Capiox RX Hollow Fiber Oxygenator with/without Hardshell Reservoir
K153213 · Terumo Cardiovascular Systems Corporation · DTZ · Dec 3, 2015 · Cardiovascular
Device Facts
| Record ID | K153213 |
| Device Name | Capiox RX Hollow Fiber Oxygenator with/without Hardshell Reservoir |
| Applicant | Terumo Cardiovascular Systems Corporation |
| Product Code | DTZ · Cardiovascular |
| Decision Date | Dec 3, 2015 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 870.4350 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The CAPIOX® RX Hollow Fiber Oxygenators with/without Hardshell Reservoir are intended to be used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery. The integral heat exchanger is used to warm or cool blood and/or perfusion fluid as it flows through the device. The (detachable) hardshell reservoir(s) is (are) used to store blood during extra-corporeal circulation from both venous line and the cardiotomy line (via gravity or vacuum assisted venous drainage procedures). The reservoir contains a venous section that is comprised of a filter and defoamer to facilitate air bubble removal. The cardiotomy section of the reservoir contains a filter to remove particulate matter and a defoamer to facilitate air bubble removal. The reservoir may also be used for Post-Operative Chest Drainage Procedures. The CAPIOX® RX Oxygenators with/without Hardshell Reservoirs can be used in procedures lasting up to 6 hours. The CAPIOX® RX15 is for use with patients when the required blood flow rate will not exceed 5.0 L/min when used with a 4 Liter Reservoir; and when the required blood flow rate will not exceed 4.0 L/min when used with a 3 Liter Reservoir). The CAPIOX® RX25 is for use with patients when the required blood flow rate will not exceed 7.0 L/min.
Device Story
Device is a hollow fiber oxygenator with integrated heat exchanger and hardshell reservoir for cardiopulmonary bypass. Inputs: venous and cardiotomy blood via gravity or vacuum-assisted drainage. Operation: blood flows through porous fiber bundle for gas exchange via diffusion; stainless steel heat exchanger regulates blood temperature; reservoir filters and defoams blood. Output: oxygenated, temperature-controlled blood returned to patient. Used in OR by perfusionists/surgeons. Modifications include flared venous inlet drop tubes to minimize air emboli, modified drainage cage, full-length graduated volume scale, and updated PVC tubing plasticizer (TOTM vs DEHP). Benefits: improved air handling, better volume visibility, and reduced safety concerns regarding plasticizers.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via in-vitro performance evaluations including air handling, hemolysis, venous reservoir pressure drop, venous defoaming, leak testing, winged adapter circulation testing, drainage cage dimensional testing, connection testing, package testing, and aged evaluation.
Technological Characteristics
Integrated oxygenator/heat exchanger module with hardshell reservoir. Materials: nylon, polycarbonate, silicone, stainless steel, PVC (with TOTM plasticizer), polyurethane, polyester, polypropylene, polyethylene, X-Coating™. Gas exchange via porous fiber membrane diffusion. Heat exchange via stainless steel interface. Mechanical filtration/defoaming. Dimensions vary by model (RX15/RX25).
Indications for Use
Indicated for patients undergoing cardiopulmonary bypass surgery requiring gas exchange, blood warming/cooling, and blood storage/filtration. RX15 model for flow rates up to 5 L/min (4 L/min for specific 3L reservoir codes); RX25 model for flow rates up to 7 L/min. Duration up to 6 hours. Includes use in vacuum-assisted venous drainage, post-operative chest drainage, and autotransfusion.
Regulatory Classification
Identification
A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”
Predicate Devices
- Capiox® RX Hollow Fiber Oxygenator with/without Hardshell Reservoir (K130333)
Reference Devices
- Capiox® FX15 Advance Oxygenator with Integrated Filter and Hardshell Reservoir (K151389)
Related Devices
- K151791 — Terumo Capiox FX15 Advance Oxygenator with Integrated Arterial Filter and Reservoir, Terumo Capiox FX25 Advance Oxygenator with Integrated Arterial Filter and Reservoir · Terumo Cardiovascular Systems Corporation · Nov 19, 2015
- K130333 — CAPIX RX HOLLOW FIBER OXYGENATOR WITH/WITHOUT HARDSHELL RESERVOIR · Terumo Cardiovascular Systems Corp. · Mar 13, 2013
- K153140 — Terumo Capiox SX18 Oxygenator/ Hardshell Reservoir, Terumo Capiox SX18 Oxygenator/ Hardshell Reservoir with Xcoating, Terumo Capiox SX25 Oxygenator/Hardshell Reservoir, Terumo Capiox SX25 Oxygenator/Hardshell Reservoir with Xcoating · Terumo Cardiovascular Systems Corp. · Dec 2, 2015
- K022115 — CAPIOX RX05 HOLLOW FIBER OXYGENATOR W/WITHOUT HARDSHELL RESERVOIR · Terumo Corp. · Sep 27, 2002
- K962667 — CAPIOX SX25 HOLLOW FIBER OXYGENATOR WITH DETACHABLE HARDSHELL RESERVOIR · Terumo Medical Corp. · Oct 7, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, which is a traditional symbol of medicine. The symbol is composed of three human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 03, 2015
Terumo Cardiovascular Systems Corporation Tierra Brown Regulatory Affairs Specialist 125 Blue Ball Road Elkton, Maryland 21921
Re: K153213
Trade/Device Name: Capiox RX Hollow Fiber Oxygenator with/without Hardshell Reservoir Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ, DTN, DTR, DTP Dated: November 4, 2015 Received: November 5, 2015
Dear Tierra Brown:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hilleman
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for Terumo Corporation. The logo consists of a stylized letter 'T' inside a circle, with the word 'TERUMO' written vertically next to it. The logo is in a teal color.
