K153080 · Optovue, Inc. · HLI · Feb 11, 2016 · Ophthalmic
Device Facts
Record ID
K153080
Device Name
RTVue XR OCT Avanti with AngioVue Software
Applicant
Optovue, Inc.
Product Code
HLI · Ophthalmic
Decision Date
Feb 11, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1570
Device Class
Class 2
Indications for Use
The RTVue XR OCT Avanti with Normative Database is an optical coherence tomography system indicated for the in vivo imaging and measurement of the retinal nerve fiber layer, and optic disc as a tool and aid in the diagnosis and management of retinal diseases by a clinician. The RTVue XR OCT Avanti with Normative Database is also a quantitative tool for the comparison of retinal nerve fiber layer, and optic disk measurements in the human eve to a database of known normal subjects. It is intended for use as a diagnostic device to aid in the detection and management of ocular diseases. The RTVue XR OCT Avanti with AngioVue Software is indicated as an aid in the visualization of vascular structures of the retina and choroid.
Device Story
RTVue XR OCT Avanti is a computer-controlled ophthalmic imaging system using low-coherence interferometry to measure retinal tissue reflectivity. Device captures cross-sectional A-scan sequences to image cornea, anterior chamber, retinal nerve fiber layer, and optic disc. AngioVue software module adds non-invasive OCT angiography to visualize retinal/choroidal microvasculature using motion-contrast techniques without intravenous dyes. System includes proprietary Motion Correction Technology (MCT) to reduce artifacts from blinks/eye motion by minimizing differences between sequential scan volumes. Operated by clinicians in clinical settings; provides 3D images and quantitative measurements. Output aids clinicians in detecting/managing ocular diseases by comparing measurements to a normative database or visualizing vascular structures. Does not replace fluorescein angiography; does not show dye leakage/staining.
Clinical Evidence
Bench testing only. Software verified and validated per IEC 62304:2006 and FDA guidance. Risk analysis performed per EN ISO 14971. Clinical utility supported by qualitative image comparisons of retinal disease cases between RTVue XR OCT Avanti with AngioVue and fluorescein angiography to demonstrate non-invasive 3D visualization of microvasculature.
Technological Characteristics
Computer-controlled ophthalmic imaging system; low-coherence interferometer; 70,000 A-lines/second line-scan camera. Hardware identical to predicate. Software-based OCT angiography using motion-contrast. Proprietary Motion Correction Technology (MCT). Connectivity via standard GUI. Power: 100-120 VAC or 200-240 VAC. Class II device.
Indications for Use
Indicated for in vivo imaging and measurement of retinal nerve fiber layer and optic disc to aid clinicians in diagnosis and management of retinal and ocular diseases. Also indicated for visualization of retinal and choroidal vascular structures.
Regulatory Classification
Identification
An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.
Special Controls
*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, possibly black or dark blue. The image is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 11, 2016
Optovue, Inc. Mr. Edward J. Sinclair VP, Regulatory and Quality Affairs 2800 Bavview Drive Fremont, CA 94538
Re: K153080
Trade/Device Name: RTVue XR OCT with Avanti with AngioVue Software Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: HLI, OBO Dated: January 13, 2016 Received: January 14, 2016
Dear Mr. Sinclair:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Kesia Alexander
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known): K153080
Device Name
RTVue XR OCT Avanti with AngioVue Software
Indications for Use (Describe)
The RTVue XR OCT Avanti with Normative Database is an optical coherence tomography system indicated for the in vivo imaging and measurement of the retinal nerve fiber layer, and optic disc as a tool and aid in the diagnosis and management of retinal diseases by a clinician. The RTVue XR OCT Avanti with Normative Database is also a quantitative tool for the comparison of retinal nerve fiber layer, and optic disk measurements in the human eve to a database of known normal subjects. It is intended for use as a diagnostic device to aid in the detection and management of ocular diseases.
The RTVue XR OCT Avanti with AngioVue Software is indicated as an aid in the visualization of vascular structures of the retina and choroid.
#### Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary (K153080)
## Submitter Information
| Company: | Optovue, Inc.<br>2800 Bayview Drive<br>Fremont, CA 94538<br>(510) 897-1575<br>Edward J. Sinclair<br>Vice President, Regulatory and Quality Affairs |
|----------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | January 12, 2016 |
## Device Name and Classification
| Common Name: | Optical Coherence Tomography | |
|----------------------|---------------------------------------------|--|
| Proprietary Name: | RTVue XR OCT Avanti with AngioVue™ Software | |
| Classification Name: | Tomography, Optical Coherence | |
| Product Code: | OBO, HLI | |
| Regulation Number: | 886.1570 | |
| Class: | II | |
## Predicate Device
Optovue, Inc. RTVue XR OCT (K120238)
## Reference Device
Carl Zeiss Meditec, Inc. CIRRUS HD-OCT with Software Version 8 (K150977)
#### Intended Use
The RTVue XR OCT Avanti with Normative Database is an optical coherence tomography system indicated for the in vivo imaging and measurement of the retinal nerve fiber layer, and optic disc as a tool and aid in the diagnosis and management of retinal diseases by a clinician. The RTVue XR OCT Avanti with Normative Database is also a quantitative tool for the comparison of retina, retinal nerve fiber layer, and optic disc measurements in the human eye to a database of known normal subjects. It is intended for use as a diagnostic device to aid in the detection and management of ocular diseases.
The RTVue XR OCT Avanti with AngioVue™ software is indicated as an aid in the visualization of vascular structures of the retina and choroid.
{5}------------------------------------------------
## Device Description
- 1. RTVue XR OCT Avanti (without AngioVue™ software module)
The RTVue XR OCT Avanti is a non-invasive diagnostic device for imaging the cornea, anterior chamber, and measurement of the retinal nerve fiber layer and optic disc with micrometer range resolution as a tool and aid in the diagnosis and management of retinal disease. Imaging and measurements include but are not limited to the internal limiting membrane (ILM), the retinal nerve fiber layer (RNFL), the ganglion cell complex (GCC), the retinal pigment epithelium (RPE), the outer retinal thickness, the total retinal thickness and optic disc structures including the cup and neuroretinal rim as an aid in the diagnosis and management of retinal disease. The measurements for the ILM and RPE are height measurements relative to the RPE reference plane. The RNFL, GCC and outer retinal thickness and total retinal thickness are thickness measurements where RNFL is the thickness of the RNFL layer, the GCC is the thickness from the ILM to the inner plexiform layer (IPL), the outer retinal thickness is the thickness from the IPL to the RPE, and total retinal thickness is the thickness from the ILM to the RPE.
The RTVue XR OCT Avanti is a computer controlled ophthalmic imaging system. The device scans the patient's eye using a low coherence interferometer to measure the reflectivity of the retinal tissue. The cross sectional retinal tissue structure is composed of a sequence of A-scans. It has a traditional patient and instrument interface like most ophthalmic devices. The computer has a graphic user interface for acquiring and analyzing the image. The line-scan camera operates at approximately 70,000 A-lines per second.
The RTVue XR OCT Avanti offers three scan types: Retina, Glaucoma, and Cornea. For the cornea and anterior eye scans, a lens must be attached to the front of the device for proper scanning. This lens is called the CAM auxiliary attachment (Cornea Anterior Module). The CAM software module provides for menu selections in the graphical user interface, which are selected by the operator to label corresponding corneal landmarks instead of those of the retina.
With the normative database (NDB), the RTVue XR OCT Avanti can compare the measured data from the GCC, the RNFL, the full retinal thickness, optic disc cup and optic disc rim measurements to the normative database. The device will provide the analysis information to he used as a clinical reference to aid in the diagnosis and management of ocular diseases.
