← Product Code [KWQ](/productcode/KWQ) · K153042

# TRYPTIK2C-plate Anterior Cervical Plate System (K153042)

_Spineart · KWQ · Mar 28, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K153042

## Device Facts

- **Applicant:** Spineart
- **Product Code:** [KWQ](/productcode/KWQ.md)
- **Decision Date:** Mar 28, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

TRYPTIK2C-Plate® Anterior Cervical Plate System is intended to be used for temporary stabilization of the cervical spine (C2-C7) during the development of solid spinal fusion in patients with instability caused by the following degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumor, spondylolisthesis, spinal stenosis, deformity (i.e., scoliosis, kyphosis, lordosis), pseudarthrosis, and failed previous fusions

## Device Story

System consists of anterior cervical plates and polyaxial bone screws; provides temporary stabilization of cervical spine (C2-C7) as adjunct to spinal fusion. Plates attach to anterior vertebral body; screws provide fixation. Used in surgical settings by orthopedic or neurosurgeons. Benefits include immobilization of spinal segments to facilitate solid fusion. Device is purely mechanical; no software or algorithm involved.

## Clinical Evidence

Bench testing only. Mechanical performance evaluated per ASTM F1717-15, including static compression bending, static tension bending, static torsion, and dynamic compression bending. Additional in-house static push-out testing performed. No clinical data provided.

## Technological Characteristics

Constructed of Titanium alloy Ti6Al4V ELI (ISO 5832.3, ASTM F136). System includes 1-4 level plates (2.4mm thick, 16.5mm wide) and polyaxial bone screws (4.0mm/4.5mm diameter). Features integrated expansive rings for anti-back out (one-step locking). Gamma sterilized. Mechanical fixation principle.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- IST Anterior Cervical Plate System ([K072650](/device/K072650.md))
- Synthes Cervical Vertabrae Plates ([K792352](/device/K792352.md))
- Synthes Anterior Cervical Vertebrae Plate ([K926453](/device/K926453.md))
- Tosca Anterior Cervical Plate System ([K043082](/device/K043082.md))
- Spider Cervical Plating (SCP) System ([K052292](/device/K052292.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and features the department's name around the perimeter. In the center is an abstract image of a bird-like figure, composed of three human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 28, 2016

SPINEART Mr. Franck Pennesi Director of Industry & Quality International Center Cointrin 20 route de pré-bois, CP1813 Geneva, 1215 SWITZERLAND

Re: K153042

Trade/Device Name: TRYPTIK2C-plate® Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWO Dated: March 3, 2016 Received: March 7, 2016

Dear Mr. Pennesi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

## Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|

**Indications for Use**

| 510(k) Number (if known)                                                             | K153042                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                      |                                                                      |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Device Name                                                                          | TRYPTIK2C-Plate® Anterior Cervical Plate System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                      |                                                                      |
| Indications for Use (Describe)                                                       | TRYPTIK2C-Plate® Anterior Cervical Plate System is intended to be used for temporary stabilization of the cervical             spine (C2-C7) during the development of solid spinal fusion in patients with instability caused by the following             degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient             history and radiographic studies), trauma (including fractures), tumor, spondylolisthesis, spinal stenosis, deformity (i.e.,             scoliosis, kyphosis, lordosis), pseudarthrosis, and failed previous fusions |                                                                                      |                                                                      |
| Type of Use (Select one or both, as applicable)                                      | <table><tr><td><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</td><td><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table>                                                                                                                                                                                                                                                                                                                                                                                                                      | <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                      |                                                                      |

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services  

Food and Drug Administration  

Office of Chief Information Officer  

Paperwork Reduction Act (PRA) Staff  

PRAStaff@fda.hhs.gov

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*

{3}------------------------------------------------

## TRADITIONAL 510k TRYPTIK2c-Plate ANTERIOR CERVICAL PLATE SYSTEM

Image /page/3/Picture/1 description: The image shows the logo for Spineart. The logo features the word "spineart" in a stylized font, with a purple butterfly-like symbol above the text. The word "spineart" is written in a modern, sans-serif font, with the letters connected in a unique way.

