OptiOne Single-Head Contrast Delivery System with Pedestal; with Ceiling Suspension;Injector, OptiOne Base System
Applicant
Liebel-Flarsheim Company, LLC
Product Code
IZQ · Cardiovascular
Decision Date
Jan 14, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1650
Device Class
Class 2
Indications for Use
The OptiOne Single-Head Contrast Delivery System is a contrast delivery system and is designed to inject radiopaque contrast media into a patient's vascular system to obtain diagnostic images with computed tomography (i.e. "CT") equipment.
Device Story
OptiOne is a motor-driven, microprocessor-controlled contrast delivery system for CT imaging; used in healthcare facilities by clinicians/technologists. System components include powerhead, console, and power supply. Operator programs flow rate, volume, pressure, and timing via touch screen console. Powerhead utilizes motor control circuits, ball screw, and syringe push ram to deliver contrast. Features include OptiBolus for optimized flow, remote start for radiation safety, and electrical safety stops. Output is controlled contrast delivery to patient vascular system; enables diagnostic image acquisition. Benefits include precise, programmable contrast administration for improved image quality.
Clinical Evidence
Bench testing only. Verification and validation activities confirmed device performance against design inputs and specifications. Compliance with recognized standards including AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, and IEC 62304 was demonstrated.
Technological Characteristics
Electromechanical syringe drive system; plastic and metal construction. Features: 4-phase protocol capability, 40-protocol storage, 98°F syringe heater, 0.1-10 mL/s flow rate, 50-325 psi pressure limit. Connectivity via relays, optical couplings, and CAN serial interface. Software lifecycle per IEC 62304. Usability per IEC 62366.
Indications for Use
Indicated for patients requiring radiopaque contrast media injection into the vascular system for diagnostic CT imaging.
Regulatory Classification
Special Controls
*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
OptiVantage DH Injector System with Enhanced Communication (K063503)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol of medicine and healing. The caduceus is composed of three human profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circle around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 14, 2016
Liebel-Flarsheim Company, LLC Craig Buehler Sr. Regulatory Affairs Specialist 2111 East Galbraith Rd Cincinnati, Ohio 45237
Re: K152361
Trade/Device Name: OptiOne Single-Head Contrast Delivery System Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector And Syringe Regulatory Class: Class II Product Code: IZQ Dated: December 14, 2015 Received: December 15, 2015
Dear Craig Buehler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152361
Device Name
OptiOne Single-Head Contrast Delivery System
Indications for Use (Describe)
The OptiOne Single-Head Contrast Delivery System is a contrast delivery system and is designed to inject radiopaque contrast media into a patient's vascular system to obtain diagnostic images with computed tomography (i.e. "CT") equipment.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Mallinckrodt Pharmaceuticals. The logo consists of a geometric shape on the left, made up of overlapping cubes in shades of green, blue, and purple. To the right of the shape is the company name, "Mallinckrodt" in a bold, dark blue font, with the word "Pharmaceuticals" underneath in a lighter blue, smaller font.
# Section 5: 510(k) Summary
### I. SUBMITTER
Liebel-Flarsheim Company LLC 2111 East Galbraith Rd. Cincinnati. OH 45237 Establishment Number: 1518293
| Contact: | Mr. Craig Buehler |
|----------------|-----------------------------------|
| Title: | Sr. Regulatory Affairs Specialist |
| Telephone: | (314) 654-3188 |
| Fax: | (314) 654-3344 |
| Date Prepared: | 19 August 2015 |
### II. DEVICE
Name of Device: OptiOne Single-Head Contrast Delivery System Common or Usual Name: Power Injector Classification Name: Injector, Contrast Medium, Automatic (21 CFR 870.1650) Regulatory Class: II Product Code: IZQ
| Model Number | Description |
|--------------|-------------------------------------------------------------------------|
| 847002 | Injector, OptiOne Base System |
| S8472 | OptiOne Single-Head Contrast Delivery System with Pedestal |
| S8473 | OptiOne Single-Head Contrast Delivery System with Ceiling<br>Suspension |
### III. PREDICATE DEVICE
OptiVantage DH Injector System with Enhanced Communication (K063503)
## IV. DEVICE DESCRIPTION:
The OptiOne Single-Head Contrast Delivery System (OptiOne) is intended to inject radiopaque contrast media into a patient's vascular system to obtain diagnostic images when used in conjunction with computed tomography equipment. Each injection is accomplished with a motor-driven syringe mechanism with microprocessor control of the flow rate, volume, pressure and timing. The OptiOne consists of the following key components: a Powerhead, a Console, and a Power Supply. The Power
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Mallinckrodt Pharmaceuticals logo. The logo consists of a geometric shape on the left and the company name on the right. The geometric shape is made up of several overlapping cubes in shades of green, blue, and purple. The company name is written in a dark blue font, with the word "Mallinckrodt" on top and the word "Pharmaceuticals" below.
