K152481 · Codman & Shurtleff, Inc. · GXQ · Oct 30, 2015 · Neurology
Device Facts
Record ID
K152481
Device Name
DURAFORM Dural Graft Implant
Applicant
Codman & Shurtleff, Inc.
Product Code
GXQ · Neurology
Decision Date
Oct 30, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5910
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DURAFORM Dural Graft Implant is intended for use in procedures where the repair or substitution of the patient's dura mater is needed.
Device Story
DURAFORM Dural Graft Implant is a single-use, sterile collagen implant derived from processed bovine tendons. Designed as a sutureless, onlay graft for dural repair or substitution; tensionless sutures optional. Used by surgeons in clinical settings. Device provides a physical barrier for dural defects. Substantial equivalence established via shelf-life extension from 12 to 30 months compared to predicate. No changes to materials, design, or manufacturing process.
Clinical Evidence
No clinical or animal studies were performed. Substantial equivalence for the shelf-life extension was supported by bench-top device stability testing and packaging integrity testing.
Technological Characteristics
Single-use, sterile collagen implant derived from bovine tendons. Sutureless onlay graft design. Packaging integrity tested per ASTM F1929-12 and ASTM F88/F88M-09. No electronic components or software.
Indications for Use
Indicated for patients requiring repair or substitution of the dura mater during surgical procedures.
Regulatory Classification
Identification
A dura substitute is a sheet or material that is used to repair the dura mater (the membrane surrounding the brain).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an emblem featuring a stylized depiction of three human profiles facing to the right, layered on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 30, 2015
Codman & Shurtleff, Inc. Ms. Jennifer Siu Regulatory Affairs Specialist 325 Paramount Drive Raynham, Massachusetts 02767
Re: K152481
Trade/Device Name: DURAFORM Dural Graft Implant Regulation Number: 21 CFR 882.5910 Regulation Name: Dura Substitute Regulatory Class: Class II Product Code: GXQ Dated: October 1, 2015 Received: October 2, 2015
Dear Ms. Siu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K152481
Device Name DURAFORM Dural Graft Implant
Indications for Use (Describe)
The DURAFORM Dural Graft Implant is intended for use in procedures where the repair or substitution of the patient's dura mater is needed.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| I. Submitter | Codman & Shurtleff, Inc.<br>325 Paramount Drive<br>Raynham, MA 02767 | | |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|--|
| | Establishment Registration Number: 1226348 | | |
| | Primary Contact: | Jennifer Siu<br>(508) 828-3288<br>(508) 977-6979 | |
| | Secondary Contact: | Jocelyn Raposo<br>(508) 828-3421<br>(508) 977-6979 | |
| | Date of Submission: | August 28, 2015 | |
| II. Device | Device Proprietary Name | DURAFORM™ Dural Graft Implant | |
| | Common Name | Dura Substitute | |
| | Classification Name | Dura Substitute (21 CFR 882.5910) | |
| | Regulatory Classification | II | |
| | Product Code | GXQ | |
| III. Predicate Device | The predicate device for this submission is the CODMAN Dural Graft<br>Implant (K033395), which was cleared on February 25, 2004. | | |
| IV. Device Description | The DURAFORM Dural Graft Implant is a single use, sterile collagen<br>implant manufactured from processed bovine tendons. The Duraform<br>Dural Graft Implant is designed to be a sutureless, onlay graft, but<br>tensionless sutures can be used if preferred by the surgeon. | | |
| V. Indications for Use | The DURAFORM Dural Graft Implant is intended for use in procedures<br>where the repair or substitution of the patient's dura mater is needed. | | |
| VI. Comparison to Predicate Device | The DURAFORM Dural Graft Implant is identical to the predicate<br>device, formerly known as Codman Dural Graft Implant (K033395), with<br>regard to indications for use, design, materials, manufacturing process, | | |
{4}------------------------------------------------
| VI. Comparison<br>to Predicate<br>Device (Cont.) | clinical utility, packaging, and sterilization; only the shelf life is different.<br>Codman is extending the shelf life for the DURAFORM Dural Graft<br>Implant from 12 months to 30 months. The appropriate spectrum of<br>stability testing has been performed to demonstrate substantial<br>equivalence to the CODMAN Dural Graft Implant with a 12-month shelf<br>life. |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| VII.<br>Performance<br>Data | There were no changes made that affect the DURAFORM Dural Graft<br>Implant's indications for use, design, materials, manufacturing process,<br>clinical utility, packaging, and sterilization. |
| | Device and packaging stability testing for device and packaging<br>characteristics that may be affected by an extended shelf life was<br>performed. All testing was performed on final sterile devices unless<br>otherwise specified. |
The following device stability testing was performed at time-zero and after 30 months of real-time aging:
| Test | Test Method Summary | Results |
|---------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Compressibility<br>(Time-Zero and<br>30-Month) | Compressibility quantifies the<br>ability of a device to be compressed.<br>Compressibility is measured by<br>comparing the compressed<br>thickness to the original thickness of<br>the device. The compressed<br>thickness is measured with a<br>thickness gage after compression at<br>120 seconds. | Both time-zero and<br>30-month real-time tests<br>passed the acceptance<br>criterion. Therefore, the<br>proposed device is<br>substantially equivalent to<br>the predicate device and<br>does not raise new issues in<br>safety and effectiveness. |
