K153204 · In2bones Sas · HTY · Feb 1, 2016 · Orthopedic
Device Facts
Record ID
K153204
Device Name
In2Bones Kirschner wire
Applicant
In2bones Sas
Product Code
HTY · Orthopedic
Decision Date
Feb 1, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The In2Bones® Kirschner wires are indicated for fixation of bone fractures, bone reconstruction, and as guide pins for insertion of other implants. The size of the In2Bones® Kirschner wire chosen should be adapted to the specific indication.
Device Story
In2Bones® Kirschner wire is a metallic orthopedic fixation device; used for bone fracture fixation, reconstruction, and as guide pins for other implants. Device consists of stainless steel wire available in four point styles (sharp, partially threaded, lanceolate, both ends sharp); diameters 0.8mm to 2.5mm; lengths 70mm to 300mm. Operated by surgeons in hospitals or outpatient surgery centers using standard surgical power tools for insertion. Device provides mechanical stabilization of bone segments; facilitates surgical procedures. Single-use only; supplied sterile (gamma radiation) or non-sterile (requires steam sterilization).
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence is supported by design, material, and size comparisons to legally marketed predicate devices and conformance to ISO 5838-1 (2013).
Technological Characteristics
Material: Stainless steel 316LVM (ISO 5832-1, ASTM F138-13). Form factor: Metallic wire, 0.8-2.5mm diameter, 70-300mm length. Point styles: Sharp, partially threaded, lanceolate, double-ended sharp. Sterilization: Gamma radiation (sterile) or steam (non-sterile). No coating. Mechanical fixation principle.
Indications for Use
Indicated for patients requiring bone fracture fixation, bone reconstruction, or guide pins for implant insertion.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Syntec-Taichung Non-sterile Kirschner Wires and Steinmann Pins (K983121)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 1, 2016
In2Bones SAS % Norman F. Estrin, Ph.D. Managing Partner Estrin Consulting Group LLC 9109 Copenhaver Drive Potomac, Maryland 20854
Re: K153204
Trade/Device Name: In2Bones® Kirschner Wire Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HTY, JDW Dated: November 1, 2015 Received: November 4, 2015
Dear Dr. Estrin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## 005_Indications for Use.pdf
## Indications for Use
| 510(k) Number (if known): | Not yet assigned K153204 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | In2Bones® Kirschner wire |
| Indications For Use: | The In2Bones® Kirschner wires are indicated for fixation of bone fractures, bone reconstruction, and as guide pins for insertion of other implants. The size of the In2Bones® Kirschner wire chosen should be adapted to the specific indication. |
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for IN2BONES. The logo is set on a gray oval background. The text "IN2BONES" is in white, with the "2" in orange.
## 510(k) SUMMARY For In2Bones® Kirschner wire
| Sponsor<br>identification | In2Bones SAS<br>28 chemin du Petit Bois<br>69130 Ecully – France<br>Phone: +33.4.72.29.26.26<br>Fax: +33.4.72.29.26.29 |
|-------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>registration number | 3010470577 |
| Date of preparation | October 28, 2015 |
| Contact person | Norman F. Estrin, Ph.D.<br>Estrin Consulting Group LLC<br>9109 Copenhaver Drive<br>Potomac, MD 20854<br>Phone: (301) 279-2899<br>Cell: 240-994-9999<br>Email: estrin@yourFDAconsultant.com |
| Authorized Agent<br>in the United States | Norman F. Estrin, Ph.D.<br>Estrin Consulting Group LLC<br>9109 Copenhaver Drive<br>Potomac, MD 20854<br>Phone: (301) 279-2899<br>Cell: 240-994-9999<br>Email: estrin@yourFDAconsultant.com |
| Proprietary Name | In2Bones® Kirschner wire |
| Common name | Osteosynthesis wire |
| Device<br>classification<br>regulation | 21 CFR 888.3040: Smooth or threaded metallic bone fixation<br>fastener<br>Class II |
| Device Product<br>Code and Panel | HTY: Pin, Fixation, Smooth<br>JDW: Pin, Fixation, Threaded<br>87 orthopedics |
