K152936 · National Dental, Inc. · EBZ · Nov 13, 2015 · Dental
Device Facts
Record ID
K152936
Device Name
Sirius Max Dental Diode Curing Light
Applicant
National Dental, Inc.
Product Code
EBZ · Dental
Decision Date
Nov 13, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6070
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Intended Use
The Sirius Max Diode Curing Light is intended for use as a source of illumination for photo initiated curing and/or activation/ polymerization of dental restorative materials and adhesives.
Device Story
Sirius Max Diode Curing Light is a battery-operated, handheld dental device used by clinicians to polymerize restorative materials and adhesives. It utilizes a solid-state LED module to emit high-intensity light (430-490nm) for curing and a separate white light LED (400-700nm) for intraoral transillumination. The device features a microprocessor-controlled user interface with an OLED display, allowing selection of six curing modes (Normal, Softstart, Pulse, High Power, Xtra Power, Orthodontics) and nine timer settings. The handpiece is constructed of aluminum with a glass lens and is used with disposable barrier sleeves for infection control. The clinician operates the device via a membrane touch pad to trigger light output. By providing controlled radiant energy, the device enables the hardening of dental materials, facilitating restorative procedures. It is recharged via a dedicated station connected to a 5vdc power source.
Clinical Evidence
Bench testing only. Comparative performance testing against the predicate device measured light output across all modes, showing an average variation of 1.4% from settings, well within the ±20% limit per IEC 60601-2-22. Additional testing included depth of cure, conformance to ANSI/ADA Specification No. 48-2, temperature control validation, cleaning/disinfection validation, and software verification/validation.
Technological Characteristics
Materials: Aluminum body, plastic control surface, glass output lens. Light source: Solid-state LED (430-490nm curing; 400-700nm transillumination). Power: Internal Li-battery, 5vdc charging station. Connectivity: Standalone. Standards: IEC 60601-1, IEC 60601-2-2, IEC 60601-2-22, ANSI/ADA 48-2. Software: Microprocessor-based control system with 6 curing modes and 9 timer settings. Infection control: Disposable barrier sleeves.
Indications for Use
Indicated for use as a source of illumination for photo-initiated curing and/or activation/polymerization of dental restorative materials and adhesives in a dental clinical setting.
Regulatory Classification
Identification
An ultraviolet activator for polymerization is a device that produces ultraviolet radiation intended to polymerize (set) resinous dental pit and fissure sealants or restorative materials by transmission of light through a rod.
K121093 — D1 LED CURING LIGHT, SPEC 3, DENTRONIX LED 3000 · Dxm Co., Ltd. · Nov 27, 2012
K173876 — Cybird LED Curing Light · Dxm Co., Ltd. · Jan 24, 2018
K140432 — HI-LIGHT, HI-LIGHT PLUS · Dentall Corporation · Aug 3, 2015
K072181 — MINI LED AUTOFOCUS · Satelec · Sep 19, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are rendered in black and have a flowing, interconnected design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 13, 2015
National Dental Inc. c/o Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, Minnesota 55313
Re: K152936
Trade/Device Name: Sirius Max Diode Curing Light Regulation Number: 21 CFR 872.6070 Regulation Name: Ultraviolet Activator for Polymerization Regulatory Class: Class II Product Code: EBZ Dated: October 29, 2015 Received: November 3, 2015
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runno DDS, MA
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K152936
Device Name SiriusMax Diode Curing Light
Indications for Use (Describe)
The Sirius Max Diode Curing Light is intended for use as a source of illumination for photo initiated curing and/or activation/ polymerization of dental restorative matcrials and adhesives.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">✔</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">❏</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
### NDI's Sirius Max Diode Curing Light
| Submitted for: | National Dental Inc., |
|-----------------|-----------------------|
| | 3-149 Collier St. |
| | Barrie, Ont. Canada |
| Phone: | 800-392-1171 |
| Facsimile: | 800-262-6888 |
| Contact Person: | Cliff Magneson |
| Date Prepared: | 25 October 2015 |
| | Device Proprietary Name(s): | Sirius Max Diode Curing Light |
|--|-----------------------------|--------------------------------------------------|
| | Common or Usual Name: | 400 nm Diode Curing Light (Class II Curing Light |
| | Product Classification: | Ultraviolet activator for polymerization |
| | Product Code: | EBZ |
| | Predicate Device(s): | VALO CORDLESS K110582 |
### Rationale for Substantial Equivalence
Both the subject and predicate Curing Light device share similar intended uses technical characteristics, features, and specifications. The Curing Light characteristics of the Sirius Max Diode Curing Light, including working wavelengths and outputs, Curing Light delivery methods, safety features, and performance specifications are similar to those of the cleared Ultradent Valo Cordless Diode Curing Light. The Curing Light operating system and controls of the subject device are similar to those used by the previously-cleared predicate device showing substantial equivalence. Safety and performance test results have been shown to satisfy applicable international standards recognized by the Agency.
