← Product Code [LHN](/productcode/LHN) · K152224

# Proton Therapy System - Proteus 235 (Proteus One, Proteus Plus, Proteus TK2) (K152224)

_Ion Beam Application S.A. · LHN · Aug 9, 2016 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K152224

## Device Facts

- **Applicant:** Ion Beam Application S.A.
- **Product Code:** [LHN](/productcode/LHN.md)
- **Decision Date:** Aug 9, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic

## Indications for Use

The Proton Therapy System - Proteus 235 (brand names: Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck.

## Device Story

Proton therapy system delivers proton beams for radiation treatment of localized tumors; utilizes either isochronous cyclotron (C230) or superconducting synchro-cyclotron (S2C2) to generate high-energy protons; beam delivery equipment directs protons to patient treatment site; used in clinical settings (fixed beam or gantry rooms); operated by trained medical personnel; output used by clinicians for targeted tumor irradiation; benefits include precise dose delivery to localized tissue while sparing surrounding healthy structures.

## Clinical Evidence

Bench testing only. System and sub-system level verification and validation demonstrated performance equivalence to the predicate. Key metrics compared include maximum beam range (≥32 cm), field size (24x20 cm), distal fall-off (≤0.25 g/cm2), spot size (≤15 mm), spot position accuracy (≤1.5 mm), irradiation time (<2 minutes for 2Gy/1L), and effective SAD (≥2m).

## Technological Characteristics

Medical charged-particle radiation therapy system. Accelerator options: 230 MeV isochronous cyclotron (resistive coils) or 230 MeV superconducting synchro-cyclotron (superconducting coils). Cooling: chilled water or Gifford-Mc Mahon cryocoolers. Beam delivery: energy selection system. Dimensions: S2C2 volume ~10 m³, weight ~42 tons. Connectivity: integrated beam line to treatment rooms. Sterilization: N/A (not patient-contacting).

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- IBA Proton Therapy System ([K132919](/device/K132919.md))
- S-250 Proton Beam Radiation Therapy system ([K120676](/device/K120676.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is a stylized image of three human profiles facing to the right, with a wing-like shape extending from the back of the heads.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 9, 2016

Ion Beam Application S.A. % Mr. Bruce Armon Partner Saul Ewing LLP Center Square West, 1500 Market Street, 38th Floor PHILADELPHIA PA 19102-2186

Re: K15224

Trade/Device Name: Proton Therapy System - Proteus 235 (Proteus One, Proteus Plus, Proteus TK2) Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: LHN Dated: July 4, 2016 Received: July 8, 2016

Dear Mr. Armon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael D'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known) K152224

#### Device Name

Proton Therapy System - Proteus 235 (Proteus One, Proteus Plus, Proteus TK2)

#### Indications for Use (Describe)

"The Proton Therapy System - Proteus 235 (brand names: Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck."

|  | <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|--|-----------------------------------------------------------------------------------------------|
|  | Over-The-Counter Use (21 CFR 801 Subpart C)                                                   |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Type of Use (Select one or both, as applicable)

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a logo for an organization, possibly a company or association. The logo features the letters 'Iba' in a stylized, lowercase font, rendered in a vibrant green color. Above the 'I' is a green circle, and a curved green line extends from the top left, adding a dynamic element to the design. The overall impression is clean, modern, and suggests a focus on nature or growth.

M-Id 49380 Rev.B

Page 1

# 510(k) SUMMARY Ion Beam Applications S.A.

24 June 2016

# Applicant

| Ion Beam Applications S.A. |                                 |  |  |  |  |  |
|----------------------------|---------------------------------|--|--|--|--|--|
| Chemin du Cyclotron, 3     |                                 |  |  |  |  |  |
| B-1348 Louvain-la-Neuve    |                                 |  |  |  |  |  |
| Belgium                    |                                 |  |  |  |  |  |
| Attention :                | Mrs. Anne-Sophie Grell          |  |  |  |  |  |
| Phone :                    | 32-10-47-58-16                  |  |  |  |  |  |
| Facsimile:                 | 32-10-47-58-10                  |  |  |  |  |  |
| E-mail:                    | anne-sophie.grell@iba-group.com |  |  |  |  |  |

