MEDICALGORITHMICS UNIFIED ARRHYTHMIA DIAGNOSIC SYSTEM
Applicant
Medicalgorithmics S.A.
Product Code
DSI · Cardiovascular
Decision Date
Oct 8, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
The indications for use for the PocketECG III monitor are as follows: 1. Patients who have a demonstrated need for cardiac monitoring. These may include but are not limited to patients who require monitoring for: a) non-life threatening arrhythmias such as supraventricular tachycardias (e.g. atrial fibrillation, atrial flutter, PACs. PSVT) and ventricular ectopy: b) evaluation of bradyarrhythmias and intermittent bundle branch block, including after cardiovascular surgery and myocardial infarction; and c) arrhythmias associated with co-morbid conditions such as hyperthyroidism or chronic lung disease; 2. Patients with symptoms that may be due to cardiac arrhythmias. These may include but are not limited to symptoms such as: a) dizziness or lightheadedness; b) syncope of unknown etiology in which arrhythmias are suspected or need to be excluded; and c) dyspnea(shortness of breath); 3. Patients with palpitations with or without known arrhythmias to obtain correlation of rhythm with symptoms; 4. Patients who require monitoring of effect of drugs to control ventricular rate in various atrial arrhythmias (e.g. atrial fibrillation); 5. Patients recovering from cardiac surgery who are indicated for outpatient arrhythmia monitoring; 6. Patients with diagnosed sleep disordered breathing including sleep apnea (obstructive, central) to evaluate possible nocturnal arrhythmias; 7. Patients requiring arrhythmia evaluation of etiology of stroke or transient cerebral ischemia, possibly secondary to atrial fibrillation or atrial flutter; 8. Data from the device may be used by another device to analyze, measure or report QT interval. The device is not intended to sound any alarms for QT interval changes.
Device Story
Ambulatory ECG monitor; captures electrographic signals; classifies heart beats; recognizes rhythm abnormalities. Device transmits ECG data and beat annotations via cellular network (CDMA) to remote server. Monitoring center staff review data. Used in outpatient settings; patient-worn. Output assists clinicians in diagnosing arrhythmias, correlating symptoms with rhythm, and monitoring drug effects. Benefits include remote cardiac monitoring without hospitalization.
Clinical Evidence
Bench testing only. Device meets requirements of IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-47, and AAMI/ANSI EC57:2012 for arrhythmia detector and alarm performance.
Technological Characteristics
Ambulatory ECG system; includes transmitter, lithium-ion rechargeable batteries, AC charger, and PC application. Connectivity via CDMA cellular network. Complies with IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-47, and AAMI/ANSI EC57 standards. Quality system conforms to ISO 13485 and ISO 9001.
Indications for Use
Indicated for patients requiring cardiac monitoring for non-life-threatening arrhythmias (e.g., AFib, flutter, PACs, PSVT, ventricular ectopy), bradyarrhythmias, bundle branch block, or symptoms (dizziness, syncope, dyspnea, palpitations). Also for post-cardiac surgery, sleep-disordered breathing, or stroke/TIA etiology evaluation. Contraindicated for patients requiring inpatient monitoring or those deemed by a physician to require hospitalization.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of human profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 8, 2015
Medicalgorithmics S.A. % Martin Jasinski Vice President Medicalgorithmics LLC 245 West 107th Street Suite 11 A New York, New York 10025
Re: K152550
Trade/Device Name: PocketECG III - Medicalgorithmics Unified Arrhythmia Diagnostic System, type PECGT-IIIV Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: DSI Dated: August 5, 2015 Received: September 8, 2015
