← Product Code [FMI](/productcode/FMI) · K152334

# Sungshim Insulin Pen Needle (K152334)

_Sung Shim Medicare Co., Ltd. · FMI · Mar 17, 2016 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K152334

## Device Facts

- **Applicant:** Sung Shim Medicare Co., Ltd.
- **Product Code:** [FMI](/productcode/FMI.md)
- **Decision Date:** Mar 17, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5570
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

This product is a single use sterile insulin pen needle intended for subcutaneous injection of insulin to patient with diabetes. It is to be used with compatible insulin injectors.

## Device Story

Single-use sterile insulin pen needle; consists of hub, needle, needle cap, sterile cap, and sterile paper. Connects to pen-type insulin injectors for subcutaneous insulin delivery. Used by patients with diabetes or healthcare providers. Device maintains sterility via primary container and paper. Bench testing confirms mechanical integrity and compatibility with standard insulin injectors (e.g., Flex Pen, Lantus, Novo Let N).

## Clinical Evidence

Bench testing only. Testing included biocompatibility, mechanical properties, sterility (ISO 11135:2007), and EO residuals (ISO 10993-7). Functional testing followed ISO 7864:1993. Compatibility with specific insulin injectors verified per ISO 11608-2.

## Technological Characteristics

Materials: Polypropylene (sterile cap, hub), Polyethylene (needle cap), Stainless steel 304 (needle). EO gas sterilization. Dimensions: 29G-32G, lengths 4mm-13mm. Connectivity: Mechanical interface with insulin pens. Standards: ISO 7864:1993, ISO 10993-7, ISO 11135:2007, ISO 11608-2.

## Regulatory Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

## Predicate Devices

- Feel Fine Insulin Pen Needle ([K080904](/device/K080904.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling a symbol of three stylized human profiles facing right, with flowing lines beneath them. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the circular seal.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 17, 2016

Sungshim Medicare Co., Ltd. c/o Mr. Peter Chung President Plus Global 300 Atwood Street Pittsburgh, Pennsylvania 15213

Re: K152334

Trade/Device Name: Sungshim Insulin Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: February 4, 2016 Received: February 10, 2016

Dear Mr. Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K152334

Device Name Sungshim Insulin Pen Needle

Indications for Use (Describe)

This product to be used with insulin injector is a single use sterile insulin pen needle to be used for injecting insulin to patient with diabetes.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

# 510(k) Summary

[as required by 807.92(c)]

# 1. Applicant

- 1) Company : Sungshim Medicare Co.,Ltd.
- 2) Address : 190, Maesil-ro, Sojeong-myeon, Sejong-si, Korea
- 3) Tel : 82-32-676-7066
- 4) Fax : 82-32-676-7063
- 5) Prepared date : Oct. 7, 2015
- 5) Contact person : Peter Chung, 412-687-3976
- 6) Contact person address : 300, Atwood Street, Pittsburgh, PA, 15213, USA
- 7) Submission date : Aug. 10, 2015
- 8) 510(k) number : K152334

### 2. Device Information

- 1) Trade name : Sungshim Insulin Pen Needle
- 2) Common name : Hypodermic single lumen needle
- 3) Classification name : Needle, Hypodermic, Single Lumen
- 4) Product code : FMI
- 5) Regulation number : 880.5570
- 6) Class of device : Class II
- 7) Panel : General Hospital

# 3. The legally marketed device to which we are claiming equivalence

K080904 Feel Fine Insulin Pen Needle

#### 4. Device description

Sungshim insulin pen needle consist of a Sterile cap, Needle, Hub and Sterile paper. The Primary container and Sterile paper function to sustain sterilization of the product. The Hub can be connected with pen type insulin syringe. The Needle cap protects the Needle. This device is single use.

#### 5. Intended Use :

This product is a single use sterile insulin pen needle intended for subcutaneous injection of insulin to patient with diabetes. It is to be used with compatible insulin injectors.