# novessaringinussipolises in 1 – koimans
| Indications for Use | 510(k) Number |
|-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Unknown at time of Submission |
| Device Name | Indications for Use |
| | Capiox® RX Hollow Fiber Oxygenator with/without Hardshell Reservoir |
| | <b>Indications for Use as presented in the Instructions for Use:</b><br>The CAPIOX RX is intended to be used during open heart surgical procedures<br>requiring cardiopulmonary bypass for periods up to 6 hours.<br>The CAPIOX RX25 is used with patients when required blood flow rate will not<br>exceed 7 L/min. The CAPIOX RX15 is for use with patients when required blood<br>flow rate will not exceed 5 L/min (4L/min if using product codes CX*RX15RW30 or<br>CX*RX15RE30). The CAPIOX RX Hardshell Reservoir is also intended for use in<br>vacuum assisted venous drainage procedure, in post-operative chest drainage and<br>autotransfusion procedures to aseptically return the blood to the patient for blood<br>volume replacement. |
| | <b>Indications for Use as described in the 510(k):</b><br>The CAPIOX® RX Hollow Fiber Oxygenators with/without Hardshell Reservoir are<br>intended to be used to exchange gases between blood and a gaseous<br>environment to satisfy the gas exchange needs of a patient during<br>cardiopulmonary bypass surgery.<br>The integral heat exchanger is used to warm or cool blood and/or perfusion fluid as<br>it flows through the device.<br>The (detachable) hardshell reservoir(s) is (are) used to store blood during extra-<br>corporeal circulation from both venous line and the cardiotomy line (via gravity or<br>vacuum assisted venous drainage procedures). The reservoir contains a venous<br>section that is comprised of a filter and defoamer to facilitate air bubble removal.<br>The cardiotomy section of the reservoir contains a filter to remove particulate<br>matter and a defoamer to facilitate air bubble removal. The reservoir may also be<br>used for Post-Operative Chest Drainage Procedures.<br>The CAPIOX® RX Oxygenators with/without Hardshell Reservoirs can be used in<br>procedures lasting up to 6 hours.<br>The CAPIOX® RX15 is for use with patients when the required blood flow rate will<br>not exceed 5.0 L/min when used with a 4 Liter Reservoir; and when the required<br>blood flow rate will not exceed 4.0 L/min when used with a 3 Liter Reservoir).<br>The CAPIOX® RX25 is for use with patients when the required blood flow rate will<br>not exceed 7.0 L/min. |
| Prescription Use<br>(Part 21 CFR 801 Subpart D)<br>XX | OR<br>Over-The-Counter Use<br>(Part 21 CFR 807 Subpart C) |
Concurrence of CDRH, Office of Device Evaluation (ODE)
(addan) il asa sahtona no sumitinos - anij siht wojiga attirim ton od asaaja)
{3}------------------------------------------------
| Image: Terumo logo TERUMO |
|---------------------------------|
| SECTION 5 – 510(k) Summary |
| Capiox® RX Oxygenator/Reservoir |
| 5.1 Submitter Information | | |
|-----------------------------------|-------------------------------------------|--|
| Date of Preparation | November 3, 2015 | |
| Name | Terumo Cardiovascular Systems Corporation | |
| Address | 125 Blue Ball Road<br>Elkton, MD 21921 | |
| Establishment Registration Number | 1124841 | |
| Name of Contact Person | Tierra A. Brown, MSRA, RAC | |
| Contact Person's Title | Regulatory Affairs Specialist | |
| Phone Number | 800-262-3304 ext. 7021 | |
| Fax Number | 410-392-7171 | |
| Email | tierra.brown@terumomedical.com | |
{4}------------------------------------------------
| (T TERUMO |
|---------------------------------|
| SECTION 5 – 510(k) Summary |
| Capiox® RX Oxygenator/Reservoir |
| 5.2 Name of Device | |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Type | Blood Oxygenator with and/or without Blood<br>Reservoir<br>Defoamer<br>Heat Exchanger |
| Device Common Name | Blood Reservoir<br>Blood Gas Oxygenator<br>Defoamer<br>Heat Exchanger |
| Device Trade Name | Capiox® RX Hollow Fiber Oxygenator with/without<br>Hardshell Reservoir |
| Predicate Device | Capiox® RX Hollow Fiber Oxygenator with/without<br>Hardshell Reservoir; Cleared by FDA with K130333 |
| Classification Regulation | Reservoir: 21 CFR § 870.4400<br>Oxygenator: 21 CFR § 870.4350<br>Defoamer: 21 CFR § 870.4230<br>Heat Exchanger: 21 CFR § 870.4240 |
| Device Risk Classification | Reservoir: Class II<br>Oxygenator: Class II<br>Defoamer: Class II<br>Heat Exchanger: Class II |
