K152765 · B. Braun Interventional Systems, Inc. · DTK · Feb 26, 2016 · Cardiovascular
Device Facts
Record ID
K152765
Device Name
VENATECH CONVERTIBLE VENA CAVA FILTER SYSTEM
Applicant
B. Braun Interventional Systems, Inc.
Product Code
DTK · Cardiovascular
Decision Date
Feb 26, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.3375
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The VenaTech® Convertible™ Vena Cava Filter System is indicated for the prevention of recurrent pulmonary embolism via placement in the vena cava in the following situations: - Pulmonary thromboembolism when anticoagulants are contraindicated; - Failure of anticoagulant therapy in thromboembolic diseases; - Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced: and - Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated. When clinically indicated after implantation, the VenaTech Convertible filter can be converted to an open configuration to discontinue filtration according to the 'Conversion Procedure' in the Instructions for Use.
Device Story
VenaTech Convertible Vena Cava Filter System is a cobalt-chromium, self-expanding, permanent filter; implanted in infrarenal inferior vena cava to trap emboli; prevents recurrent pulmonary embolism. Delivered via femoral or jugular approach using pre-loaded cartridge/introducer system. When clinically indicated, filter can be percutaneously converted to open configuration to discontinue filtration. Used in clinical settings by physicians. Provides mechanical protection against pulmonary embolism; conversion feature allows deactivation of clot-trapping function without removal.
Clinical Evidence
Multi-center, prospective, single-arm study (n=149) evaluated safety and performance. Primary objective was technical success of conversion. 149 successful implants; 96 conversion attempts; 93 actual conversions; 89 successful conversions. Technical success achieved in 92.7% of subjects. Subjects followed for 6 months post-conversion or post-implant.
Technological Characteristics
Cobalt-chromium metal alloy; self-expanding; permanent vena cava filter. Includes introducer accessories. Evaluated per ISO 25539-3 (cardiovascular implants) and ISO 10993 (biocompatibility). Features mechanical conversion mechanism to open configuration.
Indications for Use
Indicated for patients requiring prevention of recurrent pulmonary embolism due to contraindication to anticoagulants, failure of anticoagulant therapy, emergency treatment of massive pulmonary embolism, or chronic recurrent pulmonary embolism.
Regulatory Classification
Identification
A cardiovascular intravascular filter is an implant that is placed in the inferior vena cava for the purpose of preventing pulmonary thromboemboli (blood clots generated in the lower limbs and broken loose into the blood stream) from flowing into the right side of the heart and the pulmonary circulation.
Special Controls
*Classification.* Class II. The special controls for this device are:(1) “Use of International Standards Organization's ISO 10993 ‘Biological Evaluation of Medical Devices Part I: Evaluation and Testing,’ ” and
(2) FDA's:
(i) “510(k) Sterility Review Guidance and Revision of 2/12/90 (K90-1)” and
(ii) “Guidance for Cardiovascular Intravascular Filter 510(k) Submissions.”
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 26, 2016
B.Braun Interventional Systems Inc. Peter Flosdorf Engineering Manager 824 Twelfth Avenue Bethlehem. Pennsylvania 18018
Re: K152765
Trade/Device Name: VenaTech Convertible Vena Cava Filter System Regulation Number: 21 CFR 870.3375 Regulation Name: Cardiovascular Intravascular Filter Regulatory Class: Class II Product Code: DTK Dated: December 30, 2015 Received: December 31, 2015
Dear Mr. Flosdorf:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152765
#### Device Name
VenaTech® Convertible™ Vena Cava Filter System
Indications for Use (Describe)
The VenaTech® Convertible™ Vena Cava Filter System is indicated for the prevention of recurrent pulmonary embolism via placement in the vena cava in the following situations:
- · Pulmonary thromboembolism when anticoagulants are contraindicated;
- · Failure of anticoagulant therapy in thromboembolic diseases;
- · Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced: and
- · Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated.
When clinically indicated after implantation, the VenaTech Convertible filter can be converted to an open configuration to discontinue filtration according to the 'Conversion Procedure' in the Instructions for Use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) SUMMARY
#### Date Prepared: February 25, 2016
| 510(k) Submitter | | Contact |
|------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| B. Braun Interventional Systems Inc.<br>824 Twelfth Avenue<br>Bethlehem, PA 18018<br>FDA Est. Reg. #: 3006332832 | | Contact:<br>Peter Flosdorf, Engineering Manager<br>Tel: 610-997-4694; Fax: 610-849-5319<br>Peter.Flosdorf@bbrauninterventional.com |
| General Information | | |
| Trade Name | VenaTech® Convertible™ Vena Cava Filter System | |
| Common Name | Vena cava filter system | |
| Classification<br>Information | Cardiovascular Intravascular Filter (21 CFR 870.3375; Class II)<br>ProCode: DTK; Panel: Cardiovascular | |
| Predicate | VenaTech LP Vena Cava Filter System (K010485, K063217) | |
| Reference Device | VenaTech LGM Vena Cava Filter System (K881604, K901454, K925679, K932921) | |
#### Device Description
The VenaTech Convertible Vena Cava Filter is a flexible, symmetrical, cobalt-chromium metal alloy, self-expanding vena cava filter intended for permanent implantation into the infrarenal inferior vena cava to provide protection against pulmonary embolism. When clinically indicated, the clot-trapping features of the VenaTech Convertible filter can be deactivated by percutaneously converting the filter to an open configuration. The VenaTech Convertible filter is pre-loaded in a cartridge (syringe) and provided as a system with introducer accessories and instructions to accommodate delivery and implantation either using the femoral or jugular approach.
