← Product Code [KIF](/productcode/KIF) · K152956

# Pulp Canal Sealer, Pulp Canal Sealer EWT (K152956)

_Sybron Dental Specialties · KIF · May 6, 2016 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K152956

## Device Facts

- **Applicant:** Sybron Dental Specialties
- **Product Code:** [KIF](/productcode/KIF.md)
- **Decision Date:** May 6, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3820
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Pulp Canal Sealer and Pulp Canal Sealer EWT are used for permanent obturation of the root canal space with the aid of obturating points in accordance with ISO 6876 for Dental Root Canal Sealing materials.

## Device Story

Pulp Canal Sealer and Pulp Canal Sealer EWT are two-component (powder base and liquid catalyst) zinc oxide/eugenol root canal sealants. The powder base contains zinc oxide and thymol iodide; the liquid catalyst contains eugenol. The components are mixed into a homogenous paste by a licensed dental professional in a clinical setting. The paste is placed as a thin layer into the root canal space alongside obturating points (e.g., gutta percha) to fill microscopic gaps between the filler and the root canal wall. The material sets via a chelation reaction between zinc oxide and eugenol, which occurs in the presence of moisture. The device provides a permanent seal for the root canal, aiding in the success of endodontic treatment.

## Clinical Evidence

No clinical data. Bench testing only, including consistency, solubility, work/set time, film thickness, and biocompatibility testing per ISO 6876:2012 and 21 CFR 820.30.

## Technological Characteristics

Two-component zinc oxide/eugenol paste. Powder base: Zinc Oxide, Thymol Iodide. Liquid catalyst: Eugenol. Setting mechanism: Chelation reaction. Complies with ISO 6876:2012 for dental root canal sealing materials. Film thickness < 50 µm.

## Regulatory Identification

A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.

## Predicate Devices

- Pulp Canal Sealer/Pulp Canal Sealer EWT ([K945244](/device/K945244.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 6, 2016

Sybron Dental Specialties Ms. Wendy Garman VP. Regulatory Affairs 1717 W. Collins Ave. Orange, California 92867

Re: K152956

Trade/Device Name: Pulp Canal Sealer and Pulp Canal Sealer EWT Regulation Number: 21 CFR 872.3820 Regulation Name: Root Canal Filling Resin Regulatory Class: II Product Code: KIF Dated: March 7, 2016 Received: March 8, 2016

Dear Ms. Garman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina Kiang

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known) K152956

#### Device Name

Pulp Canal Sealer and Pulp Canal Sealer EWT

Indications for Use (Describe)

Pulp Canal Sealer and Pulp Canal Sealer EWT are used for permanent obturation of the root canal space with the aid of obturating points in accordance with ISO 6876 for Dental Root Canal Sealing materials.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and comblete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Sybron Dental Specialties. The logo features the letters "sds" in a bold, sans-serif font, with a curved line above the letters. Below the letters, the words "Sybron Dental Specialties" are written in a smaller, sans-serif font. The logo is blue and white.

## Pulp Canal Sealer and Pulp Canal Sealer EWT

- Submitter Information: 1.
Sybron Dental Specialties 1717 W. Collins Ave. Orange CA, 92687

| Contact Person:   | Heather Crilly |
|-------------------|----------------|
| Telephone Number: | 909-962-5664   |
| Fax Number:       | 909-962-5694   |

Date Prepared: May 05, 2016

- 2. Device Name:
.

.

●

- Proprietary Name: Pulp Canal Sealer and Pulp Canal Sealer EWT
- Classification Name: Root Canal Filling Resin 872.3820
- CFR Number: .
	- Device Class: II
- Product Code: KIF .
- 3. Predicate Device:

The Proposed Pulp Canal Sealer and Pulp Canal Sealer EWT are substantially equivalent to the legally marketed device Pulp Canal Sealer/Pulp Canal Sealer EWT (K945244) cleared on December 29, 1994, product code KIF.

- 4. Description of Device:
Pulp Canal Sealer and Pulp Canal Sealer EWT are used for permanent obturation of the root canal space with the aid of obturating points in accordance with ISO 6876 for Dental Root Canal Sealing materials. They are Zinc Oxide/Eugenol root canal sealants. The products are made up of two separate components, a powder base and a liquid catalyst. to form the final device. The liquid catalyst is the same formulation for both products. Pulp Canal Sealer is a fast setting material, while Pulp Canal Sealer EWT features an extended work time of greater than 6 hours on the pad.

- 5. Statement of Indications for Use:
Pulp Canal Sealer and Pulp Canal Sealer EWT are used for permanent obturation of the root canal space with the aid of obturating points in accordance with ISO 6876 for Dental Root Canal Sealing materials.

{4}------------------------------------------------

- Description of Substantial Equivalence 6.
Technological Characteristics:

The powder base formulation of the Proposed Pulp Canal Sealer and the predicate device, Pulp Canal Sealer/Pulp Canal Sealer EWT (K945244), are very similar. The only difference being the formula range of the ingredient Thymol Iodide in the Pulp Canal Sealer product. The predicate Pulp Canal Sealer contains a higher limit of Thymol Iodide, while the Proposed Pulp Canal Sealer contains a lower formula range. The Proposed Pulp Canal Sealer EWT powder base remains unchanged.

