← Product Code [KCT](/productcode/KCT) · K152241

# Medtronic Transportation/Sterilization Cassettes (K152241)

_Medtronic Sofamor Danek USA, Inc. · KCT · Jan 20, 2016 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K152241

## Device Facts

- **Applicant:** Medtronic Sofamor Danek USA, Inc.
- **Product Code:** [KCT](/productcode/KCT.md)
- **Decision Date:** Jan 20, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6850
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Medtronic Transportation/Sterilization Cassettes are intended for use in healthcare facilities to organize, enclose, sterilize, transport, and store medical devices and other instrumentation between surgical and other medical uses. The Medtronic Transportation/Sterilization Cassettes are not intended on their own to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterilization wrap.

## Device Story

Medtronic Transportation/Sterilization Cassettes are reusable containers designed to organize, protect, and facilitate the steam sterilization of orthopedic and neurological surgical instruments. The system consists of perforated bases, lids, trays, caddies, and locking latches. Used in healthcare facilities, the cassettes are loaded with instruments, wrapped in an FDA-cleared sterilization wrap, and processed in standard hospital autoclaves. The perforated design allows for steam penetration and air removal during sterilization cycles (gravity displacement or dynamic-air-removal). After sterilization, the cassettes protect the instruments during transport and storage. The device does not have direct patient contact; it serves as a protective housing for surgical tools, ensuring they remain organized and accessible for clinical use. By providing a standardized, validated method for reprocessing, the cassettes help maintain the integrity of surgical instruments, ultimately supporting safe and efficient surgical workflows.

## Clinical Evidence

Bench testing only. No clinical data. Performance testing included sterilization validation (achieving SAL 10^-6) per AAMI/ISO standards (e.g., ANSI/AAMI ST79, ISO 17665-1), distribution testing per ASTM D4169, and biocompatibility assessment per ISO 10993-1. Validations confirmed effectiveness for worst-case configurations (28.4 lbs) and lumen dimensions.

## Technological Characteristics

Materials: Intrinsically stable metals (aluminum, stainless steel) and thermoplastic polymers (nylon, polypropylene). Design: Perforated bases, lids, trays, and caddies with locking latches. Sterilization: Steam (Gravity Displacement and Dynamic-Air-Removal). Connectivity: None. Dimensions: Various, up to 22.75 x 11.26 x 5.51 inches. Standards: ASTM D4169, AAMI TIR39, AAMI TIR30, AAMI TIR17, AAMI TIR12, ANSI/AAMI ST81, ANSI/AAMI ST79, ISO 17665-1/2, ISO 17664, ISO 14161, ISO 11737-2, ISO 10993-1.

## Regulatory Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

## Predicate Devices

- Restore Modular Sterilization Tray System ([K131455](/device/K131455.md))
- Olympus Sterilization Trays ([K033222](/device/K033222.md))
- Metapak Multi-Purpose Instrument Tray ([K993535](/device/K993535.md))
- THS Sterilization Tray ([K120947](/device/K120947.md))

## Reference Devices

- Synthes Reusable Sterilization Container Systems ([K130720](/device/K130720.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure, represented by three overlapping profiles, suggesting a sense of community and care. The logo is presented in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 20, 2016

Medtronic Sofamor Danek Ms. Victoria Scheitlin Regulatory Affairs Specialist 1800 Pyramid Place Memphis. Tennessee 38132

Re: K152241

Trade/Device Name: Medtronic Transportation/Sterilization Cassettes Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization wrap Regulatory Class: Class II Product Code: KCT Dated: December 18, 2015 Received: December 21, 2015

Dear Ms. Scheitlin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{1}------------------------------------------------

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

*Tejashri Purohit-Sheth, M.D.*

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K152241

Device Name Medtronic Transportation/Sterilization Cassettes

### Indications for Use (Describe)

The Medtronic Transportation/Sterilization Cassettes are intended for use in healthcare facilities to organize, enclose, sterilize, transport, and store medical devices and other instrumentation between surgical and other medical uses. The Medtronic Transportation/Sterilization Cassettes are not intended on their own to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterilization wrap.

