K152507 · Dr. August Wolff GmbH & Co. KG Arzeimittel · NUC · May 20, 2016 · Obstetrics/Gynecology
Device Facts
Record ID
K152507
Device Name
Dr. Wolff Vagisan Moisturising Cream
Applicant
Dr. August Wolff GmbH & Co. KG Arzeimittel
Product Code
NUC · Obstetrics/Gynecology
Decision Date
May 20, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Vagisan® Moisturizing Cream is a personal lubricant for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex condoms and synthetic (polyurethane and polyisoprene) condoms.
Device Story
Vagisan Moisturizing Cream; non-sterile, water-based, white, non-greasy vaginal cream. Delivered via tube with reusable applicator or sachet. Used by patients for vaginal dryness; moisturizes and lubricates to enhance comfort during sexual activity. Supplements natural lubrication. Not compatible with latex or synthetic condoms. Clinician/patient use. Provides long-lasting moisture; improves comfort; supports sexual health.
Clinical Evidence
Three clinical studies conducted. 1) HRIPT (n=106): 98% showed no irritation/sensitization. 2) Prospective, multicenter, randomized, controlled study (n=117): compared to cleared lubricant; 93% of subjects and 100% of physicians rated tolerability as very good or good. 3) Prospective, non-randomized, non-controlled study (n=117) in breast cancer patients: 90% of subjects and 91% of physicians rated tolerability as very good or good. Additional bench testing included ISO 10993-5:2009 cytotoxicity testing and toxicological risk assessment.
Technological Characteristics
Water-based vaginal cream. Specifications include pH, viscosity, osmolality, and microbial limits. Biocompatibility tested per ISO 10993-5:2009. Non-sterile. Delivered via tube or sachet.
Indications for Use
Indicated for individuals requiring vaginal moisturization and lubrication to enhance comfort during intimate sexual activity and supplement natural lubrication. Not compatible with natural rubber latex, polyurethane, or polyisoprene condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 20, 2016
Dr. August Wolff Gmbh & Co. Kg Arzeimittel % Alan Curtis Regulatory Affairs Consultant AEC Medical Device Consulting LLC 8300 E. Dixileta Drive 285 Scottsdale, AZ 85266
Re: K152507
> Trade/Device Name: Vagisan® Moisturizing Cream Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: Class II Product Code: NUC Dated: April 15, 2016 Received: April 18, 2016
Dear Alan Curtis,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image shows the text "Herbert P. Lerner -S" in a large, bold, sans-serif font. The text is black against a white background. The letters are evenly spaced and easy to read. The overall impression is clean and professional.
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152507
Device Name Vagisan® Moisturizing Cream
#### Indications for Use (Describe)
Vagisan® Moisturizing Cream is a personal lubricant for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex condoms and synthetic (polyurethane and polyisoprene) condoms.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "DR.AUGUST.WOLFF" in a bold, sans-serif font. A curved line is drawn beneath the text, creating a visual underline. The text and line are centered in the image.
# SECTION 5 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
(Prepared August 31, 2015)
# REGULATORY AUTHORITY
This 510(k) Summary is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990, 21 CFR 807.92
#### COMPANY NAME
Dr. August Wolff GmbH & Co. KG Arzneimittel Sudbrackstrasse 56 33611 Bielefeld, Germany
#### Contact Person:
Alan Curtis, RAC Scottsdale, AZ
#### NAME OF DEVICE
| Trade Name: | Vagisan® Moisturizing Cream |
|------------------------|--------------------------------------|
| Common Name: | Personal Lubricant |
| Device Product Code: | NUC |
| Classification Name: | Condom (21 CFR 884.5300) |
| Device Panel: | Obstetrical and Gynecological Device |
| Device Classification: | Class II |
### PREDICATE DEVICE
- Replens® Vaginal Moisturizer (K101098) ●
# DEVICE DESCRIPTION
Vagisan Moisturizing Cream is a non-sterile, water-based, white, non-irritating, non-greasy, non-staining vaginal cream delivered in a tube with a reusable applicator or sachet (sample size) as a long-lasting moisturizer for vaginal dryness.