Dear Martin Jasinski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known): K152550 Device Name: PocketECG III -Medicalgorithmics Unified Arrhythmia Diagnostic System, type PECGT-IIIV
#### Indications For Use:
The indications for use for the PocketECG III monitor are as follows:
- 1. Patients who have a demonstrated need for cardiac monitoring. These may include but are not limited to patients who require monitoring for: a) non-life threatening arrhythmias such as supraventricular tachycardias (e.g. atrial fibrillation, atrial flutter, PACs. PSVT) and ventricular ectopy: b) evaluation of bradyarrhythmias and intermittent bundle branch block, including after cardiovascular surgery and myocardial infarction; and c) arrhythmias associated with co-morbid conditions such as hyperthyroidism or chronic lung disease;
- 2. Patients with symptoms that may be due to cardiac arrhythmias. These may include but are not limited to symptoms such as: a) dizziness or lightheadedness; b) syncope of unknown etiology in which arrhythmias are suspected or need to be excluded; and c) dyspnea(shortness of breath);
- 3. Patients with palpitations with or without known arrhythmias to obtain correlation of rhythm with symptoms;
- 4. Patients who require monitoring of effect of drugs to control ventricular rate in various atrial arrhythmias (e.g. atrial fibrillation);
- 5. Patients recovering from cardiac surgery who are indicated for outpatient arrhythmia monitoring;
- 6. Patients with diagnosed sleep disordered breathing including sleep apnea (obstructive, central) to evaluate possible nocturnal arrhythmias;
- 7. Patients requiring arrhythmia evaluation of etiology of stroke or transient cerebral ischemia, possibly secondary to atrial fibrillation or atrial flutter;
- 8. Data from the device may be used by another device to analyze, measure or report QT interval. The device is not intended to sound any alarms for QT interval changes.
#### Contradictions:
- 1. Patients with potentially life-threatening arrhythmias who require inpatient monitoring;
- 2. Patients who the attending physician thinks should be hospitalized.
Prescription Use AND/OR AND/OR AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (CONTINUE ON A SEPARATE PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
## 510(k) Summary
## August 4, 2015
| 1. Submitter Name and Address | |
|-------------------------------|------------------------------------------------------------------------|
| Medicalgorithmics LLC | 245 West 107th St., Suite 11A<br>New York, NY 10025, USA |
| Contact Person | Martin Jasinski, phone (917) 9419581,<br>fax (817) 5829527 |
| 2. Device | |
| Trade name: | PocketECG III – Medicalgorithmics Unified Arrhythmia Diagnostic System |
| Type | PECGT-IIIV |
| Classification name: | Arrhythmia Detector and Alarm |
| Product code: | DSI |
| Regulation no: | 870.1025 |
| Class: | Class II, Special Controls |
#### 3. Substantial Equivalence
The PocketECG III - Medicalgorithmics Unified Arrhythmia Diagnostic System, type PECGT-IIIV is a modification of a legally marked Medicalgorithmics predicate device:
- . PocketECG v3 - Medicalgorithmics Unified Arrhythmia Diagnostic System, type PECGT-III; K124060 (Reg. no. 870.1025)
Note: For commercial reasons Medicalgorithmics S.A. launched the medical device onto the U.S. market under the name of PocketECG III - Medicalgorithmics Unified Arrhythmia Diagnostic System, and has continued to distribute the device under the same name. The modified device is the latest evolution of this predicate one.
#### 4. Device Description
PocketECG III - Medicalgorithmics Unified Arrhythmia Diagnostic System, type PECGT-IIIV is an ambulatory ECG monitor which analyzes electrographic signal, classifies all detected heart beats and recognizes rhythm abnormalities. All detection results, including annotations for every detected heart beat and the entire ECG signal are transmitted via cellular telephony network to a remote server accessible by a Monitoring Center for reviewing by trained medical staff.
