Blue Sky Bio Zygomatic Implant System is indicated for surgical installation in the zygoma region, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. Blue Sky Bio zygomatic implants are recommended for the posterior (pre-molar/molar) region, one implant on each side, with at least two standard dental implants in the anterior region to support a fixed restoration. Blue Sky Bio zygomatic implants may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Device Story
System consists of threaded root-form titanium alloy dental implants and mating abutments for zygomatic bone placement. Implants feature internal hexagon or tapered internal hexagon connections; diameters 4.3mm–5.0mm; lengths 35mm–55mm. Abutments provide 17° or 30° angulation for screw-retained overdenture prosthetic support. Used by dental surgeons in clinical settings. Device provides structural foundation for fixed dental restorations in patients with severe jaw resorption. Benefits include restoration of chewing function and esthetics in cases where standard dental implants are insufficient due to bone loss.
Clinical Evidence
Bench testing only. Evidence includes engineering analysis, dimensional analysis, and dynamic compression-bending fatigue testing per ISO 14801 on worst-case constructs (smallest diameter implant with largest angulation). Biocompatibility data referenced from previous submissions (K102034, K060957). No clinical data included.
Indicated for patients with severe jaw resorption requiring surgical installation of implants in the zygoma region to restore esthetics and chewing function. Recommended for posterior (pre-molar/molar) region, one per side, with at least two anterior standard dental implants to support fixed restorations. Suitable for immediate loading if primary stability and appropriate occlusal loading are achieved.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a symbol of three stylized human profiles facing to the right, with three wave-like lines below them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 29, 2016
Blue Sky Bio, LLC c/o Kevin A. Thomas, Ph.D. Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K153064
Trade/Device Name: Blue Sky Bio Zygomatic Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: March 31, 2016 Received: March 31, 2016
Dear Dr. Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Tina Kiang S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K153064
Device Name
Blue Sky Bio Zygomatic Implant System
Indications for Use (Describe)
Blue Sky Bio Zygomatic Implant System is indicated for surgical installation in the zygoma region, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. Blue Sky Bio zygomatic implants are recommended for the posterior (pre-molar) region, one implant on each side, with at least two standard dental implants in the anterior region to support a fixed restoration. Blue Sky Bio zygomatic implants may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription (Part 21 CFR 201.5 Subpart D) | □ Over-The-Counter (Part 21 CFR 201.6 Subpart E) |
|--------------------------------------------------------------------------------------------------------|--------------------------------------------------|
|--------------------------------------------------------------------------------------------------------|--------------------------------------------------|
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary Blue Sky Bio, LLC Blue Sky Bio Zygomatic Implant System
April 28, 2016
## ADMINISTRATIVE INFORMATION
| Manufacturer Name | Blue Sky Bio, LLC | |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|
| | 888 E Belvidere Road, Suite 212 | |
| | Grayslake, IL 60030 | |
| | Telephone | +1 718-376-0422 |
| | Fax | +1 888-234-3685 |
| Official Contact | Michele Vovolka | |
| | Vice President of RA/QA | |
| Representative/Consultant | Kevin A. Thomas, PhD<br>Linda K. Schulz, BSDH, RDH<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130 | |
| | Telephone: | +1 (858) 792-1235 |
| | Fax: | +1 (858) 792-1236 |
| | Email: | kthomas@paxmed.com<br>lschulz@paxmed.com |
# DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | Blue Sky Bio Zygomatic Implant System |
|----------------------------|-----------------------------------------------------------------|
| Common Name | Endosseous dental implant<br>Endosseous dental implant abutment |
| Classification Regulations | 21 CFR 872.3640 |
| Product Code | DZE<br>NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
{5}------------------------------------------------
#### PREDICATE DEVICE INFORMATION
Primary predicate device:
K141777, Neodent Implant System, JJGC Indústria e Comércio de Materiais Dentários SA
Reference predicate devices:
K102034, Blue Sky Bio Dental Implant System, Blue Sky Bio, LLC K060957, Blue Sky Bio Dental Implant System, Blue Sky Bio, LLC K093562, Zygomatic Implant System, Southern Implants, Inc.
