← Product Code [NHV](/productcode/NHV) · K153084

# Nephros S100 (K153084)

_Nephros, Inc. · NHV · Apr 8, 2016 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K153084

## Device Facts

- **Applicant:** Nephros, Inc.
- **Product Code:** [NHV](/productcode/NHV.md)
- **Decision Date:** Apr 8, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5665
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The S100 Point of Use Filter is intended to be used to filter EPA quality drinking water. The filter retains bacteria. By retaining bacteria in water for washing and drinking, the filter may aid in infection control. The filter produces water that is suitable for cleaning of equipment used in medical procedures and washing of surgeon's hands. The filter is not intended to provide water that can be used as a substitute for USP sterile water.

## Device Story

S100 Point of Use Filter is a hollow fiber microfiltration device; utilizes size exclusion to retain bacteria from municipal tap water. Used in clinical settings for washing hands and cleaning medical equipment; aids in infection control. Device consists of polysulfone hollow fiber membrane in ABS housing. Operates via water pressure (up to 75 psi); flow rate 4 L/min at 20 psi. Provides filtered water for non-sterile medical applications; not a substitute for USP sterile water. Used by healthcare personnel; provides benefit by reducing bacterial load in water used for clinical tasks.

## Clinical Evidence

Bench testing only. Performance data includes flow rate vs. pressure drop, simulated use life, bacteria challenge testing (>10^9 reduction), and biocompatibility testing.

## Technological Characteristics

Polysulfone hollow fiber microfiltration membrane; ABS plastic housing. Size exclusion mechanism. Max inlet pressure 75 psi. Flow rate 4 L/min at 20 psi. Use life up to 3 months. Non-powered, mechanical filtration.

## Regulatory Identification

A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

## Predicate Devices

- DSU-H and SSU-H Ultrafilters ([K141731](/device/K141731.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to create a sense of depth and connection.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 8, 2016

Nephros, Inc. James Summerton Director Product Development 41 Grand Ave River Edge, NJ 07661

Re: K153084

> Trade/Device Name: S100 Point of Use Filter Regulation Number: 21 CFR§ 876.5665 Regulation Name: Water Purification System for Hemodialysis Regulatory Class: II Product Code: NHV Dated: March 7, 2016 Received: March 10, 2016

Dear James Summerton,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

## Joyce M. Whang -S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known): K153084 Device Name: S100 Point of Use Filter

Indications For Use:

The S100 Point of Use Filter is intended to be used to filter EPA quality drinking water. The filter retains bacteria. By retaining bacteria in water for washing and drinking, the filter may aid in infection control. The filter produces water that is suitable for cleaning of equipment used in medical procedures and washing of surgeon's hands. The filter is not intended to provide water that can be used as a substitute for USP sterile water.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use ____________ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

| Submitter:                                                                  | Nephros Inc.<br>41 Grand Ave<br>River Edge, NJ 07661<br>Establishment Registration # 3003337893                                                                                                                                                                                                                                                                                                                                                                                |                                | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |               | Subject Device              | Predicate Device                   |
|-----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|-----------------------------|------------------------------------|
| Contact Person                                                              | Jim Summerton, Director Product Development<br>41 Grand Ave<br>River Edge, NJ 07661<br>201-343-5202 (p)<br>201-343-5207 (f)<br>summerton@nephros.com                                                                                                                                                                                                                                                                                                                           | 510(k)<br>Number               | K153084                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | K141731                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Materials     |                             |                                    |
| Date Prepared                                                               | March 8, 2016                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Manufacturer<br>Name           | Nephros Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Nephros Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Casing        | ABS                         | ABS                                |
| Trade Name                                                                  | S100 Point of Use Filter                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Indications<br>for Use         | The S100 Point of Use Filter is<br>intended to be used to filter EPA<br>quality drinking water. The filter<br>retains bacteria. By retaining<br>bacteria in water for washing and<br>drinking, the filter may aid in<br>infection control. The filter<br>produces water that is suitable for<br>cleaning of equipment used in<br>medical procedures and washing<br>of surgeon's hands. The filter is<br>not intended to provide water that<br>can be used as a substitute for<br>USP sterile water. | The DSU-H and SSU-H Ultrafilters<br>are intended to be used to filter EPA<br>quality drinking water. The filters<br>retain bacteria, viruses and<br>endotoxin. By providing ultrapure<br>water for washing and drinking, the<br>filters aid in infection control. The<br>filters produce water that is suitable<br>for wound cleansing, cleaning of<br>equipment used in medical<br>procedures and washing of<br>surgeon's hands. The filters are not<br>intended to provide water that can be<br>used as a substitute for USP sterile<br>water. | Filter        | Polysulfone Microfiltration | Polysulfone Ultrafiltration Hollow |
| Proposed Class                                                              | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Operation                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Element(s)    | Hollow Fiber                | Fiber                              |
| Classification Name<br>and Number                                           | 21 CFR Part 876.5665 Water Purification Device, General<br>Medical Use                                                                                                                                                                                                                                                                                                                                                                                                         | Feed Water<br>Source           | Municipal tap water                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Municipal tap water                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Removal       |                             |                                    |
| Product Code                                                                | NHV                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Use Life                       | Up to 3 months                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Up to 3 Months (SSU-H)<br>Up to 6 Months (DSU-H)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Bacteria      | > 109                       | > 1011                             |
| Predicate Device                                                            | DSU-H and SSU-H Ultrafilters – K141731                                                                                                                                                                                                                                                                                                                                                                                                                                         | Maximum<br>Inlet Pressure      | 75 psi                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 75 psi (SSU-H)<br>100 psi (DSU-H)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Reduction     |                             |                                    |
| Device Description                                                          | The S100 Point of Use Filter is a hollow fiber microfilter that<br>retains bacteria from water used for washing and drinking.                                                                                                                                                                                                                                                                                                                                                  | Flow Rate vs.<br>Pressure Drop | 4 L/min at 20 psi                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 4.5 L/min at 20 psi (SSU-H)<br>4 L/min at 20 psi (DSU-H)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Virus removal | n/a                         | > 108                              |
| Intended Use                                                                | The S100 Point of Use Filter is intended to be used to filter EPA<br>quality drinking water. The filter retains bacteria. By retaining<br>bacteria in water for washing and drinking, the filter may aid in<br>infection control. The filter produces water that is suitable for<br>cleaning of equipment used in medical procedures and washing<br>of surgeon's hands. The filter is not intended to provide water<br>that can be used as a substitute for USP sterile water. | Endotoxin                      | n/a                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | > 105                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |               |                             |                                    |
| Summary of the<br>Technological                                             | The proposed device contains a polysulfone hollow fiber<br>membrane encased in an ABS housing that retains bacteria.                                                                                                                                                                                                                                                                                                                                                           | Reduction                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |               |                             |                                    |
| Characteristics                                                             | Removal of contaminants is through size exclusion.                                                                                                                                                                                                                                                                                                                                                                                                                             |                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |               |                             |                                    |
| Assessment of Non-clinical Performance<br>Data / Substantial<br>Equivalence | The S100 Point of Use Filter has been tested for performance.<br>The tests conducted include Flow Rate versus Pressure Drop,<br>Simulated Use Life, Bacteria Challenge Test and<br>Biocompatibility. The filters were found to be substantially<br>equivalent to the predicate DSU-H and SSU-H Ultrafilters<br>(K141731).                                                                                                                                                      |                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |               |                             |                                    |

{4}------------------------------------------------

{5}------------------------------------------------

## SUBJECT TO PREDCATE DEVICE COMPARISON TABLE

{6}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/device/K153084](https://fda.innolitics.com/device/K153084)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