### Intended Uses and Indications for Use
The Sirius Max Diode Curing Light is intended for use as a source of illumination for photo initiated curing and/or activation/ polymerization of dental restorative materials and adhesives.
{4}------------------------------------------------
### Device Description
The Sirius Max Diode Curing Light is shipped as a system consisting of the charging station ("charger"), the handpiece, 50 SiriusMax Barrier Sleeves, an orange protective eye shield, and screwdriver for removal of the battery. The charger allows for placement of the handpiece in a holder with electrical connection for battery charging and interface with the 5vdc power source. The handpiece is constructed of aluminum with plastic control surface and glass output lens and is non-sterile utilizing the barrier sleeves for infection control. The handpiece houses the control circuitry with microprocessor and user interface for selection of mode and curing light output providing visual indications of power settings and of the unit's status. This interface also features the unit's power ON/OFF switch as well as the button to activate the white light intraoral illumination light to aid in the visualization of dental anatomy. The unit is battery operated and contains an internal battery pack with Li batteries. The battery pack is service removable with the use of the screwdriver provide. An instruction for use is also included inside the packaging. The Instructions For Use detail the function of the device and describe the 6 different curing modes. They are Normal mode 1200mw/cm2, Softstart mode 1400mw/cm2, Pulse mode 1400mw/cm2, High Power mode 1300 mw/cm2, Xtra Power mode, 3000mw/cm2, and Orthodontics mode 1400mw/cm2. The curing LED module provides high intensity light in the range of 430-490nm. The white light transillumination LED module (400-700nm) is provided as an aid for illuminating anatomy in the oral cavity.
### Conformity to International Standards
The Sirius Max Diode Curing Light complies with the performance requirements as the predicate device: IEC 60601-1, IEC 60601-2-2, , and IEC 60601-2-22, ANSI/ADA 48-2. The following tests were performed in addition to bench tests described in the 510(k); depth of cure, conformance to ANSI/ADA Specification No. 48-2, Temperature Control Validation, cleaning and disinfection validation, software verification and Validation and IEC 60601-1 Electrical Safety.
### Comparative Performance Data
The Sirius Max Diode Curing Light has been tested side-by-side against the predicate device. Measurements of the subject devices output ranging from 0.1 to 5.0W output all modes were shown to vary from the unit's settings by an average of only 1.4% in all modes. The intended performance of these devices, based on IEC 60601-2-22, is that Curing Light output should vary from the device's setting by less than ± 20% of the setting. Both the subject and predicate devices have been shown to satisfy this standard.
