K152836 · Nobel Biocare AB · DZE · Apr 11, 2016 · Dental
Device Facts
Record ID
K152836
Device Name
TREFOIL System
Applicant
Nobel Biocare AB
Product Code
DZE · Dental
Decision Date
Apr 11, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TREFOIL System is used to restore chewing function in fully edentulous mandibles. The three implants of the TREFOIL System are placed between the mental foramina in fully edentulous mandibles in a 1-stage surgical technique combined with an immediate function loading protocol, provided sufficient primary stability for the selected technique is achieved. In cases where sufficient primary stability for two implants or more is not reached, the implants along with the Framework may also be used with an early or delayed loading protocol. The following prerequisites must be fulfilled: - Adequate quantity of bone (minimum height of 13 mm and minimum width of 6-7 mm). - Adequate mouth opening (minimum 40 mm) to accommodate the guided surgery instruments. - Implant-supported prosthetics seated directly on dedicated implants
Device Story
TREFOIL System restores chewing function in fully edentulous mandibles; utilizes three threaded CP4 titanium endosseous implants placed between mental foramina. System includes surgical tooling and pre-designed titanium-vanadium alloy prosthetic bar. Implants feature internal conical abutment connection and TiUnite surface treatment. Surgical procedure involves 1-stage technique with immediate function loading protocol; early/delayed loading used if primary stability insufficient. Clinician uses guided surgery instruments to place implants; framework bar provides compensation mechanism for passive fit. System benefits patients by providing fixed prosthetic restoration. Bench testing confirms angular/translational accuracy and fatigue strength per ISO 14801.
Clinical Evidence
Bench testing only. Accuracy of surgical protocol evaluated via cone beam CT on bone models. Passive fit of framework confirmed via 3D modeling. Fatigue testing conducted per ISO 14801, demonstrating higher fatigue strength than predicate. Biocompatibility, sterilization (ANSI/AAMI/ISO 11137/17665-1), and shelf-life validated via predicate data and real-time aging.
Technological Characteristics
Implants: CP4 titanium, parallel-walled, internal conical connection, TiUnite surface (anodic oxidation). Bar: Titanium-vanadium alloy. Surgical tools: Stainless steel (DIN EN 10027-2 1.4197), DLC coated. Sterilization: Gamma radiation (sterile) or moist heat (non-sterile). Connectivity: None (mechanical system).
Indications for Use
Indicated for fully edentulous patients requiring restoration of chewing function in the mandible. Requires adequate bone (min 13mm height, 6-7mm width) and mouth opening (min 40mm). Contraindicated if insufficient primary stability is achieved for immediate loading, though early/delayed loading protocols may be used.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 11, 2016
Nobel Biocare AB c/o Ms. Charlemagne Chua Senior Regulatory Affairs Manager Nobel Biocare USA LLC 22715 Savi Ranch Parkway Yorba Linda. California 92887
Re: K152836
Trade/Device Name: TREFOIL System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA, DZI Dated: March 2, 2016 Received: March 3, 2016
Dear Ms. Chua:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Tina Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
| Form Approved: OMB No. 0910-0120 |
|-----------------------------------|
| Expiration Date: January 31, 2017 |
| See PRA Statement below. |
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
**Indications for Use**
| 510(k) Number (if known) | K152836 |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | TREFOIL System |
| Indications for Use (Describe) | The TREFOIL System is used to restore chewing function in fully edentulous mandibles. |
| | The three implants of the TREFOIL System are placed between the mental foramina in fully edentulous mandibles in a 1-stage surgical technique combined with an immediate function loading protocol, provided sufficient primary stability for the selected technique is achieved. In cases where sufficient primary stability for two implants or more is not reached, the implants along with the Framework may also be used with an early or delayed loading protocol. |
| | The following prerequisites must be fulfilled: |
| | - Adequate quantity of bone (minimum height of 13 mm and minimum width of 6-7 mm). |
| | - Adequate mouth opening (minimum 40 mm) to accommodate the guided surgery instruments. |
| | - Implant-supported prosthetics seated directly on dedicated implants |
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;"><b>Prescription Use</b></span> (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
MARCH 2016
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
K152836-S001: Al Response (11-20-2015)
Page 20 of 270
10
{4}------------------------------------------------
## I. SUBMITTER
Nobel Biocare AB Vastra Hamngatan 1 Goteborg, SE-411 17 Sweden
Submitted by Nobel Biocare USA LLC 22715 Savi Ranch Parkway Yorba Linda, CA 92887
Contact Person: Charlemagne Chua, Senior Regulatory Affairs Manager Phone: (714) 282-4800 x 7830 Fax: (714) 998-9348
Date Prepared: April 7, 2016
II. DEVICE
Name of Device: TREFOIL System Common or Usual Name: Endosseous Dental Implant Classification Name: Endosseous Dental Implant (21 CFR 872.3640) Regulatory Class: II Primary Product Code: DZE Other Product Code(s): NHA. DZI
## III. PREDICATE DEVICE
Primary predicate: Nobel Biocare - Branemark Novum (K000018) This predicate has not been subject to a design-related recall.