| Resilience<br>(Time-Zero and<br>30-Month) | Resilience quantifies the capability<br>of a device returning back to its<br>original state. Resilience is<br>measured by comparing the<br>recovered thickness after<br>compression to the original<br>thickness of the device. The<br>recovered thickness is measured<br>after the device has been<br>compressed for 120 seconds and left<br>to recover to its original state for<br>30 minutes. | Both time-zero and<br>30-month real-time tests<br>passed the acceptance<br>criterion. Therefore, the<br>proposed device is<br>substantially equivalent to<br>the predicate device and<br>does not raise new issues in<br>safety and effectiveness. |
| Test | Test Method Summary | Results |
| Imbibition<br>(Time-Zero and<br>30-Month) | The imbibition quotient measures<br>how much fluid a material is able to<br>absorb. In order to determine<br>imbibition quotient, the dividend of<br>the dry weight of a DURAFORM<br>Dural Graft Implant sample is taken<br>from wet weight of the sample after<br>10 seconds of irrigation, the<br>procedure representing clinical use. | Both time-zero and<br>30-month real-time tests<br>passed the acceptance<br>criterion. Therefore, the<br>proposed device is<br>substantially equivalent to<br>the predicate device and<br>does not raise new issues in<br>safety and effectiveness. |
| Apparent<br>Density<br>(Time-Zero and<br>30-Month) | Apparent density measures the<br>density of a dural substitute<br>material. Apparent density is<br>measured using the dry weight of<br>the device. | Both time-zero and<br>30-month real time tests<br>passed the acceptance<br>criterion. Therefore, the<br>proposed device is<br>substantially equivalent to<br>the predicate device and<br>does not raise new issues in<br>safety and effectiveness. |
| Visual<br>Inspection<br>(Time-Zero and<br>30-Month) | A visual inspection is performed for<br>the samples to be white in color<br>with no stains visible. | Both time-zero and<br>30-month real-time tests<br>passed the acceptance<br>criterion. Therefore, the<br>proposed device is<br>substantially equivalent to<br>the predicate device and<br>does not raise new issues in<br>safety and effectiveness. |
| Tensile<br>Strength<br>(Time-Zero and<br>30-Month) | Tensile strength testing is performed<br>on the DURAFORM Dural Graft<br>Implant per FDA's guidance on<br>physical and mechanical properties<br>testing for dura substitutes, as<br>outlined in the Guidance Document<br>for Dura Substitute Devices:<br>Guidance for Industry.<br>In order to test tensile strength, each<br>sample is wetted with water, placed<br>into an Instron instrument, and<br>rewetted to ensure the sample has<br>not dried out. Then, the instrument<br>pulls the sample and captures the<br>peak load when failure occurs. | Time-zero testing passed<br>the acceptance criteria;<br>30-month testing results<br>were deemed acceptable.<br>Therefore, the proposed<br>device is substantially<br>equivalent to the predicate<br>device and does not raise<br>new issues in safety and<br>effectiveness. |
| Test | Test Method Summary | Results |
| Visual<br>Inspection<br>(120-Month<br>and<br>132-Month) | A visual inspection is performed<br>for any damage to the packaging. | Both 120-month and<br>132-month tests passed the<br>acceptance criterion.<br>Therefore, packaging integrity<br>will be maintained for the<br>duration of the proposed shelf<br>life at 30 months. |
| Seal Integrity<br>(Dye<br>Penetration)<br>(120-Month<br>and<br>132-Month) | Seal integrity (dye penetration)<br>testing was tested in accordance<br>with ASTM F1929-12 Standard<br>Test Method for Detecting Seal<br>Leaks in Porous Medical<br>Packaging by Dye Penetration. | Both 120-month and<br>132-month tests passed the<br>acceptance criterion.<br>Therefore, packaging integrity<br>will be maintained for the<br>duration of the proposed shelf<br>life at 30 months. |
| Seal Strength<br>(120-Month<br>and<br>132-Month) | Seal strength testing was tested<br>in accordance with ASTM<br>F88/F88M-09 Standard Test<br>Method for Seal Strength of<br>Flexible Barrier Materials. | Both 120-month and<br>132-month tests passed the<br>acceptance criterion.<br>Therefore, packaging integrity<br>will be maintained for the<br>duration of the proposed shelf<br>life at 30 months. |
{5}------------------------------------------------
VII.
Performance Data (Cont.)
{6}------------------------------------------------
## VII. Performance Data (Cont.)
The following packaging stability testing, performed at 120 months and 132 months of real-time aging, was leveraged to support packaging integrity for the proposed shelf life at 30 months:
## Animal Testing:
No animal studies were performed as appropriate validation of the shelf life extension was achieved based on the comparison to the predicate device and from the results of the device and packaging stability studies.
## Clinical Testing:
No clinical studies were performed as appropriate validation of the shelf life extension was achieved based on the comparison to the predicate device and from the results of the device and packaging stability studies.
## Conclusion:
Based upon the indications for use, design, materials, packaging, comparison to the currently marketed device, and shelf life testing performed by Codman, it is concluded that the DURAFORM Dural Graft Implant with a 30-month shelf life is substantially equivalent to the predicate CODMAN Dural Graft Implant with a 12-month shelf life and therefore, does not raise any new issues of safety and effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.