Sterilization validations for the worst case Medtronic Transportation Cassette (22.75 x 11.26 x 5.5 inches) included implants and common surgical instruments such as rasps, drivers, trials, handles, inserters, probes, drills, etc. The validated total weight was 28.4lbs. The validated worst case loading configurations of the Medtronic Transportation/ Sterilization Cassette included the following worst case lumen dimensions.

- 363 x 1.575mm - 247.5 x 4.1mm

Device List:

1850060 Case - Triple Generic Outer Base (22.74 x 11.260 x 5.040 inches) 1850064 Lid - Generic Outer Lid (22.75 x 11.260 x 0.470 inches) 7022101L Tray Lid (21 x 10.13 x 0.075 inches) P1850061 Tray 1 (20.75 x 9.79 x 1.32 inches) P1850062 Tray 2 (21 x 10.13 x 1.69 inches) P1850063 Tray 3 (21 x 10.13 x 1.38 inches) 7059532 Large Caddy (9.47 x 6.37 x 1.3 inches) 7059532L Large Lid (5.85 x 4.725 x 0.095 inches) P9213018 Small Caddy (2 x 1.5 x 1.025 inches) P9213018 Small Lid (2 x 1.29 x 0.095 inches)

Sterilization Parameters:

| Cycle                                              | Temperature   | Exposure time | Minimum dry time |
|----------------------------------------------------|---------------|---------------|------------------|
| Gravity Displacement                               | 250°F (121°C) | 30 Minutes    | 30 Minutes       |
| Gravity Displacement                               | 270°F (132°C) | 15 Minutes    | 30 Minutes       |
| Gravity Displacement                               | 275°F (135°C) | 10 Minutes    | 30 Minutes       |
| Dynamic-Air-Removal<br>(4 Pre-conditioning pulses) | 270°F (132°C) | 4 Minutes     | 30 Minutes       |
| Dynamic-Air-Removal<br>(4 Pre-conditioning pulses) | 275°F (135°C) | 3 Minutes     | 30 Minutes       |

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

## 510(k) SUMMARY

MEDTRONIC Sofamor Danek Medtronic Transportation/Sterilization Cassettes January 19, 2015

| Company:             | Medtronic Sofamor Danek<br>1800 Pyramid Place<br>Memphis, Tennessee 38132<br>(800) 876-3133                                                                                                                                                                                                    |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:             | Victoria Scheitlin<br>Regulatory Affairs Specialist<br>Telephone: (901) 344-0706                                                                                                                                                                                                               |
| Date Prepared:       | Fax: (901) 346-9738<br>January 19, 2015                                                                                                                                                                                                                                                        |
| Device:              |                                                                                                                                                                                                                                                                                                |
| Name of Device:      | Medtronic Transportation/Sterilization Cassettes                                                                                                                                                                                                                                               |
| 510(k) Number:       | K152241                                                                                                                                                                                                                                                                                        |
| Classification Name: | Sterilization Wrap Containers, Trays, Cassettes & Other Accessories (21 CFR 880.6850)                                                                                                                                                                                                          |
| Class:               | II                                                                                                                                                                                                                                                                                             |
| Product Code:        | KCT                                                                                                                                                                                                                                                                                            |
| Predicate Devices:   | <ul><li>K131455 (Primary Predicate) – Restore Modular Sterilization Tray System (S.E. 08/30/2013)</li><li>K033222 – Olympus Sterilization Trays (S.E. 08/02/2004)</li><li>K993535 – Metapak Multi-Purpose Instrument Tray (S.E. 12/07/2001)</li><li>K120947 – THS Sterilization Tray</li></ul> |

- K120947 THS Sterilization Tray . (S.E. 10/17/2012)
Additionally, this submission contains a reference device K130720 -Synthes Reusable Sterilization Container Systems (S.E. 08/14/2014). The Synthes Reusable Sterilization Container Systems are fabricated from similar materials to those used in the Medtronic Transportation/Sterilization Cassettes, specifically nylon and polypropylene. The reference device also has a similar sterilization method, design configuration options and sizes. The predicates have not been subject to a design related recall that

Medtronic is aware.

