The Hudson RCI® AquaPak® Sterile Prefilled Nebulizer adds sterile water or saline solution in aerosol form to a patient's breathing gases. It may be used with pediatric (ages 2 years and above) and adult patients.
Device Story
Device is a jet nebulizer with adjustable air entrainment; consists of a sterile adaptor (type 028 or 033) and a prefilled sterile reservoir (440ml, 760ml, or 1070ml). Operates by connecting to a 50 PSI oxygen source via flow meter; oxygen stream entrains room air to achieve specific oxygen concentrations (FiO2). Produces aerosolized sterile water or saline (1.5-3 micron particle size) for inhalation therapy. Used in hospitals, sub-acute/long-term care facilities, and home environments; operated by clinicians or patients. Single-use device. Benefits patient by providing humidified or medicated aerosol therapy to breathing gases.
Clinical Evidence
Bench testing only. Testing included lift pressure, oxygen entrainment, nebulization rate, and particle size distribution, all showing results identical to the predicate. Packaging and shelf-life validated per ASTM F1980-07 and ASTM F2096. Sterility validated per ISO 11137-1:2006 and ISO 11137-2:2013.
Technological Characteristics
Jet nebulizer with adjustable air entrainment window. Components: sterile adaptor (gamma sterilized) and prefilled reservoir (sterile water or saline). Dimensions: 440ml, 760ml, 1070ml reservoirs. Energy source: 50 PSI oxygen gas. Connectivity: None. Standards: ASTM F1980-07 (accelerated aging), ASTM F2096 (leak detection), ISO 11137-1/2 (gamma sterilization).
Indications for Use
Indicated for pediatric (ages 2+) and adult patients requiring aerosolized sterile water or saline solution added to breathing gases.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 26, 2016
Teleflex Medical, Inc. Brian Gall Senior Regulatory Affairs Specialist 3015 Carrington Mill Blvd. Morrisville, North Carolina 27560
Re: K153010
Trade/Device Name: Hudson RCI® AquaPak® Sterile Prefilled Nebulizers Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer (Direct Patient Interface) Regulatory Class: Class II Product Code: CAF Dated: June 16, 2016 Received: June 17, 2016
Dear Mr. Brian Gall:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | |
|-----------------------------------------|--|
| Food and Drug Administration | |
| <b>Indications for Use</b> | |
| Form Approved: OMB No. 0910-0120 | |
|-----------------------------------|--|
| Expiration Date: January 31, 2017 | |
| See PRA Statement below. | |
| 510(k) Number (if known) | K153010 |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Hudson RCI® AquaPak® Sterile Prefilled Nebulizers |
| Indications for Use (Describe) | The Hudson RCI® AquaPak® Sterile Prefilled Nebulizer adds sterile water or saline solution in aerosol form to a patient's breathing gases. It may be used with pediatric (ages 2 years and above) and adult patients. |
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <div style="display:inline-block"><span style="text-decoration: underline;">Prescription Use</span> (Part 21 CFR 801 Subpart D)</div> | <div style="display:inline-block">Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
| FORM FDA 3881 (8/14) | Page 1 of 1 |
|----------------------|-------------|
|----------------------|-------------|
PSC Publishing Services (301) 443-6740 EF
,
s
1 1 1 1 1
{3}------------------------------------------------
### Section 31 - Updated 510(k) Summary
## 510(k) SUMMARY
## K153010 – Hudson RCI® AquaPak® Sterile Prefilled Nebulizers
#### A. Name, Address, Phone and Fax Number of Applicant
Teleflex Medical, Incorporated 3015 Carrington Mill Blvd. Morrisville, NC 27560 USA Phone: 919-228-4350 Fax: 919-361-3939
#### B. Contact Person
Brian Gall Senior Requlatory Affairs Specialist, Respiratory Division
#### C. Date Prepared
15 June 2016
#### D. Device Name
#### Trade Name: Hudson RCI® AquaPak® Sterile Prefilled Nebulizers
Common Name: Nebulizer
Classification Name: Nebulizer (Direct Patient Interface), CFR - 868.5630, Class II
#### E. Device Description
Hudson RCI® AquaPak® Sterile Prefilled Nebulizers provide sterile water or sterile saline for inhalation therapy. Nebulizers generate aerosol, a fine mist of liquid water (or sodium chloride solution) that is suspended in the gas to be inhaled by the patient during aerosol therapy.
The Hudson RCI® AquaPak® Sterile Prefilled Nebulizers consist of two components, a sterile adaptor and a prefilled sterile reservoir. The adaptor may be type 028 or type 033. The standard model with a yellow collar (type 028) and the quiet model with a blue collar (type 033) feature an adjustable air entrainment window, which provides a specific oxygen concentration by entraining room air into the oxygen stream. Prefilled sterile reservoirs for the Hudson RCI® AquaPak® Sterile Prefilled Nebulizers come in three sizes; 440ml, 760ml and 1070ml, and may be sterile water or sterile sodium chloride solution. The sterile reservoirs are cleared under K141214, and are presented here for reference only. Each reservoir, when combined with an adaptor component, provides nebulizer functionality when connected to a gas source.
The primary purpose of this Traditional 510(k) submission is to change the nebulizer adaptor cleared in K141214 from non-sterile, and update the packaging materials and shelf life to maintain sterility.
{4}------------------------------------------------
#### Section 31 - Updated 510(k) Summary
#### F. Indications for Use
The Hudson RCI® AquaPak® Sterile Prefilled Nebulizer adds sterile water or saline solution in aerosol form to a patient's breathing gases. It may be used with pediatric (ages 2 years and above) and adult patients.
