← Product Code [MQL](/productcode/MQL) · K152904

# SYDNEY IVF EMBRYO BIOPSY MEDIUM (K152904)

_William A. Cook Australia Pty. , Ltd. · MQL · Dec 3, 2015 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K152904

## Device Facts

- **Applicant:** William A. Cook Australia Pty. , Ltd.
- **Product Code:** [MQL](/productcode/MQL.md)
- **Decision Date:** Dec 3, 2015
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 884.6180
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

Sydney IVF Embryo Biopsy Medium is intended for use in assisted reproduction technologies to facilitate the aspiration of blastomeres for pre-implantation genetic diagnosis.

## Device Story

Sydney IVF Embryo Biopsy Medium is a bicarbonate-based, calcium- and magnesium-free culture medium used in assisted reproduction laboratories. During pre-implantation genetic diagnosis (PGD), embryologists place embryos into the medium for approximately five minutes to facilitate the breakdown of gap junctions between blastomeres. This process allows for the removal of one or two blastomeres for genetic analysis. Following biopsy, the embryo is returned to standard culture media. The device is provided as a sterile, single-use liquid in glass vials.

## Clinical Evidence

No clinical data. Bench testing only. Stability studies validated a 20-week shelf-life at 2-8°C, assessing endotoxin, Mouse Embryo Assay (MEA), pH, osmolality, sterility, and concentrations of proline, pyruvate, and ammonia.

## Technological Characteristics

Bicarbonate-based medium; calcium and magnesium-free. Contains Human Serum Albumin (5 mg/mL) and Gentamicin (0.01 mg/mL). Osmolality: 285-295 mOsm/kg; Endotoxin: < 0.40 EU/mL. Packaging: Borosilicate type 1 glass vials with Fluorotec-coated rubber stoppers. Sterilization: Aseptic filtration.

## Regulatory Identification

Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).

## Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

## Predicate Devices

- Sydney IVF Embryo Biopsy Medium ([K023850](/device/K023850.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized eagle or bird in flight, composed of three interconnected human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 3, 2015

William A. Cook Australia Pty, Ltd. Gordana Pozvek Senior Regulatory Affairs Specialist 95 Brandl Street Brisbane Technology Park, Eight Mile Plains Brisbane, QLD 4113 AU

Re: K152904

> Trade/Device Name: Sydney IVF Embryo Biopsy Medium Regulation Number: 21 CFR 884.6180 Regulation Name: Reproductive Media and Supplements Regulatory Class: Class II Product Code: MOL Dated: September 29, 2015 Received: October 1, 2015

Dear Gordana Pozvek,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K152904

Device Name Sydney IVF Embryo Biopsy Medium

Indications for Use (Describe)

Sydney IVF Embryo Biopsy Medium is intended for use in assisted reproduction technologies to facilitate the aspiration of blastomeres for pre-implantation genetic diagnosis.

| Type of Use (Select one or both, as applicable)                                  |                                                                                 |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span>      Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span>      Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in large, white, sans-serif letters at the top, and the word "MEDICAL" is written in smaller, white, sans-serif letters at the bottom.

WILLIAM A. COOK AUSTRALIA PTY. LTD. 95 BRANDL STREET BRISBANE TECHNOLOGY PARK, EIGHT MILE PLAINS BRISBANE, QLD 4113, AUSTRALIA PHONE: 1800.777.222 FAX: +61.7.3841.1288 WWW.COOKMEDICAL.COM ABN 79 005 526 723

# 510(k) Summary

#### SUBMITTED BY:

William A. Cook Australia Pty Ltd 95 Brandl Street Eight Mile Plains OLD 4113 Australia

| Contact Person: | Gordana Pozvek Ph.D.           |
|-----------------|--------------------------------|
| Tel:            | +61 (7) 3841 1188              |
| Fax:            | +61 (7) 3841 3905              |
| E-mail:         | Gordana.Pozvek@CookMedical.com |

Date Prepared: September 29, 2015

#### DEVICE IDENTIFICATION:

| Trade Name:       | Sydney IVF Embryo Biopsy Medium (Product code: K-SIEB-20) |
|-------------------|-----------------------------------------------------------|
| Common Name:      | IVF Culture Media                                         |
| Regulation No:    | 21 CFR 884.6180, Reproductive Media & Supplements         |
| Regulatory Class: | II                                                        |
| Product Code:     | MQL - Media, Reproductive                                 |

### PREDICATE DEVICE:

Sydney IVF Embryo Biopsy Medium (K023850), cleared January 22, 2003.

## DEVICE DESCRIPTION:

Sydney IVF Embryo Biopsy Medium is bicarbonate based, free of calcium and magnesium to facilitate the aspiration of blastomeres for pre-implantation genetic diagnosis of the embryo. Embryos are placed in this medium for approximately five minutes to break down gap junctions between blastomeres. One or two blastomeres are removed, and the embryo is then returned to Cleavage Medium or Blastocyst Medium for further culture.

Sydney IVF Embryo Biopsy Medium contains Human Serum Albumin (5 mg/mL) and Gentamicin (0.01 mg/mL). The device is available as a 20 mL fill only.

{4}------------------------------------------------

The Sydney IVF Embryo Biopsy Medium is provided in glass vials with Fluorotec® coated rubber stoppers held in place with a tamper evident seal. These products are single use, sterile (aseptic filtration) devices.

#### INDICATIONS FOR USE:

Sydney IVF Embryo Biopsy Medium is intended for use in assisted reproduction technologies to facilitate the aspiration of blastomeres for pre-implantation genetic diagnosis.

#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE:

Sydney IVF Embryo Biopsy Medium and the predicate device (K023850) have the same fundamental technology and similar technological characteristics including the following:

- Similar chemical formulation
- Similar performance specifications:
	- Osmolality 285 295 mOsm/kg —
	- Endotoxin < 0.40 EU/mL
	- । A Mouse Embryo Assay (MEA) is used to screen the product for embryo toxicity.
- Same method of manufacturing process aseptic filtration. ●
- Same packaging borosilicate type 1 vials with FluroTec coated stopper and tamper . evident seals.

The modification that was made to the predicate device was a change in shelf-life from 8 weeks at 2-8°C (for predicate device) to 20 weeks at 2-8°C. In addition, minor changes were made to the formulation and specifications.

The technological characteristics of Sydney IVF Embryo Biopsy Medium are comparable to the predicate device.

#### PERFORMANCE DATA:

The shelf-life of Sydney IVF Embryo Biopsy Medium has been validated in stability studies to 20 weeks at 2 - 8°C. Stability tests included endotoxin. MEA. pH. osmolality. sterility and the concentrations of proline (amino acid), pyruvate and the HSA by-product ammonia.

#### CONCLUSION:

The results of the testing provide reasonable assurance that the Sydney IVF Embryo Biopsy Medium is as safe and effective as the predicate device and supports a determination of substantial equivalence.

---

**Source:** [https://fda.innolitics.com/device/K152904](https://fda.innolitics.com/device/K152904)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).