K152301 · Natus Medical Incorporated Dba Excel-Tech Ltd. (Xltek) · OMA · Jun 3, 2016 · Neurology
Device Facts
Record ID
K152301
Device Name
Background Pattern Classification [BPc(TM)]
Applicant
Natus Medical Incorporated Dba Excel-Tech Ltd. (Xltek)
Product Code
OMA · Neurology
Decision Date
Jun 3, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Intended Use
The Background Pattern classification algorithm is intended for: Neonatal patients, defined as from birth to 28 days post-delivery, and corresponding to a post-conceptual age of 37 to 46 weeks, in clinical environments such as the intensive care unit, operating room, and for clinical research. To analyze and identify background patterns in aEEG, including continuous and discontinuous activity, burst suppression, low voltage, and inactive patterns. The aEEG must be obtained from a pair of parietal electrodes located at positions corresponding with P3 and P4 of the International 10/20 System. The output of the background pattern classification algorithm must be reviewed and interpreted by qualified clinical practitioners. The device does not provide any diagnostic conclusion about the patient's condition.
Device Story
BPc™ is a software-only tool for neonatal aEEG analysis. It processes aEEG signals derived from two parietal electrodes (P3-P4). The algorithm applies a set of rules to classify background patterns into five categories: Continuous, Discontinuous, Burst-suppression, Low voltage, and Inactive/flat. The software marks regions of the aEEG trace with the identified pattern. It is used in clinical environments like the NICU or OR. The output is presented to a qualified clinician who reviews, accepts, or discards the automated markings. The device does not provide diagnostic conclusions or treatment recommendations; it serves as an aid to clinical interpretation. By automating the identification of background patterns, it assists clinicians in monitoring the neurological state of neonates.
Clinical Evidence
Clinical validation compared BPc™ algorithm performance against a gold standard (panel of 3 board-certified neurophysiologists). Study included 64 neonates (36 female/28 male, mean gestational age 39.3 weeks). Primary endpoints were Positive Percent Agreement (PPA) and False Detection Rate (FDR). Overall PPA was 77% (95% CI: 72-82) and overall FDR was 2.5 FD/h (95% CI: 1.6-3.5). Performance was shown to be equivalent to the expert panel.
Technological Characteristics
Software-only algorithm for aEEG pattern classification. Operates on processed EEG signals from two parietal electrodes (P3-P4). Rule-based classification logic. Compliant with ISO 14971, ISO 13485, and IEC 62304 standards.
Indications for Use
Indicated for neonatal patients (birth to 28 days, 37-46 weeks post-conceptual age) in clinical settings (NICU, OR, research) to analyze and identify aEEG background patterns (continuous, discontinuous, burst suppression, low voltage, inactive). Requires review by qualified practitioners.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
K161027 — Cadwell AmpliScan · Cadwell Industries, Inc. · Nov 8, 2016
K120485 — NIHON KOHDEN QP-160 AK TREND PROGRAM WITH THE ADDITION OF DSA ASYMMETRY TREND AND FFT P · Nihon Kohden America, Inc. · Mar 16, 2012
K252070 — Ceribell Infant Seizure Detection Software · Ceribell, Inc. · Nov 21, 2025
K231068 — autoSCORE · Holberg Eeg AS · Jan 7, 2024
K163644 — Nihon Kohden QP-160AK EEG Trend Program · Nihon Kohden Corporation · May 19, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 3, 2016
Natus Medical Incorporated DBA Excel-tech Ltd. (Xltek) Sanjay Mehta Senior Manager Quality and Regulatory Affairs 2568 Bristol Circle Oakville, CA L6H5S1
Re: K152301
Trade/Device Name: Background Pattern Classification (BPc 114) Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OMA, ORT Dated: April 22, 2016 Received: May 5, 2016
Dear Sanjay Mehta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J.Hoffmann -A
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152301
Device Name
Background Pattern Classification
#### Indications for Use (Describe)
The Background Pattern Classification algorithm is intended for:
· Neonatal patients, defined as from birth to 28 days post-delivery, and corresponding to a post-conceptual age of 37 to 46 weeks, in clinical environments such as the intensive care unit, operating room, and for clinical research.
• To analyze and identify background patterns in aEG, including continuous and discontinuous activity, burst suppression, low voltage, and inactive patterns. The aEEG must be obtained from a pair of parietal electrodes located at positions corresponding with P3 and P4 of the International 10/20 System. The background pattern classification algorithm must be reviewed and interpreted by qualified clinical practitioners.