## 510(k) SUMMARY

|                                       | SPINEART                                                                           |
|---------------------------------------|------------------------------------------------------------------------------------|
| Submitted by                          | International Center Cointrin                                                      |
|                                       | 20 route de pré-bois                                                               |
|                                       | CP1813                                                                             |
|                                       | 1215 GENEVA 15                                                                     |
|                                       | SWITZERLAND                                                                        |
| Contacts                              | Franck PENNESI Director of Industry & Quality                                      |
|                                       | Phone : +41 22 570 1246 Fax : +41 22 799 40 26                                     |
|                                       | Mail : fpennesi@spineart.com                                                       |
|                                       | Regulatory contact : Dr Isabelle DRUBAIX (Idée Consulting)                         |
|                                       | Mail : idrubaix@nordnet.fr                                                         |
| Date Prepared                         | March 3rd 2016                                                                     |
| Common Name                           | Anterior Cervical Plate System                                                     |
| Trade Name                            | TRYPTIK2C-Plate Anterior Cervical Plate System                                     |
| Classification Name                   | Spinal intervertebral body fixation orthosis                                       |
| Class                                 | II                                                                                 |
| Product Code                          | KWQ Appliance, Fixation, Spinal Intervertebral Body                                |
| CFR section                           | 888.3060 Spinal intervertebral body fixation orthosis                              |
| Device panel                          | ORTHOPEDIC                                                                         |
|                                       | Primary predicate: IST Anterior Cervical Plate System manufactured by Innovative   |
| Legally marketed<br>predicate devices | Spinal Technologies Inc (K072650)                                                  |
|                                       | Other predicates: Synthes Cervical Vertabrae Plates manufactured by Depuy          |
|                                       | Synthes (K792352); Synthes Anterior Cervical Vertebrae Plate manufactured by       |
|                                       | Synthes USA (K926453); Tosca Anterior Cervical Plate System manufactured by        |
|                                       | Signus Medizintechnik GMBH (K043082);Spider Cervical Plating (SCP) System          |
|                                       | manufactured by X-SPINE SYSTEMS, INC. (K052292)                                    |
| Indications for use                   | TRYPTIK2C-Plate Anterior Cervical Plate System is intended to be used for          |
|                                       | temporary stabilization of the cervical spine (C2-C7) during the development of    |
|                                       | solid spinal fusion in patients with instability caused by the following           |
|                                       | degenerative disc disease (as defined by neck pain of discogenic origin with       |
|                                       | degeneration of the disc confirmed by patient history and radiographic studies),   |
|                                       | trauma (including fractures), tumor, spondylolisthesis, spinal stenosis, deformity |
|                                       | (i.e., scoliosis, kyphosis, lordosis), pseudarthrosis, and failed previous fusions |

{4}------------------------------------------------

| Description of the device          | The TRYPTIK2C-Plate Anterior Cervical Plate System consists of a range of anterior cervical plates and polyaxial (variable angle) bone screws. The plate attaches to the anterior portion of the vertebral body of the cervical spine (C2-C7) and functions as an adjunct to fusion to provide immobilization and stabilization of cervical segments of the spine.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics   | The TRYPTIK2C-Plate Anterior Cervical Plate System consists of 1-level, 2-level, 3-level and 4-level plates with cancellous and cortical variable-angle bone screws. The Plate configurations are ranging in total lengths from 20mm (1-level) to 92mm (4-level). The plate is 2.4mm thick and 16.5mm wide and incorporates integrated expansive rings for anti-back out of the bone screws that functions as a one-step locking mechanism. The Screw range comes in two diameters, Ø 4.0 and Ø 4.5, and length is ranging from 12mm to 18mm with 2mm increment.<br>The TRYPTIK2C-Plate Anterior Cervical Plates and screws are all made of Titanium alloy Ti6Al4V ELI conforming to ISO 5832.3 and ASTM F136. The TRYPTIK2C-Plate Anterior Cervical Plates and screws are delivered sterile (gamma sterilization). The TRYPTIK2C-Plate Anterior Cervical Plate System is supplied with all the surgical instruments required for its installation. |
| Discussion of Testing              | The following non-clinical tests were conducted according to ASTM F1717-15: Static Compression Bending, Static Tension Bending, Static Torsion and Dynamic Compression Bending. Results demonstrate comparable mechanical properties to the predicate devices. Additionally, static push-out testing has been conducted according to an in-house protocol.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Comparison to Predicate<br>Devices | The indications for use, design features and materials of the TRYPTIK2C-Plate Anterior Cervical Plates and screws are substantially equivalent to those of the predicate devices. The substantial equivalence of the TRYPTIK2C-Plate Anterior Cervical Plates and screws is supported by the performance testing, materials information, and data analysis provided within this Premarket Notification                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Conclusion                         | Based on the above information, the TRYPTIK2C-Plate Anterior Cervical Plates and screws can be considered as substantially equivalent to its predicate devices in terms of intended use, material, design, mechanical properties and function.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

---

**Source:** [https://fda.innolitics.com/device/K153042](https://fda.innolitics.com/device/K153042)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
