← Product Code [KPR](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR) · K152788

# EOS System (K152788)

_Eos Imaging · KPR · Oct 21, 2015 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR/K152788

## Device Facts

- **Applicant:** Eos Imaging
- **Product Code:** [KPR](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR.md)
- **Decision Date:** Oct 21, 2015
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

EOS is intended for use in general radiographic examinations and applications, excluding the evaluation of lung nodules and examinations involving fluoroscopy, and mammography. EOS allows the radiographic acquisition of either one or two orthogonal X-ray images for diagnostic purposes, in one single scan, of the whole body or a reduced area of investigation of a patient in the upright or seated position. The Micro Dose feature is indicated for imaging with a patient entrance dose of 10 to 90 µGy for assessing global skeletal deformities in follow-up pediatric examinations. Micro Dose is not indicated for focal skeletal abnormalities and/or other pediatic abnormalities. Micro Dose is not indicated for use in patients with a Body Mass Index over 30.

## Device Story

Digital radiography system; uses two orthogonal xenon gas-filled detectors and X-ray tubes to acquire frontal and lateral images simultaneously in one scan. Used in clinical settings; operated by trained personnel. Features 'Micro Dose' mode for low-dose pediatric follow-up. System stores images in local database; displays on high-resolution medical monitors for physician diagnosis. Supports DICOM 3.0 transmission for archiving/printing. Provides Dose Structured Reports and Reject/Repeat Analysis. Benefits include reduced radiation exposure for pediatric patients during skeletal deformity monitoring.

## Clinical Evidence

Bench testing only. Performance and functional testing confirmed equivalent performance to predicate, including verification of dosing, image quality, and software validation. Testing demonstrated that optimized acquisition protocols maintain or improve image quality while reducing entrance dose.

## Technological Characteristics

Digital radiography system; two orthogonal xenon gas-filled detectors and X-ray tubes. DICOM 3.0 compatible. Conforms to IEC 60601-1. Features include Micro Dose acquisition, Dose Structured Reporting, and Reject/Repeat Analysis. Software modifications include spectral filter monitoring and image processing contrast/dynamic range optimization.

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- EOS System ([K142773](/device/K142773.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.

October 21, 2015

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

EOS Imaging % John J. Smith, M.D., J.D Partner Hogan Lovells US LLP 555 Thirteenth Street, NW WASHINGTON DC 20004

Re: K152788 Trade/Device Name: EOS System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR Dated: September 25, 2015 Received: September 25, 2015

Dear Dr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert Oals

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known)

# K152788

Device Name

#### EOS System

Indications for Use (Describe)

EOS is intended for use in general radiographic examinations and applications, excluding the evaluation of lung nodules and examinations involving fluoroscopy, and mammography. EOS allows the radiographic acquisition of either one or two orthogonal X-ray images for diagnostic purposes, in one single scan, of the whole body or a reduced area of investigation of a patient in the upright or seated position.

The Micro Dose feature is indicated for imaging with a patient entrance dose of 10 to 90 µGy for assessing global skeletal deformities in follow-up pediatric examinations. Micro Dose is not indicated for focal skeletal abnormalities and/or other pediatic abnormalities. Micro Dose is not indicated for use in patients with a Body Mass Index over 30.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

# EOS imaginq's EOS System

EOS imaqing 10 rue Mercoeur PARIS F-75011 FRANCE

Phone: + 33 1 55 25 60 60 Facsimile: + 33 1 55 25 60 61

> Contact Person: Mathias Breuil, Regulatory Affairs Project Leader

Date Prepared: October 13, 2015

#### Name of Device and Name/Address of Sponsor:

EOS System EOS imaging 10 rue Mercoeur PARIS F-75011 FRANCE Common or Usual Name: Digital Radiography System 21 CFR 892.1680; Stationary X-ray System Classification Name: Product Code KPR - System, X-Ray, Stationary Predicate Devices: K142773 cleared on January 22, 2015 Trade/Device Name: EOS System Regulation: 21 CFR 892.1680; Stationary X-Ray System Regulatory Class: II; Product Code: KPR, MQB

## Intended Use/ indication for use

EOS is intended for use in general radiographic examinations and applications, excluding the evaluation of lung nodules and examinations involving fluoroscopy, angiography, and mammography. EOS allows the radiographic acquisition of either one or two orthogonal X-ray images for diagnostic purposes, in one single scan, of the whole body or a reduced area of investigation of a patient in the upright or seated position.