{5}------------------------------------------------
## Comparison of Features and Characteristics
Table 1, following, lists key Features and Characteristics of the subject and predicate devices. Note that the slight change in verbage of Indications for Use is intended to more accurately describe the process of photo-activation of dual cured materials where curing is initiated but not completed by the light activation process. The SiriusMax also has 9 different timer settings for the curing modes compared to 8 found on the predicate device however, output intensities are in the same range and testing was done to validate the depth of cure and confirm substantial equivalence. The SiriusMax has a single Peak wavelength at 460nm whereas the predicate device has a primary peak at 460nm and a small peak at 410nm. The low enerqy of the small peak at 410nm combined with the fact that the vioilet wavelengths have lower penetration depth than blue wavelenghths makes the 460nm the predominant activator and therefore the subject and predicate are substantially equivalent.
| | NDI Sirius Max Diode Curing Light | Ultradent Products Inc. Valo Cordless |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | The Sirius Max Diode Curing Light is intended<br>for use as a source of illumination for photo<br>initiated curing and/or activation/polymerization<br>of dental restorative materials and adhesives. | Source of illumination for curing photo-activated<br>dental restorative materials and adhesives. |
| Product Code | EBZ | EBZ |
| Regulation Number | 21 CFR 872.6070 | 21 CFR 872.6070 |
| Device Class | II | II |
| Classification<br>Panel | Dental | Dental |
| Common Name of<br>Classification | Ultraviolet activator<br>for polymerization | Ultraviolet activator for<br>polymerization |
| Light Source | Solid State Light Emitting Diode | Solid State Light Emitting Diode |
| Power<br>Requirements | 10w5vDC supplied from 110-240VAC 50-<br>60Hz | 100-240VAC@50-60Hz |
| Peak Wavelength | 460nm | 460nm<br>small peak at 410nm |
| Maximum Power | 3000 mW/cm2 | 3,200 mW/cm2 |
| Light Intensity and<br>Pulse<br>Characteristics | Normal 1200mw/cm2 10,20 sec<br>SoftStart 1400mw/cm2 15 sec<br>Pulse 1400mw/cm2 5,10 sec<br>HighPower1400mw/cm2 5 sec<br>XtraPower 3000mw/cm2 1, 3 sec<br>Orthodontics 1400mw/cm2, 10 x 3 sec | Standard Power (1000mw/cm2) 5,10,15,20<br>seconds<br>High Power (1400mw/cm2)<br>1,2,3,4 seconds<br>XtraPower Mode (3200mw/cm2) 3 seconds only |
| Cooling System | Convection Cooled | Convection Cooled |
| Pulse Control | Digital Emission Control | Digital Emission Control |
| Curing Light Source | Solid-state Light Emitting Diode | Solid-state Light Emitting Diode |
| Materials | Aluminum body, glass lens, plastic controls | Aluminum body, glass lens, plastic controls |
| User Interface | Membrane Touch Pad, OLED Display | Membrane Touch Pad, LED Display |
| Sterile Barrier | FDA Cleared Disposable Sleeve/sheath | FDA Cleared Disposable Sleeve/sheath |
| 510k Number | Pending | K110582 |
{6}------------------------------------------------
# Conclusion
The subject device shares the same principle of operation as the predicate device. Both are LED solid state diode Curing Lights that emit radiant energy at an approximate peak wavelength of 460 nm (nom) with outputs that range from 0.1 to 5.0W. Both deliver Curing Light energy to subject target material(s) controlled by trained, experienced clinicians. They share the same indications for use. Both have been found to satisfy international safety standards relating to electrical medical devices in qeneral and dental Curing Lights in particular. Thev share similar safety labeling, and associated safety features. Both the subject and predicate device's outputs were measured and compared to their settings to determine the accuracy of the devices controls. Both met international standards pertaining to accuracy of output of photonic emissions.
The Sirius Max Diode Curing Light device shares intended uses, principle of operation, technical attributes, functional capabilities, biocompatibility standards and performance characteristics with the listed predicate device. Both the subject and predicate device have been shown to comply with applicable Federal and international safety and performance standards. Our conclusion is that the Sirius Max Curing Light is substantially equivalent to the listed predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.