| FDA Product Codes | Reservoir: DTN<br>Oxygenator: DTZ<br>Defoamer: DTP<br>Heat Exchanger: DTR |
| Reason for 510(k) Submission | Modification to Hardshell Reservoir |
| Substantial Equivalence Statement | The device submitted in this 510(k) maintains<br>characteristics that are substantially equivalent in<br>intended use, design, technology/principles of<br>operation, materials and specifications to the<br>predicate device. |
{5}------------------------------------------------
# Device Information
# 5.3 Intended Use:
The intended use remains the same for the modified device and predicate device.
The CAPIOX® RX Hollow Fiber Oxygenators with/without Hardshell Reservoir are intended to be used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery.
The integral heat exchanger is used to warm or cool blood and/or perfusion fluid as it flows through the device.
The (detachable) hardshell reservoir(s) is (are) used to store blood during extra-corporeal circulation from both venous line and the cardiotomy line (via gravity or vacuum assisted venous drainage procedures). The reservoir contains a venous section that is comprised of a filter and defoamer to facilitate air bubble removal. The cardiotomy section of the reservoir contains a filter to remove particulate matter and a defoamer to facilitate air bubble removal. The reservoir may also be used for Post-Operative Chest Drainage Procedures.
The CAPIOX® RX Oxygenators with/without Hardshell Reservoirs can be used in procedures lasting up to 6 hours.
The CAPIOX® RX15 is for use with patients when the required blood flow rate will not exceed 5.0 L/min when used with a 4 Liter Reservoir; and when the required blood flow rate will not exceed 4.0 L/min when used with a 3 Liter Reservoir).
The CAPIOX® RX25 is for use with patients when the required blood flow rate will not exceed 7.0 L/min.
# 5.4 Principles of Operation and Technology:
The principles of operation and technology remain the same for the modified device and predicate device.
The modified and predicate Capiox® RX device oxygenator utilizes a porous fiber technology to facilitate the transfer of gases between a blood-phase environment and a gas-phase environment for the intent of satisfying the gas exchange needs of a patient during cardiopulmonary bypass surgery. A wound fiber bundle offers the porous membrane surface to sufficiently permit the movement of gases through the walls of the hollow fibers via diffusion.
The modified and predicate Capiox® RX device have an integrated heat exchanger that is comprised of stainless steel encased in a polycarbonate housing. The stainless steel acts as a heat transfer material that permits heat that is generated from a temperature controlled external water bath to transverse across the walls of the stainless steel to effect the necessary temperature change upon circulating blood.
With respect to the filtration of blood, the modified and predicate Capiox® RX Reservoir relies upon mechanical entrapment of particulates and emboli within the filter mesh as a means to remove those particulates from the blood.
{6}------------------------------------------------
# 5.5 Design and Materials:
.
The design of the modified and predicate Capiox® RX devices both utilize an integrated oxygenator/heat exchanger module that provides for temperature control of blood as it enters the oxygenation phase. Each of the devices also utilizes a hardshell reservoir that is used to collect and store blood during a cardiopulmonary bypass procedure. The reservoirs each provide filtration of venous and cardiotomy blood as it enters the reservoir.
With respect to the design of the Oxygenator, the modified Capiox® RX Oxygenator device is unaffected by the changes being incorporated at this time. The subject of this Special 510(k) is a modification being made to the Hardshell Reservoir.
# Differences in the design of the Capiox® RX devices are as follows:
- The design of the modified Capiox® RX device will include:
- flared venous inlet drop tubes o
- o The RX15 3-Liter Reservoir's drainage cage, that houses the flared venous inlet drop tubes, will be modified
- A full-length graduated side scale label (decal) o
- A modified 1/4" x 3/8" winged adapter (adapter is for the RX15 3-Liter Reservoir o only)
- The intent of the flared venous inlet drop tubes is to help minimize air emboli generation . in the venous section of the reservoir.