The Micro Dose feature is indicated for imaging with a patient entrance dose of 10 to 90 µGv for assessing global skeletal deformities in follow-up pediatric examinations. Micro Dose is not indicated for focal skeletal abnormalities and/or other pediatric abnormalities. Micro Dose is not indicated for use in patients with a Body Mass Index over 30.

## Technological Characteristics

EOS is a digital radiography system in which two sets of xenon gas filled digital detectors and X-ray tubes are positioned orthogonally to generate frontal and lateral images simultaneously by scanning the patient over the area of interest. An acquisition feature named Micro Dose allows image acquisition with a patient entrance dose of 10 to 90 uGv for assessing global skeletal deformities in follow-up pediatric exams. The diagnostic images are stored in a local database and are displayed on a high-resolution, medical-quality monitor, where the diagnosis is performed. The diagnostic image can be transmitted through a DICOM 3.0 compatible digital network for printing and archiving.

{4}------------------------------------------------

# Performance Data

EOS is designed to conform to IEC 60601-1 and collateral standards. A CB (certification body) test certificate has been issued. Software verification and validation testing was also conducted. Additional performance and functional testing has confirmed the equivalent performance of the modified EOS compared to the cleared predicate EOS. This testing included bench testing to confirm appropriate dosing and image quality.

# Substantial Equivalence

The modified EOS has the same intended use and similar indications, principles of operation, and technological characteristics as the cleared predicate device.

|                                                                  | Modified EOS System<br>(K152788)                                                                                                                                                                                                 | Comparison with cleared Predicate EOS System<br>(K142773)                                                                                                                                                                     |
|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use/<br>Indications for Use                             | Both devices have the same intended use.                                                                                                                                                                                         |                                                                                                                                                                                                                               |
| Technological<br>Characteristics /<br>Principles of<br>Operation | The fundamental technological characteristics of the modified EOS are unchanged from<br>the cleared EOS. The modified EOS contains the following minor modifications that have<br>been implemented following the last clearance: |                                                                                                                                                                                                                               |
| • Hardware<br>Modifications                                      | Component Supplier Changes                                                                                                                                                                                                       | The change in suppliers has not resulted in any<br>changes to the design or manufacturing<br>specifications of the system. Performance and<br>functional testing has confirmed non-regression.                                |
|                                                                  | Modification to electrical<br>components following<br>obsolescence                                                                                                                                                               | Components have been replaced with iso-functional<br>components.                                                                                                                                                              |
|                                                                  | Mirror Sticker Accessory                                                                                                                                                                                                         | New optional accessory kit                                                                                                                                                                                                    |
| • Software<br>Modifications                                      | Software modification related to a<br>recall action.                                                                                                                                                                             | The acquisition software has been modified in order<br>to improve the detection of a dysfunction of the<br>spectral filter and to prevent the use of the system<br>after the alert message on spectral filtration<br>appears. |
|                                                                  | Optimization of default acquisition<br>protocols                                                                                                                                                                                 | Performance testing has demonstrated that this<br>modification allows reducing the entrance dose for<br>the changed protocols, with maintaining equivalent<br>or better image quality than the cleared EOS.                   |
|                                                                  | Improvement of image<br>processing.                                                                                                                                                                                              | The aim of this optimization is to increase the image<br>contrast and dynamic during the image processing.                                                                                                                    |
|                                                                  | Addition of a new features like<br>Dose Structured Report and<br>Reject and Repeat Analysis<br>(                                                                                                                                 |                                                                                                                                                                                                                               |

## Conclusions

The minor differences in the modified EOS's technological characteristics do not raise any new questions of safety or effectiveness. Performance data demonstrates that the modified EOS is as safe and effective as the cleared predicate device. Thus, the modified EOS is substantially equivalent to its predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR/K152788](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR/K152788)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
