The IvyAIR Meniscus System is intended for use as a suture retention device to facilitate percutaneous or endoscopic soft tissue procedures. The IvyAIR Meniscus System is indicated for use in meniscal repairs and allograft transplant procedures. The IvyAIR Meniscus System device is intended to be used for anchoring the allograft to the meniscal rim during allograft transplant procedures.
Device Story
All-inside meniscus repair device; consists of two non-absorbable polymer implants pre-tied with 2-0 non-absorbable suture; pre-loaded into needle delivery system. Available in curved, straight, and reverse-curved configurations for surgical access to specific meniscus zones. Used by surgeons during percutaneous or endoscopic procedures to anchor soft tissue or allografts to the meniscal rim. Device is sterile and single-use. Provides mechanical fixation to facilitate tissue healing/transplant stability.
Clinical Evidence
No clinical data. Bench testing only. Testing included implant strength to failure, dynamic and static tensile loading, and comprehensive biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, reactivity, systemic toxicity, pyrogenicity, genotoxicity, and intramuscular implantation). All acceptance criteria met.
Technological Characteristics
Non-absorbable polymer implants; 2-0 non-absorbable suture; needle delivery system. Available in curved, straight, and reverse-curved configurations. Biocompatibility per ISO 10993-1. Sterile, single-use.
Indications for Use
Indicated for patients undergoing meniscal repair or meniscal allograft transplant procedures requiring soft tissue anchoring to the meniscal rim.
Regulatory Classification
Identification
Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Predicate Devices
Smith & Nephew FAST-FIX 360 Meniscal Repair System (K121861)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird design with three profiles facing right, suggesting a sense of forward movement or progress. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 14, 2016
Ivy Sports Medicine, LLC Mr. John Dichiara Senior Vice President Quality Assurance, Clinical and Regulatory Affairs 545 Penobscot Drive Redwood City, California 94063
Re: K153087
Trade/Device Name: Ivyair Meniscus System, Curved, Ivyair Meniscus System, Straight, Ivyair Meniscus System, Reverse Curved Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable Poly(Ethylene Terephthalate) Surgical Suture Regulatory Class: Class II Product Code: GAT Dated: October 21, 2015 Received: October 26, 2015
Dear Mr. Dichiara:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Ivy Sports Medicine, LLC. The logo consists of a stylized green image of three leaves on the left, followed by the text "Ivy Sports Medicine, LLC" in black. The text is in a bold, sans-serif font.
SECTION 5
## INDICATIONS FOR USE
510(k) Number (if known): K153087
Device Name: IvyAIR Meniscus System
#### Indications for Use:
The IvyAIR Meniscus System is intended for use as a suture retention device to facilitate percutaneous or endoscopic soft tissue procedures. The lvyAIR Meniscus System is indicated for use in meniscal repairs and allograft transplant procedures. The lvyAlR Meniscus System device is intended to be used for anchoring the allograft to the meniscal rim during allograft transplant procedures.
Prescription Use
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
### SECTION 6
## 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### 510(k) Number: K
#### Applicant Information:
| Owner Name: | Ivy Sports Medicine, LLC |
|-------------|-----------------------------------------------|
| Address: | 545 Penobscot Drive<br>Redwood City, CA 94063 |
| Main Phone: | 650-562-0800 |
| Fax: | 650-562-0808 |
Establishment Registration Number: 2956141
| Contact Person: | John Dichiara |
|-----------------|---------------------|
| | Phone: 917-439-2597 |
| | Fax: 650-562-0808 |
Date Prepared: October 21, 2015
#### Device Information:
- Classification: Class II
- Trade Name: IvyAIR Meniscus System
- Common name: Meniscus Repair Device
- Classification name: Nonabsorbable poly(ethylene terephthalate) surgical suture
- Regulation number: 21 CFR §878.5000
- Product Code: GAT
#### Predicate Device:
The IvyAlR Meniscus System is substantially equivalent in Intended Use, Fundamental Scientific Technology and Performance to the following legally marketed device in commercial distribution: Smith & Nephew FAST-FIX 360 Meniscal Repair System cleared under K121861.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for Ivy Sports Medicine, LLC. The logo features a stylized green graphic to the left of the text. The text "Ivy Sports Medicine, LLC" is written in a bold, sans-serif font.
