Ambu SPUR II Adult Resuscitator, Ambu SPUR II Pediatric Resuscitator, Ambu SPUR II Infant Resuscitator
Applicant
Ambu A/S
Product Code
BTM · Anesthesiology
Decision Date
Aug 29, 2016
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 868.5915
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Ambu SPUR II resuscitator is a single patient use resuscitator intended for pulmonary resuscitation. The range of application for each version is: -Adult: Adults and children with a body weight more than 30 kg (66lbs). -Paediatric: infants and Children with a body weights up to 30 kg (66lbs). -Infant: Neonates and infant with a body weight up to 10kg (22lbs).
Device Story
Manual, single-patient-use resuscitator for pulmonary resuscitation. Device consists of patient valve, bag, inlet valve, reservoir bag/tube, and oxygen tube. Operator manually compresses flexible bag to deliver air/oxygen to patient airway via face mask, endotracheal tube, or laryngeal mask. Patient valve directs inspiratory flow to patient and expiratory flow to atmosphere. Optional M-Port in valve housing allows drug delivery or side-stream EtCO2 monitoring. Pressure limiting valve (40 cmH2O) included in Pediatric/Infant versions; optional in Adult. Used in clinical settings by healthcare providers. MR Conditional. Benefits include portable, disposable airway support.
Clinical Evidence
No clinical data. Bench testing only. Compliance with ISO 10651-4, ISO 5356-1, and ISO 10993-1 (biocompatibility) demonstrated. Performance testing included mechanical shock, drop test, immersion, ventilatory requirements (delivered volume, resistance, pressure limitation), and M-Port CO2 monitoring accuracy. MR Conditional status verified per ASTM F2503-05.
Technological Characteristics
Manual, operator-powered resuscitator. Components: patient valve, flexible bag, inlet valve, reservoir. Materials: biocompatible polymers (ISO 10993-1). Standards: ISO 10651-4:2009, ISO 5356-1:2004, EN 13544-2:2009. MR Conditional. Optional M-Port for EtCO2/medication. Pressure limiting valve (40 cmH2O).
Indications for Use
Indicated for pulmonary resuscitation in adults and children >30kg (Adult version), infants and children ≤30kg (Pediatric version), and neonates and infants ≤10kg (Infant version).
Regulatory Classification
Identification
A manual emergency ventilator is a device, usually incorporating a bag and valve, intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, layered on top of each other to create a sense of depth and unity.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 29, 2016
Ambu A/S % Mr. Sanjay Parikh Director, QA/RA Ambu Inc. 6230 Old Dobbin Lane, Suite 250 Columbia, Maryland 21045
Re: K152931
Trade/Device Name: Ambu SPUR II Adult Resuscitator Ambu SPUR II Pediatric Resuscitator Ambu SPUR II Infant Resuscitator Regulation Number: 21 CFR 868.5915 Regulation Name: Manual Emergency Ventilator Regulatory Class: Class II Product Code: BTM Dated: June 26, 2016 Received: July 27, 2016
Dear Mr. Sanjay Parikh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K152931
Device Name Ambu SPUR II Adult Resuscitator Ambu SPUR II Pediatric Resuscitator Ambu SPUR II Infant Resuscitator
Indications for Use (Describe)
The Ambu SPUR II resuscitator is a single patient use resuscitator intended for pulmonary resuscitation.
The range of application for each version is:
-Adult: Adults and children with a body weight more than 30 kg (66lbs).
-Paediatric: infants and Children with a body weights up to 30 kg (66lbs).
-Infant: Neonates and infant with a body weight up to 10kg (22lbs).