#### 5. Indications for Use and contradictions
The indications for use for the PocketECG III, type PECGT-IIIV monitor are as follows:
{4}------------------------------------------------
1. Patients who have a demonstrated need for cardiac monitoring. These may include but are not limited to patients who require monitoring for: a) non-life threatening arrhythmias such as supraventricular tachycardias (e.g. atrial fibrillation, atrial flutter, PACs, PSVT) and ventricular ectopy; b) evaluation of bradyarrhythmias and intermittent bundle branch block, including after cardiovascular surgery and mvocardial infarction: and c) arrhythmias associated with co-morbid conditions such as hyperthyroidism or chronic lung disease:
- 2. Patients with symptoms that may be due to cardiac arrhythmias. These may include but are not limited to symptoms such as: a) dizziness or lightheadedness; b) syncope of unknown etiology in which arrhythmias are suspected or need to be excluded; and c) dyspnea (shortness of breath);
- 3. Patients with palpitations with or without known arrhythmias to obtain correlation of rhythm with symptoms;
- 4. Patients who require monitoring of effect of drugs to control ventricular rate in various atrial arrhythmias (e.g. atrial fibrillation);
- 5. Patients recovering from cardiac surgery who are indicated for outpatient arrhythmia monitoring:
- 6. Patients with diagnosed sleep disordered breathing including sleep apnea (obstructive, central) to evaluate possible nocturnal arrhythmias;
- 7. Patients requiring arrhythmia evaluation of etiology of stroke or transient cerebral ischemia, possibly secondary to atrial fibrillation or atrial flutter;
- 8. Data from the device may be used by another device to analyze, measure or report QT interval. The device is not intended to sound any alarms for OT interval changes.
## Contradictions:
- 1. Patients with potentially life-threatening arrhythmias who require inpatient monitoring.
- 2. Patients who the attending physician thinks should be hospitalized.
## 6. Predicate Device Information
#### To K124060 - PocketEG v3 - Medicalgorithmics Unified Arrhythmia Diagnostic System, type PECGT-III
- Similarities:
- o The subject device and the predicate device have the same fundamental scientific technology and intended use:
- o The subject device and the predicate device have the same components: transmitter, lithium-ion rechargeable batteries, AC plug-in battery charger and PC application:
- o The subject device and the predicate device analyze electrographic signal, classify all detected heart beats and recognize rhythm abnormalities:
- o The subject device and the predicate device send all detection results, including annotations for every detected heart beat and the entire ECG signal via cellular telephony network to a remote server;
- Differences: -
{5}------------------------------------------------
- The predicate device uses a Global System for Mobile Communication (GSM) o protocol, whereas the subject device uses Code Division Multiple Access protocol (CDMA) for data transmission to a remote server.
## 7. Referenced standards
The PocketECG III - Medicalgorithmics Unified Arrhythmia Diagnostic System, type PECGT-IIIV meets the requirements of following performance standards in accordance with FDA Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm like the predicate device.
- IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) "Medical electrical . equipment - Part 1: General requirements for basic safety and essential performance", with National Differences for United States (US) ANSI/AAMI ES 60601-1:2005/A2:2010 "Medical electrical equipment - Part 1: General requirements for basic safety and essential performance",
- IEC 60601-1-2:2007 Medical electrical equipment Part 1-2: General . requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests,
- . IEC 60601-1-11:2010 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment,
- IEC 60601-2-47:2012 Medical electrical equipment Part 2-47: Particular . requirements for the basic safety and essential performance of ambulatory electrocardiographic systems.
- . AAMI / ANSI EC57:2012 Testing and Reporting Performance Results of Cardiac Rhythm and ST-Segment Measurement Algorithms.
Quality management system - Medical devices is in conformance with the standards: EN ISO 13485:2012 + EN ISO 13485:2012/AC:2012, ISO 13485:2003 under CMDCAS and EN ISO 9001:2008.
### 8. Substantial Equivalence Conclusion
PocketECG III - Medicalgorithmics Unified Arrhythmia Diagnostic System, type PECGT-IIIV is safe, effective and substantially equivalent to the predicate device as supported by the descriptive information and the performance testing. The subject device is composed of off-the-shelf, certified devices and components fully complying with the US safety and EMC standards.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.