The liquid catalyst is the same formulation used for both the Proposed Pulp Canal Sealer and Pulp Canal Sealer EWT and the predicate device, Pulp Canal Sealer/Pulp Canal Sealer EWT (K945244). The liquid catalyst remains unchanged.

Pulp Canal Sealers are intended to fill the microscopic gaps between the root canal filler/obturator product (i.e. Gutta Percha) and the root canal wall. The powder base contains Zinc Oxide and the liquid catalyst contains Eugenol. The base and catalyst are combined and mixed into a homogenous paste which is then placed directly into the root canal as a thin layer. Once the Zinc Oxide and Eugenol are mixed, the Eugenol is consumed and converted into a solid complex. The sealer paste will set due to a chelation reaction between the Zinc Oxide and Eugenol. The hardening of the thin layer occurs within the root canal in the presence of moisture.

Thymol Iodide has no significance within the chemical reaction that takes place in the actual formation of the root canal sealant. The only known effect of the addition of Thymol Iodide is that it may accelerate set times. The difference between the predicate limit and the proposed limit of Thymol Iodide is a minor change, and no effect to the formulation could be measured or recorded.

The formulation utilizing the lowered formulation range of Thymol Iodide has been on the market for a number of years in Europe. Based on post market data and design verification testing, the performance characteristics of this material results are substantially equivalent to those of the predicate. The slight variation in Thymol Iodide does not change the performance characteristics of the device.

- 7. Non-Clinical Performance Data:
Verification and validation activities were performed in accordance with design control requirements as specified in 21 CFR 820.30, ISO 13485:2012 Medical Devices- Quality Management Systems, and ISO 6876 Dental - Root Canal Sealing Materials, and the results demonstrated substantial equivalence to the predicate. The Tables 5.1 and 5.2 below depict the modifications associated with Pulp Canal Sealer and Pulp Canal Sealer EWT.

Non-clinical performance data included testing for Consistency, Solubility, Work/Set Time, Film Thickness, and Biocompatibility testing (refer to Table 6.2). The data analyzed from the various tests substantiate that the Proposed Pulp Canal Sealer and Pulp Canal Sealer EWT are substantially equivalent to the predicate devices. The following standard was utilized for non-clinical performance testing of the proposed:

{5}------------------------------------------------

- ISO 6876:2012 Dentistry Root canal sealing materials .
# Clinical Performance Data:

Clinical performance testing has not been performed for Pulp Canal Sealer and Pulp Canal Sealer EWT.

| Element                                                            | Predicate Device-<br>Pulp Canal Sealer/Pulp Canal<br>Sealer EWT (K945244)                                                                                                                                                      | Proposed Pulp Canal Sealer<br>and Pulp Canal Sealer EWT<br>(K152956)                                                                                                                                                                           |
|--------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name                                                         | Pulp Canal Sealer/Pulp Canal<br>Sealer EWT                                                                                                                                                                                     | Pulp Canal Sealer and Pulp<br>Canal Sealer EWT                                                                                                                                                                                                 |
| Target Users                                                       | Licensed dental professionals                                                                                                                                                                                                  | Licensed dental professionals                                                                                                                                                                                                                  |
| Device Description                                                 | Pulp Canal Sealer and Pulp<br>Canal Sealer EWT are used for<br>permanent obturation of the root<br>canal space with the aid of<br>obturating points in accordance<br>with ISO 6876 for Dental Root<br>Canal Sealing materials. | The Proposed Pulp Canal Sealer<br>and Pulp Canal Sealer EWT are<br>used for permanent obturation of<br>the root canal space with the aid<br>of obturating points in<br>accordance with ISO 6876 for<br>Dental Root Canal Sealing<br>materials. |
| Common Name                                                        | Root Canal Sealer                                                                                                                                                                                                              | Root Canal Sealer                                                                                                                                                                                                                              |
| Classification Name                                                | Root Canal Filling Resin<br>per 21 CFR § 872.3820                                                                                                                                                                              | Root Canal Filling Resin<br>per 21 CFR § 872.3820                                                                                                                                                                                              |
| Class                                                              | II                                                                                                                                                                                                                             | II                                                                                                                                                                                                                                             |
| Product Code                                                       | KIF                                                                                                                                                                                                                            | KIF                                                                                                                                                                                                                                            |
| Powder Base Formula<br>for Pulp Canal Sealer<br>(PN 24875) Only    | Contains:<br>Thymol Iodide                                                                                                                                                                                                     | Contains:<br>Decreased range of Thymol<br>Iodide                                                                                                                                                                                               |
| Powder Base Formula<br>for Pulp Canal Sealer<br>EWT(PN 24746) Only | Formula per K945244                                                                                                                                                                                                            | No Change from K945244                                                                                                                                                                                                                         |
| Catalyst Formula (Same<br>for both products)                       | Formula per K945244                                                                                                                                                                                                            | No Change from K945244                                                                                                                                                                                                                         |
| Manufacturing Location                                             | Romulus, MI (USA)                                                                                                                                                                                                              | Scafati, Italy                                                                                                                                                                                                                                 |
| Legal<br>Manufacturer/Spec<br>Developer                            | Kerr Corporation<br>Romulus, MI                                                                                                                                                                                                | SybronEndo<br>Glendora, CA                                                                                                                                                                                                                     |
| Packaging                                                          | Powder:<br>Cap - Polypropylene<br>Bottle – Polystyrene<br><br>Liquid:<br>Cap – Low Density Polyethylene<br>Bottle – Polypropylene                                                                                              | Powder:<br>Cap - Polypropylene<br>Bottle – Polystyrene<br><br>Liquid:<br>Cap –Low Density Polyethylene<br>Bottle – Polypropylene                                                                                                               |