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995,
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
# Section 5 - 510(k) Summary
This 510(k) summary has been prepared in accordance with 21 CFR 807.87(h) and the 510(k) has been prepared in accordance with 21 CFR 807.92
### 5.1 Submitter
Ambu A/S Baltorpbakken 13 DK-2750 Ballerup Denmark Tel.: +45 7225 2000 Fax.: +45 7225 2055
### 5.2 Contact Person
Name: Line Kagenow Svenstrup Job Title: Regulatory Affairs Professional Address: Ambu A/S, Baltorpbakken 13, DK-2750 Ballerup Telephone number: +45 7225 2217 Fax number: +45 7225 2055
### 5.3 Date summary Prepared
August 22, 2016
# 5.4 Device Trade name
Ambu SPUR II Adult Resuscitator Ambu SPUR II Pediatric Resuscitator Ambu SPUR II Infant Resuscitator
# 5.5 Device Common Name
Manual Resuscitator (Single Patient Use)
# 5.6 Classification name
Ventilator, Emergency, Manual (Resuscitator)
| Device Class: | II |
|----------------------------|-----------------|
| Classification regulation: | 21 cfr 868.5915 |
| Product code: | BTM |
| The panel is: | Anesthesiology |
{4}------------------------------------------------
### 5.7 Legally Marketed Devices to which the device is substantially equivalent:
### Ambu SPUR II Adult Resuscitator:
| Manufacturer | Trade Name | 510k<br>number | Product<br>code |
|--------------|----------------------|----------------|-----------------|
| AMBU A/S | Ambu® SPUR® II Adult | K042682 | BTM |
### Ambu SPUR II Pediatric Resuscitator:
| Manufacturer | Trade Name | 510k<br>number | Product<br>code |
|--------------|--------------------------|----------------|-----------------|
| AMBU A/S | Ambu® SPUR® II Pediatric | K042843 | BTM |
#### Ambu SPUR II Infant Resuscitator:
| Manufacturer | Trade Name | 510k<br>number | Product<br>code |
|--------------|-----------------------|----------------|-----------------|
| AMBU A/S | Ambu® SPUR® II Infant | K042843 | BTM |
### 5.8 Description of the Device
Ambu SPUR II Resuscitator is intended for manual pulmonary resuscitation. It is a disposable device intended for Single Patient Use only.
Ambu SPUR II Resuscitator is available in three sizes: Adult, Pediatric and Infant.
The main components of the product are a patient valve, a bag, an inlet valve, a reservoir bag/tube and an oxygen tube.
The patient valve directs the ventilation air through a patient connector into the patient airway and directs the patient expiration air through an expiration connector. The patient connector is constructed to turn around its own axis enabling the resuscitator to turn in relation to connected masks or endotracheal tubes.
A pressure limiting valve (40 cmHzO) with an override mechanism is placed in the patient valve housing of all Pediatric and Infant resuscitators. For the Adult resuscitators a pressure limiting valve is optional.
M-Port is mounted in the patient valve housing and provides access to the inspiratory and expiratory gas flow allowing connection of a syringe for drug delivery or a qas sampling line for measuring side stream EtCO2. When the M-Port is not used it is
{5}------------------------------------------------
sealed with a cap and the resuscitator operates as a resuscitator without the M-Port. The M-Port is an optional feature.
The bag of the resuscitator is made of a flexible and elastic material. which ensures the bag to have sufficient compression and recoiling properties. When the bag is compressed, air is delivered via the patient valve to the patient.
The inlet valve allows ambient air to flow into the bag and prevents air flowing backwards from the bag through the inlet valve during ventilation.
The reservoir consists of either a reservoir bag (closed reservoir) or a reservoir tube (open reservoir) attached to the inlet valve.
Supplementary oxygen can be supplied through the oxygen tube attached to the oxygen connector (reservoir bag) or mounted inside the inlet valve assembly (reservoir tube). Supplementary oxygen flows into the reservoir attached to the inlet valve assembly.
A Face mask, Endotracheal Tube or Larynqeal Mask/Combitube can be connected to the patient connector of the resuscitator to ensure contact to the patient's airway.
The Ambu SPUR II Resuscitator complies with ISO 10651-4: Particular requirements for operator powered resuscitators.
### 5.9 Intended Use
The Ambu SPUR II resuscitator is a sinqle patient use resuscitator intended for pulmonary resuscitation.
The range of the application for each version is:
- Adult: Adults and children with a body weight more than 30kg (66lb). -
- -Pediatric: Infant and children with a body weight up to 30kg (66lb).
- -Infant: Neonates and infant with a body weight up to 10kg (22lb).
### 5.10 Summary of the technological characteristics in comparison to the predicate devices
Ambu SPUR II Resuscitator is substantially equivalent to:
- Ambu® SPUR® II Adult Resuscitator (K042682); ●
- . Ambu® SPUR® II Infant & Pediatric Resuscitator (K042843)
There are the following differences between the predicate devices and the subject devices:
#### Ambu SPUR II Resuscitator, All Sizes
- The intended use of the subject devices does not include "emergency . support" This use is thus covered by the term "pulmonary resuscitation" and the change will not affect the use of the products.
- The material of the bag has been changed. The change has been verified through bench testing and Biocompatibility Evaluation and does not affect the safety and performance of the resuscitators.
- . Differences in standards used - ASTM F920-93 has been withdrawn without replacement. ISO 8382 has been replaced by ISO 10651-4:2009. ASTM
{6}------------------------------------------------
F1054-87 has been replaced by ISO 5356-1:2004. The New Ambu SPUR II Resuscitators also complies with EN 13544-2:2009.
