K152678 · Tyrx, Inc. · FTL · Nov 17, 2015 · General, Plastic Surgery
Device Facts
Record ID
K152678
Device Name
TYRX Neuro Antibacterial Envelope
Applicant
Tyrx, Inc.
Product Code
FTL · General, Plastic Surgery
Decision Date
Nov 17, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TYRX Neuro Non-Absorbable Antibacterial Envelope is intended to hold a vagus nerve stimulator, a spinal cord neuromodulator, a deep brain stimulator or a sacral nerve stimulator securely in order to create a stable environment when implanted in the body. TYRX Neuro Non-Absorbable Antibacterial Envelope contains the antimicrobial agents rifampin and minocycline, which have been shown to reduce infection in an in vivo model of bacterial challenge following surgical implantation of a pulse generator. This device is intended to be used in conjunction with vagus nerve stimulators or deep brain stimulators implanted in the infraclavicular fossa, or in conjunction with spinal cord neuromodulators or sacral nerve stimulators implanted laterally to the body midline and slightly superior to the gluteal region. TYRX Neuro Non-Absorbable Antibacterial Envelope is intended for single-patient, one-time use only.
Device Story
Device is a sterile, dual-component (absorbable/non-absorbable) surgical mesh envelope; constructed of knitted polypropylene filaments coated with an absorbable polyarylate polymer containing rifampin and minocycline. Used in healthcare facilities by surgeons to house implantable pulse generators (VNS, DBS, SCN, SNS). Envelope provides mechanical stability for the stimulator and releases antimicrobial agents to reduce surgical site infection risk. No electronic components; no data input/output. Benefits include reduced infection risk and improved device stabilization post-implantation.
Clinical Evidence
No human clinical trials. Evidence includes biocompatibility testing per ISO 10993, in vitro antimicrobial activity against various Staphylococci and other bacteria, and an in vivo sheep study. The sheep study evaluated minocycline/rifampin diffusion into plasma and cerebrospinal fluid (CSF) following implantation with a neuromodulator; results showed no quantifiable concentrations of antibiotics in plasma or CSF up to 7 days post-implant, indicating no significant risk for CNS side effects.
Technological Characteristics
Knitted polypropylene mesh substrate; coated with absorbable polyarylate polymer containing rifampin and minocycline (102ug/cm2 each). Available in two sizes (2.5" x 2.7" and 2.9" x 3.3"). Sterile, single-use. No energy source, connectivity, or software.
Indications for Use
Indicated for patients requiring implantation of a vagus nerve stimulator, deep brain stimulator, spinal cord neuromodulator, or sacral nerve stimulator. Used to hold the device securely and create a stable environment. Contraindications not specified.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 17, 2015
Tyrx, Inc. % Regina Novak Senior Regulatory Affairs Specialist Medtronic TYRX, Inc. 1 Deer Park Drive Suite G Monmouth Junction, New Jersey 08852
Re: K152678
Trade/Device Name: Tyrx Neuro Antibacterial Envelope Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTL Dated: September 17, 2015 Received: September 18, 2015
Dear Ms. Novak:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# William J. Heetderks -S
for
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| Form Approved: OMB No. | 0910-0120 |
|------------------------|--------------------------|
| Expiration Date: | January 31, 2017 |
| | See PRA Statement below. |
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
**Indications for Use**
| 510(k) Number (if known) | K152678 |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | TYRX Neuro Non-Absorbable Antibacterial Envelope |
| Indications for Use (Describe) | TYRX Neuro Non-Absorbable Antibacterial Envelope is intended to hold a vagus nerve stimulator, a spinal cord neuromodulator, a deep brain stimulator or a sacral nerve stimulator securely in order to create a stable environment when implanted in the body.<br>TYRX Neuro Non-Absorbable Antibacterial Envelope contains the antimicrobial agents rifampin and minocycline, which have been shown to reduce infection in an in vivo model of bacterial challenge following surgical implantation of a pulse generator. This device is intended to be used in conjunction with vagus nerve stimulators or deep brain stimulators implanted in the infraclavicular fossa, or in conjunction with spinal cord neuromodulators or sacral nerve stimulators implanted laterally to the body midline and slightly superior to the gluteal region.<br>TYRX Neuro Non-Absorbable Antibacterial Envelope is intended for single-patient, one-time use only. |
Type of Use (Select one or both, as applicable)
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------|-----------------------------------------------------------|
|------------------------------------------------------------|-----------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
일
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Medtronic logo. The logo features a stylized human figure in motion, depicted in a circular frame. Below the figure is the word "Medtronic" in a bold, blue font.
