The ISEN-Goggles are used to record and visually display movements of the eyes including nystagmus.
Device Story
ISEN-Goggles are opaque, head-mounted goggles used to record and display eye movements; includes two internal video cameras and infrared (IR) LEDs (940nm) for illumination. Device features green LEDs (573nm) to provide optional focal points for patients. Signals from cameras are processed by an internal image processor, allowing near-simultaneous viewing of both eyes on an integrated LCD screen. Data is stored on an internal SD-type memory card and can be transferred via a micro-USB port. Operated by clinicians in a clinical setting to observe nystagmus; output allows providers to visually assess eye movement patterns to aid in diagnostic decision-making. System is battery-powered and portable, representing a miniaturized, integrated version of traditional video eye movement recorders.
Clinical Evidence
No clinical data. Bench testing included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), lamp safety (IEC 62471), FCC compliance (47 CFR Part 15), battery safety (IEC 62133), biocompatibility (ISO 10993), and hardware/software verification and validation.
Technological Characteristics
Opaque goggles with adjustable headband, two IR-sensitive video cameras, 940nm IR LEDs, 573nm green LEDs, integrated LCD display, SD-type memory, and micro-USB port. Powered by internal rechargeable Li-ion battery. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62471, IEC 62133, ISO 10993, FCC Part 15. Software is embedded for image processing and display.
Indications for Use
Indicated for recording and visual display of eye movements, including nystagmus, in patients requiring diagnostic evaluation of ocular motility.
Regulatory Classification
Identification
A nystagmograph is a device used to measure, record, or visually display the involuntary movements (nystagmus) of the eyeball.
Predicate Devices
ICS Medical VEMR Video Eye Movement Recorder (K972631)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing right, stacked on top of each other, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 26, 2016
Isen Tech & Trading Company, Ltd. % Linda Bovard, RAC President Bovard Consulting LLC 378 Hardy Avenue Eugene, Oregon 97404
Re: K152727
Trade/Device Name: Isen-goggles Regulation Number: 21 CFR 882.1460 Regulation Name: Nystagmograph Regulatory Class: Class II Product Code: GWN Dated: March 22, 2016 Received: March 28, 2016
Dear Ms. Bovard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
William J. Heetderks -A
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K152727
Device Name ISEN-Goggles
Indications for Use (Describe)
The ISEN-Goggles are used to record and visually display movements of the eyes including nystagmus.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
# Contact Details
| Applicant Name: | ISEN Tech & Trading Co., Ltd. |
|-----------------|-----------------------------------------------------------|
| | Rm509, Bldg A, Cyber Tower, No. 2 Zhongguancun South Ave. |
| | Beijing, 100086, China |
| | +86-10-5162-7073 |
| Contact: | Martin Liu, CEO |
| | martin@isen.com.cn |
| Date Prepared: | March 20, 2016 |
## Device Name
Trade Name: ISEN-Goggles
Common Name: IR/Video Eye Movement Viewer/Recorder
Classification Name: 882.1460 Nystagmograph; GWN Nystagmograph
## Legally Marketed Predicate Device(s)
| 510(k) Number | Product Code | Trade Name | Applicant |
|---------------|--------------|-------------------------------------------------|-------------------|
| K972631 | GWN | ICS Medical VEMR Video<br>Eye Movement Recorder | ICS Medical Corp. |
## Device Description
This device consists of patient eye goggles with an adjustable, elastic headband, an LCD display, two green LED lights to provide optional focal points, eight infrared LED lights to provide illumination for the internal cameras, an internal SD-type memory, and buttons to operate the device's function. There is also a micro-USB port to facilitate transfer of data captured and stored in the device's SD-type memory.
The patient eye goggles are an opaque, facially worn device fitted with two small video cameras, and two wavelengths of light emitting diodes (LEDs). One set of lights provides illumination for the cameras. The other allows the user to provide a focus point for the patient, if this is desired. These signals are processed by the image processor that allows for nearsimultaneous viewing of the video images from both eyes on the attached LCD screen.
{4}------------------------------------------------
# Intended Use/Indications for use
The ISEN-Goggles are used to record and visually display movements of the eyes including nystagmus.
# Substantial Equivalence Comparison
The predicate device for this submission is the ICS Medical VEMR Video Eye Movement Recorder (ICS VEMR) from ICS Medical that was cleared under K972631.
#### Table 5-1 Summary of Similarities and Differences
| Category | ISEN-Goggles | ICS VEMR<br>K972631 |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Goggles fitted with IR<br>sensitive video cameras | Yes | Yes |
| IR Source | IR LEDs<br>Peak Wavelengths ( $λ$ ) = 940nm (red) | IR LEDs<br>Peak Wavelength ( $λ$ ) = 940nm (red) |
| Lights for visual focus point | Yes – 573nm (green) | No |
| Light blocking opaque<br>goggles | Yes | Yes |
| Energy Source | Internal Rechargeable Battery | External Power Supply |
| Components | Goggles<br>Mask 1 and 2<br>Battery Charger<br>Rechargeable Li-ion Battery (not<br>supplied but required) | Goggles<br>VCR and Remote<br>Separate 17" Display Monitor |
| Weight | 17.6 oz. (mask plus goggles) | 12-15oz – Goggles<br>9lb 4oz – VCR and Remote<br>37lb 5oz – 17" Display Monitor<br>TOTAL: 47lb 5oz – 47lb 8 oz |
| Interface to control device | Buttons located on goggle-mounted<br>LCD. | Remote Control |
| Images Splitter/Power<br>supply interface | Image Splitter not necessary with this<br>technology | Yes |
| Video Recording Capability | Yes (built-in SD-type memory<br>drive). A micro-USB port is<br>provided to enable data transfer. | Yes (built-in VCR) |
| Video-Goggle interface | An LCD display is attached to the<br>goggles, allowing for simultaneous<br>display of both eyes from two video<br>cameras within the mask. | An interface that allows for display<br>of both eyes on video monitor and<br>provides the power for the video<br>cameras and the LEDs |
Both devices are essentially goggles with internal cameras and LED lights. There are some differences. For example, the ICS VEMR product has only one type of LED that displays red light (940 nm) while the ISEN-Goggles has red (940nm) LEDs for illuminating the image
{5}------------------------------------------------
being recorded and green (573nm) LEDs to provide the option of a focal point for the patient during the test.
Many of the other differences relate to the miniaturization of components since the marketing of the ICS VEMR. The ISEN-Goggles combines many of the separate features of the ICS VEMR into one device instead of three separate devices that work together. This contributes to the reduction in overall weight and size of the system. The ISEN-Goggles weigh only 17.6 oz while the ICS VEMR system weighed 471b 5oz to 471b 8oz for the ICS VEMR system. The energy sources are also different.
However, these differences do not affect the substantial equivalence; they are still using the same technological basis to accomplish the same purpose.
# Non-clinical Testing
Non-clinical testing for the ISEN-Goggles included electrical safety testing of the Goggles and battery charger to IEC 60601-1, electromagnetic compatibility testing of the Goggles and battery charger to IEC 60601-1-2, lamp safety to IEC 62471, compatibility with FCC regulations per 47 CFR Part 15, battery testing to IEC 62133, and biocompatibility testing to ISO 10993. In addition, hardware and software verification and validation was completed.
# Clinical Testing
No clinical testing is included in this 510(k).
#### Summarv
The ISEN-Goggles are substantially equivalent to the ICS VEMR in intended use, device features, and use parameters.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.