Reference predicates: Nobel Biocare - NobelActive (K142260) This predicate has not been subject to a design-related recall
Nobel Biocare - NobelProcera Overdenture Bar (K132749) This predicate has not been subject to a design-related recall
## IV. DEVICE DESCRIPTION
The TREFOIL System is a method of placing three dental implants in predetermined positions (between the mental foramina) and using a pre-designed prosthetic bar to act as a screw-retained framework seated on the implants. The TREFOIL System restores chewing function and esthetics in the mandible in completely edentulous patients.
The TREFOIL System consists of dental implants, surgical components necessary to place the implants in predetermined positions, and prosthetic components that are included in the prosthetic bar or are used in the dental lab during the creation of the prosthetic bar.
A.4.
{5}------------------------------------------------
The dental implants are threaded endosseous implants made of CP4 titanium. The implant is parallel walled and has an internal conical abutment connection. The implant is available in 11.5 mm length with two available collar lengths (4.5, 6.0 mm). The apex of the implants have cutting chamber allowing for self-tapping. The implant bone interface has the TiUnite implant surface treatment. The TREFOIL System bar is made of titanium vanadium allov. The TREFOIL System surgical tooling is made of stainless steel.
## V. INDICATIONS FOR USE
The TREFOIL System is used to restore chewing function in fully edentulous mandibles.
The three implants of the TREFOIL System are placed between the mental foramina in fully edentulous mandibles in a 1-stage surgical technique combined with an immediate function loading protocol, provided sufficient primary stability for the selected technique is achieved. In cases where sufficient primary stability for two implants or more is not reached, the implants along with the Framework may also be used with an early or delayed loading protocol.
The following prerequisites must be fulfilled:
- Adequate quantity of bone (minimum height of 13mm and minimum width of 6-7mm).
- Adequate mouth opening (minimum 40 mm) to accommodate the guided surgery instruments.
- Implant-supported prosthetics seated directly on dedicated implants
{6}------------------------------------------------
#### VI. Comparison of Technological Characteristics
## Comparison of indication for use statement
| | Subject Device | Primary Predicate |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | TREFOIL System | Branemark Novum (K000018) |
| Indication for use statement | The TREFOIL System is used to restore chewing function<br>in fully edentulous mandibles.<br><br>The three implants of the TREFOIL System are placed<br>between the mental foramina in fully edentulous mandibles<br>in a 1-stage surgical technique combined with an<br>immediate function loading protocol, provided sufficient<br>primary stability for the selected technique is achieved.<br>In cases where sufficient primary stability for two implants or<br>more is not reached, the implants along with the Framework<br>may also be used with an early or delayed loading protocol.<br><br>The following prerequisites must be fulfilled:<br>- Adequate quantity of bone (minimum height of<br>13mm and minimum width of 6-<br>7mm).<br>- Adequate mouth opening (minimum 40 mm) to<br>accommodate the guided surgery instruments.<br>- Implant-supported prosthetics seated directly<br>on dedicated implants | Totally edentulous mandibles with a minimum height of 13 mm and a minimum width of 6mm. Patient must be subject to dental treatment with endosseous implants. For use in a single stage procedure where the implants are immediately loaded. |
Comparison of Indications for Use: The intended use of the TREFOL System is substantially equivalent to the Branemark Novum (K000018) predicate. The prerequisites for minimum bone with remain the same, while the TREFOL System provides additional safety prerequisites for mouth opening minimum size and implant supporting of seated prosiderations from the labeling of the Branemark Novum (K00018) predicate have been added to the indications for use of the TREFOL System regarding number of implants for placement, functional loading, and primary stability, which do not affect the intended use.