#### Device Description:
The IvyAIR Meniscus System is an all-inside meniscus repair device. Each device includes two non-absorbable polymer implants, pre-tied with 2-0 non-absorbable suture and pre-loaded into a needle delivery system. The IvyAIR System is available in three needle configurations: Curved. Straight and Reverse Curve to allow the surgeon a choice for access to the desired zone of the meniscus. The IvyAIR Meniscus System is provided sterile for single use only.
#### Intended Use:
The IvyAIR Meniscus System is intended for use as a suture retention device to facilitate percutaneous or endoscopic soft tissue procedures. The IvyAIR Meniscus System is indicated for use in meniscal repairs and allograft transplant procedures. The IvyAIR Meniscus System device is intended to be used for anchoring the allograft to the meniscal rim during allograft transplant procedures.
#### Comparison to Predicate Device(s):
The IvyAIR Meniscus System has a similar intended use and similar technological characteristics to the predicate device: Smith & Nephew FAST-FIX 360 Meniscal Repair System. Any differences identified do not raise new questions of safety or effectiveness, as supported by biomechanical, biocompatibility and bench testing.
#### Technological Characteristics/Performance Data:
The IvyAIR Meniscus System is substantially equivalent to the predicate device in intended use, fundamental scientific technology, and performance specifications. Design verification testing was performed to verify that the performance of the IvvAlR Meniscus System remains substantially equivalent to the predicate device. The device has been tested for biocompatibility according to ISO10993-1 and was determined to be biocompatible. Testing performed on the IvyAIR Meniscus System included the following:
| Bench Testing | Biocompatibility Testing |
|------------------------------------|----------------------------------------------------------|
| Implant strength test to failure | MEM Elution Using L-929 Mouse Fibroblast Cells |
| Dynamic and static tensile loading | Guinea Pig Maximization Sensitization Test |
| | Intracutaneous Reactivity Test |
| | Acute Systemic Injection Test |
| | Material mediated pyrogen |
| | Subacute (14 Day) Intraperitoneal Toxicity Study In Rats |
| | Subchronic (14 Day) Intravenous Toxicity Study In Rats |
| | including histopathology |
| | Bacterial Mutagenicity Test (Ames Assay) |
| | In Vitro Mouse Lymphoma Assay |
| | In Vivo Mouse Micronucleus Assay |
| | Intramuscular Implantation Tests (2 weeks and 13 weeks) |
All of the pre-determined acceptance criteria were met with passing results.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Ivy Sports Medicine, LLC. The logo consists of a stylized green image of leaves on the left, followed by the text "Ivy Sports Medicine, LLC" in black. The text is in a bold, sans-serif font.
#### Clinical Testing:
Clinical evaluation is not required for this device.
#### Substantial equivalence:
The IvyAIR Meniscus System has the following similarities to the Smith & Nephew FAST-FIX 360 Meniscal Repair System predicate device cleared under K121861.
- has the same intended use and indication for use,
- . has the same technological characteristics,
- . has the same principles of operation,
- . has equivalent performance characteristics, and
- . has the same sterilization process.
#### Summary
In summary, the IvyAIR Meniscus System subject to this submission is as safe and effective as the Smith & Nephew FAST-FIX 360 Meniscal Repair System. It has the same indication for use, the same technological characteristics, and the same principles of operation as the Smith & Nephew FAST-FIX 360 Meniscal Repair System. The differences between the IvyAIR Meniscus System and the Smith & Nephew FAST-FIX 360 Meniscal Repair System raise no new issues of safety or effectiveness. Performance data demonstrate that the IvyAlR Meniscus System is as safe and effective as the Smith & Nephew FAST-FIX 360 Meniscal Repair System and is therefore substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.