- . The Ambu SPUR II Resuscitators are MR Conditional. Magnetic Field Interaction Testing had been performed to confirm that the Resuscitator can be used in a MRI environment under certain conditions.
#### Ambu SPUR II Resuscitator Adult
- The M-port of the subject device has an additional function of EtCO2 (end . tidal CO2) side stream measurement. On the predicate device, the M-port was used only for medication delivery. The product risk evaluation and verification of this design change indicate that adding of this function does not introduce any new risk to the product and will not affect the safety and effectiveness of the device.
- . The following changes are only changes to the specifications given to the user; they do not have any impact on the design itself. The following specifications have been updated to show compliance with ISO 10651-4:2009:
- o The stroke volume specification
- The dead space specification o
- The operating temperature O
#### Ambu SPUR II Resuscitator Pediatric and Ambu SPUR II Resuscitator Infant
- · The following change is only a change to the specifications given to the user; it does not have any impact on the design itself:
- o The expiratory resistance specification. This specification has been adjusted in order to fulfil the requirements from the Resuscitator Standard ISO 10651-4:2009.
The above differences have been evaluated and the conclusion is that they do not affect the performance and safety of the subject devices, the New Ambu SPUR II Resuscitator Adult, Pediatric and Infant, compared to the predicate devices in any undesirable way. The operation and technological characteristics of Ambu SPUR II Resuscitators are considered substantial equivalent to the predicate devices operation and technological characteristics. All devices have the same intended use.
Ambu concludes that the Ambu SPUR II Resuscitators are substantially equivalent to the predicate devices.
### 5.11 Performance Data - Bench
Ambu has demonstrated conformity with to the following recognized consensus standards;
- ISO 10651-4 Lung ventilators Part 4: Particular requirements for operator-. powered resuscitators
- ISO 5356-1 Anaesthetic and respiratory equipment Conical connectors -● Part 1: Cones and sockets
All tests have been performed as described in the standards. Result: All tests were passed.
Connectors
- Patient connection portconnector, .
- Expiratory port connector for breathing gases .
- Face mask connectors (connectors of facemask) ●
{7}------------------------------------------------
- . Oxygen tube connector and pressure gauge connector
Operation requirements
- Patient valve function after contamination with vomits .
- Mechanical shock ●
- Drop test ●
- Immersion in water ●
Ventilatory requirements
- Supplementary oxygen and delivered oxyqen concentration ●
- Expiratory resistance ●
- Inspiratory resistance
- Patient valve malfunction ●
- Patient valve leakage Forward leakage ●
- Resuscitator deadspace and rebreathing ●
Ventilation performance
- Minimum delivered volume .
- Pressure limitation .
Storage and Operation, tested at time zero and after accelerated aging
- Inspiratory/Expiratory resistance test ●
- Stroke volume test
- Patient valve malfunction test ●
- Supplementary oxygen and delivery oxygen concentration ●
- . Pressure limitation test
Additional performance data was submitted to document the following properties of Ambu SPUR II:
CO2 monitoring
- Suitability of the M-port to be used for CO2 monitoring (accept criteria: . Maximum deviation of 5% at a confidence level of 95% in comparison to the Philips Side Stream Adapter).
Result: Passed with at least 0.4 liter of tidal volume.
MR Conditional
- . MR Conditional properties (Magnetic Field Interaction Testinq was performed in according to the terminology specified in the American Society for Testing and Materials (ASTM) International, Designation: F2503-05 Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment)
Result: Passed, Ambu SPUR II is MR Conditional and can be used in a MRI environment under specified conditions.
Test reports for Biocompatibility shows that the devices complies with the requirements of ISO 10993-1:
- Cytotoxicity (ISO 10993-5) .
- Sensitization (ISO 10993-10) ●
- . Intracutaneous reactivity test (ISO 10993-10)
In addition Extractable and Leachable (E&L) testing has been performed.
Result: All tests were passed
{8}------------------------------------------------
Since the device is in compliance with the listed standards and has passed the listed performance tests, it is concluded that technological characteristics of Ambu SPUR II is as safe and as effective as the currently marketed predicate devices.
### 5.12 Performance Data - Clinical
No Clinical tests are performed.
### 5.13 Conclusion
Based on the Intended use, operation and technological characteristics, and performance testing, it has been concluded that the Ambu SPUR II Adult Resuscitator, Ambu SPUR II Pediatric Resuscitator and Ambu SPUR II Infant Resuscitator is substantial equivalent to the predicate devices.
It is concluded that the Ambu SPUR II Adult Resuscitator, Ambu SPUR II Pediatric Resuscitator and Ambu SPUR II Infant Resuscitator are as safe and as effective as the legally marketed predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.