The quantitative formulation is considered confidential commercial information.
The specifications for the subject device include appearance, odor, pH, viscosity, osmolality, antimicrobial effectiveness, total microbial count, fungal/yeast/mold limits, and absence of pathogenic organisms.
{4}------------------------------------------------
# INDICATION FOR USE STATEMENT
Vagisan Moisturizing Cream is a personal lubricant for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex condoms and synthetic (polyurethane and polyisoprene) condoms.
# SUBSTANTIAL EQUIVALENCE COMPARISON
# Comparison to Predicate Devices
The predicate device (K101098) has the same intended use as the subject device. The subject and predicate devices are water-based lubricants with similar primary ingredients. However, since the formulations of the subject and predicate devices are not identical, the devices do not have the same technological characteristics. The different technological characteristics of the subject device do not raise different types of safety and effectiveness questions.
# PERFORMANCE DATA
Results of pre-clinical and clinical evaluations were used to demonstrate that Vagisan Moisturizing Cream is substantially equivalent to the predicate device and meets safety, and effectiveness criteria.
Biocompatibility Testing: The results of three clinical studies conducted on the subject lubricant, cytotoxicity testing, and a toxicological risk assessment were used to demonstrate the biocompatibility of the subject lubricant. The following clinical studies were conducted on the subject lubricant and applicator:
- Human repeat insult patch test (HRIPT) was conducted on 106 ● human subjects. The study consisted of a 14 day induction phase followed by a 7 day rest period before a 48 hour challenge phase for delayed hypersensitivity. Visual assessments of the treatment site were conducted on six days during the induction phase and following the challenge phase. 104 of 106 (98%) subjects did not show any signs or symptoms of cutaneous irritation. The HRIPT results demonstrated that the subject device is non-irritating and non-sensitizing.
- A prospective, multicenter, controlled, randomized study was conducted on 117 human subjects in whom the subject lubricant was compared to a previously cleared lubricant. The subjects were given either the subject lubricant or control lubricant for a 28 day therapy period followed by a 14 day rest period before
#### CONFIDENTIAL
{5}------------------------------------------------
another 28 day therapy period with the second lubricant. Vaginal examinations were conducted on days 1 and 29 of both therapy periods for a total of four assessments per subject. The subjects also recorded self-assessed symptoms daily in a patient diary. The subject lubricant tolerability was assessed as very good or good by 93% of study subjects and 100% of physicians. The results demonstrated the subject lubricant is non-irritating and non-sensitizing.
- A prospective, non-randomized, non-controlled study was also ● conducted on 117 human subjects undergoing breast cancer treatment. The study subjects were given the subject lubricant for a 28 day therapy period. Vaginal examinations were conducted on days 1 and 29 for each subject. The study subjects also selfreported symptoms in a patient diary. The subject lubricant tolerability was assessed as very good or good by 90% of study subjects and 91% of physicians. The results demonstrated good tolerability of the subject lubricant.
Cytotoxicity testing was also conducted on the subject lubricant in accordance with ISO 10993-5:2009. Although the results showed a concentration-dependent cytotoxic effect, this is not uncommon for lubricants tested neat (worse-case exposure conditions) and does not preclude clearance of a personal lubricant. The subject lubricant also performed better than several other previously cleared lubricants. Based on the results from the clinical studies, the in vitro cytotoxicity testing does not raise significant safety concerns.
A toxicological risk assessment was also performed on the subject lubricant based on the total amount of each ingredient to assess for systemic toxicity. The toxicological risk assessment demonstrated that the total amount of each ingredient in the lubricant is less than the tolerable exposure values obtained from a literature review.
Stability Testing: The results of a real time aging study demonstrate that the subject lubricant maintains its specifications over 24 months.
### CONCLUSION
Vagisan Moisturizing Cream is substantially equivalent to the proposed predicate device.
CONFIDENTIAL
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.