#### INTENDED USE
Blue Sky Bio Zygomatic Implant System is indicated for surgical installation in the zygoma region, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. Blue Sky Bio zygomatic implants are recommended for the posterior (pre-molar/molar) region, one implant on each side, with at least two standard dental implants in the anterior region to support a fixed restoration. Blue Sky Bio zygomatic implants may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
#### DEVICE DESCRIPTION
Blue Sky Bio Zygomatic Implant System submission includes threaded root-form dental implants and mating abutments designed for placement into the zygomatic bone. The zygomatic implants are provided with an internal hexagon connection and a tapered internal hexagon interface for connection to the subject abutments. The internal hexagon connection implants are provided with a body diameter of 4.7 mm and platform diameters of 3.5 mm and 4.5 mm. The tapered internal hexagon connection implants are provided with a body diameter of 4.3 mm and a narrow platform (NP) connection, and with a body diameter of 5.0 mm with a regular platform (RP) connection. All implants are provided in multiple overall threaded lengths ranging from 35 mm to 55 mm. This submission includes mating abutments with platform diameters of 3.5, 4.3, and 4.5 mm. and each abutment diameter is provided with 17° and 30° of angulation. All subject device abutments are for support of screw-retained overdenture prosthetic restorations. The abutment screws compatible with the subject device abutments were cleared in K060957 and K102034. The subject device zygomatic implants are made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401). The previously cleared abutment screws also are made of material conforming to ASTM F136.
#### PERFORMANCE DATA
Non-clinical data submitted, referenced, or relied upon to demonstrate substantial equivalence include: biocompatibility (referenced from K102034 and K060957), engineering analysis, dimensional analysis, and dynamic compression-bending testing of the Taper Hex 4.3 mm body diameter implant with the NP platform according to ISO 14801 Dentistry - Implants - Dynamic fatigue test for endosseous dental implants. No clinical data were included in this submission.
{6}------------------------------------------------
### EQUIVALENCE TO MARKETED DEVICE
Blue Sky Bio, LLC submits the following information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices, the subject device is substantially equivalent in indications and design principles to the following legally marketed predicate devices:
K141777, Neodent Implant System, JJGC Indústria e Comércio de Materiais Dentários SA;
K102034, Blue Sky Bio Dental Implant System, Blue Sky Bio, LLC;
K060957, Blue Sky Bio Dental Implant System, Blue Sky Bio, LLC; and
K093562, Zygomatic Implant System, Southern Implants, Inc.
A comparison of the technological characteristics of the subject device and the primary predicate device K141777 is provided in the following table.
| | Subject Device | Primary Predicate Device |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison | Blue Sky Bio, LLC<br>Blue Sky Bio Zygomatic Implant System<br>K153064 | JJGC Indústria e Comércio de Materiais Dentários SA<br>Neodent Implant System<br>K141777 |
| Indications for Use | Blue Sky Bio Zygomatic Implant System is indicated<br>for surgical installation in the zygoma region, in cases<br>of severe jaw resorption, in order to restore patient<br>esthetics and chewing function. Blue Sky Bio<br>zygomatic implants are recommended for the posterior<br>(pre-molar/molar) region, one implant on each side,<br>with at least two standard dental implants in the<br>anterior region to support a fixed restoration. Blue Sky<br>Bio zygomatic implants may be loaded immediately<br>when good primary stability is achieved and with<br>appropriate occlusal loading. | Zygomatic Implants are indicated for surgical<br>installation in the zygoma region, in cases of severe<br>jaw resorption, in order to restore patient esthetics and<br>chewing function. Zygomatic Implants are<br>recommended for the posterior (pre-molar/molar)<br>region, one implant on each side, with at least two<br>standard dental implants in the anterior region to<br>support a fixed restoration. Zygomatic Implants may<br>be loaded immediately when good primary stability is<br>achieved and with appropriate occlusal loading. |
| Implants | | |
| Design | Threaded root-form implant to be used with mating<br>abutments | Threaded root-form implant to be used with mating<br>abutments |
| Implant body Ø<br>and<br>Platform Ø | Internal hex connection<br>Implant body Ø: 4.7 mm<br>Abutment platform Ø: 3.5 mm, 4.5 mm<br>(implant body tapers at apical portion)<br>Tapered internal hex connection<br>Implant body Ø: 4.3 mm with NP 3.5 mm<br>abutment platform<br>Implant body Ø: 5.0 mm with RP 4.3 mm<br>abutment platform<br>(implant body tapers at apical portion) | External hex and Morse taper<br>Implant body Ø: 4.4 mm tapering to 3.9 mm<br>Platform Ø: 4.1 mm |