#### IV. Description:

The Medtronic Transportation/Sterilization Cassettes include cases, trays, lids, caddies, modules, or brackets fabricated from a 

{5}------------------------------------------------

variety of materials commonly used to enclose, protect, and organize Medtronic orthopedic or neurological non-sterile devices which meet national or international specifications. The Medtronic Transportation/Sterilization Cassettes are not intended on their own to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterilization wrap. The Medtronic Transportation/Sterilization Cassettes design consists of multiple components designed to be integrated into a single unit, which protects the interior components during transportation, sterilization, and storage. These components consist of: an outer base, internal individual trays with or without lids, caddies and an outer locking lid. The components of the Medtronic Transportation/Sterilization Cassettes are fabricated from intrinsically stable metals and thermoplastic polymers. All of the components of the Medtronic Transportation/Sterilization Cassettes are perforated with an evenly distributed hole pattern, and are designed to be used for sterilization via steam sterilization. Since the Medtronic Transportation/Sterilization Cassettes are perforated, an FDA cleared wrap must be used for sterilization purposes and to maintain the sterility of the contents. The Medtronic Transportation/Sterilization Cassettes are designed to be used with standard autoclaves used in hospitals and healthcare facilities. Thus allowing the Medtronic Transportation/Sterilization Cassettes to be effective for sterilization and are designed such that they withstand the environment of repeated steam sterilization cycles.

#### V. Indications For Use:

The Medtronic Transportation/Sterilization Cassettes are intended for use in healthcare facilities to organize, enclose, sterilize, transport, and store medical devices and other instrumentation between surgical and other medical uses. The Medtronic Transportation/Sterilization Cassettes are not intended on their own to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterilization wrap.

Sterilization validations for the worst case Medtronic Transportation/Sterilization Cassette (22.75 x 11.26 x 5.5 inches) included implants and common surgical instruments such as rasps, drivers, trials, handles, inserters, probes, drills, etc. The validated total weight was 28.4lbs. The validated worst case loading configurations of the Medtronic Transportation/Sterilization Cassette included the following worst case lumen dimensions:

- 363 x 1.575mm - 247.5 x 4.1mm

Device List: 1850060 Case - Triple Generic Outer Base (22.74 x 11.260 x 5.040 inches) 1850064 Lid - Generic Outer Lid (22.75 x 11.260 x 0.470 inches)

{6}------------------------------------------------

7022101L Tray Lid (21 x 10.13 x 0.075 inches) P1850061 Tray 1 (20.75 x 9.79 x 1.32 inches) P1850062 Tray 2 (21 x 10.13 x 1.69 inches) P1850063 Tray 3 (21 x 10.13 x 1.38 inches) 7059532 Large Caddy (9.47 x 6.37 x 1.3 inches) 7059532L Large Lid (5.85 x 4.725 x 0.095 inches) P9213018 Small Caddy (2 x 1.5 x 1.025 inches) P9213018 Small Lid (2 x 1.29 x 0.095 inches)

Sterilization Parameters:

| Cycle                                                     | Temperature   | Exposure time | Minimum dry time |
|-----------------------------------------------------------|---------------|---------------|------------------|
| Gravity<br>Displacement                                   | 250°F (121°C) | 30 Minutes    | 30 Minutes       |
| Gravity<br>Displacement                                   | 270°F (132°C) | 15 Minutes    | 30 Minutes       |
| Gravity<br>Displacement                                   | 275°F (135°C) | 10 Minutes    | 30 Minutes       |
| Dynamic-Air-<br>Removal<br>(4 Pre-conditioning<br>pulses) | 270°F (132°C) | 4 Minutes     | 30 Minutes       |
| Dynamic-Air-<br>Removal<br>(4 Pre-conditioning<br>pulses) | 275°F (135°C) | 3 Minutes     | 30 Minutes       |

### VI. Comparison of Technological Characteristics and Performance Data with the Predicate Devices:

Table 2 and Table 3 display the comparison of the Medtronic Transportation/Sterilization Cassettes compared against the predicates. The comparison analysis was performed to support the substantial equivalence of the Medtronic Transportation/Sterilization Cassettes and trays to the listed predicates.

The Medtronic Transportation/Sterilization Cassettes were validated to be the greatest challenge (worst case). The Medtronic Transportation/Sterilization Cassettes described within this submission were determined to be the worst case configuration for sterilization and reprocessing due to the weight, percent perforation, and the number of trays and caddies included within the final configuration in the entirety.

