Edwards Balloon Catheter

K153069 · Edwards Lifesciences · OMZ · Jan 7, 2016 · Cardiovascular

Device Facts

Record IDK153069
Device NameEdwards Balloon Catheter
ApplicantEdwards Lifesciences
Product CodeOMZ · Cardiovascular
Decision DateJan 7, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Edwards Balloon Catheter is indicated for balloon pulmonic valvuloplasty.

Device Story

The Edwards Balloon Catheter is a percutaneous catheter used for balloon pulmonic valvuloplasty. The device features a nylon balloon mounted on a multi-durometer braided shaft with platinum/iridium radiopaque markers and a polycarbonate y-connector for inflation and guidewire access. It is operated by a physician in a clinical setting to dilate the pulmonic valve. The physician navigates the catheter to the target site under fluoroscopic guidance using the radiopaque markers. Inflation of the balloon exerts radial force to dilate the valve, potentially improving blood flow and reducing stenosis. The device is provided sterile for single use.

Clinical Evidence

Bench testing only. Testing included biocompatibility (ISO 10993-1), ethylene oxide sterilization validation, package qualification, and functional testing (visual inspection, radiopacity, balloon diameter, insertion force, inflation/deflation time, compliance, kink test, retrieval force, fatigue, burst, leakage, and bond testing).

Technological Characteristics

Materials: nylon balloon, thermoplastic elastomer (polyether block amide) multi-durometer braided shaft, platinum/iridium markers, polycarbonate y-connector. Dimensions: 130 cm working length; 4 cm balloon length; 16, 20, 23, 25 mm diameters. Energy: mechanical (balloon inflation). Sterilization: ethylene oxide.

Indications for Use

Indicated for balloon pulmonic valvuloplasty in patients requiring dilation of the pulmonic valve.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing right, stacked on top of each other, with flowing lines beneath them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 07, 2016 Edwards Lifesciences LLC Mr. Chris Kennelly Regulatory Affairs Associate II One Edwards Way Irvine, California 92614 Re: K153069 Trade/Device Name: Edwards Balloon Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: OMZ Dated: October 21, 2015 Received: October 22, 2015 Dear Mr. Kennelly: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, M.A. Hillebrenner for for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K153069 Device Name Edwards Balloon Catheter Indications for Use (Describe) The Edwards Balloon Catheter is indicated for balloon pulmonic valvuloplasty. Type of Use (Select one or both, as applicable) | <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |-------------------------------------------------------------------------------------------------------| | <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary (K153069) | Submitter: | Edwards Lifesciences LLC<br>One Edwards Way<br>Irvine, CA 92614 | |-------------------|-----------------------------------------------------------------------------------| | Contact: | Chris Kennelly<br>Phone: 949-250-2019<br>Fax: 949-809-5655 | | Prepared: | October 20, 2015 | | Trade Name: | Edwards Balloon Catheter | | Common Name: | Balloon valvuloplasty catheter | | Classification: | Pulmonary (Pulmonic) Valvuloplasty Catheter, 21 CFR 870.1250<br>Product Code: OMZ | | Predicate Device: | NuMed Z-MEDTM Balloon Dilatation Catheter (K040830) | ## Device Description: The Edwards Balloon Catheter is used for balloon pulmonic valvuloplasty. The device consists of a nylon balloon, a thermoplastic elastomer (polyether block amide) multi-durometer braided shaft with 130 cm working length, platinum/iridium radio-detectable markers, and a polycarbonate y-connector that consists of a balloon inflation port and guidewire lumen. The effective length of the balloon is 4 cm and is offered in 16 mm, 20 mm, 23 mm and 25 mm diameters. ## Indications for Use: The Edwards Balloon Catheter is indicated for balloon pulmonic valvuloplasty. ## Comparison to Predicate: The Edwards Balloon Catheter is substantially equivalent to the predicate device in intended use, design, technology and performance. The Edwards Balloon Catheter differs from the predicate device in introducer size compatibility, rated burst pressure, catheter length, material composition and inflation method. The differences between the subject and predicate devices do not have an adverse impact on safety or effectiveness, as demonstrated by bench testing. {4}------------------------------------------------ # Summary of Non-Clinical Testing: Non-clinical testing was completed to demonstrate that the Edwards Balloon Catheter meets the established performance characteristics, and to verify that design requirements are satisfied. Testing included biocompatibility evaluation per ISO 10993-1, ethylene oxide sterilization validation, and package qualification. Device functional testing included surface/visual Inspection, dimensional inspection, radiopacity, balloon diameter, insertion force into sheath, balloon inflation/deflation time, balloon compliance, catheter kink test, balloon catheter retrieval force, balloon fatigue and burst, leakage test, and bond testing. ## Conclusion: The Edwards Balloon Catheter is substantially equivalent to the predicate device, the NuMed Z-MED™ Balloon Dilatation Catheter.