CAMLOG and CONELOG Abutments for Screw-retained Restorations (ASR)
Applicant
Altatec GmbH
Product Code
NHA · Dental
Decision Date
Jan 12, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
CAMLOG® and CONELOG® Abutments for Screw-retained Restorations (ASR) are intended to support occlusal screw-retained prosthetic restorations in the upper jaw and lower jaw in conjunction with CAMLOG® and CONELOG® Implants.
Device Story
ASR abutments are dental components used to support screw-retained prosthetic restorations (crowns, bridges, bars) on CAMLOG or CONELOG implants. Available in straight and angled (17°/30°) configurations with various platform diameters and gingival heights; includes Type A/B anti-rotation cam options. Components are fabricated from titanium alloy (ASTM F136). Used in dental clinics by clinicians to restore chewing function. The device acts as a mechanical interface between the endosseous implant and the final prosthesis. Benefits include restoration of oral function and aesthetics for patients with missing teeth.
Clinical Evidence
No clinical data submitted. Substantial equivalence is supported by bench testing, including static and dynamic compression-bending fatigue testing per ISO 14801, and validation of sterilization processes per ISO 11137 and ISO 17665. Biocompatibility was evaluated per ISO 10993-1.
Technological Characteristics
Materials: Titanium-6 Aluminum-4 Vanadium ELI alloy (ASTM F136). Design: Straight and angled (17°, 30°) abutments; multiple platform diameters (3.3-6.0 mm) and gingival heights. Interface: CAMLOG or CONELOG implant/abutment connections. Sterilization: Radiation and moist heat. No software or electronic components.
Indications for Use
Indicated for patients requiring support for occlusal screw-retained prosthetic restorations in the upper or lower jaw in conjunction with CAMLOG® or CONELOG® dental implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an abstract design of an eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 12, 2016
ALTATEC GmbH c/o Ms. Linda K. Schulz Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K152509
Trade/Device Name: CAMLOG® and CONELOG® Abutments For Screw-retained Restorations (ASR) Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: December 10, 2015 Received: December 11, 2015
Dear Ms. Schulz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### Device Name
CAMLOG® and CONELOG® Abutments for Screw-retained Restorations (ASR)
#### Indications for Use (Describe)
CAMLOG® and CONELOG® Abutments for Screw-retained Restorations (ASR) are intended to support occlusal screw- retained prosthetic restorations in the upper jaw and lower jaw in conjunction with CAMLOG® and CONELOG® Implants.
Type of Use (Select one or both, as applicable)
|X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
# Altatec GmbH CAMLOG® and CONELOG® Abutments for Screw-retained Restorations (ASR)
January 11, 2016
### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Altatec GmbH<br>Maybachstrasse 5<br>D-71299 Wimsheim, Germany | | | |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|--|--|
| | Telephone: | +49 7044 9445 0 | | |
| | Fax: | +49 7044 9445 723 | | |
| Official Contact | Colleen Boswell<br>Director of RA/QA Ortho Organizers, Inc. | | | |
| Representative/Consultant | Linda K. Schulz, BSDH, RDH<br>Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130 | | | |
| | Telephone: | +1 (858) 792-1235 | | |
| | Fax: | +1 (858) 792-1236 | | |
| | Email: | lschulz@paxmed.com<br>flarson@paxmed.com | | |
### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | CAMLOG® and CONELOG® Abutments for Screw-retained<br>Restorations (ASR) |
|----------------------------|-------------------------------------------------------------------------|
| Common Name | Endosseous dental implant abutment |
| Classification Name | Endosseous dental implant abutment |
| Classification Regulations | 21 CFR 872.3630, Class II |
| Product Code | NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
### INTENDED USE
CAMLOG® and CONELOG® Abutments for Screw-retained Restorations (ASR) are intended to support occlusal screw-retained prosthetic restorations in the upper jaw and lower jaw in conjunction with CAMLOG® and CONELOG® Implants.
{4}------------------------------------------------
#### PREDICATE INFORMATION
The primary predicate device is K133421.
Reference Predicates
K141871 - Straumann® Screw Retained Abutments K073553 - CAMLOG Implant System Abutments K083496 - CAMLOG Implant System Modified Implants and Abutments K113779 - CONELOG® Implant System
### DEVICE DESCRIPTION
Abutment for Screw-retained Restorations (ASR) is available with the Camlog or the Conelog implant/abutment interface. ASR Straight is provided in five platform diameters (3.3. 3.8. 4.3. 5.0 and 6.0 mm) and three gingival heights (0.5, 2.0 and 4.0 mm), for a total of 15 abutment options. ASR Angled is provided in four platform diameters (3.3, 3.8, 4.3, and 5.0 mm), four gingival heights (2.5, 3.5, 4.0 and 5.0 mm), and two angles (17° and 30°). Each of the angled designs is provided in a Type A option (anti-rotation cam alignment away from the abutment cone angle) or Type B option (anti-rotation cam alignment in the direction of the abutment cone angle), for a total of 32 abutment options. A healing cap, and a titanium cap for temporary or permanent restoration fabrication are available for all sizes. All ASR components are made from titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401).
