K152898 · Beijing Sincoheren Science and Tech. Development Co., Ltd. · GEX · Feb 9, 2016 · General, Plastic Surgery
Device Facts
Record ID
K152898
Device Name
DIODE LASER THERAPY SYSTEMS
Applicant
Beijing Sincoheren Science and Tech. Development Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Feb 9, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Diode Laser Therapy Systems are intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI). Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
Device Story
Diode Laser Therapy System (Model SDL-B) emits 808nm laser light via handpiece to target chromophores in hair follicles. System includes console, treatment handpiece, and foot switch. Operates by delivering controlled pulse duration and energy density to induce thermal damage to target tissue while utilizing water-based epidermal cooling to protect surrounding skin. Used in clinical settings by trained operators. Output parameters (fluence, pulse duration, frequency) are adjusted by the clinician to achieve hair reduction. Benefits include long-term, stable reduction of hair regrowth.
Clinical Evidence
Bench testing only. No clinical data provided. Performance supported by biocompatibility testing (cytotoxicity, sensitization, irritation per ISO 10993), electrical safety, and EMC testing.
Technological Characteristics
Diode laser system emitting 808nm wavelength. Features sapphire optical window, water cooling system, and foot switch control. Dimensions: 12mm x 10mm spot size. Electrical safety per IEC 60601-1 and IEC 60825-1; EMC per IEC 60601-1-2. Software classified as moderate level of concern.
Indications for Use
Indicated for hair removal and permanent hair reduction on all skin types (Fitzpatrick I-VI).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 9, 2016
Beijing Sincoheren Science And Tech. Development Co., Ltd % Mr. Mike Gu Regulatory Affairs Manager Guangzhou Osmunda Medical Device Technology Service Co., Ltd 7th Floor, Jingui Business Building, No.982 Congyun Rd. Baiyun District Guangzhou, 510420 CN
Re: K152898
Trade/Device Name: Diode Laser Therapy Systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: September 20, 2015 Received: October 1, 2015
Dear Mr. Gu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152898
Device Name
Diode Laser Therapy Systems, Model SDL-B
### Indications for Use (Describe)
The Diode Laser Therapy Systems are intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI). Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a close-up of a logo or symbol set against a green background. The logo features stylized white lines that form an abstract design. The lines appear to be interconnected, creating a sense of unity or network. The overall impression is modern and clean, with the white lines contrasting sharply against the solid green backdrop.
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
#### l. SUBMITTER
Beijing Sincoheren Science and Technology Development Co., Ltd. Room 305, No.43, Xizhimen North Street, Haidian District, Beijing, 100044, China Phone: +86-(0)10-57734966
Fax: +86-(0)10-82290038
| Primary Contact Person: | Mike Gu |
|---------------------------|-------------------------------------------------------|
| | Regulatory Affairs Manager |
| | OSMUNDA Medical Device Consulting Co., Ltd |
| | Tel: (+86) 20-6232 1333 |
| | Fax: (+86) 20-8633 0253 |
| Secondary Contact Person: | Xin Wang |
| | Quality Manager |
| | Beijing Sincoheren Science and Technology Development |
| | Co., Ltd. |
| | Tel: (+86) 10-82294249-8002 |
| | Fax: (+86) 10-82294249-8007 |
II. DEVICE
| Name of Device: | Diode Laser Therapy Systems, Model SDL-B |
|-----------------|------------------------------------------|
|-----------------|------------------------------------------|
| Name of Device: | Diode Laser Therapy Systems, Model SDL-B |
|-----------------------|-----------------------------------------------------|
| Common/Usual Name: | Diode Laser Therapy Systems |
| Classification Names: | Powered Laser Surgical Instrument (21 CFR 878.4810) |
| Regulation Class: | II |
| Product Code: | GEX |
PREDICATE DEVICE lll.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a green square with a white design on it. The design appears to be abstract, possibly representing a stylized plant or organic form. The white color contrasts sharply with the green background, making the design stand out. The overall impression is simple and modern.