### Table 5.1: Predicate Pulp Canal Sealer/Pulp Canal Sealer EWT (K945244) and Proposed Pulp Canal Sealer and Pulp Canal Sealer EWT Device Comparison Table

{6}------------------------------------------------

Table 5.2: Predicate Pulp Canal Sealer/Pulp Canal Sealer EWT (K945244) and Proposed Pulp Canal Sealer and Pulp Canal Sealer EWT Physical and Mechanical Properties Comparison

| PHYSICAL & MECHANICAL PROPERTIES COMPARISON |                                                             |                                                       |                                                             |                                                       |
|---------------------------------------------|-------------------------------------------------------------|-------------------------------------------------------|-------------------------------------------------------------|-------------------------------------------------------|
|                                             | Predicate Device (K945244)                                  |                                                       | Proposed Device (K152956)                                   |                                                       |
| Element                                     | Pulp Canal<br>Sealer                                        | Pulp Canal<br>Sealer EWT                              | Pulp Canal<br>Sealer                                        | Pulp Canal<br>Sealer EWT                              |
| Chemistry of<br>Setting Reaction            | Chelation between zinc oxide & eugenol                      |                                                       | Chelation between zinc oxide & eugenol                      |                                                       |
| Consistency                                 | 29 mm                                                       | 28.5 mm                                               | 28 mm                                                       | 25 mm                                                 |
| Solubility                                  | 0.09%                                                       | 0.47%                                                 | 0.34%                                                       | 0.36%                                                 |
| Work/Set Time                               | Work Time:<br>> 45 Minutes<br>Set Time:<br>< 60 Minutes Max | Work Time:<br>> 6 Hours<br>Set Time:<br>< 2 Hours Max | Work Time:<br>> 45 Minutes<br>Set Time:<br>< 60 Minutes Max | Work Time:<br>> 6 Hours<br>Set Time:<br>< 2 Hours Max |
| Film Thickness<br>(µm)                      | 22.6 µm                                                     | 23.6 µm                                               | 20.66 µm                                                    | 33.87 µm                                              |
| Safety                                      | Pass                                                        |                                                       | Pass                                                        |                                                       |

Conclusion as to Substantial Equivalence:

Pulp Canal Sealer/Pulp Canal Sealer EWT (K945244) is an existing device which was granted FDA market clearance in 1994. Sybron Dental Specialties seeks only to slightly modify the existing device cleared under K945244. Additionally, the manufacturer name will change to SybronEndo, the device will be produced at a sister facility in Scafati, Italy (Kerr Italia), and the product code will be updated to KIF. Tables 5.1 and 5.2 depict the modifications associated with the Proposed Pulp Canal Sealer and Pulp Canal Sealer EWT.

The film thickness of the proposed device is roughly 10 microns thicker than the predicate device as shown in Table 5.2 above. Table 5.2 demonstrates that the difference in film thickness between the predicate and proposed devices does not affect the substantial equivalence of the subject and predicate devices. The sealer was tested to meet ISO 6876 standards which state that a sealer shall have a film thickness of not more than 50 um. The Proposed Pulp Canal Sealer and Pulp Canal Sealer EWT is still below the maximum 50 um allowance and will flow sufficiently to fill the voids, as the film thickness remains within the proper specifications of the ISO standard requirements.

The modification of Pulp Canal Sealer (PN 24875) to lower the formulation range of the ingredient Thymol Iodide, and the move of the manufacturing location to Scafati, Italy do not affect the intended use nor does it alter the fundamental scientific technology of the device. The nonclinical testing demonstrates that the Proposed Pulp Canal Sealer and Pulp Canal Sealer EWT perform as well as the predicate device. The Proposed Pulp Canal Sealer and Pulp Canal Sealer EWT are substantially equivalent to the predicate Pulp Canal Sealer/Pulp Canal Sealer EWT (K945244).

---

**Source:** [https://fda.innolitics.com/device/K152956](https://fda.innolitics.com/device/K152956)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
