HELIXAR ELECTROSURGICAL GENERATOR WITH ARGON BEAM COAGULATION, HELIXAR MOBILE PEDESTAL

K152860 · Conmed Corporation · GEI · Dec 29, 2015 · General, Plastic Surgery

Device Facts

Record IDK152860
Device NameHELIXAR ELECTROSURGICAL GENERATOR WITH ARGON BEAM COAGULATION, HELIXAR MOBILE PEDESTAL
ApplicantConmed Corporation
Product CodeGEI · General, Plastic Surgery
Decision DateDec 29, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ConMed HelixAR ABC System is intended to deliver argon gas as well as high frequency electrical current for the cutting and/or coagulation of tissue.

Device Story

The HelixAR ABC System is an electrosurgical generator with argon beam coagulation (ABC) capabilities, mounted on a mobile pedestal. It accepts inputs from monopolar, bipolar, and ABC handpieces and footswitches (wired or wireless). The system transforms electrical energy and argon gas flow into high-frequency current and ionized argon gas for tissue cutting and coagulation. It features a GUI for selecting modes, power, and flow settings, and includes an argon tank pressure monitor. Used in clinical settings by healthcare providers, the device provides visual and audible alerts and Automatic Return Monitoring (ARM) for patient safety. The output allows surgeons to perform precise tissue management, potentially reducing blood loss and improving surgical outcomes. The system supports smoke evacuation integration.

Clinical Evidence

Benchtop, ex-vivo, and in-vivo simulated use testing were conducted. Testing verified compliance with AAMI/ANSI ES60601-1, IEC 60601-2-2, and IEC 60601-1-2. Risk management was performed per ISO 14971. No human clinical trial data was required or presented.

Technological Characteristics

Electrosurgical generator with argon beam coagulation; GUI display; monopolar, bipolar, and ABC modes; ABC pulse modes; wireless footswitch compatibility; Automatic Return Monitoring (ARM). Complies with AAMI/ANSI ES60601-1, IEC 60601-2-2, and IEC 60601-1-2. Mobile pedestal houses D-size argon cylinders.

Indications for Use

Indicated for cutting and/or coagulation of tissue using argon gas and high frequency electrical current. Intended for prescription use.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized representation of three human profiles facing to the right, stacked on top of each other. The profiles are depicted in a simple, flowing line style. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 29, 2015 CONMED Corporation Ms. Lisa Anderson Manager, Regulatory Affairs 525 French Road Utica, New York 13502 Re: K152860 Trade/Device Name: HelixAR ABC System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: September 29, 2015 Received: September 30, 2015 Dear Ms. Anderson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related {2}------------------------------------------------ adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # David Krause -S - for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page. #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use 510(k) Number (if known) K152860 Device Name HelixAR ABC System Indications for Use (Describe) The CONMED HelixAR ABC System is intended to deliver argon gas as well as high frequency electrical current for the cutting and/ or coagulation of tissue. Type of Use (Select one or both, as applicable) × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) {4}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ ## 510(k) Summary of Safety and Effectiveness CONMED HelixAR ABC System In accordance with the requirements of the Safe Medical Device Act of 1990 and 21CFR 807.92, ConMed Corporation is hereby submitting the 510(k) Summary of Safety and Effectiveness for 510(k) number K152860 as of December 2, 2015. #### A. Submitter ConMed Corporation 525 French Road Utica, NY 13502 Establishment Registration: 1320894 ### B. Company Contact Lisa Anderson Manager, Regulatory Affairs T: (315) 624-3371 F: (315) 624-3225 ### C. Device Name | Proprietary Name: | HelixAR | |----------------------|-------------------------------------------------------| | Common Name: | Electrosurgical Generator with Argon Beam Coagulation | | Classification Name: | Electrosurgical, Cutting & Coagulation & Accessories | | Regulation Number: | 878.4400 | | Product Code: | GEI | | Regulatory Class: | II | | Panel: | General and Plastic Surgery | #### D. Predicate Device | Primary Device Name: | System 7500-B (currently System 7550) | |----------------------|---------------------------------------| | Company Name: | CONMED Corporation | | 510(k): | K050161 | This predicate has not been subject to a design-related recall. | Secondary Device Name: | System 5000 | |------------------------|--------------------| | Company Name: | CONMED Corporation | | 510(k): | K020186 | This predicate has not been subject to a design-related recall. {6}------------------------------------------------ ### E. Device Description The HelixAR ABC System is composed of an electrosurgical generator with argon beam coagulation (ABC) and a mobile pedestal. The generator is designed with a graphic user interface (GUI) display which allows the user to select monopolar, bipolar, and ABC modes of operation, choose power and argon flow settings, adjust user settings options, and create, edit, or delete user programs. The HelixAR mobile pedestal (cart) houses two D-size argon gas cylinders and provides connection ports for the use of wired footswitches. The mobile pedestal is designed with an argon tank capacity monitor that provides the user with feedback on the argon gas cylinder pressure. ### F. Intended Use / Indications for Use The ConMed HelixAR ABC System is intended to deliver argon gas as well as high frequency electrical current for the cutting and/or coagulation of tissue. ### G. Technological Characteristics The HelixAR ABC System is similar to the predicate devices in that the system includes an electrosurgical generator used in conjunction with monopolar and bipolar handpieces and footswitches and an accessory cart. The HelixAR ABC System output panel provides receptacles for monopolar and bipolar hand- and footcontrolled devices. Footswitch receptacles for monopolar, bipolar, and ABC footswitches are provided on the accessory cart which also accommodates the use of a smoke evacuation system. The HelixAR ABC System provides Automatic Return Monitoring (ARM) for dual dispersive electrodes as well as visual and audible alerts similar to those used in the predicate designs. The HelixAR ABC System differs from the predicate devices in that the generator is designed with a GUI display and is compatible with the CONMED Wireless Footswitch Kit. The Wireless Footswitch Kit provides the same functionality as corded footswitches and the convenience of wireless devices. The HelixAR generator design includes the monopolar, bipolar, and ABC modes present in the predicate devices plus the addition of ABC pulse modes. The ABC pulse modes provide the same output characteristics of the other ABC modes with a pulsed application of energy. #### H. Performance Testing Benchtop and ex-vivo comparison testing as well as in-vivo simulated use testing demonstrate the HelixAR ABC System is substantially equivalent to the ConMed System 7550 and System 5000 with regard to intended use, materials, technology, and performance. Mechanical, electrical, and ex-vivo tissue verification activities and software validation testing demonstrate the devices comply with design requirements and the applicable sections of AAMI/ANSI ES60601-1. IEC 60601-2-2, and IEC 60601-1-2. Risk management activities in accordance with ISO 14971 demonstrate the risks associated with the use of the HelixAR ABC System are mitigated to an {7}------------------------------------------------ acceptable level. Analyses of these activities conclude the benefits associated with the use of the HelixAR ABC System outweigh the residual risks. #### Substantial Equivalence - The differences between the predicate devices and the proposed device do not raise any new risks of safety or efficacy. Supporting information per this premarket submission confirms that the ConMed HelixAR ABC System is safe and effective for its intended use and is substantially equivalent in design, intended use, principals of operation, and technical characteristics to the CONMED System 7550 and CONMED System 5000.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...