K152501 · Neuropro Spinaljaxx, Inc. · MAX · Jun 29, 2016 · Orthopedic
Device Facts
Record ID
K152501
Device Name
Spinal Jaxx Interbody Fusion Device
Applicant
Neuropro Spinaljaxx, Inc.
Product Code
MAX · Orthopedic
Decision Date
Jun 29, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Spinal Jaxx interbody fusion device is indicated for spinal fusion for patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2 – S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The indicated patient population is skeletally mature patients who have had six (6) months of non-operative treatment. The Spinal Jaxx interbody fusion device must be used with autogenous bone graft material and with supplemental fixation.
Device Story
Lumbar intervertebral spacer; provides structural stability in disc space; height-adjustable by surgeon to restore disc height. Used in surgical spinal fusion procedures; requires autogenous bone graft and supplemental fixation. Implants constructed from PEEK Optima LT1, titanium alloy, CP titanium, and Nitinol. Instruments made from stainless steel, aluminum, Nitinol, and silicone rubber. Device provides mechanical support to facilitate fusion in patients with degenerative disc disease.
Clinical Evidence
Bench testing only. Mechanical testing performed per Class II Special Controls Guidance Document: Intervertebral Fusion Device (June 12, 2007), including static/dynamic compression and shear (ASTM F2077), subsidence (ASTM F2267-04), static expulsion (ASTM Draft F-04.25.02.02), wear debris analysis (ASTM F1877-05), and cyclic potentiodynamic polarization (ASTM F2129-15).
Technological Characteristics
Intervertebral body fusion device; materials include PEEK Optima LT1, titanium alloy, CP titanium, and Nitinol. Instruments utilize stainless steel, aluminum, Nitinol, and silicone rubber. Mechanical testing conducted per ASTM F2077, F2267-04, F1877-05, and F2129-15. Device is a mechanical implant; no software or electronic components.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2–S1), including those with up to Grade 1 spondylolisthesis or retrolisthesis, who have failed six months of non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized image of a bird or abstract shape formed by three overlapping profiles of human faces.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 29, 2016
NeuroPro Spinal Jaxx, Incorporated c/o Rich Jansen. Pharm.D. Consultant Silver Pine Consulting, LLC 11821 Bramble Cove Drive Fort Myers, Florida 33905
Re: K152501
Trade/Device Name: Spinal Jaxx Interbody Fusion Device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: June 3, 2016 Received: June 6, 2016
Dear Dr. Jansen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152501
Device Name Spinal Jaxx Interbody Fusion Device
#### Indications for Use (Describe)
The Spinal Jaxx interbody fusion device is indicated for spinal fusion for patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2 – S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). The indicated patient population is skeletally mature patients who have had six (6) months of non-operative treatment. The Spinal Jaxx interbody fusion device must be used with autogenous bone graft material and with supplemental fixation.
#### Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY Interbody Fusion Syst
| Spinal Jaxx Interbody Fusion System | |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | June 24, 2016 |
| Company: | NeuroPro Spinal Jaxx, Inc.<br>4707 Greenleaf Ct., Ste. C<br>Modesto, CA 95356 |
| Contact: | John Green, COO<br>Jgreen@neuroprotech.com<br>571-334-7424 |
| Regulatory<br>Contact | Rich Jansen, Pharm. D.<br>Silver Pine Consulting |
| Device Name: | Spinal Jaxx Interbody Fusion Device |
| Classification: | Per 21 CFR 888.3080<br>Intervertebral Body Fusion Device<br>Product Code: MAX<br>Regulatory Class II<br>Panel Code 87 |
| Predicate<br>Devices: | The primary predicate is the CALIBER™ Spacer (K102293). Additional<br>predicate devices include the Rise Spacer (K113447), Brantigan Cage (AKA<br>LT Cage & Jaguar Cage) (P960025), Ray Threaded Fusion Cage<br>(P950019), Patriot Spacer (K072970) and the Nanovis Foricore device<br>(K140280) |
### Purpose:
The purpose of this submission is clearance of the Spinal Jaxx Interbody Fusion Device as a new medical device that is substantially equivalent to the predicate devices.
## Device Description:
The Spinal Jaxx Interbody Fusion device is a lumbar intervertebral spacer intended to provide structural stability in a skeletally mature individual based on the surgical intervention by the surgeon. The physician can insert it into the diseased disc space and adjust the height of the device as desired to create near normal disc height. The implants are made from PEEK Optima LT1, a medical grade titanium alloy, CP titanium and Nitinol. Implants are available in various sizes to accommodate patient anatomy. Instruments are made from stainless steel, aluminum, Nitinol, and silicone rubber.
#### Indications for Use:
The Spinal Jaxx interbody fusion device is indicated for spinal fusion for patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2 – S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. Patients may also have up to Grade 1 spondylolisthesis at the involved level(s). The indicated patient population is skeletally mature patients who have had six (6) months of non-operative treatment. The Spinal Jaxx interbody fusion device must be used with autogenous bone graft material and with supplemental fixation.
#### Performance Data:
To demonstrate substantial equivalence to the predicate devices, mechanical testing was conducted in accordance with "Class II Special Controls Guidance Document: Intervertebral Fusion Device, June 12, 2007" including the ASTM standards:
{4}------------------------------------------------
- Static Compression and Static Compression Shear testing per ASTM F2077 ●
- Dynamic Axial Compression and Dynamic Compression Shear testing per ASTM F2077
- Subsidence Testing per ASTM F2267-04
- Static Expulsion/Industry Standard per ASTM Draft Standard F-04.25.02.02
- Wear Debris analysis per ASTM F1877-05
- Cyclic Potentiodynamic Polarization Testing per ASTM F2129-15
## Technological Characteristics:
NeuroPro has compared the Spinal Jaxx Interbody Fusion System to the predicate devices in regards to indications for use, materials, function, sizes and mechanical test results. These comparisons demonstrate substantial equivalence to the predicate devices.
## Conclusions:
The Spinal Jaxx Interbody Fusion System is substantially equivalent to the predicate devices in terms of indications for use, design, material, and function.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.