# Contact and Agent for Ion Beam Applications S.A.

| Bruce D. Armon                  |                 |
|---------------------------------|-----------------|
| Saul Ewing LLP                  |                 |
| 1500 Market Street              |                 |
| Centre Square West – 38th Floor |                 |
| Philadelphia, PA 19102          |                 |
| Phone:                          | (215) 972-7124  |
| Facsimile:                      | (215) 972-1906  |
| E-mail:                         | barmon@saul.com |

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for "Iba" in green. To the left of the logo is the text "M-Id 49380 Rev.B Page 2". The text appears to be a document identifier and page number.

## Classification Name

Medical Charged-Particle Radiation Therapy Systems. (21 C.F.R. \$892.5050, Product Code LHN)

Trade Name: Proton Therapy System - Proteus 235, Proteus One, Proteus Plus, Proteus TK2.

Common Name: Proton Therapy System

### Predicate Device

The IBA Proton Therapy System – Proteus 235 (PTS) with the addition of the IBA Compact Beam Production System is substantially equivalent to the previously cleared IBA Proton Therapy System (K132919) and to the S-250 Proton Beam Radiation Therapy system (K120676) from Mevion Medical Systems that includes a super conducting synchrocyclotron accelerator. The current PTS and its predicate devices have the same intended use and principles of operation, and are substantially equivalent in terms of technological characteristics.

### Indication for Use

The intended use of the product (wth the addition of the IBA Compact Beam Production Systems (CBPS) is not changed in respect to the previously cleared intended use (including no change in labelling that would affect it):

"The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck."

### Pre-market notification description

The change introduced in the IBA Proton Therapy System - Proteus 235 and described in this filing is the introduction of an optionally new IBA Compact Beam Production System (CBPS). Compact Beam Production System (CBPS) feature consists of the introduction, as an option, of a SuperConducting Synchro-Cyclotron (S2C2), offering a compact and more affordable alternative to the isochronous cyclotron. Both synchrocyclotron and isochronous cyclotron are dedicated to the generation of high energy protons for protontherapy. This is accompanied by an adaptation of the beam delivery equipment, which directs the proton beam to the patient's treatment site, to process the beam produced by the S2C2.

### Description of the device modification

This PTS corresponds to the previously cleared IBA Proton Therapy System (K132919) with the addition of IBA Compact Beam Production System (CBPS) based on a Superconducting Synchro-cyclotron (S2C2), which produces and delivers a proton beam for the treatment of patients.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows a logo for an organization, possibly named "Iba". The logo is primarily green and features a stylized, cursive font. Above the text, there is a curved line that extends from the left, ending with a circular dot, giving the impression of a check mark or a stylized accent mark. The overall design is simple, clean, and suggests a focus on nature or health due to the green color.

M-Id 49380 Rev.B

Page 3

A Proteus 235 Proton Therapy System (PTS) includes one proton accelerator for the production of the beam which can be shared between several treatment rooms in the Proteus Plus configurations and directed towards one only room in the Proteus One configurations. With the actual change, the accelerator can be either the isochronous cyclotron (C230) as in the previously cleared IBA PTS, or a Superconducting Synchro-Cyclotron (S2C2). As before, the PTS can be equipped with treatment rooms of several types according to the range of beam incidences they offer: e.g. a fixed beam treatment room or a gantry treatment room. These rooms are connected to the beam line transporting the protons produced by the accelerator (C230 or S2C2).

This change has no impact on:

- . The intended use(including no changes in labelling that would affect it);
- The fundamental scientific technology of the Proteus 235 including operating ● principles and mechanism of action;
- The product's clinical efficacy.

