← Product Code [LIT](/productcode/LIT) · K152873

# Crosstella RX Balloon Dilatation Catheter (K152873)

_KANEKA Corporation · LIT · Jan 22, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K152873

## Device Facts

- **Applicant:** KANEKA Corporation
- **Product Code:** [LIT](/productcode/LIT.md)
- **Decision Date:** Jan 22, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Crosstella RX PTA Balloon Dilatation Catheter is intended to dilate stenosis in the iliac, femoral, popliteal, infra-popliteal and renal arteries, and for the treatment of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.

## Device Story

Crosstella RX is a rapid exchange PTA balloon dilatation catheter; consists of distal/proximal tubes, balloon, guidewire transition tube, radiopaque markers, hub, and strain relief. Operates by inserting guidewire (max 0.018 inch) through distal lumen; balloon inflated/deflated via inflation device connected to proximal hub. Used in clinical settings by physicians for peripheral angioplasty and stent post-dilatation. Facilitates vessel patency by dilating stenotic lesions or arteriovenous fistulae. Includes flushing needle for heparinized saline preparation. Bench testing confirms performance and biocompatibility.

## Clinical Evidence

Bench testing only. No clinical data. Testing included dimensional verification, balloon burst pressure, fatigue, compliance, inflation/deflation time, bond strength, flexibility, kink test, torque, radiopacity, coating integrity, particulate evaluation, and shelf-life. Biocompatibility testing included cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemocompatibility, and genotoxicity.

## Technological Characteristics

Rapid exchange PTA balloon catheter. Materials: similar to predicates. Dimensions: 0.018" guidewire compatibility; 2.0-6.0 mm balloon diameter; 20-200 mm working length; 900-2000 mm catheter length. Sensing/Actuation: mechanical balloon inflation. Connectivity: none. Sterilization: not specified. Software: none.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- Metacross RX PTA Balloon Dilatation Catheter ([K150865](/device/K150865.md))
- PowerCross™ .018" OTW PTA Dilatation Catheter ([K093286](/device/K093286.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized symbol featuring a series of overlapping human profiles facing to the right, with a flowing element beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 22, 2016

Kaneka Corporation % Mr. Christopher Sloan Principal Consultant Ouintiles Consulting 1801 Rockville Pike, Suite 300 Rockville, Maryland 20852

Re: K152873

Trade/Device Name: Crosstella RX PTA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: December 18, 2015 Received: December 21, 2015

Dear Mr. Sloan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

510(k) Number (if known) K152873

#### Device Name

Crosstella RX PTA Balloon Dilatation Catheter

#### Indications for Use (Describe)

The Crosstella RX PTA Balloon Dilatation Catheter is intended to dilate stenosis in the iliac, femoral, popliteal, infra-popliteal and renal arteries, and for the treatment of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(K) SUMMARY

# Crosstella RX PTA Balloon Dilatation Catheter

## 510(k) Submitter

Kaneka Corporation Medical Device Division 1-12-32, Akasaka, Minato-ku Tokyo, Japan Contact Person: Toshihiko Motomine Telephone: +81-3-5574-8023 Email: Toshihiko.Motomine@kaneka.co.jp

## Official Correspondent

Christopher M. Sloan Principal Consultant Quintiles Consulting 1801 Rockville Pike, Suite 300 Rockville, MD 20852 Phone: (301) 272-3114 Facsimile: (301) 272-0004 Email: chris.sloan@quintiles.com

Date Prepared: September 10, 2015

{5}------------------------------------------------

## Subject Device Name:

| Trade Name           | Crosstella RX                                                  |
|----------------------|----------------------------------------------------------------|
| Common or usual name | Percutaneous (peripheral) transluminal<br>angioplasty catheter |
| Classification name  | Percutaneous catheter [21 CFR 870.1250]<br>product code (LIT)] |
| Class                | II                                                             |
| Classification Panel | Cardiovascular (74)                                            |

## Predicate Devices:

- Metacross RX PTA Balloon Dilatation Catheter [K150865 (Kaneka Corporation)]
- PowerCross™ .018" OTW PTA Dilatation Catheter or "PowerCross" [K093286 (ev3 ● Inc.)]

