This device is intended to measure the blood pressure and pulse rate on upper arms of adults.
Device Story
The Digital Blood Pressure Monitor is an arm-type, fully automatic device for home use. It utilizes the oscillometric method to measure blood pressure and pulse rate. An internal air pump inflates a cuff to extrude the artery; a pressure sensor collects cuff pressure data and converts it into digital signals for the CPU. Software processes these signals to calculate systolic and diastolic blood pressure and pulse rate, which are displayed on an LCD. The device is powered by internal dry batteries. It is intended for use by adults in a home environment. The output provides users with blood pressure and pulse rate readings, assisting in monitoring cardiovascular health.
Clinical Evidence
Clinical evaluation was performed according to ANSI/AAMI/ISO 81060-2 standards to validate the accuracy of the blood pressure and pulse rate measurements.
Technological Characteristics
Oscillometric measurement principle; internal air pump and valve for inflation/deflation; LCD digital display; powered by 4 AA alkaline batteries. Complies with AAMI/ANSI ES 60601-1 (electrical safety), IEC 80601-2-30 (performance), IEC 60601-1-2 (EMC), IEC 60601-1-11 (home-use), and ISO 10993-5/10 (biocompatibility).
Indications for Use
Indicated for adults for the non-invasive measurement of blood pressure and pulse rate on the upper arm.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Arm Automatic Blood Pressure Monitor, model BF1115 (K130325)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is an abstract symbol that resembles an eagle or bird with three human profiles incorporated into its design.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 24, 2016
Suzhou Sunspiri Co., Ltd. % Long Yang Coo Shenzhen Hlongmed Biotech Company Limited R1508, East Building, Yihai Plaza, Chuangye Road, Nanshan District Shenzhen, 518054 CN
Re: K153033
Trade/Device Name: Digital Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: May 13, 2016 Received: May 19, 2016
Dear Long Yang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{2}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
#### Device Name
Digital Blood Pressure Monitor
Models: JWS Series, including: JDS-189, JDS-500A, IDS-500E, JDS-600, JDS-700, IDS-703, IDS-704, JDS-800, BM20
Indications for Use (Describe)
This device is intended to measure the blood pressure and pulse rate on upper arms of adults.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
#### Date Prepared: June 22, 2016
#### 1. Submitter
SUZHOU SUNSPIRIT CO., LTD.
No.72, Hengshan Rd., Suzhou New District, P. R. China, 215009 Establishment Registration Number: 3010271644 Tel: 0086-512-68238996 Fax: 0086-512-68234280
#### 2. Submission Correspondent
1) Ms. Jing Zhang (QA manager) SUZHOU SUNSPIRIT CO., LTD. No.72, Hengshan Rd., Suzhou New District, P. R. China, 215009 2) Mr. Long Yang (COO) Shenzhen Hlongmed Biotech Company Limited R1508, East Building, Yihai Plaza, Chuangye Road, Nanshan District,Shenzhen, P.R. China,518054
Tel: 0086-755-86664986
Fax: 0086-755-86664933
E-mail: yanglong@hlongmed.com
## 3. Proposed Device Information
Trade name: Digital Blood Pressure Monitor Model: JDS Series, including: JDS-189, JDS-500, JDS-500A, JDS-500E, JDS-600, JDS-700, JDS-703, JDS-704, JDS-800, BM20 Common name: Digital Blood Pressure Monitor Classification name: Noninvasive Blood Pressure Measurement System Review Panel: Cardiovascular System Devices Panel (74) Product Code: DXN Regulation Class: II Regulation Number: 870.1130
{5}------------------------------------------------
## 4. Predicate Device Information
Company Name: Shenzhen Pump Medical System Co., Ltd. Device Name: Arm Automatic Blood Pressure Monitor, model BF1115 510(k) number: K130325
## 5. Device Description
The Digital Blood Pressure Monitor is Arm-type fully automatic Digital Blood Pressure Monitor and used the oscillometric method for non-invasive measurement of blood-pressure at home.
The equipment is non-transit-operable, portable equipment, not used at the conditions of patient movement in normal use. The unit is powered by internal dry battery.