# Intended Use / Indications
The VenaTech® Convertible™ Vena Cava Filter System is indicated for the prevention of recurrent pulmonary embolism via placement in the vena cava in the following situations:
- Pulmonary thromboembolism when anticoagulants are contraindicated:
- Failure of anticoagulant therapy in thromboembolic diseases; ●
- Emergency treatment following massive pulmonary embolism where . anticipated benefits of conventional therapy are reduced; and
- . Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated.
When clinically indicated after implantation, the VenaTech Convertible filter can be converted to an open configuration to discontinue filtration according to the 'Conversion Procedure' in the Instructions for Use.
{5}------------------------------------------------
B. Braun Interventional Systems Inc. VenaTech® Convertible™ Vena Cava Filter System
# Substantial Equivalence Comparison
The VenaTech Convertible Vena Cava Filter System has substantially equivalent indications and principle of operation to the currently marketed VenaTech LP predicate system. The VenaTech Convertible filter design, materials, introducer accessories, sterile package configuration, and sterilization process are substantially similar when compared to the currently marketed predicate filter. The VenaTech Convertible system has design features that permit the filter elements to convert to an open configuration when clinically indicated to no longer trap emboli. Like the predicates, the VenaTech Convertible filter is intended for permanent implantation and is not to be removed following implantation.
### Performance Data
# Non-clinical Evaluations (Bench)
The VenaTech Convertible system has been evaluated through ISO 14971-compliant risk analysis and design verification and validation (DV&V) testing in accordance with ISO 25539-3 Cardiovascular implants - Endovascular devices - Part 3: Vena cava filters and in consideration of the FDA guidance for cardiovascular intravascular filter 510(k) submissions. Results from the DV&V testing. which was completed both on the bench and in animals, shows the VenaTech Convertible filter is sufficiently robust for its intended use as a permanent vena cava filter and demonstrates the mechanical integrity and performance of the filter's convertible feature. The ISO 25539-3 compliant DVT included the below evaluations:
- Cavography ability ●
- Clot trapping test ●
- Conversion force
- Dilator spinlock resistance
- Dimension of introduction ● system
- Fatigue resistance (fracture, cracks, and corrosion)
- Filter dimension
- Filter head resistance ●
- Filter radio-opacity
- Filter welding resistance
- Head assembly resistance ●
- Hook crimping strength
- Introduction sheath radioopacity
- . Introduction system traction resistance
- Introduction sheath water tightness
- Migration resistance ●
- (Packaging) Transportation ●
- (Packaging) Fall (drop)
- Passage in sheath ●
- Radial force
- Radial pressure (contact ● pressure)
- Magnetic resonance (MR) testing
Additionally, the VenaTech Convertible filter was subjected to chemical characterization evaluations performed in accordance with ISO 10993 Biological Evaluation of Medical Devices - Part 18: Chemical characterization of materials.
{6}------------------------------------------------
## Animal Studies
Simulated use testing was performed in a sheep model. Results of this animal testing demonstrated that the VenaTech Convertible filter performed according to its intended use, and thus supports substantial equivalence.
## Clinical Evaluation
A multi-center, prospective, single-arm, historical (literature) controlled clinical study was performed to evaluate the VenaTech Convertible filter safety and performance in a minimum of 75 subjects in whom the filter has been implanted, converted and followed for 6-months. Converted subjects were followed at 30-days, 3-months, and 6-months post conversion. Subjects that were unable to undergo conversion (due to clinical condition based on investigator discretion) were followed at 6-months post filter implant (these subjects are referred to as 'permanent filtration subjects'). In this study, eleven (11) sites enrolled a total of 149 subjects. The VenaTech Convertible filter was successfully implanted in all 149 subjects. There were 96 attempted conversions of the filter, 93 actual conversions, and 89 successful conversions with no technical complications. The primary objective, technical success, was achieved in 92.7% of subjects, therefore the primary objective was met. Refer to the IFU for clinical study information.
### Conclusions
The VenaTech Convertible Vena Cava Filter System is similar in indications, principle of operation, and design to the currently marketed VenaTech LP Vena Cava Filter System predicate. The few design differences, which allow the filter to be converted to no longer trap emboli, were subjected to non-clinical and clinical performance evaluations in accordance with FDA guidance and industry standards. Based on the predicate comparison and performance evaluation information provided in this 510(k), it can be concluded that the VenaTech Convertible Vena Cava Filter System raises no new questions of safety or effectiveness compared to the predicate devices and is, therefore, substantially equivalent.
- End of 510(k) Summary -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.