{7}------------------------------------------------

| Feature |                                         | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | K131455<br>(Primary<br>Predicate)                                                                                                                                                                                                                                                                                                                                              | K033222                                                                                                                                                                                                                                                                                                                                                          | K993535                                                                                                                                                                                                                                                                                                                                                                  | K120947                                                                                                                                                                                                                                                                                                                                                                       |
|---------|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|         | 1. Trade Name                           | Medtronic Transportation/Sterilization Cassettes                                                                                                                                                                                                                                                                                                                                                                                                                                           | Restore™<br>Modular<br>Sterilization<br>Tray System                                                                                                                                                                                                                                                                                                                            | Olympus<br>Sterilization<br>Trays                                                                                                                                                                                                                                                                                                                                | Metapak<br>Multi-Purpose<br>Instrument<br>Tray                                                                                                                                                                                                                                                                                                                           | THS<br>Sterilization<br>Tray                                                                                                                                                                                                                                                                                                                                                  |
| 2.      | Fundamental<br>Scientific<br>Technology | Sterilization Cassette                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                                                                                          |
| 2.      | Intended Use                            | The Medtronic Transportation/Sterilization Cassettes<br>are intended for use in healthcare facilities to organize,<br>enclose, sterilize, transport, and store medical devices<br>and other instrumentation between surgical and other<br>medical uses. The Medtronic<br>Transportation/Sterilization Cassettes are not intended<br>on their own to maintain sterility; it is intended to be<br>used in conjunction with a legally marketed, validated,<br>FDA-cleared sterilization wrap. | The Restore™<br>Modular<br>Sterilization<br>Tray System is<br>intended for<br>use in<br>healthcare<br>facilities to<br>organize,<br>enclose,<br>sterilize,<br>transport, and<br>store medical<br>devices and<br>other<br>instrumentation<br>between<br>surgical and<br>other medical<br>uses when<br>used In<br>conjunction<br>with a FDA<br>cleared<br>sterilization<br>wrap. | The Olympus<br>sterilization<br>trays are<br>intended to<br>be used to<br>enclose<br>Olympus<br>medical<br>devices<br>including<br>hand<br>instruments,<br>trocars,<br>camera<br>heads,<br>adapter, and<br>endoscopes<br>to be<br>sterilized by a<br>health care<br>provider. It is<br>intended to<br>allow steam<br>sterilization of<br>the enclosed<br>medical | The Riley<br>Metapak<br>Multi-Purpose<br>Instrument<br>Tray is used<br>for loading<br>surgical<br>instruments in<br>order to<br>conveniently<br>organize,<br>sterilizes,<br>transports<br>and store the<br>instruments<br>between use.<br>The Riley<br>Metapak<br>Multi-Purpose<br>Instrument<br>Tray can be<br>used in pre-<br>vacuum<br>steam and<br>ethylene<br>oxide | The THS<br>Sterilization<br>Tray is used<br>to enclose,<br>protect, and<br>organize the<br>THS scopes,<br>diagnostic<br>sheath, and<br>associated<br>accessory<br>components,<br>and to<br>facilitate the<br>sterilization<br>process by<br>allowing<br>sterilant<br>penetration<br>and air<br>removal when<br>used in<br>conjunction<br>with an<br>approved<br>sterilization |
|         | Feature                                 | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | K131455<br>(Primary<br>Predicate)                                                                                                                                                                                                                                                                                                                                              | K033222                                                                                                                                                                                                                                                                                                                                                          | K993535                                                                                                                                                                                                                                                                                                                                                                  | K120947                                                                                                                                                                                                                                                                                                                                                                       |
|         |                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                | device.                                                                                                                                                                                                                                                                                                                                                          | sterilization                                                                                                                                                                                                                                                                                                                                                            | wrap.                                                                                                                                                                                                                                                                                                                                                                         |