# Technological Characteristics of the modified device

The following table gives a high level overview of the main characteristics of the IBA Superconducting Synchro-cyclotron (S2C2), compared to the predicate conventional IBA isochronous cyclotron (C230), as in the previously cleared PTS, and the predicate synchro-cyclotron from Mevion Medical Systems already used in Proton Therapy.

| Characteristic                               | IBA Isochronous<br>cyclotron C230             | IBA Superconducting<br>synchrocyclotron S2C2         | Mevion Medical Systems<br>synchrocyclotron S-250 |
|----------------------------------------------|-----------------------------------------------|------------------------------------------------------|--------------------------------------------------|
| Accelerator                                  | 230 MeV isochronous<br>cyclotron              | 230 MeV superconducting<br>synchrocyclotron          | 250 MeV superconducting<br>synchrocyclotron      |
| Particle                                     | Proton                                        | Proton                                               | Proton                                           |
| Energy range                                 | up to 230 MeV with<br>energy selection system | up to 230 MeV with<br>energy selection system        | Up to 250 MeV                                    |
| Ion Source                                   | Hot cathode PIG ion<br>source                 | Cold cathode PIG source                              | Cold cathode PIG source                          |
| Type of coils                                | Resistive coils                               | Superconducting coils                                | Superconducting coils                            |
| Volume (m³)                                  | ~30                                           | ~10                                                  | 3-5                                              |
| Weight (tons)                                | ~230                                          | ~42                                                  | ~20                                              |
| Estimated<br>Electrical power<br>consumption | 2258 MWh/year                                 | 358 MWh/year                                         | 100 MWh/year                                     |
| Cooling method                               | Chilled water                                 | Chilled water and<br>Gifford-Mc Mahon<br>cryocoolers | Gifford-Mc Mahon<br>cryocoolers                  |

Table 1: Overview of the main characteristics of the synchro-cyclotron versus the IBA Isochronous cyclotron and the Mevion synchrocyclotron

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows a logo for an organization, possibly named "Iba". The logo is primarily green and features stylized lettering. Above the letters, there is a curved line that extends from the left, arching over the letters, and ending on the right. A green dot is positioned above the left side of the letters, completing the design.

M-Id 49380 Rev.B

Page 4

The Proteus 235 with the CBPS feature has the same intended use and indications, principles of operation, and technological characteristics as IBA Proton Therapy System (K132919) and to the S-250 Proton Beam Radiation Therapy system (K120676) from the Mevion Medical Systems for the synchrocyclotron accelerator.

# Summary of Performance testing

Testing has been performed at the system and sub-system level and demonstrated that the IBA Proton Therapy System - Proteus 235 with the CBPS new feature is substantially equivalent to the predicate device. Thus, this device is as safe, as effective and performs as well as the predicate device. From those testing, the following results have been obtained:

| Characteristic                                        | IBA Proteus Compact Gantry<br>Beam Line with Superconducting<br>synchrocyclotron (CBPS)<br>(K152224) | IBA Proteus Compact Gantry<br>Beam Line with C230 cyclotron<br>(K 132919) |
|-------------------------------------------------------|------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| Maximum<br>Beam Range in<br>Patient (Tissue<br>depth) | > or = 32 cm                                                                                         | > or = 32 cm                                                              |
| Field Size                                            | 24 cm x 20 cm                                                                                        | 24 cm x 20 cm                                                             |
| Beam Distal<br>Fall Off                               | ≤0.25 g/cm2 above the physical limit                                                                 | ≤0.25 g/cm2 above the physical limit                                      |
| Spot Size                                             | ≤ 15 mm on the range of energies                                                                     | ≤ 15 mm on the range of energies                                          |
| Spot Position<br>Accuracy                             | ≤ 15 % of beam sigma or < to 1.5 mm                                                                  | ≤ 15 % of beam sigma or < to 1.5 mm                                       |
| Irradiation<br>Time                                   | <2 minutes for delivering<br>2Gy to 1L volume<br>(10 x 10 x 10 cm3)                                  | <2 minutes for delivering<br>2Gy to 1L volume<br>(10 x 10 x 10 cm3)       |
| Effective SAD                                         | ≥ 2m                                                                                                 | ≥ 2m                                                                      |

Table 2. Overview of the clinical beam performance characteristics of the IBA Proton Therapy System using a cyclotron versus synchrocyclotron

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for IBA, which is a green stylized text. To the left of the logo is the text "M-Id 49380 Rev.B Page 5". The text is black and appears to be a document identifier and page number.

# Conclusion

The verification and validation activities ensure that the device is as safe, as effective, and performs as well as the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K152224](https://fda.innolitics.com/device/K152224)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