There are two device configurations available, depending on the input voltage requirements at the facility intended for installation of the device:
| Model Number | Power Requirements | | |
|----------------|--------------------|-----------|---------|
| | Voltage | Frequency | Current |
| RTVue XR 100-1 | 100-120 VAC | 50/60 Hz | 8.33 A |
| RTVue XR 100-2 | 200-240 VAC | 50/60 Hz | 4.38 A |
{6}------------------------------------------------
- 2. RTVue XR OCT Avanti with AngioVue™ angiography software module
The RTVue XR OCT with AngioVue™ has an additional software module to aid in the visualization of vascular structures of the retina and choroid using a motion-contrast techniques without the need for intravenous dyes.
## Summary of Technological Characteristics
The RTVue XR OCT Avanti with AngioVue Software uses the same hardware as the previously cleared RTVue XR OCT. The RTVue XR OCT Avanti with AngioVue Software uses the previously cleared RTVue XR software with the addition of the AngioVue software module for OCT angiography scans. The AngioVue software uses motion contrast techniques to aid in the visualization of vascular structures of the retina and choroid.
The AngioVue scan performs sequential OCT cross-sectional scans to detect the motion of scattering particles such as red blood cells within the eye's vasculature. The software module also provides the ability to visualize the three-dimensional microvasculature of the retina and choroid. AngioVue image processing can employ a proprietary motion correction technology (MCT) to reduce potential artifacts caused by blinks and eye motion during scan acquisition.
## Performance Testing
The RTVue XR OCT device with normative database was previously verified for performance and functionality to assure conformance to the requirements for its basic intended use. Biocompatibility, electromagnetic compatibility, and electrical safety testing was conducted on the previously cleared RTVue XR OCT device. Since there were no hardware changes, these performance and safety tests were not required to be repeated.
The RTVue XR OCT Avanti with AngioVue Software has the additional OCT angiography software module. Software documentation was prepared and submitted for a "moderate" Level of Concern device in accordance with FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. Device software was verified and validated to support the proposed indications for use according to IEC 62304:2006 Medical device software - Software life cycle processes and FDA's General Principles of Software Validation; Final Guidance for Industry and FDA Staff.
## Risk Analysis
The new software module was assessed to make sure all the risks were sufficiently mitigated and that no new hazards were introduced as a result of the new software features according to the intended use. Identification of the associated hazards has been performed in order to evaluate, estimate and control the associated risks in accordance with EN ISO 14971 Application of risk analysis to medical devices.
## Summary of Changes
{7}------------------------------------------------
The new AngioVue™ OCT angiography software module can be installed on the previously cleared RTVue XR OCT device to provide a three-dimensional image of the microvasculature of the retina and choroid. The AngioVue software features include data acquisition options as well as data review reports.
Additionally, a proprietary motion correction technology (MCT) will be used in conjunction with AngioVue OCT angiography images to minimize potential artifacts caused by blinks and eye motion during scan acquisition. The MCT performs motion correction based on minimization of the overall difference between the two scan volumes acquired during OCT angiography image capture.
## Conclusions From Data
All testing deemed necessary was conducted and the results demonstrated that the RTVue XR OCT Avanti with AngioVue Software is as safe and as effective as the predicate device for the intended use. Testing, risk analysis and image comparison to fluorescein angiography also confirmed that no new questions of safety or effectiveness were identified. Previously submitted testing that originally supported use of the RTVue XR OCT remains sufficient and unchanged in the modified device software.
Based upon no significant differences between the proposed device and the predicate device, the RTVue XR OCT Avanti with AngioVue Software is substantially equivalent in design, features, performance, fundamental scientific technology, and is appropriate for the proposed indications for use.
In case examples of a variety of retinal diseases, RTVue XR OCT Avanti with AngioVue Software cube scans were compared with fluorescein angiography images. The images demonstrate that the RTVue XR OCT Avanti angiography in combination with OCT intensitybased information can give non-invasive three-dimensional information regarding retinal microvasculature in the retina and choroid. RTVue XR OCT Avanti angiography is not intended as a substitute for fluorescein angiography. Vascular findings on fluorescein angiography may be absent, poorly defined, or variably defined on RTVue XR OCT Avanti with AngioVue Software angiography scans. Additionally, dye leakage, staining, and pooling are not features of OCT angiography.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.