| 3.      | Product Code                            | KCT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | KCT                                                                                                                                                                                                                                                                                                                                                                            | KCT                                                                                                                                                                                                                                                                                                                                                              | KCT                                                                                                                                                                                                                                                                                                                                                                      | KCT                                                                                                                                                                                                                                                                                                                                                                           |
| 4.      | Material<br>Composition                 | Thermoplastic polymers, aluminum, and stainless steel                                                                                                                                                                                                                                                                                                                                                                                                                                      | Thermoplastic<br>polymers,<br>aluminum, and<br>stainless steel                                                                                                                                                                                                                                                                                                                 | Thermoplasti<br>c polymers<br>and stainless<br>steel                                                                                                                                                                                                                                                                                                             | Thermoplasti<br>c polymers,<br>aluminum,<br>and stainless<br>steel                                                                                                                                                                                                                                                                                                       | Thermoplasti<br>c polymers,<br>aluminum,<br>and stainless<br>steel                                                                                                                                                                                                                                                                                                            |
| 5.      | Design                                  | A base, a lid with a locking latch, and individual inserts                                                                                                                                                                                                                                                                                                                                                                                                                                 | A base, a lid,<br>and individual<br>inserts                                                                                                                                                                                                                                                                                                                                    | A bottom, a<br>lid with<br>locking tabs,<br>and individual<br>inserts                                                                                                                                                                                                                                                                                            | A base, a lid<br>with latches,<br>and individual<br>inserts                                                                                                                                                                                                                                                                                                              | A base, a lid,<br>and individual<br>inserts                                                                                                                                                                                                                                                                                                                                   |
| 6.      | Dimensions                              | The greatest challenge dimension was assessed to be:<br>22.75 x 11.26 x 5.51 inches<br>The inserts are offered in different sizes                                                                                                                                                                                                                                                                                                                                                          | 10 x 20 x 4<br>inches or<br>a 10 x 10 x 4<br>inches tray with<br>a lid.<br>The inserts<br>are offered in<br>different sizes<br>(not specified<br>in summary.)                                                                                                                                                                                                                  | A wide variety<br>of dimensions<br>and<br>configurations<br>ranging from:<br>405 x 95 x<br>200mm (16 x<br>4 x 8 inches)<br>to<br>665 x 52 x<br>204mm (26 x<br>2 x 8 inches)                                                                                                                                                                                      | Bases and<br>lids are<br>offered in 3<br>sizes:<br>• 22 x 10 x 6<br>inches<br>• 22 x 10 x 4<br>inches<br>• 26 x 10 x 6<br>inches<br>Insert trays<br>are offered in<br>2 sizes:<br>• 21 x 9.5 x<br>2.5 inches<br>• 10 x 9.5 x<br>2.5 inches<br>The inserts<br>are offered<br>in different<br>sizes (not<br>specified in<br>summary.)                                      | The "base"<br>tray is<br>17.3 x 7.25 x<br>4 inches<br>The inserts<br>are offered in<br>different sizes<br>(not specified<br>in summary.)                                                                                                                                                                                                                                      |