The device does not provide any diagnostic conclusion about the patient's condition.
Type of Use (Select one or both, as applicable)
| <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------|
| <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "xltek, a division of natus". The logo features a stylized human figure in shades of green and blue, followed by the company name in a teal sans-serif font. Below the name, in a smaller, darker font, is the text "a division of natus".
# 510K Summarv
Date: June 3rd ., 2016
Submitted by: Natus Medical Incorporated DBA Excel-Tech Ltd. (XLTEK) 2568 Bristol Circle Oakville, Ontario Canada L6H5S1
#### Contact Person:
Sanjay Mehta Senior Manager, Quality Assurance and Regulatory Affairs Natus Medical Incorporated Tel.: (905) 829-5300 ext 388 Fax .: (905) 829-5304 E-mail: Sanjay.mehta@natus.com
Propietary Name : Background Pattern Classification Algorithm (BPc™)
Common Name: aEEG software
Classification Name (Number): Amplitude Integrated Electroencephalograph (882.1400), Burst Suppression Detection Software for Electroencephalograph( 882.1400).
Product code: OMA; ORT Device Class: II
Predicate Devices: QP-160AK Trend program (K092573)
## Description
BPc™ is a software only product that identifies background patterns seen on aEEG signal recorded from a pair of parietal electrodes (P3-P4) in neonates, defined as from birth to 28 days post-delivery, and corresponding to a post-conceptual age of 37 to 46 weeks. The classification of aEEG background pattern into one of five different classes is done in accordance with the scoring scheme described in the following table:
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for "xltek, a division of natus". The logo features a stylized, abstract symbol in shades of green and blue, resembling a stylized flower or star. To the right of the symbol is the word "xltek" in a sans-serif font, also in shades of blue and green. Below "xltek" is the phrase "a division of natus" in a smaller, sans-serif font.
## Classification of aEEG Patterns in Term Neonates
Describes the dominating type of electrocortical activity in the aEEG trace.
- 1. Continuous (C): Continuous activity with lower (minimum) amplitude around (5 to) 7 to 10 µV and maximum amplitude of 10 to 25 (to 50) µV.
- 2. Discontinuous (DC): Discontinuous background with minimum amplitude variable, but below 5 µV, and maximum amplitude above 10 µV.
- 3. Burst-suppression (BSA): Discontinuous background with minimum amplitude without variability at 0 to 1 (2) µV and bursts with amplitude >25 µV. BS+ denotes burst density >100 bursts/h, and BS- means burst density <100 bursts/h.
- 4. Low v oltage (LV): Continuous background pattern of very low voltage (around or below 5 µV).
- 5. Inactive, flat (FT): Primarily inactive (isoelectric tracing) background below 5 uV.
Similar to basic EEG interpretation, pattern recognition forms the basis of aEEG interpretation. The classification scheme takes in consideration variations in the amplitude for the lower and upper margin of the aEEG signal. The BPc™ algorithm applies a set of rules to estimate the background pattern based on upper and lower margins of the aEEG signal.
The output of the device consists in marked regions with the corresponding background pattern name and a list of detected patterns in the signal. These detections (marked regions) are then reviewed. accepted or discarded by the qualified medical practitioner. The software does not make any final decisions that result in any automatic diagnosis or treatment. None of the components of the device is responsible for data acquisition, review or any other function different from analysis.
Image /page/4/Figure/11 description: The image shows a screenshot of a medical device interface, specifically a CFM (Cerebral Function Monitor). The interface displays aEEG (amplitude-integrated EEG) data over time, with annotations indicating different background patterns such as 'Continuous' and 'Discontinuous'. The interface also includes controls for navigating the data, adding annotations, and classifying background patterns. The time is 9:45:45 AM and the source is Bedside, Automatic.
BPc™ Output as presented to the end-user (top panel), BPc " edit scoring overlay (lower panel)
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "xltek, a division of natus". The logo features a stylized, abstract graphic to the left of the text "xltek". The graphic is composed of several geometric shapes in shades of green and blue, arranged to resemble a stylized flower or star. Below "xltek" is the text "a division of natus" in a smaller, sans-serif font.
## Indications for Use
The Background Pattern classification algorithm is intended for:
- Neonatal patients, defined as from birth to 28 days post-delivery, and corresponding to a post-conceptual age of 37 to 46 weeks, in clinical environments such as the intensive care unit, operating room, and for clinical research.