Supply delivers power to the Powerhead and Console. The Console provides remote control and programmability of the injection through a touch screen display system. The Powerhead contains the components to perform the actual injection such as motor control circuits, ball screw and syringe push ram.
The OptiOne is intended for use by doctors, radiology technologists and other licensed medical practitioners in a healthcare facility or hospital with computed tomography equipment.
The OptiOne incorporates the same materials as its predicate device and will not contact the patient under normal operation conditions.
#### V. INDICATIONS FOR USE
The OptiOne Single-Head Contrast Delivery System is a contrast delivery system and is designed to inject radiopaque contrast media into a patient's vascular system to obtain diagnostic images when used with computed tomography (i.e. "CT") equipment.
The OptiOne and the OptiVantage predicate device both have the same indication for use.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The OptiOne Single-Head Contrast Delivery System (OptiOne) was developed using the predicate device, OptiVantage DH Injector System with Enhanced Communication (OptiVantage), as the basis for the injector design. Many of the same components and accessories found in and used with the OptiVantage are also found and used with the OptiOne. The main difference between the OptiOne and its predicate is the OptiVantage is a dual head injector and the OptiOne is a single head injector.
The OptiOne has the same technological characteristics as the predicate OptiVantage. These similarities and the differences reducing the powerhead from the predicate device's dual head to a single head are described in the table beginning on the next page.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Mallinckrodt Pharmaceuticals. The logo consists of a geometric design on the left, made up of overlapping cubes in shades of green, blue, and purple. To the right of the geometric design is the company name, "Mallinckrodt," in a bold, dark blue font, with the word "Pharmaceuticals" underneath in a smaller, lighter blue font.
| Feature | OptiOne Single-Head Contrast<br>Delivery System | OptiVantage DH Injector System<br>with Enhanced Communication<br>(K063503) |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| Indication for Use | is a contrast delivery system and is<br>designed to inject radiopaque<br>contrast media into a patient's<br>vascular system to obtain<br>diagnostic images when used with<br>computed tomography (i.e. "CT")<br>equipment. | Same |
| Number of Phases | 4 phases per protocol maximum | 6 phases per protocol maximum |
| Protocol Storage | 40 protocols | Same |
| Syringe Sizes | 125 mL, 100 mL, 75 mL, 50 mL<br>pre-filled<br>200 mL empty | Same |
| Single or Dual Syringe<br>Modality | Single Syringe Capability | Single and Dual Syringe Capability |
| Syringe Drive System | Electromechanical | Same |
| Syringe Heater | 98° ± 6° F (37° ± 3° C) nominal.<br>Maintains the temperature of pre-<br>heated contrast (not intended to<br>heat cold/room temperature<br>contrast). | Same |
| Auto Syringe Fill Rate | 7.0 mL/s<br>Tolerance: +/-5% | Same |
| Manual Fill Rate | 2.0 - 15 mL/s<br>Tolerance: +/- 1 mL/s | Same |
| Flow Rate | 0.1 - 10 mL/s adjustable in<br>increments of 0.1 mL/s<br>Tolerance: +/-0.05 mL/s for Flow<br>Rates less than 1 mL/s<br>+/-5 % for Flow Rates 1 mL/s to<br>10 mL/s | Same |
| Peak Pressure Limit | Pounds per Square Inch (psi):<br>50 - 325 adjustable in 5 psi<br>increments<br>kPa: 345 – 2240 adjustable in<br>34 kPa increments | Same |
| Feature | OptiOne Single-Head Contrast<br>Delivery System | OptiVantage DH Injector System<br>with Enhanced Communication<br>(K063503) |
| Remote Start | Use of the remote hand switch<br>allows the operator to perform<br>injections from outside the area of<br>direct radiation. | Same |
| Scan Delay | 0 - 600 seconds adjustable in<br>increments of 1 second | Same |
| Phase Delay | 0 – 600 seconds adjustable in<br>increments of 1 second or<br>permanent pause | Same |
| Inject Delay | 0 - 600 seconds adjustable in<br>increments of 1 second | Same |
| OptiBolus | Yes - The OptiBolus feature is<br>used to deliver an exponentially<br>decaying flow rate injection that<br>optimizes the contrast usage and<br>provides an extended period of<br>uniform enhancement of the area<br>of interest. | Same |
| Universal Console | 12.25 W x 2.5 H x 8.5 D inches<br>(311 W x 64 H x 216 D mm)<br>5.8 lbs (2.6 kg) | Same |