Table 2: Substantial Equivalence - Comparison of Medtronic Transportation/Sterilization Cassettes with Predicates on Technological Characteristics

{8}------------------------------------------------

{9}------------------------------------------------

| Feature                         | Subject Device                                                                                                                                                                                                           | K131455<br>(Primary<br>Predicate)                                                                               | K033222                                                                                             | K993535                                                                                                  | K120947                                                                                                                                                                                                                         |                                                                                                                 |                                      |                                                                                                                                                                                         |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 7. Configuration                | Perforated bases, lids, and inserts                                                                                                                                                                                      | Perforated<br>bases, lids,<br>and inserts                                                                       | • Perforated<br>or Solid<br>Lids<br>• Perforated<br>Bottoms                                         | Perforated<br>bases, lids,<br>and inserts                                                                | Perforated<br>bases, lids,<br>and inserts                                                                                                                                                                                       |                                                                                                                 |                                      |                                                                                                                                                                                         |
| 8. Air Permeance                | Yes                                                                                                                                                                                                                      | Yes                                                                                                             | Yes                                                                                                 | Yes                                                                                                      | Yes                                                                                                                                                                                                                             |                                                                                                                 |                                      |                                                                                                                                                                                         |
| 9. Percent<br>Perforation       | Evenly distributed hole pattern.                                                                                                                                                                                         | Evenly<br>distributed hole<br>pattern.                                                                          | Evenly<br>distributed<br>hole pattern.                                                              | Evenly<br>distributed<br>hole pattern.                                                                   | Evenly<br>distributed<br>hole pattern.                                                                                                                                                                                          |                                                                                                                 |                                      |                                                                                                                                                                                         |
| 10. Sterilization<br>Method     | • Pre-Vacuum<br>• Gravity Displacement                                                                                                                                                                                   | High Vacuum<br>Steam                                                                                            | Pre-<br>vacuum<br>Steam                                                                             | • Pre-vacuum<br>Steam<br>• Ethylene<br>Oxide<br>Sterilization                                            | •Pre-Vacuum<br>•Gravity<br>Displacement                                                                                                                                                                                         |                                                                                                                 |                                      |                                                                                                                                                                                         |
| 11. Sterilization<br>Parameters | Cycle<br>Gravity<br>Displacement<br>Gravity<br>Displacement<br>Gravity<br>Displacement<br>Dynamic-Air-<br>Removal<br>(4 Pre-<br>conditioning<br>pulses)<br>Dynamic-Air-<br>Removal<br>(4 Pre-<br>conditioning<br>pulses) | Temperature<br>250°F<br>(121°C)<br>270°F<br>(132°C)<br>275°F<br>(135°C)<br>270°F<br>(132°C)<br>275°F<br>(135°C) | Exposure<br>time<br>30<br>Minutes<br>15<br>Minutes<br>10<br>Minutes<br>4<br>Minutes<br>3<br>Minutes | Minimum<br>dry time<br>30<br>Minutes<br>30<br>Minutes<br>30<br>Minutes<br>30<br>Minutes<br>30<br>Minutes | High Vacuum<br>(pre-vacuum,<br>three pulse,<br>standard):<br>Temp: 270° F<br>Exposure<br>Time: 4<br>Minutes<br>Cycle Dry Time<br>(wrapped):<br>20 Minutes<br>(minimum)<br>Cool Time:<br>Varies<br>according to<br>load contents | Prevacuum<br>Steam: 132°C<br>- 134°C<br>5 minutes<br>minimum<br>Drying Time:<br>10 –20<br>minutes as<br>needed. | Not provided<br>in 510(k)<br>Summary | Pre-Vacuum<br>Parameters:<br>270°F for 4<br>minutes<br>Pre-Vacuum<br>Dry Time: 30<br>minutes<br>Gravity<br>Parameters:<br>250°F for 30<br>minutes<br>Gravity Dry<br>Time: 30<br>minutes |
| 12. Reusable                    | Yes                                                                                                                                                                                                                      | Yes                                                                                                             | Yes                                                                                                 | Yes                                                                                                      | Yes                                                                                                                                                                                                                             |                                                                                                                 |                                      |                                                                                                                                                                                         |

{10}------------------------------------------------

|    | Feature                                                      | Subject Device                                              | K131455<br>(Primary<br>Predicate)                        | K033222                                                  | K993535                                 | K120947                                 |
|----|--------------------------------------------------------------|-------------------------------------------------------------|----------------------------------------------------------|----------------------------------------------------------|-----------------------------------------|-----------------------------------------|
| 1. | Sterilant<br>Penetration                                     | Yes                                                         | Yes                                                      | Yes                                                      | Yes                                     | Yes                                     |
| 2. | Microbial<br>Barrier<br>Properties                           | To be used with an<br>FDA approved<br>sterilization wrap    | To be used with an<br>FDA approved<br>sterilization wrap | To be used with an<br>FDA approved<br>sterilization wrap | No claims are<br>made in the<br>summary | No claims are<br>made in the<br>summary |
| 3. | Material<br>Compatibility<br>with<br>Sterilization<br>Method | Materials are<br>compatible with<br>sterilization<br>method | Yes                                                      | Compatible to<br>Steam Sterilization<br>at 132°C         | Yes                                     | Yes                                     |
| 4. | Toxicological<br>Properties                                  | Materials are<br>biocompatible                              | Materials are<br>biocompatible                           | Materials are<br>biocompatible                           | Not stated in<br>summary                | Not stated in<br>summary                |

Table 3: Substantial Equivalence - Comparison of Medtronic Transportation/Sterilization Cassettes with Predicates on Performance Data

No new issues of safety or effectiveness have been raised, based on the nonclinical tests performed. There are no differences between the subject and predicate devices. The subject device is as safe, as effective and performs at least as safely and effectively as the legally marketed predicate devices.

{11}------------------------------------------------

#### VII. Performance Testing

The testing of the Medtronic Transportation/Sterilization Cassettes is supported by packaging testing, sterilization validations, and biocompatibility testing. . The performance testing performed on the Medtronic Transportation/Sterilization Cassettes was tested and validated to the greatest challenge (worst case).

