial coordinates, providing graphical and numerical representation of instruments and anatomy on the user display screen. The OrthAlign Plus® System is being updated to include navigation functionality for unicompartmental knee arthroplasty, tibial transverse resections, as in the predicate device Aesculap OrthoPilot Next Generation (K141694). All other features and principles of operation remain unchanged.
## PERFORMANCE DATA:
Device performance testing confirms that the OrthAlign Plus® System can be used according to its intended use. The OrthAlign Plus® System has been verified and validated according to OrthAlign's procedures for product design and development. Performance testing addressed only the added Unicompartmental instruments and surgical procedure steps. Performance testing included:
- . System hardware verification/validation testing to ensure the instruments meet their mechanical requirements.
- Instrumentation cleaning, sterilization and shipping validations for the specified processes.
- . Navigation device sterilization, packaging, shelf life, environmental conditions and shipping validations for the specified ranges of conditions involved in each process. (Summary data for the identical predicate device is referenced.)
- System components biocompatibility assessment per ISO 10993-1 (2009). ●
- Customer requirements / usability validation in cadaver with an advising surgeon to . validate the system meets design input requirements for its functions in a simulated use environment.
- . System accuracy testing: bench testing with mechanical fixtures and foam models to verify navigated resection plane angular and depth accuracy.
This testing regime demonstrates that the subject device is as safe and effective as the predicate devices. This testing regime demonstrates that the subject device is substantially equivalent to the legally marketed predicate devices, for its intended use in the accurate navigation of tibial resection planes in unicompartmental knee arthroplasty.
The information provided by OrthAlign in this 510(k) application confirms that the OrthAlign Plus® System is substantially equivalent to predicate devices such as the OrthAlign Plus® System (K140331) and Aesculap OrthoPilot Next Generation (K141694).
{7}------------------------------------------------
# BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE:
A technological comparison and bench and cadaver testing demonstrate the substantial equivalence of the OrthAlign Plus® System to the predicate devices.
The subject device is identical to the predicate OrthAlign Plus® System (K140331), with the following exceptions:
- . The subject device allows for the navigation of tibial resection planes in unicompartmental knee arthroplasty.
The subject device is substantially equivalent in terms of unicompartmental knee arthroplasty function to the predicate device Aesculap OrthoPilot Next Generation (K141694).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.