Digital Blood Pressure Monitor is based on pressure vibration method. Blood pressure cuff use the air pump to inflate, then the arteries are extruded by the cuff with pressure. Pressure sensor collects the pressure in the cuff, and then converts it to digital signal to the CPU. Then the software calculates the systolic and diastolic blood pressure and pulse rate.
Digital Blood Pressure Monitor, JDS-189, JDS-500A, JDS-500E, JDS-600,
JDS-700, JDS-703, JDS-704, JDS-800, BM 20 in term of intended use measurement method, structure, and specifications are totally same, only the appearance, memory space has difference.
#### 6. Intended use/Indications for use:
This device is intended to measure the blood pressure and pulse rate on upper arms of adults.
## 7. Testing data and clinical study
Laboratory testing was conducted to validate and verify that Digital Blood Pressure Monitor met all requirements of related international standards, including electrical safety, EMC, biocompatibility, specification. Results of these tests demonstrate compliance to the requirements of the bellow consensus standards.
Clinical study has been evaluated according to ANSI/AAMI/ISO 81060-2.
#### Applied Standard:
{6}------------------------------------------------
- Electrical Safety requirement: AAMI / ANSI ES 60601-1:2005
- Performance requirement: IEC80601-2-30 ●
- Electromagnetic Compatibility Requirements: IEC 60601-1-2 ●
- Home-used medical equipment requirements: IEC 60601 -1 -11
- Biocompatibility Evaluation for NIBP Cuff: ISO 10993-5, ISO 10993-10 ●
- Clinical Evaluation: ANSI/AAMI/ISO 81060-2
# 8. Comparison to Predicate Device
| Trade Name | Digital Blood Pressure Monitor | Arm Automatic Blood Pressure Monitor |
|--------------------------|------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Model | JDS-189, JDS-500, JDS-500A, JDS-500E, JDS-600, JDS-700, JDS-703, JDS-704, JDS-800, BM 20 | BF1115 |
| Indications for use | This device is intended to measure the blood pressure and pulse rate on upper arms of adults. | It is intended for measuring adult blood pressure and pulse rate. |
| Target population | Adult | Adult |
| Environment of use | Home | Home |
| Technology | Oscillometric method | Oscillometric method |
| Measurement Method | Non-invasive | Non-invasive |
| Measurement localization | Upper arm | Upper arm |
| Memory Space | 2×30(BM20, JDS-189, JDS-500, JDS-500A, JDS-500E, JDS-600) / 1×99(JDS-700)/ 2×60(JDS-703, JDS-704)/ 2×50(JDS-800) | 50 sets of measurement values( blood pressure and pulse rate) for each user( Memory 1 and Memory 2) |
| Measuring parameters | blood pressure (systolic and diastolic) and heart rate | blood pressure (systolic and diastolic) and heart rate |
{7}------------------------------------------------
| Measurement<br>range | Pressure: 0 – 300 mmHg<br>Pulse: 40 –180 beats/min | Pressure: 0 to 280 mmHg<br>Pulse: 40 to 180 pulses/min |
|-----------------------------|----------------------------------------------------------|----------------------------------------------------------|
| Measurement<br>accuracy | Pressure: ± 3 mmHg<br>Pulse: ± 5% of the displayed value | Pressure: ± 3 mmHg<br>Pulse: ± 5% of the displayed value |
| Inflation Mode | Automatic by Internal Pump | Automatic by Internal Pump |
| Deflation<br>Mode | Automatic by Valve | Automatic by Valve |
| Measurable<br>circumference | 22-32cm | 22-36cm |
| Display | LCD, digital display | LCD, digital display |
| Energy source | 4 AA Alkaline batteries | 4 AA Alkaline batteries or AC<br>adaptor(AC 100~ 240V) |
| Operating<br>environment | +10~+40 °C , ≤85% RH | +5~+40°C , ≤93%RH |
| Storage<br>environment | -5 ~ +50 °C, ≤85% RH | -25~+70°C,10~95%RH |
The subject device is Substantially Equivalent (SE) to the predicate device which is US legally market device.
# 9. Conclusion
Digital Blood Pressure Monitor has the same intended use, the same technological characteristics as the predicate device. Moreover, non-clinical testing & clinical testing contained in this submission demonstrated that any difference in their technological characteristics does not raise any new issues of safety and effectiveness.
In a word, Digital Blood Pressure Monitor is substantial equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.