- To analyze and identify background patterns in aEEG, including continuous and discontinuous activity, burst suppression, low voltage, and inactive patterns. The aEEG must be obtained from a pair of parietal electrodes located at positions corresponding with P3 and P4 of the International 10/20 System. The output of the background pattern classification algorithm must be reviewed and interpreted by qualified clinical practitioners.
The device does not provide any diagnostic conclusion about the patient's condition.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for "xltek", which is written in a sans-serif font in a light blue color. Above the text is a stylized graphic of a person with their arms raised, with the graphic being in shades of green and blue. Below the text is the phrase "a division of natus" in a smaller, sans-serif font.
# Predicate Comparison
The substantial equivalence of the BPc™ algorithm is based on its similarities to the cleared QP-160AK Trend program (K092573).
| Device Feature | Subject Device<br>BPc™ Algorithm | Predicate<br>QP-160AK EEG Trend<br>Program (K092573) | Comparison |
|---------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Device Class | Class II | Class II | Same |
| Common Name | Amplitude Integrated<br>electroencephalograph | Amplitude-integrated<br>electroencephalograph | Same |
| Intended Use | | | |
| Purpose and function: | aEEG monitoring | EEG/aEEG monitoring | Different. Subject<br>device only works on<br>aEEG recordings. No<br>safety/effectiveness<br>concern. |
| Patient population | Neonatal | Neonatal | Same |
| Environment of use | Clinical environments<br>(NICU, research) | Clinical environments<br>(NICU, research) | Same |
| Intended User | qualified medical<br>practitioners | qualified medical<br>practitioners | Same |
| Signal Processing | | | |
| Input Signal | aEEG | EEG | Different. Subject<br>device only works on<br>aEEG recordings. No<br>safety/effectiveness<br>concern. |
| Number of Electrodes<br>and location | 2 electrodes located<br>according at P3-P4 of<br>the International 10-20<br>System | ≥16 electrodes located<br>according to the<br>International 10-20<br>System | Different. (see<br>Discussion) |
| Environment of use: | Clinical environments<br>(NICU, research) | Clinical environments<br>(NICU, research) | Same |
| Parameters and Performance | | | |
| Burst-Suppression | Yes | Yes | Same |
| Additional background<br>patterns (i.e<br>Continous,<br>discontinuous, flat<br>trace) | Yes | No | Different. (see<br>Discussion) |
| Device Feature | Subject Device<br>BPcTM Algorithm | Predicate<br>QP-160AK EEG Trend<br>Program (K092573) | Comparison |
| PPA % (bootstrap CI) | C- 86 (77 - 94)<br>D- 64 (51 - 77)<br>BS- 89 (78 - 99)<br>LV- 66 (50 - 83)<br>FT- 80 (63 - 96)<br>Overall- 77 (72 – 82) | Performance data Not<br>Available | Unknown |
| FDR (false detection<br>rate, False<br>Positive/hour) | C- 0.3<br>D- 0.1<br>BS- 4.4<br>LV- 4.2<br>FT- 4.2<br>Overall- 2.5 | Performance data Not<br>Available | Unknown |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for "xltek, a division of natus". The logo features a stylized human figure with outstretched arms above the text "xltek". The text "a division of natus" is in a smaller font size and is located below the word "xltek".
Continuous(C)
Discontinuous(D) Burst suppression (BS) Low voltage (LV) Inactive, flat (FT)
## Discussion
Both devices are intended for aEEG monitoring in the neonatal population and to be used in same clinical environments. The predicate device however, has the addition of EEG. Even though EEG analysis is not a feature of the subject device, we believe this to have no impact on the safety and effectiveness of the subject device when used as labeled. Moreover, aEEG alone (as in the subject device) has been long established by the scientific community and accepted by the American Academy of Neurology (Shellhaas et al 2011) as a safe and effective monitoring tool for the neonatal population. In line with current medical practice and scientific evidence, the FDA has long determined product codes and cleared for market aEEG only devices (i.e FDA product code OMA, K071449, K031149). Hence, this difference does not affect the safety and effectiveness of the subject device as compared to the predicate.