# 510(k) Summary
Medtronic, Inc. Medtronic TYRX 1 Deer Park Drive, Suite G Monmouth Junction, NJ 08852 www.tvrx.com
el 732.246.8676 ax 732.246.8677
In accordance with 21 CFR 807.92, TYRX, Inc. provides this summary of the safety and effectiveness information available for TYRX™ NeuroNon-Absorbable Antibacterial Envelope, as well as the substantial equivalence decision making process used for the TYRX™ Neuro Antibacterial Envelope subject device.
Sponsor/Applicant Name and Address:
TYRX Inc. DBA Medtronic TYRX, Inc. 1 Deer Park Drive Suite G Monmouth Junction, N.J. 08852
Establishment Registration Number:
Sponsor Contact Information:
Date of preparation of 510(k) Summary
New Device Trade/Proprietary Name:
Device Common Name:
3005619263
Regina Novak Senior Regulatory Affairs Specialist Phone: 732-964-1492 732-246-8677 Fax: Email:regina.novak(@medtronic.com
Alternate Contact: Susan Olinger, JD Director, Regulatory Affairs Phone: 732-964-1110 763-246-8677 Fax: Email:susan.olinger@medtronic.com
September 17, 2015
TYRX™ Neuro Non-Absorbable Antibacterial Envelope
Surgical Mesh
{4}------------------------------------------------
Regulatory Classification: PROCODE:
Class II FTL
Predicate Device Name and 510(k) Number
Primary Predicate AIGISRx N Antibacterial Envelope K132699
TYRX™ Neuro Absorbable Antibacterial Envelope K150291
#### Device Description:
TYRX TM Neuro Non-Absorbable Antibacterial Envelope is a dual component (absorbable and non-absorbable) sterile device designed to hold a vagus nerve stimulator (VNS), a deep brain stimulator (DBS), a spinal cord neuromodulator (SCN) or a sacral nerve stimulator (SNS) securely to create a stable environment when implanted in the body. The device is available in 2 sizes, Medium (2.5" x 2.7" ) and Large (2.9" x 3.3"). The device is constructed of knitted filaments of polypropylene (mesh substrate) that are coated with an absorbable polyarylate polymer mixture containing the antimicrobial agents rifampin and minocvcline. Rifampin and minocycline have been shown to reduce infection in an in vivo model of bacterial challenge following surgical implantation of an implantable pulse generator. This device is to be used in a healthcare facility/hospital by personnel experienced in the implantation of VNS, DBS, SCN, or SNS.
#### Indications for Use
TYRX TM Neuro Non-Absorbable Antibacterial Envelope is intended to hold a vagus nerve stimulator, deep brain stimulator, spinal cord neuromodulator, or sacral nerve stimulator securely in order to create a stable environment when implanted in the body.
TYRX TM Neuro Non-Absorbable Antibacterial Envelope contains the antimicrobial agents rifampin and minocycline which have been shown to reduce infection in an in vivo model of bacterial challenge following surgical implantation of a pulse generator. This device is intended to be used in conjunction with vagus nerve stimulators or deep brain stimulators implanted in the
{5}------------------------------------------------
infraclavicular fossa and spinal cord neuromodulators or sacral nerve stimulators implanted laterally to the body midline and slightly superior to the gluteal region.
TYRX 110 Neuro Non-Absorbable Antibacterial Envelope is intended for single-patient, one-time use only.