### 6. Performance data:

- (1) Bench test were performed. Bench testing included biocompatibility, mechanical testing, sterility testing including EO residues. The tests demonstrated that the device performs in a substantially equivalent manner to the predicate device. The following bench testing is performed to demonstrate the functionality is substantially equivalent.
Test criteria : ISO 7864:1993 Sterile hypodermic needles for single use

{4}------------------------------------------------

| Test item                                    | Requirements                                                                                                                                                                      | Results |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Inner/out side<br>and structure              | When it was tested accordingly to the standard specifications for<br>medical device-"sterile hypodermic needles for single use" clause 2,<br>it should satisfy the requirements.  | Pass    |
| Measurement                                  | When it was tested accordingly to the standard specifications for<br>medical device-"sterile hypodermic needles for single use" clause 3,<br>it should satisfy the requirements.  | Pass    |
| Elasticity test                              | When it was tested accordingly to the standard specifications for<br>medical device-"sterile hypodermic needles for single use" clause 4,<br>it should satisfy the requirements.  | Pass    |
| Draw test                                    | When it was tested accordingly to the standard specifications for<br>medical device-"sterile hypodermic needles for single use" clause 6,<br>it should satisfy the requirements.  | Pass    |
| Package                                      | When it was tested accordingly to the standard specifications for<br>medical device-"sterile hypodermic needles for single use" clause 11,<br>it should satisfy the requirements. | Pass    |
| Ethylene oxide<br>sterilization<br>residuals | According to ISO 10993-7<br>ETO≤ 25ppm / ECH≤ 25ppm / EG≤250ppm                                                                                                                   | Pass    |
| Sterility test                               | When it was tested accordingly to ISO 11135:2007                                                                                                                                  | Pass    |
| Extraction test                              |                                                                                                                                                                                   |         |
| Appearance                                   | The sample solution shouled be colorless and transparent and<br>there should not be foreign materials.                                                                            | Pass    |
| pH                                           | Difference in pH≤1.5                                                                                                                                                              |         |
| KMnO4 Reducing<br>adents                     | Difference in titiers≤ 2.0ml                                                                                                                                                      |         |
| Evaporating residue                          | Difference in extractable≤ 1.0mg                                                                                                                                                  |         |
| heavy Pb,Fe,Sn,Zn<br>metal Cd                | Not greater than a comined total of 5mg/L of Pb, Fe, Sn and Zn<br>Shall be less than 0.1mg/L of Cd                                                                                |         |
| UV-vis spectrum                              | Difference in absorbance (250nm~350nm)≤ 0.1                                                                                                                                       |         |

Compatibility testing was performed according to ISO 11608-2 and passed with the following injectors:

- Novo Nordisk A/S, Flex Pen / 3ml / Insulin sprat .
- Sanofi-Aventis Deutschland GmbH. LANTUS / 3ml / Insulin glargine ●
- Novo Nordisk A/S, Novo Let N / 3ml / Isophane insulin .

The performance tests demonstrated that Sungshim insulin pen needle performs in a substantially equivalent manner to the predicate device.