The difference on input signal is also related to the aforementioned difference on signal analysis. While the predicate device input signal is the raw EEG the subject device uses the aEEG only. aEEG is a form of processed EEG signal, that is, aEEG is derived after collection of the raw EEG data, in this regard the subject device analyzes processed EEG. Even though the predicate device uses the raw EEG data, it actually requires some type of EEG processing in order to carry its intended function. Both devices use processed EEG as part of their function. Detection of background patterns on raw EEG versus aEEG is roughly equivalent (Toet et al 2002) although the difference on input signal dictates the method used on the detection algorithm. Safety and effectiveness of the subject device for the detection of background pattern has been established through clinical testing, that is comparison of device performance versus the Gold standard in clinical practice therefore we believe this difference does not affect the safety and effectiveness of the subject device as compared to the predicate.
The predicate device allows recording of aEEG from 16 channels or more while the subject device only uses the aEEG form two electrode locations (P3-P4). The restricted number of electrodes might
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the text 'K152301: BPC Algorithm 4th. SUPPLEMENT PAGE 6 of 10'. The text is in bold and is centered on the page. The text is likely a header or title for a document.
Image /page/8/Picture/1 description: The image shows the logo for "xltek, a division of natus". The logo features a stylized human figure in light green and blue, with the word "xltek" in teal next to it. Below the word "xltek" is the phrase "a division of natus" in a smaller, sans-serif font.
only provide a snapshot into the neonate's neurological state, however the background pattern classification based on P3-P4 has been previously established as valid, safe and effective given the intended use of the device (Spitzmiller et al 2007, al Nageeb et al 1999). In acknowledgment of these facts FDA has previously cleared similar devices where only biparietal electrodes were available to monitor the state of the brain (K791580, K983229, K031149). In addition the background classification scheme adopted by the subject device is that accepted by the medical community and is based only on P3-P4 electrodes. Hence, the difference in the number of electrodes between the subject and predicate do not raise new concerns of safety/effectiveness for the subject device.
Performance data for the predicate device was not available aty the time of this submission. Instead of comparing performance to that of the predicate we decided to established performance of the BPc™ algorithm on its own merit as compared to the gold standard of care, which is performance of a panel of 3 medical experts carrying out the same task, Based on the results of the Clinical Validation we believe that subject device performance is equivalent to that of the gold standard (i.e medical experts). Furthermore, the subject device intended use and accompanying labeling clearly restrict the use of the device to qualified clinical practitioners who will "review(ed)" the output of the subject device. Therefore, provided that all clinical results are available to the users plus the set restrictions for device use are in place, we believe that the use of the device as intended is safe and effective and equivalent to the predicate.
## IFU Comparison and Discussion
Note: Highlighted in GREEN are the components we claim equivalence to. In GRAY are components to which we do not claim equivalence.
#### Predicate device IFU
The QP-160AK Trend program is a software only device intended to be installed on the EEG-1200A series electroencephalograph to record, calculate, and display EEG data obtained from the EEG-1200A system. This device is intended to be used by qualified medical practitioners, trained in Electroencephalography, who will exercise professional judgment when using the information.
The intended use for each of this software's output is as follows:
- The EEG and aEEG waveforms are intended to help the user monitor the state of the brain.
- The user defined Fast Fourier Transform (FFT) parameters of this software (FFT power) are intended to help the user analyze the EEG waveform.
- The burst suppression parameters of this software (interburstinterval and busrts per minute) are intended to aid in the identification and characterization of areas of burst suppression pattern in the EEG.
This device does not provide any diagnostic conclusion about the patient's condition to the user.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the logo for "xltek, a division of natus". The logo features a stylized, abstract symbol to the left of the text "xltek". Below "xltek" is the text "a division of natus" in a smaller font size. The colors used in the logo are primarily shades of blue and green.
## Subject device IFU
The Background Pattern classification algorithm is intended for:
- Neonatal patients, defined as from birth to 28 days post-delivery, and corresponding to a post-conceptual age of 37 to 46 weeks, in clinical environments such as the intensive care unit, operating room, and for clinical research.
- To analyze and identify background patterns in aEEG. including continuous and discontinuous activity, burst suppression, low voltage, and inactive patterns. The aEEG must be obtained from a pair of parietal electrodes located at positions corresponding with P3 and P4 of the International 10/20 System. The output of the background pattern classification algorithm must be reviewed and interpreted by qualified clinical practitioners.
The device does not provide any diagnostic conclusion about the patient's condition.
#### We claim equivalence to the aEEG monitoring and to the identification of burst suppression pattern.
Similarities:
- Both are software only.