{7}------------------------------------------------
## Implant Comparison
| Technological<br>characteristics | Subject Device | Primary Predicate | Reference Predicate | |
|----------------------------------|-------------------------|----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|-----------------------------------------|
| Implant Design Features | TREFOIL System | Branemark Novum (K000018) | NobelActive (K142260) | |
| | Implant<br>Body Design | Parallel walled | Parallel walled | Expanding Taper |
| | Implant Tip<br>Design | Tapered self cutting<br>3.8 mm apex w/9.5° angle | Tapered self cutting with through hole<br>3.5 mm apex w/15° angle | Drilling blades on apex<br>5.35 mm apex |
| | Implant<br>Length | 16.0 mm overall (11.5 mm body w/ 4.5 mm<br>collar)<br>17.5 mm overall (11.5 mm body w/ 6.0 mm<br>collar) | 17.5 mm overall (11.5 mm body w/ 6.0 mm<br>collar) | 6.5, 8, 9.5, 11, 12.5 mm |
| | Implant<br>Diameter | 4.93 mm | 5.0 mm | 5.5 |
| | Platform<br>Diameter | 4.5 mm | 4.5 mm | 5.5 |
| | Thread<br>angle/pitch | 60°/.8mm | 60°/.8mm | variable |
| | Connection<br>Type | Internal Hex with snap feature | External hex | Internal Hex |
| | Device<br>Material | CP titanium | CP titanium | CP titanium |
| | Surface<br>modification | TiUnite (anodic oxidation) | Machined titanium | TiUnite (anodic oxidation) |
{8}------------------------------------------------
## Framework Bar Comparison
| Technological<br>characteristics | Subject Device | Primary Predicate | Reference Predicate | |
|----------------------------------|-------------------------------|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Bar Design Features | TREFOIL System | Branemark Novum (K000018) | NobelProcera Overdenture Bar<br>(K132749) | |
| | Bar Design | Preshaped single piece design screw<br>retained to implant. | Preshaped two piece design. Lower bar<br>screw retained to implants directly. Upper<br>prosthetic bar screw retained to lower bar. | |
| | Intended<br>platform | TREFOIL implant (intenal conical connection<br>with snap fit) | Branemark Novum implant (external hex) | |
| | Materials | Titanium vanadium alloy | CP titanium | |
| | Surface<br>modification | Machined titanium | Machined titanium | |
| | Compensation<br>mechanism | Round abutment and corresponding<br>framework and screw disks | Implant collar partially shaped by pre-<br>torquing technique. | |
| | Prosthetic<br>media | Acrylic | Acrylic | |
| | Fixed<br>cantilever<br>length | 18.8 mm | 19.7 | |
| Technological<br>characteristics | | | Subject Device | Predicate |
| | | | TREFOIL System | Branemark Novum (K000018) |
| Surgical tooling type<br>Drill | Diameter | 2.0, 3.0, 3.8, 4.0, 4.2, 4.4 mm | 2.0, 2.5, 3.0, 3.5, 3.8, 4.0, 4.2, 4.3, 4.4 mm | |
| | Flute length | 16.0 mm (including tip) | 18.0, 24.0 mm (excluding tip) | |
| | Material | Stainless Steel<br>DIN EN 10027-2 – 1.4197<br>DLC coated | Stainless Steel<br>440F Hardened to MN 500 HV2<br>No coating | |
| | Connection | ISO 1797 Type 1 | ISO 1797 Type 1 | |
| | Markings | All drills marked at 11.5 mm and 13.5 mm | 18.0 mm length single marked for 11.5 mm implant thread<br>24.0 mm length marked for both 11.5 mm implant thread | |
| | Flute design | Two flutes | Three flutes | |
| | Tip design | 90° (2mm)<br>130° (3.0, 3.8, 4.0, 4.2, 4.4 mm) | 120° | |
| | Intended use | The drills are used together with corresponding Templates<br>for drills for the preparation of the implant sites. | The drills are used together with corresponding Templates for<br>drills for the preparation of the implant sites. | |
| Screw Tap | Flute length | 17.0 mm | 15.5 mm | |
| | Diameter | 5.0 mm | 4.5 mm | |
| | Materials | Stainless Steel<br>DIN EN 10027-2 – 1.4197<br>DLC coated | Titanium vanadium alloy | |
| | Connection | ISO 1797 Type 1 | 2.4 mm square | |