#### G. Target Population
The Hudson RCI® AquaPak® Sterile Prefilled Nebulizer may be used with pediatric (ages 2 years and above) and adult patients.
#### H. Environments of Use
This device is intended for hospital, sub-acute facilities, long-term care facilities, and in a home care environment.
This product is single use only.
#### l. Contraindications
There are no known contraindications.
#### J. Comparative Characteristics
The proposed Hudson RCI® AquaPak® Sterile Prefilled Nebulizer is substantially equivalent to the nebulizer in the predicate device:
| Comparative<br>Characteristics | Proposed Device:<br>Hudson RCI® AquaPak®<br>Sterile Prefilled Nebulizer | Predicate Device:<br>Hudson RCI® AquaPak®<br>Prefilled Nebulizer |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|
| Manufacturer | Teleflex Medical, Inc. | Same |
| 510(k) Number | K153010 | K141214 |
| Indications for<br>Use | The Hudson RCI®<br>AquaPak® Sterile Prefilled<br>Nebulizer adds sterile<br>water or saline solution in<br>aerosol form to a<br>patient's breathing gases.<br>lt may be used with<br>pediatric (ages 2 years<br>and above) and adult<br>patients. | Same |
| Principle of<br>Operation | Jet nebulizer with<br>adjustable air<br>entrainment | Same |
| Gas source | 50 Psi oxygen regulated<br>via a flow meter | Same |
| Flow rate and | 028 Model: | Same |
{5}------------------------------------------------
Section 31 - Updated 510(k) Summary
| FiO2 Control<br>Capabilities | 28% at 5 LPM<br>35% at 8 LPM<br>40% - 98% at 10 LPM<br><br>033 Model:<br>33% - 35% at 8 LPM<br>40% - 98% at 10 LPM | |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------|---------------------|
| Aerosol Particle<br>Size Delivery | 1.5 to 3 micron range | Same |
| Sterilization | Adaptor sterilized by<br>Gamma Sterilization | Non-sterile adaptor |
| Single Use | Single Use only device | Same |
| Shelf Life | Adaptor – 5 years from<br>date of manufacture | Adaptor – N/A |
| Packaging | Adaptors – 50/case, 10 or<br>20/case when packaged<br>with water/ saline | Same |
## K. Non-clinical Comparative Performance Testing
Bench testing has been performed to verify that the performance of the proposed Hudson RCI® AquaPak® Sterile Prefilled Nebulizer is substantially equivalent to the predicate device, and that the Hudson RCI® AquaPak® Sterile Prefilled Nebulizer will perform as intended.
There are no new patient contacting materials, so no biocompatibility testing was required for this submission.
The 'Lift Pressure', 'Oxygen Entrainment', 'Nebulization Rate', and 'Particle Size Distribution' testing is identical to the testing performed in the predicate K141214. The results of this testing showed that the subject device met all of the acceptance criteria established by the predicate. The 'Packaging configuration / Shelf Life' and 'Sterility Testing' ensure the sterile packaging and sterility are maintained after sterilization and shelf life testing per the referenced standards. The packaging for the subject device met the acceptance criteria of the standards cited below.
| Test<br>Performed | Reference to Standard (if<br>applicable) | Principle of Test |
|-------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Packaging<br>Configuration<br>/ Shelf life<br>Testing | ASTM F1980-07, Standard Guide for<br>Accelerated Aging of Sterile Barrier<br>Systems for Medical Devices. | Establishes integrity of the package materials<br>and sterile barrier after shelf life testing, visual<br>inspections and nebulization function test. |
| | ASTM F2096, Standard Test Method<br>for Detecting Gross Leaks in<br>Packaging by Internal Pressurization<br>(Bubble Test) | The packaging passed all of the acceptance<br>criteria for the tests. |
| Sterility<br>Testing | ISO 11137-1:2006 Sterilization of<br>health care products Radiation Part<br>1: Requirements for development,<br>validation and routine control of a | Validates the correct dose of gamma radiation<br>to ensure a sterile product at the point of use<br>when manufactured, stored, and handled<br>according to the instructions for use. |
{6}------------------------------------------------
# Section 31 – Updated 510(k) Summary
| | sterilization process for medical<br>devices<br>ISO 11137-2:2013 (third edition)<br>Sterilization of health care products -<br>Radiation Part 2: Establishing the<br>sterilization dose | The packaging passed all of the acceptance<br>criteria for the tests. |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Lift Pressure | N/A | Measures the amount of negative pressure<br>required to pull liquid up for nebulization.<br>The device passed all of the acceptance<br>criteria for the tests. |
| Oxygen<br>Entrainment | N/A | Measures the percentage of oxygen/air mixture<br>when adjusted to preset oxygen percentage.<br>The device passed all of the acceptance<br>criteria for the tests. |
| Nebulization<br>Rate | N/A | Determines the basic rate at which the liquid is<br>aerosolized and emitted in mL/min.<br>The device passed all of the acceptance<br>criteria for the tests. |
| Particle Size<br>Distribution | N/A | Determines the particle size (MMAD) and<br>geometric standard deviation (GSD) of the<br>aerosolized liquid. |
## L. Substantial Equivalence
The proposed Hudson RCI® AquaPak® Sterile Prefilled Nebulizer is substantially equivalent in intended use, design, performance and principles of operation to the identified predicate device cleared under 510(k) K141214. The differences between the proposed Hudson RCI® AquaPak® Sterile Prefilled Nebulizer and the predicate device are minor and raise no new issues of safety and efficacy.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.