| Device Description | The In2Bones® Kirschner wire is a metallic wire available in four<br>point styles: sharp, partially threaded, lanceolate, both ends sharp.<br>One part is fixed on standard surgical power tool equipment for<br>insertion. |
| | Sizes: The In2Bones® Kirschner wire is available in various<br>diameters (0.8mm to 2.5mm) and length (70mm to 300mm). |
| | Material: The In2Bones® Kirschner wire is manufactured from<br>stainless steel 316LVM, according to ISO 5832-1 and ASTM<br>F138-13. It does not have any coating. |
| | Single use: The In2Bones® Kirschner wire is designed for single<br>use only. |
| | Sterilization: The In2Bones® Kirschner wire is supplied sterile and<br>non-sterile. The sterile In2Bones® Kirschner wire is sterilized<br>using gamma radiation. The non-sterile In2Bones® Kirschner wire<br>must be steam sterilized before use. |
| | Place of use: The In2Bones® Kirschner wire is indicated for use in<br>a hospital, or outpatient surgery center, where sterile field may be<br>created and maintained. |
| Predicate Devices | NEWDEAL K WIRE (K022599)<br>Syntec-Taichung Non-sterile Kirschner Wires and Steinmann Pins<br>(K983121) |
| Indications for use: | The In2Bones® Kirschner wires are indicated for fixation of bone<br>fractures, bone reconstruction, and as guide pins for insertion of<br>other implants. The size of the In2Bones® Kirschner wire chosen<br>should be adapted to the specific indication. |
| Comparison of the<br>indications for use<br>with the predicate<br>devices: | The indications for use for the In2Bones® Kirschner wire are<br>similar to the predicate devices Syntec Taichung Non-sterile<br>Kirschner Wires and Steinmann Pins (K983121) and NEWDEAL<br>K WIRE (K983121). |
| Comparison of<br>Technological<br>characteristics with<br>the predicate<br>devices: | The In2Bones® Kirschner wire is similar to the predicate devices<br>in regards to insertion, design, size ranges, and materials:<br>- They all are intended for surgical implantation into bone for<br>longer than 30 days.<br>- The In2Bones® Kirschner wire has identical design as the<br>Syntec-Taichung Non-sterile Kirschner Wires and Steinmann Pins<br>(K983121) predicate, except for the partially threaded self-drilling<br>tip. The In2Bones® Kirschner wire has the identical design as the<br>NEWDEAL K WIRE (K022599) predicate, except for the<br>lanceolate self-drilling tip<br>- The In2Bones® Kirschner has similar size range when compared<br>to the predicate devices.<br>- The In2Bones® Kirschner wire has similar raw material, when<br>compared to the NEWDEAL K WIRE (K022599) and Syntec-<br>Taichung Non-sterile Kirschner Wires and Steinmann Pins<br>(K983121): all are manufactured from the same type of stainless<br>steel and meet appropriate ASTM standards. |
| Substantial<br>Equivalence<br>Summary: | The In2Bones® Kirschner wire has similar technological<br>characteristics when compared to the predicate devices<br>The indications for use, design and sizes available are similar and<br>any differences do not impact safety and effectiveness. |
| Summary<br>Performance Data | The In2Bones® Kirschner wires are stainless steel wires with<br>diameters and lengths comparable to those featured in the<br>predicate device systems. The design and indications of the<br>In2Bones® Kirschner wires are substantially equivalent to the<br>predicate devices identified in the 510(k) submission. No new<br>materials or processes are used in the development of this implant.<br>Moreover, the In2Bones® Kirschner wires conform to the<br>international standard ISO 5838-1 (2013). Testing, therefore, is not<br>needed to demonstrate that the subject devices are substantially<br>equivalent to other legally marketed Kirschner wires. |
| CONCLUSION | Based on the evaluations performed, the design and indications of<br>the In2Bones® Kirschner wires are substantially equivalent to the<br>predicate devices identified in the 510(k) submission. No new<br>materials or processes are used in the development of this implant.<br>The In2Bones® Kirschner wires are acceptable for the applications<br>requested. |
{4}------------------------------------------------
{5}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.