| | Intended use | The screw tap is used when dense bone is present to<br>prepare for the threaded implant | The screw tap is used when dense bone is present to prepare for<br>the threaded implant | |
| Stabilizing<br>screw | Core diameter | 1.8 | 1.8 | |
| | Length | 14.0 mm (to stop)<br>19.3 mm (overall) | 13.0 mm (to stop)<br>20.0 mm (overall) | |
| | Materials | 316 Stainless Steel | 316 Stainless Steel | |
| | Intended use | The stabilizing screw is used to temporarily connect the V-<br>template to the alveolar ridge | Short and long screws used to stabilize the V-gage template to<br>the alveolar ridge | |
{9}------------------------------------------------
# Surgical Tooling Comparison
{10}------------------------------------------------
## VII. PERFORMANCE DATA
Laboratory testing was performed to establish the accuracy of the TREFOIL System surgical protocol. This testing included placement of the Trefoil implants into bone models following the Trefoil Instructions for Use and utilizing Trefoil surgical tooling. The results were evaluated by cone beam computed tomography. This testing established the angular and translational accuracy errors generated by the surgical procedure.
The ability of the Trefoil framework compensation mechanism to adapt to the expected angular and translational accuracy errors qenerated by the surgical procedure was established through the use of three-dimensional modeling. The modeling confirmed that the Trefoil compensation mechanism will ensure a passive fit of the framework bar on the Trefoil implants.
Fatigue testing was performed to establish that the TREFOIL System will withstand foreseeable mastication forces. Fatigue testing was done by using a test scenario developed for splinted implant applications. Fatigue testing was conducted and results analyzed in accordance with ISO 14801. The results demonstrate higher fatigue strength than the predicate Branemark Novum (K000018).
## Summary of Non-Clinical Testing:
Since the subject device does not represent a new worst case, data from the predicate device was leveraged in the following aspects of the 510(k).
- . Sterile Device Information
- The sterilization method for the subiect devices provided sterile is the same as o the predicate. The sterilization method is Gamma radiation and has been validated in accordance with ANSI/AAMI/ISO 11137. Therefore, no additional testing was required.
- The sterilization methods for the subject devices provided non-sterile and to be o end user sterilized are pre-vacuum and gravity moist heat sterilization and have been validated in accordance with ANSI/AAMI/ISO ISO 17665-1.
- . Device Packaging
- The packaging for the subject device is the same as the predicate. This is a o titanium cylinder placed in a plastic vial with PVC shrink-wrap and tamper resistant strip. Therefore, no additional testing was required.
- . Shelf Life
- The packaging for the subject device is the same as the predicate and is labeled o with a 5 year expiration date. Real time aging was used to determine the expiration dating. Therefore, no additional testing was required.
- . Biocompatibility
- The subject device is manufactured from the same material using the same o manufacturing method as the predicate, has the same intended use, and the same patient contact type and duration. Therefore, no additional testing was required
- . Surface Treatment
- The Trefoil implants have the same surface treatment (TiUnite) as the predicate O NobelActive implants. The TiUnite surface treatment was qualified through Auger analysis, IR spectrum analysis, and cytotoxicity testing. Therefore, no additional testing was required.
## VIII. CONCLUSIONS
The information provided in this submission demonstrates that the device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.