- The intent of the modified RX15 3-Liter Reservoir's drainage cage is to accommodate . the modified flared venous inlet drop tubes.
- . The full-length graduated side scale label (decal) enhancement offers greater visibility of blood volume during cardiopulmonary bypass.
- The modified 1/4" x 3/8" winged adapter (adapter for RX15 3-Liter Reservoir only) is ● used to provide attachment to the 1/4" arterial and venous line connections.
- The design of the predicate Capiox® RX device has: .
- straight venous inlet drop tubes o
- The RX15 3-Liter Reservoir's drainage cage is the original size from initial FDA o clearance.
- A half-length graduated side scale label (decal) O
- A 1/4" x 3/8" adapter o
# Difference in the materials of the Capiox® RX devices are as follows:
- o The materials used in the construction of the modified Capiox® RX Hollow Fiber Oxygenator with/without Hardshell Reservoir device's venous inlet drop tubes along with the auxiliary drop tube, quick disconnect tubing, sampling manifold tubing, and purge line tubing will be PVC tubing with Tris-(2-Ethylhexyl) Trimellitat, identified as TOTM plasticizer.1
- . The materials used in the construction of the predicate Capiox® RX Hollow Fiber Oxygenator with/without Hardshell Reservoir device's venous inlet drop tubes along with the auxiliary drop tube, quick disconnect tubing, sampling manifold tubing, and
{7}------------------------------------------------
### Device Information
### 5.5 Design and Materials continued:
- purge line tubing is PVC tubing with Di-(2-ethylhexyhl) phthalate, identified as DEHP ● plasticizer.
- The intent of the PVC tubing plasticizer change to TOTM plasticizer is to address o potential safety concerns associated with the use of DEHP as a plasticizer.
The materials used in the construction of the modified and predicate Capiox® RX Hollow Fiber Oxygenator/Hardshell Reservoir devices include, but are not limited to, nylon, polycarbonate, silicone, stainless steel, polyvinyl chloride, polyurethane, polyester, polypropylene, polyethylene and X-Coating™.
Terumo Cardiovascular Systems concludes that the differences between the modified device and the predicate device do not affect the intended use of the device nor do they affect safety and effectiveness of the device when used as labeled.
### 5.6 Performance Evaluations:
Clinical studies involving patients are not necessary to demonstrate substantial equivalence of the modified Capiox® RX device to the predicate Capiox® RX device. Substantial equivalence is demonstrated with the following in-vitro performance evaluations:
- Air Handling Performance ●
- Hemolysis ●
- Venous Reservoir Pressure Drop .
- Venous Defoaming ●
- Leak Testing ●
- Winged Adapter Circulation Testing .
- 3-Liter Drainage Cage Dimensional Testing .
- Connections Testing ●
- Package Testing ●
- Aged Evaluation .
<sup>1</sup> PVC tubing with Tris-(2-Ethylhexy) Trimellitate, identified as TOTM plasticizer, is used with Terumo's Capiox® FX15 Advance Oxygenator with Integrated Filter and Hardshell Reservoir, K151389.
{8}------------------------------------------------
### Device Information
### 5.7 Substantial Equivalence Comparison:
In demonstrating substantial equivalence of the modified Capiox® RX Hollow Fiber Oxygenator with/without Hardshell Reservoir to the predicate Capiox® RX Hollow Fiber Oxygenator with/without Hardshell Reservoir, Terumo has considered the following aspects to ensure equivalence between the predicate and the subject devices:
- Intended use .
- Target Population .
- Duration of use ●
- Product labeling ●
- Product design
- Materials .
- Principles of Operation and Technology ●
- Device Performance ●
## 5.8 Substantial Equivalence Statement:
The modified Capiox® RX Hollow Fiber Oxygenator with/without Hardshell Reservoir is substantially equivalent in intended use, target population, duration of use, labeling, design, materials, principles of operation and technology, and performance to the predicate Capiox® RX Hollow Fiber Oxygenator with/without Hardshell Reservoir. Any noted differences between the modified device and the predicate devices do not raise new issues of safety and effectiveness.
# 5.9 510(k) Summary Conclusion
### Conclusion
Based upon the comparative studies and analyses, Terumo Cardiovascular Systems concludes that the modified Capiox® RX Hollow Fiber Oxygenator with/without Hardshell Reservoir is substantially equivalent to the predicate Capiox® RX Hollow Fiber Oxygenator with/without Hardshell Reservoir. It is further concluded that any recognized differences noted during the assessments do not raise new issues of patient/user safety or product effectiveness.