#### Device Description:

Crosstella RX PTA Balloon Dilatation Catheter (Crosstella RX) is a rapid exchange balloon dilation catheter, which consists of a distal tube, proximal tube, balloon, guidewire transition tube, radiopaque markers, hub, and strain relief. A balloon is attached to the distal end of the catheter and it can be inflated and deflated using the inflation device connected to the hub at the proximal end. Crosstella RX has a lumen (guidewire transition tube) at the distal end of the catheter through which a guidewire can be inserted, and also an opening along the balloon and distal tube to the guidewire port for the introduction of a guidewire.

The maximum compatible diameter of a guidewire used together with Crosstella RX in a PTA procedure is 0.018 inches; hence, sheath introducers with a minimum diameter of 4 or 5 Fr have been deemed to be compatible with Crosstella RX. The nominal inflated balloon diameters range from 2.0 mm to 6.0 mm with balloon working lengths of 20 mm to 200 mm. Three different catheter working lengths will be provided: 900, 1500 and 2000 mm.

The Crosstella RX is provided with the following accessory device:

- . Flushing Needle: provided per one catheter device for the purpose of flushing and filling of the guidewire lumen with heparinized saline before use.
## Indications for Use

The Crosstella RX is intended to dilate stenosis in the iliac, femoral, ilio-femoral, popliteal, infra-popliteal and renal arteries, and for the treatment of obstructive lesions of native or

{6}------------------------------------------------

synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.

#### Comparison of Indications for Use to Predicate Devices

The indications for use of the Crosstella RX is identical to that of the Metacross RX. The Crosstella RX has the same intended use (percutaneous transluminal angioplasty) and same indications for use as the PowerCross. Therefore, the subject device, Crosstella RX, may be considered substantially equivalent to predicate devices.

#### Comparison of Technological Characteristics to Predicate Devices

#### Technological Comparison to Predicate Devices

The Crosstella RX has the following similarities to the predicate devices:

- Same intended use (same as Metacross RX and PowerCross)
- . Same indications for use (same as Metacross RX and PowerCross)
- Same target population (same as Metacross RX and PowerCross)
- . Same operating principle (same as Metacross RX and PowerCross)
- Similar materials (similar to Metacross RX and PowerCross)
- . Same fundamental scientific technology (same as Metacross RX and PowerCross)
- . Same sterility assurance level and method of sterilization (same as Metacross RX and PowerCross)

#### Performance Testing

To demonstrate substantial equivalence of Crosstella RX to the predicate devices, the technological characteristics and performance criteria were evaluated using the bench testing recommendations outlined in the FDA Guidance Document "Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters" dated September 8, 2010. The following in vitro tests were performed on the subject device:

Dimensional Verification, Balloon Preparation, Deployment and Retraction, Balloon Rated Burst Pressure, Balloon Fatigue (Repeat Balloon Inflations), Balloon Compliance, Balloon Inflation and Deflation Time, Catheter Bond Strength, Flexibility and Kink Test, Torque Strength, Radiopacity, Coating Integrity, Particulate Evaluation, Balloon Rated Burst Pressure (in Stent), and Balloon Fatigue (Repeat Balloon Inflations; in Stent), Transportation and Shelf Life Testing.

{7}------------------------------------------------

The results from these tests demonstrate that the technological characteristics and performance criteria of the Crosstella RX are adequate for the intended use of the device and that the device can perform in a manner equivalent to devices currently on the market with the same intended use.

#### Biocompatibility:

To demonstrate the biological safety of the body-contacting materials and substantial equivalence of the Crosstella RX to the predicate devices, the following biocompatibility testing was performed in accordance with ISO 10993-1:

Cytotoxicity, Sensitization, Intracutaneous reactivity (irritation), Systemic toxicity (acute), Pyrogenicity, Hemocompatibility (thrombogenicity, hemolysis, and immunology), and Genotoxicity (bacterial gene mutation assay, in vitro mammalian genotoxicity assay, and in vivo cytogenetics assay)

The results from these tests demonstrate that the Crosstella RX is biocompatible for its intended use similar to the predicate devices.

#### Conclusions:

The Crosstella RX met all the predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. The Crosstella RX PTA Balloon Dilatation Catheter is substantially equivalent to legally marketed predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K152873](https://fda.innolitics.com/device/K152873)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