- · Both analyze brain electrical activity
- · Both are used to monitor brain electrical activity.
- · Both detect background patterns of the brain electrical activity.
- · Both are meant to be used by qualified medical practitioners.
- · None of the devices provide diagnostic conclusions.
#### Differences:
- Predicate uses aEEG and EEG.
- · Predicate detects Burst suppression areas only and derives numeric parameters from the detected areas.
- · Subject device detects background patterns other than burst-suppression.
## IFU differences discussion
- Predicate uses aEEG and EEG.
The predicate uses aEEG and EEG. EEG analysis is not a feature of the subject device. aEEG is a form of processed EEG signal, that is, aEEG is derived from raw EEG data, in this regard the subject device analyzes processed EEG. Even though the predicate device uses the raw EEG data, it actually requires EEG processing in order to carry its intended function (i.e to derive parameters that will help the "identification and characterization of areas of burst suppression pattern"). On this regard we claim that both devices use processed EEG as part of their function. Detection of background patterns on raw EEG versus aEEG is equivalent although the difference on input signal dictates the method used for the detection algorithm. Safety and effectiveness of the subject device for the detection of
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for "xltek, a division of natus". The logo features a stylized human figure with arms raised above its head. The figure is colored in shades of blue and green. The text "xltek" is written in a sans-serif font to the right of the figure. Below the text is the phrase "a division of natus" in a smaller font.
background pattern has been established through clinical testing, that is comparison of device performance versus the Gold standard in clinical practice. Moreover, aEEG alone (as in the subject device) has been long established by the scientific community and accepted by the American Academy of Neurology (Shellhaas et al 2011) as a safe and effective monitoring tool. In line with current medical practice and scientific evidence, the FDA has long determined product codes and cleared for market aEEG only devices (i.e FDA product code OMA, K071449, K031149). We then believe that this difference has no impact on the safety and effectiveness of the subject device when used as labeled.
- · Predicate detects Burst suppression areas only and derives numeric parameters from the detected areas.
The subject device, as the predicate, detects Burst-suppression areas, however it does not derives any numeric parameter from this detection of burst-suppression areas (carried by both devices) precedes the parameter calculation and the effectiveness of the subject device in the detection of burst-suppression areas was established trough clinical testing and shown to be equivalent to the gold standard of clinical practice. Hence, we believe that this feature of the subject device to be equivalent to the predicate. As to the derivation of numeric parameters (as in the predicate) it requires additional computation beyond that required for detection only. This additional step does not only increase the probability of errors of the predicate device but also requires that the derived values are properly validated trough clinical studies. We have no information as to the accuracy of the derived parameters on the predicate device, and given that the subject device refrains for any additional calculations we saw no need for such comparison in our clinical study. Calculation of (interburst interval and bursts per minute could also be done using the subject device but in our case that remains the responsibility of the qualified practitioner who has to do it manually at his own discretion. We therefore believe that this difference between devices has no impact on safety and effectiveness for the subject device when used as labeled.
## · Subject device detects background patterns other than burst-suppression.
In addition to detection of Burst-suppression areas (as in the predicate) the subject device detects other types of background patterns. Detection of such other areas is carried based on general rules that includes and go beyond the burst-suppression. These rules were long established by the scientific community and the effective application of those rules on our device performance were established through clinical testing and shown to be equivalent to the gold standard of clinical practice. In addition the intended user of the device is informed in detailed of the device performance characteristics and limitations. We therefore believe that this difference between the subject and predicate device has been properly addressed and related risks mitigated raising no new questions about the safety and effectiveness of the subject device.
## Conclusion
Based on the rationale discussed above we believe that; in spite of the differences in technological characteristics between the subject and the predicate device, the use of the BPc" algorithm is safe and effective for the intended use and substantially equivalent to the predicate.
{11}------------------------------------------------
## Brief Summary of Non-Clinical and Clinical Performance Tests
All functionalities and performance of the Background Pattern Classification (BPc™) Algorithm have been verified and validated through bench and clinical performance tests according to the intended use and user of the device.
Non-Clinical: The BPc™ device is compliant with all currently accepted safety standards for medical devices of its class which was demonstrated through testing, verification of all components.
- 21 CFR part 820 Quality System Requirements .
- Canadian Medical Device Regulations ●
- ISO 14971:2000, Medical Devices Application of Risk Management to Medical Devices ●
- ISO13485: 2003, Medical devices Quality management systems Requirements for ● regulatory purposes.