{5}------------------------------------------------

#### 7. Predicate device comparison table

| Manufacturer                     |  | Sungshim Medicare Co.,Ltd.                                                                                                                                                                       |                                                                                                                                        | Feel Tech                     |  |  |  |  |  |  |  |  |                                                                                                                       |  |  |  |  |  |  |  |  |  |  |
|----------------------------------|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|--|--|--|--|--|--|--|--|-----------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|
| 510(k) No.                       |  |                                                                                                                                                                                                  | K080904                                                                                                                                |                               |  |  |  |  |  |  |  |  |                                                                                                                       |  |  |  |  |  |  |  |  |  |  |
| Indication for use               |  | This product is a single use sterile insulin<br>pen needle intended for subcutaneous<br>injection of insulin to patient with<br>diabetes. It is to be used with compatible<br>insulin injectors. | These disposable sterile insulin pen<br>needles are intended for subcutaneous<br>injection of insulin in the treatment of<br>diabetes. |                               |  |  |  |  |  |  |  |  |                                                                                                                       |  |  |  |  |  |  |  |  |  |  |
| Product name                     |  | Hypodermic single lumen needle                                                                                                                                                                   | Hypodermic single lumen needle                                                                                                         |                               |  |  |  |  |  |  |  |  |                                                                                                                       |  |  |  |  |  |  |  |  |  |  |
| Trade name                       |  | Sungshim Insulin Pen Needle                                                                                                                                                                      | Feel Fine Insulin Pen Needle                                                                                                           |                               |  |  |  |  |  |  |  |  |                                                                                                                       |  |  |  |  |  |  |  |  |  |  |
| Model/type                       |  | 20 model codes including 29Gx4mm                                                                                                                                                                 | Feel Fine Insulin Pen Needle-29<br>Feel Fine Insulin Pen Needle-30<br>Feel Fine Insulin Pen Needle-31                                  |                               |  |  |  |  |  |  |  |  |                                                                                                                       |  |  |  |  |  |  |  |  |  |  |
| Appearance                       |  | Image: Sungshim needle                                                                                                                                                                           | Image: Feeltech needle                                                                                                                 |                               |  |  |  |  |  |  |  |  |                                                                                                                       |  |  |  |  |  |  |  |  |  |  |
| Product<br>configuration         |  | Hub<br>Needle cap<br>Primary container<br>Needle<br>Sterile paper                                                                                                                                | Hub<br>Needle cap<br>Sterile cap<br>Needle<br>Sterile paper                                                                            |                               |  |  |  |  |  |  |  |  |                                                                                                                       |  |  |  |  |  |  |  |  |  |  |
| Material                         |  | Part Material   Sterile cap Polypropylene   Needle cap Polyethylene   Needle Stainless steel 304   Hub Polypropylene                                                                             |                                                                                                                                        |                               |  |  |  |  |  |  |  |  | Part Material   Sterile cap Polypropylene   Needle cap Polypropylene   Needle Stainless steel 304   Hub Polypropylene |  |  |  |  |  |  |  |  |  |  |
| Length of parts (mm)             |  |                                                                                                                                                                                                  |                                                                                                                                        |                               |  |  |  |  |  |  |  |  |                                                                                                                       |  |  |  |  |  |  |  |  |  |  |
| Total length of<br>assembly      |  | 29.8                                                                                                                                                                                             | 28.8                                                                                                                                   |                               |  |  |  |  |  |  |  |  |                                                                                                                       |  |  |  |  |  |  |  |  |  |  |
| Side length of<br>assembly(max)  |  | 15.2                                                                                                                                                                                             | 15.5                                                                                                                                   |                               |  |  |  |  |  |  |  |  |                                                                                                                       |  |  |  |  |  |  |  |  |  |  |
| Inner diameter of<br>sterile cap |  | 11.2                                                                                                                                                                                             | 12.95                                                                                                                                  |                               |  |  |  |  |  |  |  |  |                                                                                                                       |  |  |  |  |  |  |  |  |  |  |
| Gauge                            |  | 29G, 30G, 31G, 32G                                                                                                                                                                               | 29, 30, 31G                                                                                                                            |                               |  |  |  |  |  |  |  |  |                                                                                                                       |  |  |  |  |  |  |  |  |  |  |
| Length of needle                 |  | 4 mm, 5 mm, 6 mm, 8 mm, 13 mm                                                                                                                                                                    | 5 mm, 8 mm, 12.7 mm                                                                                                                    |                               |  |  |  |  |  |  |  |  |                                                                                                                       |  |  |  |  |  |  |  |  |  |  |
| Sterilization                    |  | EO Gas sterilization                                                                                                                                                                             | EO Gas sterilization                                                                                                                   |                               |  |  |  |  |  |  |  |  |                                                                                                                       |  |  |  |  |  |  |  |  |  |  |
| Packagine                        |  | Sterile cap(PP)+sterile paper                                                                                                                                                                    |                                                                                                                                        | Sterile cap(PP)+sterile paper |  |  |  |  |  |  |  |  |                                                                                                                       |  |  |  |  |  |  |  |  |  |  |

Although the Indications for Use is not identical to that of the predicate device, it does not change the intended use because both are single use sterile needles intended for subcutaneous injection of insulin for diabetes patients.

### 9. Conclusion:

Comparison results demonstrate that the specifications and performance of the device are substantially equivalent to the legally marketed predicate device.

Therefore, it is concluded that Sungshim insulin pen needle is substantially equivalent to the legally marketed predicate device.

---

**Source:** [https://fda.innolitics.com/device/K152334](https://fda.innolitics.com/device/K152334)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