| Implant Lengths | All implants:<br>35, 37.5, 40, 42.5, 45, 47.5, 50, 52.5, 55 mm | External hex:<br>30, 35, 40, 45, 47.5, 50, 52.5 mm<br>Morse taper:<br>30, 35, 40, 42.5, 45, 47.5, 50, 52.5 mm |
| Implant-Abutment Interface | | |
| Type | Internal hex with 12° taper<br>Internal hex with 45° bevel | External hex<br>Morse taper |
| Implant body-abutment<br>connection angle | n/a | 45° |
| Abutments | | |
| Platform Diameter | Internal hex connection: 3.5 mm, 4.5 mm<br>Tapered internal hex connection: 4.3 mm | 4.1 mm |
| Abutment Angle | 17°, 30° | 0° (straight) |
{7}------------------------------------------------
| | Subject Device | Primary Predicate Device |
|-----------------|-----------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Comparison | Blue Sky Bio, LLC<br>Blue Sky Bio Zygomatic Implant System<br>K153064 | JJGC Indústria e Comércio de Materiais Dentários SA<br>Neodent Implant System<br>K141777 |
| Materials | | |
| Implants | Ti-6Al-4V | F67 unalloyed titanium, grade 4 |
| Implant Surface | Grit blasted and acid etched | Machined |
| Abutments | Ti-6Al-4V | F136 Ti-6A1-4V ELI |
| Abutment Screws | Ti-6Al-4V | External hex: F136 Ti-6A1-4V ELI |
The subject device and the primary predicate K141777 have substantially equivalent indications for use, including restoration of patient esthetics and chewing function, and immediate loading when used with standard dental implants placed in the anterior region to support a fixed restoration. The implants of the subject device and the predicate device K141777 have similar designs and dimensions, including lengths appropriate for zygomatic placement.
The subject zygomatic implants are very similar in design to endosseous dental implants from Blue Sky Bio cleared in K102034 and K060957. except for a slight change in the endosseous thread and the longer lengths appropriate for zygomatic placement. The subject zygomatic implants are provided in a range of implant body diameters (maximum endosseous thread diameters) of 4.3 mm to 5.0 mm, the same as the predicate implants cleared in K102034 and K060957. The smaller subject device implant body (4.3 mm diameter) as compared to the primary predicate K141777 (4.4 mm diameter) is supported by the reference predicate K093562 (4.05 mm body diameter), and by and dynamic compression-bending testing according to ISO 14801.
The subject device implants are provided in the same range of overall lengths as the primary predicate K141777 (35 mm to 52.5 mm), and the subject device 55 mm length implant is supported by the reference predicate K093562.
This submission includes abutments with an internal hexagon and tapered internal hexagon interfaces for connection to the subject zygomatic implants. The subject device abutments are provided with platform diameters of 3.5. 4.3. and 4.5 mm. and angulation of 17° or 30°. The subject device abutments are substantially equivalent to the abutments cleared in K102034 and K060957 in terms of implant-abutment connections, platform diameters, angulation, and materials. The subject abutments are to be used with compatible abutment screws cleared in K102034 and K060957.
Differences between the subject device and the primary predicate K141777 include the location of angulation correction (subject device includes abutments with angulation, versus K141777 implant designs included angulation at the abutment-implant connection), and the amount of angulation (subject device 17° or 30°, versus K141777 only 45°). These differences between the subject device and the primary predicate are supported by dynamic compression-bending testing according to ISO 14801. Dynamic testing of worst case subject device constructs consisting of the smallest diameter implant (Taper Hex 4.3 mm body diameter) and largest angulation (30°)
{8}------------------------------------------------
demonstrated fatigue performance substantially equivalent to that of the primary predicate K141777.
The subject device implants and abutments are made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401), the same as these components cleared in K102034 and K060957. The endosseous surface finish of the subject device implants is the same as the predicate implants cleared in K102034 and K060957.
The subject device implants have similar packaging and are sterilized using the same materials and processes as described in K102034 and K060957. The subject device abutments are provided nonsterile in similar packaging and are to be sterilized using the same processes as described in K102034 and K060957.
#### CONCLUSION
The subject device and the predicate devices have intended use, have similar technological characteristics, and are made of similar materials. The subject device and predicate devices encompass the same range of physical dimensions, including diameter and length of the implants, and the diameter and angulation of the abutments. The subject and predicate devices are packaged in similar materials and sterilized using similar methods.
The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.