Medtronic Transportation/Sterilization Cassettes were evaluated on the ability to withstand the distribution environment. The distribution testing concluded that the Medtronic Transportation/Sterilization Cassettes were able to with stand the distribution environment. The testing was conducting in accordance with ASTM D4169: 2009 – "Standard Practice for Performance Testing of Shipping Containers and Systems".

To establish substantial equivalence with a predicate device, the performance testing was conducted to confirm that SAL (10 %) sterility level was achieved at the validated sterilization parameters. The testing was conducting in accordance with the following standards:

- . AAMI TIR39: 2009 "Guidance On Selecting A Microbial Challenge And Inoculation Sites For Sterilization Validation Of Medical Devices"
- AAMI TIR30: 2011 "A Compendium Of Processes, Materials, Test ● Methods, And Acceptance Criteria For Cleaning Reusable Medical Devices"
- AAMI TIR17: 2008 "Compatibility Of Materials Subject To ● Sterilization"
- AAMI TIR12: 2010 "Designing, Testing, And Labeling Reusable . Medical Devices For Reprocessing In Health Care Facilities: A Guide For Medical Device Manufacturers"
- ANSI/AAMI ST81: 2004 "Sterilization Of Medical Devices-. Information To Be Provided By The Manufacturer For The Processing Of Resterilizable Medical Devices"
- ANSI/AAMI ST79: 2010 (R) 2014 "Comprehensive Guide To Steam ● Sterilization And Sterility Assurance In Health Care Facilities -Incorporates "
- ANSI/AAMI/ISO TIR17665-2: 2009 "Sterilization Of Health Care ● Products - Moist Heat - Part 2"
- ANSI/AAMI/ISO 17665-1: 2006 "Sterilization Of Health Care ● Products-Moist Heat-Part 1: Requirements For The Development, Validation, And Routine Control Of A Sterilization Process For Medical Devices"
- ISO 17664: 2004 "Sterilization Of Medical Devices Information To ● Be Provided By The Manufacturer For The Processing Of Resterilizable Medical Device"

{12}------------------------------------------------

- ANSI/AAMI/ISO 14161: 2009 "Sterilization Of Health Care Products ● — Biological Indicators — Guidance For The Selection, Use And Interpretation Of Results "
- ANSI/AAMI/ISO 11737-2: 2009 "Sterilization Of Medical Devices ● Microbiological Methods — Part 2: Tests Of Sterility Performed In The Definition, Validation And Maintenance Of A Sterilization Process "

The Medtronic Transportation/Sterilization Cassettes do not have direct patient contact. The Medtronic Transportation/Sterilization Cassette components are fabricated from a variety of intrinsically stable metals and thermoplastics. Based on the testing performed, should substances transfer from the Medtronic Transportation/Sterilization Cassettes they would not present a chemical hazard. The assessment was conducted based on ISO 10993-1:2009 "Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process".

#### VIII. Conclusion:

Results of performance testing, and the similarities with legally marketed predicate devices it can be concluded that the Medtronic Transportation/Sterilization Cassettes will perform to the intended use. Medtronic believes that the subject Transportation/Sterilization Cassettes is substantially equivalent to the following predicates:

- Predicate 1 Restore Modular Sterilization Tray System ● K131455 (S.E. 08/30/2013)
- Predicate 2 Olympus Sterilization Trays K033222 (S.E. ● 08/02/2004)
- Predicate 3 Metapak Multi-Purpose Instrument Tray . K993535 (S.E. 12/07/2001)
- Predicate 4 THS Sterilization Tray K120947 (S.E. ● 10/17/2012)
- Reference Device K130720 Synthes Reusable ● Sterilization Container Systems (S.E. 08/14/2014) for the fabrication of nylon and polypropylene as materials.

No new issues of safety or effectiveness have been raised, based on the nonclinical tests performed. The subject device is as safe, as effective and performs at least as safely and effectively as the legally marketed predicate devices. Therefore, the Medtronic Transportation/Sterilization Cassettes are considered to be substantially equivalent to the predicate devices, Class II (21 CFR 880.6850), Product code KCT.

---

**Source:** [https://fda.innolitics.com/device/K152241](https://fda.innolitics.com/device/K152241)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