## Emvera Diolux, K123257
This predicate has not been subject to a design-related recall. No reference devices were used in this submission.
#### IV. DEVICE DESCRIPTION
The Diode Laser Therapy Systems mainly consist of console, treatment hand piece and a foot switch. The diode laser emits wavelength of 808nm from the treatment hand piece, the wavelength effectively penetrates deep into and absorbed by the target chromophore. The adequate pulse duration, energy density and epidermal cooling ensure an adequate thermal damage to the target tissue without damaging the surrounding tissue to achieve effective hair removal.
#### V. INDICATION FOR USE
The Diode Laser Therapy Systems are intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI).
Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
| Specification | Predicate device | Proposed device | Discussion of<br>Differences |
|---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| K number | K123257 | -- | |
| Manufacturer | Emvera Technologies, LLC. | Beijing Sincoheren Science and<br>Technology Development Co., Ltd. | |
| Device name | Emvera Diolux | Diode Laser Therapy Systems | |
| Intended Use | The Emvera Diolux is intended for<br>hair removal, permanent hair<br>reduction on all skin<br>types (Fitzpatrick skin type I-VI).<br><br>Permanent hair reduction is<br>defined as the long-term, stable<br>reduction in the number of hairs<br>regrowing when measured at 6, 9<br>and 12 months after the<br>completion of a treatment<br>regime. | The Diode Laser Therapy Systems<br>are intended for hair removal,<br>permanent hair reduction on all skin<br>types (Fitzpatrick skin type I-VI).<br><br>Permanent hair reduction is defined<br>as the long-term, stable reduction in<br>the number of hairs regrowing when<br>measured at 6, 9 and 12 months<br>after the completion of a treatment<br>regime. | same |
#### VI. SUBSTANTIAL EQUIVELENCE
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a green square with a white design on it. The design appears to be a stylized representation of a plant or a symbol. The white design is abstract and fills most of the square, with some parts extending to the edges.
| Specification | Predicate device | Proposed device | Discussion of Differences |
|-----------------------------|-------------------------------------------------------------------------------------------|----------------------------------------------------|------------------------------------------------------------------------------------|
| Wavelength<br>(nm) | 808 | 808 | Same |
| Output power<br>(W) | 600 | 500 | Less than the<br>output of<br>predicate<br>device. |
| Fluence (Energy<br>Density) | <120 J/cm² | <120 J/cm² | Same |
| Pulse Duration | 5ms-625ms | 5ms-400ms | The<br>maximum<br>pulse<br>duration is<br>smaller than<br>the predicate<br>device. |
| Frequency | 1Hz-10Hz | 1Hz-10Hz | Same |
| Spot Size | 12mm × 12mm | 12mm × 10mm | Smaller than<br>the predicate<br>device. |
| Material | Sapphire | Sapphire | Same |
| Cooling | Water cooling | Water cooling | Same |
| Anatomical Sites | Axilla, Face, bikini, Upper limb<br>and lower limb, brow, upper lip,<br>neck, chest, back | Axilla, Face, bikini, Upper limb and<br>lower limb | Less than the<br>predicate<br>device. |
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility testing
The biocompatibility evaluation for the device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The laser output window of testing included the following tests:
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image is a close-up of a logo or symbol set against a green background. The logo appears to be white and stylized, possibly representing a letter or a combination of abstract shapes. The white design contrasts sharply with the green, making it stand out. The image is cropped, focusing on the central part of the logo.
- Cytotoxicity
- Sensitization
- Irritation
Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the device. The system complies with the IEC 60601-1 and IEC 60825-1 standards for safety and the IEC 60601-1-2 standard for EMC.
## Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure of latent design flaw could directly result in minor injury to the patient or operator.
#### VIII. CONCLUSION
The Diode Laser Therapy Systems and its application comply with standards as detailed in section 9, 11 and 17 of this premarket notification. Non-clinical tests determined that the Diode Laser Therapy Systems to be as safe, as effective and performance is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.