| Powerhead | 7 W x 6 H x 12.25 D inches<br>(178 W x 152 H x 311 D mm)<br>10.4 lbs (4.7 kg) | 12.5 W x 6 H x 8 D inches<br>(318 W x 152 H x 203 D mm)<br>14.5 lbs (6.57 kg) |
| Power Supply | 10 W x 4 H x 9 D inches<br>(267 W x 102 H x 228 D mm)<br>6 lbs (2.7 kg) | Same |
| Post Injection<br>Readout | Displayed on Powerhead and<br>Console | Same |
| OEM Interface | Available Via Relays and Optical<br>Couplings / Serial (CAN) | Same |
| Safety Stop<br>Mechanism | Electrical Stop When Injection<br>Parameters are Out of<br>Specification | Same |
| Volume Remaining<br>Display | Displayed on Powerhead and<br>Console | Same |
| Materials | Plastic and Metal | Same Plastic and Metal |
| Target Population | Humans | Same |
| Feature | OptiOne Single-Head Contrast<br>Delivery System | OptiVantage DH Injector System<br>with Enhanced Communication<br>(K063503) |
| Sterility (Syringe) | Injectors are not sterile products,<br>but syringes and disposables are<br>provided sterile. | Same |
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Mallinckrodt Pharmaceuticals. The logo consists of a geometric shape made up of overlapping cubes in shades of green, blue, and purple. To the right of the geometric shape is the company name, "Mallinckrodt" in a bold, dark blue font, with the word "Pharmaceuticals" underneath in a lighter, thinner font.
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Mallinckrodt Pharmaceuticals. On the left is a geometric design of overlapping cubes in shades of green, blue, and purple. To the right of the design is the company name, "Mallinckrodt" in large, bold, navy blue font, with the word "Pharmaceuticals" underneath in a smaller, lighter blue font.
#### VII. PERFORMANCE DATA
Extensive verification and validation activities were performed to ensure the OptiOne Single-Head Contrast Delivery System design outputs met all design inputs and the customer requirements were successfully met. The testing confirms the OptiOne Single-Head Contrast Delivery System meets the required specifications.
The OptiOne Single-Head Contrast Delivery System was also tested in conformance to the following recognized standards:
| Standard |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC ISO 14971 Second edition 2007-03-01 Application of Risk Management to<br>Medical Devices |
| ISO 15223-1 Second Edition 2012-07-01 Medical Devices - Symbols to be<br>Used with Medical Device Labels, Labelling, and Information to be Supplied -<br>Part 1 |
| AAMI / ANSI ES60601-1:2005/(R)2012 and C1:2009/(R)2012 and,<br>a2:2010/(r)2012 Medical Electrical Equipment -- Part 1: General Requirements<br>for Basic Safety and Essential Performance |
| IEC 60601-1-2 Edition 4.0 2014-02 Medical Electrical Equipment - Part 1-2:<br>General Requirements for Basic Safety and Essential Performance - Collateral<br>Standard: Electromagnetic Disturbances - Requirements and Tests |
| IEC 60601-1-6 Edition 3.1 2013-10 Medical Electrical Equipment - Part 1-6:<br>General Requirements for Basic Safety and Essential Performance - Collateral<br>Standard: Usability |
| IEC 62304 First Edition 2006-05 Medical Device Software - Software Life<br>Cycle Processes. (Software/Informatics) |
| IEC 62366 Edition 1.1 2014-01 Medical Devices - Application of Usability<br>Engineering to Medical Devices. (General I) |
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the Mallinckrodt Pharmaceuticals logo. The logo consists of a geometric design on the left and the company name on the right. The geometric design is made up of overlapping cubes in shades of green, blue, and purple. The company name is written in a dark blue font, with "Mallinckrodt" on the top line and "Pharmaceuticals" on the bottom line.
## VIII. CONCLUSIONS
The OptiOne Single-Head Contrast Delivery System was developed under the Liebel-Flarsheim Company LLC Design Control processes. The verification and validation activities ensure the OptiOne meets all customer requirements and product specifications. The OptiOne Single-Head Contrast Delivery System has the same intended use as its predicate. The powerhead and software was modified from the predicate device to reflect the clinical use of a single head contrast delivery system from a dual head injector. Based on the data contained in this submission, Liebel-Flarsheim Company LLC believes the OptiOne Single-Head Contrast Delivery System is substantially equivalent to its predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.