### PERFORMANCE DATA
Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence included: sterilization validation according to : ISO 11137-1 Sterilization of health care products -- Radiation -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices and ISO 11137-2 Sterilization of health care products -- Radiation -- Part 2: Establishing the sterilization dose, ISO 17665-1 Sterilization of health care products – Moist heat – Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices; ISO 17665-2 Sterilization of health care products – Moist heat – Part 2: Guidance on the application of ISO 17665-1; ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process, and static and dynamic compression-bending testing according to ISO 14801 Dentistry - Implants - Dynamic fatigue test for endosseous dental implants.
Clinical data were not submitted in this premarket notification.
{5}------------------------------------------------
# EQUIVALENCE TO MARKETED DEVICE
The subject device is substantially equivalent in indications and design principles predicate devices shown above. Below are summary tables showing technical comparison between the subject device and the primary predicate devices.
| | Indications for Use |
|---------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device | |
| Altatec Gmbh<br><br>CAMLOG® and<br>CONELOG® Abutments<br>for Screw-retained<br>Restorations (ASR) | CAMLOG® and CONELOG® Abutments for Screw-retained Restorations (ASR)<br>are intended to support occlusal screw-retained prosthetic restorations in the upper<br>jaw and lower jaw in conjunction with CAMLOG® and CONELOG® Implants. |
| Primary Predicate Device | |
| Straumann USA, LLC<br><br>Straumann® Magellan™<br>Abutment System<br><br>K133421 | The Straumann® Magellan™ abutments are indicated to be placed into Straumann®<br>dental implants to provide support for prosthetic reconstructions such as crowns,<br>bridges and bars.<br>The final processed devices have the purpose of restoring chewing function.<br>Magellan™ abutments are indicated for screw-retained restorations. |
| | Subject<br>Device | Primary<br>Predicate | Reference Predicate Devices | | |
|-----------------------------------------|----------------------------------------------------------------------------------------|-----------------------------------------------|----------------------------------------------|------------------------------------------|-------------------------------|
| | Altatec Gmbh | Straumann<br>USA, LLC | Straumann<br>USA, LLC | Altatec<br>GmbH | Altatec Gmbh |
| | CAMLOG®<br>and<br>CONELOG®<br>Abutments for<br>Screw-retained<br>Restorations<br>(ASR) | Straumann®<br>Magellan™<br>Abutment<br>System | Straumann®<br>Screw<br>Retained<br>Abutments | CAMLOG<br>Implant<br>System<br>Abutments | CONELOG®<br>Implant<br>System |
| | | K133421 | K141871 | K073553 | K113779 |
| Design | | | | | |
| Prosthesis<br>Attachment | Screw-retained | Screw-retained | Screw-retained | Screw-retained | Screw-retained |
| Abutment<br>Platform<br>Diameter,<br>mm | 3.3 - 6.0 | 3.5 - 4.6 | 4.6 | 3.3 - 6.0 | 3.3 - 5.0 |
| Abutment<br>Angle | Straight to 30° | Straight to 30° | Straight to<br>30° | Straight to<br>20° | Straight to<br>20° |
| Material | | | | | |
| Abutment | Ti-6AL-4V<br>ELI | Ti-6AL-7Nb | Ti-6AL-7Nb | Ti-6AL-4V<br>ELI | Ti-6AL-4V<br>ELI |
{6}------------------------------------------------
Abutment for Screw-retained Restorations (ASR) is substantially equivalent to the predicate devices in K133421 and K141871. K133421 is the primary predicate having straight and angled abutments (17° and 30°), multiple gingival heights (1.0 - 4.0 mm) and Type A and B indexing orientations. Straumann® Screw Retained Abutments K141871 adds angled abutments for the NC diameter implant. Both ASR and Straumann Screw-retained Abutment have a titanium cap designed for temporary or permanent restorations. The subject device and primary predicate have slightly different Indications for Use language. However, the difference in language does not change the intended use of abutments. Both abutments are being used in conjunction with dental implants for the purpose of supporting screw-retained dental prostheses.
## CONCLUSIONS
The subject device and the predicate devices have intended use, have similar technological characteristics, and are made of similar materials. The subject device and predicate devices encompass the same range of physical dimensions, including length, diameter, height, and angle of the abutment. The subject and predicate devices are packaged in similar materials and sterilized using similar methods.
The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.