## Technological Characteristics
The technological characteristics of the TYRX Neuro Non-Absorbable Antibacterial Envelope are identical to the predicate (K132699). Both devices have intended use which is to create a stable environment for the implanted neurostimulator.
The only difference between the subject and predicate device (K132699) is that the subject device will have an expanded Indications for Use to include deep brain stimulators and sacral nerve stimulators, the same indications for use as K1250291.
The TYRX 114 Neuro Non-Absorbable Antibacterial Envelope is constructed of knitted filaments of polypropylene and coated with an absorbable polyarylate polymer coating containing the antibiotics rifampin and minocycline each in concentrations of 102ug/cm2.
The physical, chemical and mechanical properties of the TYRX Neuro Non-Absorbable Antibacterial Envelope, subject device, are the same as the predicate device (K132699). There is no technological difference between the subject and predicate device, and there are no design changes to the predicate device. The difference is only an expansion of the Indications for Use to include use with Deep Brain Stimulators and/or Sacral Nerve Stimulators, the same as the Indications for Use for K150291.
### Biocompatability Results
Biocompatibility testing in accordance to the current ISO 10993 series was conducted and the results indicate that the device is biocompatible, per the standards.
An in vivo functionality study showed that TYRX devices do not alter or interfere with an implantable pulse generator.
{6}------------------------------------------------
## Animal Studies
In vitro studies referenced in the predicate devices K132699 and K150291 demonstrated antimicrobial activity against methicillin resistant Staphylococcus aureus (MRSA), Staphylococcus aureus, Staphylococcus epidermidis, Staphylococcus lugdunensis, Escherichia coli, Acinetobacter baumannii, Enterobacter aerogenes, and Proteus Mirabilis,
In vivo efficacy testing referenced in the predicate devices K132699 and K150291, demonstrated effectiveness in reducing infections. The bacteria tested were methicillin-resistant Staphylococcus aureus (MRSA), Staphylococcus aureus, Staphylococcus epidermidis, Acinetobacter baumannii, Escherichia coli and Staphylococcus lugdunensis. It should be noted that the in vivo and in vitro activity of the TYRX TM Neuro Non-Absorbable Antibacterial Envelope antimicrobials is variable against non-epidermidis and non-lugdunensis strains of coagulase-negative Staphylococci.
To provide additional evidence on the safety of TYRX™ Neuro Antibacterial Envelopes when implanted with neuromodulators, a study in sheep was conducted to address the question of minocycline diffusion with the TYRX™ Neuro Antibacterial Envelope (absorbable and non absorbable) and possible central nervous system (CNS) effects. The purpose of the study was to determine the concentration of minocycline and rifampin in the plasma and cerebrospinal fluid of sheep implanted with a neuromodulator and a TYRX Neuro Antibacterial Envelope (absorbable (K150291) or non absorbable (K132699), with or without a lead.
The study demonstrated that the TYRX Neuro Antibacterial Envelope was safe in the sheep model as assessed by the absence of adverse clinical signs. There were no quantifiable concentrations of minocycline or rifampin in plasma and cerebrospinal fluid samples collected up to 7 days after implantation of the neuromodulator enclosed in the TYRX Neuro Antibacterial Envelope, with or without leads. This study demonstrated that the amount of minocycline in CSF and plasma, as a result of the TYRX Neuro Antibacterial Envelope implantation is not detectable and does not pose a significant risk for CNS side effects.
# CONCLUSION:
In summary, the TYRX Neuro Non-Absorbable Antibacterial Envelope is identical in terms of physical construct to the cleared primary predicate, AIGIS Neuro Antibacterial Envelope,(K132699). The polymer coating concentration of the antibiotics rifampin and
{7}------------------------------------------------
minocycline remains unchanged. The device is supplied sterile. The only difference is that the subject device has an expanded Indications for Use to include use with Deep Brain Stimulators and Sacral Nerve Stimulators, which is identical to predicate K150291.
Based on the 510(k) summaries and the information provided, we conclude that TYRX TM Neuro Non-Absorbable Antibacterial Envelope is safe and effective for its intended use, and is substantially equivalent to the primary predicate device (K132699).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.