- IEC 62304:2006. Medical Device Software. Software Life cycle processes. ●
Clinical: Natus conducted an extensive clinical test to: 1) Evaluate the positive percent agreement (i.e., detection sensitivity) and false detection rate of the BPc" algorithm, and to 2) Demonstrate equivalence of the performance, in terms of positive percent and false detection rates, of the BPc™ algorithm as compared to the gold standard, that is, background pattern as classified by a panel of 3 EEG board certified medical professionals.
## BPc™ Clinical Validation
#### Dataset Description:
| Gestational age at<br>birth (Mean ± SD) | 39.3 (± 1.9) |
|-----------------------------------------|--------------|
| GENDER<br>(Female/Male) | 36/28 |
#### Analysis Method
EEG studies were de-identified, randomized and provided to board certified neurophysiologists that independently, blindly and manually marked background pattern states according to the classification scheme detailed below (see table) in the same manner they would normally do in clinical practice.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for "xltek, a division of natus". The logo features a stylized, abstract symbol in shades of blue and green, resembling a person with outstretched arms or a stylized flower. The text "xltek" is written in a sans-serif font, with the "x" in green and the rest of the letters in blue. Below "xltek" is the text "a division of natus" in a smaller, sans-serif font.
#### Classification of aEEG Patterns in Term Neonates.
Describes the dominating type of electrocortical activity in the aEEG trace.
- 1. Continuous (C): Continuous activity with lower (minimum) amplitude around (5 to) 7 to 10 µV and maximum amplitude of 10 to 25 (to 50) µV.
- 2. Discontinuous (D): Discontinuous background with minimum amplitude variable, but below 5 µV, and maximum amplitude above 10 µV.
- 3. Burst-suppression (BS): Discontinuous background with minimum amplitude without variability at 0 to 1 (2) µV and bursts with amplitude >25 µV. BS+ denotes burst density >100 bursts/h, and BS- means burst density <100 bursts/h.
- 4. Low v oltage (LV): Continuous background pattern of very low voltage (around or below 5 µV).
- 5. Inactive, flat (FT): Primarily inactive (isoelectric tracing) background below 5 µV.
Recordings were also independently submitted for analysis using the BPc 100 algorithm.
# Results
#### Inter Rater Performance
Inter-rater Positive Percent Agreement and False Detection / hour
| | Rev1 (vs23) | | Rev2 (vs13) | | Rev3 (vs12) | |
|---------|---------------|---------------|---------------|---------------|---------------|---------------|
| | PPA (%) | FDR<br>(FD/h) | PPA (%) | FDR<br>(FD/h) | PPA (%) | FDR<br>(FD/h) |
| C | 80 (70 - 90)* | 1.2 | 68 (51 - 88) | 0.7 | 96 (91 - 100) | 5.5 |
| D | 73 (62 - 85) | 3.3 | 67 (50 - 83) | 5.5 | 46 (32 - 60) | 4.4 |
| BS | 79 (61 - 93) | 1.6 | 80 (63 - 100) | 4.6 | 43 (24 - 61) | 0.0 |
| LV | 68 (39 - 96) | 3.0 | 67 (39 - 94) | 1.5 | 90 (67 - 100) | 5.6 |
| FT | 92 (72 - 100) | 2.0 | 79 (55 - 100) | 0.0 | 91 (70 - 100) | 1.9 |
| Overall | 78 (71 - 84) | 2.3 | 71 (63 - 80) | 3.2 | 66 (59 - 74) | 3.5 |
*Bootstrap 95% Cl
#### Algorithm Performance Comparison
| | PPA (%) | FDR<br>(FD/h) |
|---------|---------------|------------------|
| C | 86 (77 - 94)* | 0.3 (0.1 - 0.7)* |
| D | 64 (51 - 77) | 0.1 (0.1 - 0.3) |
| BS | 89 (78 - 99) | 4.4 (1.5 - 5.0) |
| LV | 66 (50 - 83) | 4.2 (2.3 - 4.8) |
| FT | 80 (63 - 96) | 4.2 (1.2 - 4.8) |
| Overall | 77 (72 - 82) | 2.5 (1.6 - 3.5) |
*Bootstrap 95% Cl
## Conclusions
Based on the results of the clinical and non-clinical testing we have found BPc™ algorithm to be substantially equivalent to the predicate and safe and effective for its intended use.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.